Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K241895

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-09-26

    (90 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K201454

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cannulated PsiFGuard is an accessory to systems intended for sacroiliac joint fusion for conditions including sacrolliac joint disruptions and degenerative sacroilitis. It is indicated for use during sacroiliac joint guidewire placement to provide feedback to the surgeon via visual and audible alerts that indicate a change in conductivity at the tip of the probe and may indicate contact of the tip with bone and possible cortex perforation.

    Device Description

    The Cannulated PsiFGuard is a device that assists surgeons to place a guidewire in the sacroiliac joint during posterior sacroiliac fusion surgery. The Cannulated PsiFGuard device serves to alert the surgeons prior to a possible cortex perforation during quidewire placement, by analyzing the electrical conductivity of the surrounding tissues. The Cannulated PsiFGuard consists of a stainless steel needle component and electronic handle (DSG Handle or the DSG Connect Handle).

    The Cannulated PsiFGuard has exactly the same principles of operation as the previously cleared systems, i.e. the tip of the instrument simultaneously penetrates the tissue for guidewire placement and senses possible cortex perforation or contact with bone. The devices then provide a visual and audible feedback to the surgeon to indicate changes in conductivity associated with possible bone perforation. If the DSG Connect handle is used, the device also allows for the signal generated by the device to be transferred to a tablet, to visually reproduce the audio signal over time.

    AI/ML Overview

    The provided document does not contain detailed information about the acceptance criteria and a specific study that proves the device meets those criteria for the Cannulated PsiFGuard. The "Performance Data" section mentions a cadaver study and mechanical performance evaluation, but it lacks the specific metrics, sample sizes, and detailed outcomes that would constitute clear acceptance criteria and proof of meeting them.

    However, based on the limited information available and inferences from typical FDA submissions, here's a structured response addressing the requested points:


    Acceptance Criteria and Device Performance for Cannulated PsiFGuard

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criterion (Inferred)Reported Device Performance (From document)
    Guidewire placement accuracy (in sacroiliac joint fusion procedures with the device vs. standard technique)Demonstrated in a cadaver study. (Specific metrics of accuracy, e.g., % accurate placement, error margin, were not provided).
    Ability to measure tissue impedance and provide alertsDemonstrated in a cadaver study. (Specific sensitivity/specificity or other performance metrics were not provided).
    Mechanical robustness under cantilever bending loadsWithstood mechanical action induced by sacroiliac joint guidewire placement. (Specific thresholds or pass/fail criteria were not provided).
    Functionality as intended"In all instances, the Cannulated PsiFGuard functioned as intended and the performance observed was as expected."

    2. Sample size used for the test set and the data provenance

    • Sample Size: The document mentions a "cadaver study" but does not specify the number of cadavers or individual guidewire placements analyzed.
    • Data Provenance: Prospective (cadaveric study setting). The country of origin is not specified, but the applicant (SpineGuard S.A.) is based in Vincennes, France.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document.

    4. Adjudication method for the test set

    This information is not provided in the document.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study involving human readers/surgeons and AI assistance is not described in the document. The device provides "visual and audible alerts" to a surgeon, but the study described is a cadaver study for device performance, not a study evaluating human performance with and without the device's assistance. The device is not an AI/Software as a Medical Device (SaMD) in the typical sense that would necessitate such a study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device is explicitly designed to provide real-time feedback (visual and audible alerts) to a surgeon during guidewire placement. Therefore, it's an "in-the-loop" device, and a standalone algorithm-only performance assessment would not be applicable or relevant to its intended use. The cadaver study investigated the device's ability to measure impedance and provide alerts, which is its core standalone function within the human-assisted workflow.

    7. The type of ground truth used

    For the cadaver study:

    • For guidewire placement accuracy: Likely determined by imaging (e.g., fluoroscopy, CT scans) after placement, or direct anatomical visualization, to verify correct guidewire position relative to the sacroiliac joint.
    • For cortex perforation/bone contact detection: Likely determined by direct visual inspection during disarticulation/dissection of the cadaveric specimens, or confirmation via imaging, combined with the device's alert.

    8. The sample size for the training set

    The document does not mention a training set in the context of device performance studies. The Cannulated PsiFGuard's principle of operation is based on established electrical conductivity sensing technology, as described by its reference device (K201454 Cannulated PediGuard with DSG Connect technology, SpineGuard S.A.). This suggests the core technology was already developed and potentially "trained" (in an engineering sense) or validated in prior devices, rather than requiring a new "training set" for a machine learning algorithm.

    9. How the ground truth for the training set was established

    As no training set is mentioned in the context of this submission, this information is not applicable/not provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222106

    Validate with FDA (Live)

    Device Name
    Waypoint GPS
    Date Cleared
    2023-05-12

    (298 days)

    Product Code
    Regulation Number
    874.1820
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K201454

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Waypoint GPS™ is indicated for use during pedicle screw pilot hole drilling to provide visual feedback to the surgeon of changes in color to the tissue at the tip of the probe that may indicate contact with soft tissues and possible vertebral cortex perforation. The Waypoint GPS™ is indicated for use in both open and percutaneous (MIS) procedures.

    Device Description

    The Waypoint GPS™ is comprised of a single use, disposable device and a software application that is downloaded onto the provided tablet. The system is intended to surgeons during pilot hole drilling for pedicle screw placement. Waypoint GPS™ probe is single use disposable and consists of a handle containing electronics and a stainless-steel shaft with distal LED and fiber optics for visualizing the tissues immediately in contact with the distal end of the probe. The device produces visual signals to indicate changes in color associated with possible verforation. It uses light to provide visual alerts to indicate a change in anatomical structure to prevent placement in soft tissues or vertebral cortex perforation.

    AI/ML Overview

    The Waypoint Guided Pedicle System (GPS) is indicated for use during pedicle screw pilot hole drilling to provide visual feedback to the surgeon of changes in color to the tissue at the tip of the probe that may indicate contact with soft tissues and possible vertebral cortex perforation.

    Here's an analysis of the acceptance criteria and the study results:

    1. Table of Acceptance Criteria and Reported Device Performance

    Although explicit "acceptance criteria" values (e.g., minimum accuracy percentage) are not provided in a single, clear table within the document, we can infer the criteria from the "Test Method Summary" and "Results" columns, especially for the GLP Animal Study which directly compares performance.

    Acceptance Criteria (Inferred)Reported Device Performance
    Biocompatibility: Device materials are non-cytotoxic, non-sensitizing, non-irritating, non-toxic, and non-pyrogenic.Passed. Non-cytotoxic, non-sensitizing, non-irritating, non-toxic, non-pyrogenic.
    Insertion Force: Device can complete a minimum of 15 insertions into Sawbones at a specified depth without mechanical failure.Passed. Each shaft was able to successfully complete 15x insertions without mechanical failure.
    Torque on the shaft (endurance): Device can withstand 15x torque cycles on the shaft without mechanical failure while inserted in Sawbones.Passed. Each shaft was able to successfully withstand 15x torque on a shaft cycle without mechanical failure.
    Torque to failure: Device shaft can withstand a target torque without failure.Passed. Each shaft was able to successfully withstand the target torque.
    Simulated Use Testing: Device can successfully complete pairing with tablet, color calibration, color testing, and battery monitoring after endurance testing. Users can perform procedure and interpret results.Passed. Each device successfully completed all procedure steps. Users can perform the device procedure and correctly interpret the device output.
    Visualization and Color: Device output (RGB values and color change) reliably reflects contact with different colored materials.Passed. Each device successfully completed the test cycle and displayed correct colors and RGB values.
    Electrical Safety and EMC: Device complies with IEC 60601-1 and IEC 60601-1-2 standards.Passed.
    Software Verification and Validation: Software complies with FDA guidance for medical device software (moderate level of concern).Passed.
    Sterilization Validation: Sterilization process achieves sterility.Passed.
    Distribution and Aging: Device maintains performance (insertion force, torque, simulated use, visualization, color) after climactic conditioning, accelerated aging, and package performance testing. Sterile barrier is maintained.Passed. Performance tests were successful on both unaged and aged devices. Sterile barrier was maintained.
    GLP Animal Study (Non-Inferiority): The subject device (Waypoint GPS System) is non-inferior to the predicate PediGuard device in correctly identifying tip location (cortical bone, cancellous bone, breach). Users can adequately interpret results.Passed. Users correctly identified GPS device tip location in 100% of procedures. Users successfully distinguished between cortical bone, cancellous bone, and a breach by the change in color and RGB values. Users correctly identified PediGuard tip location in 95% of procedures. Demonstrated with statistical significance (0.050) that the subject device is non-inferior to the predicate.

    2. Sample Size Used for the Test Set and the Data Provenance

    • GLP Animal Study: The study involved multiple users performing pilot hole procedures in ovine lumbar vertebrae. The exact number of individual procedures or specific animal count is not given, but it states "The total number GPS procedures (including transverse processes) demonstrated the device performs as intended and users adequately interpret results with 95/95." This implies a total of 95 GPS procedures were evaluated.
      • Provenance: This was an in vivo animal study using ovine (sheep) lumbar vertebrae. The geographical origin of the animals or the study site is not explicitly stated. It is a prospective study design for the animal study component.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • GLP Animal Study: "Fluoroscopic CT was used to verify pilot hole depth, GPS, and PediGuard results." This indicates that imaging (Fluoroscopic CT) served as the primary, objective ground truth. It is not explicitly stated how many experts (e.g., radiologists) interpreted these CT scans or their qualifications. However, in such studies, interpretation of CT by qualified personnel (likely veterinarians or radiologists with expertise in animal anatomy) would be standard.
    • Cadaver Study: The document mentions "a cadaver," implying human cadavers were used for a portion of the simulated use testing. Again, the number and qualifications of experts establishing ground truth (likely anatomical dissection for confirmation) are not specified.

    4. Adjudication Method for the Test Set

    • GLP Animal Study: The primary ground truth was established by Fluoroscopic CT. The interpretation of this CT data (which would inherently involve expert adjudication if there were ambiguities, though not explicitly detailed) served as the objective measure. The device's output (color description for GPS, auditory for PediGuard) was then compared against this objective ground truth. It doesn't describe a specific adjudication method like "2+1" for human reader discrepancies, as the ground truth was derived from imaging.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Yes, a comparative study was done for the GLP Animal Study. It was a side-by-side comparison of the Waypoint GPS device and the predicate PediGuard device involving "multiple users" (readers).
    • Effect Size of Human Readers' Improvement with AI vs. without AI Assistance: This specific metric (improvement of human readers with AI assistance) is not directly provided or applicable in the requested format.
      • The study compares two devices: Waypoint GPS (subject device) which provides visual feedback (color changes) and PediGuard (predicate device) which provides auditory feedback (change in cadence and pitch). Both are "assistance" devices.
      • The study aims to show non-inferiority of the GPS device to the PediGuard device, not necessarily the improvement of human readers with either device compared to without any device.
      • However, the results show:
        • Users correctly identified GPS device tip location in 100% of procedures.
        • Users correctly identified PediGuard tip location in 95% of procedures.
      • This demonstrates that with the assistance of the GPS device, users were able to correctly identify tip location more consistently (100%) than with the predicate PediGuard device (95%) in the animal model. The study concludes statistical non-inferiority (p=0.050).

    6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study

    • No, a standalone study was not performed or described. The Waypoint GPS device is designed to provide "visual feedback to the surgeon" and its performance is evaluated in the context of "user interpretation" of its outputs. The "human-in-the-loop" is an integral part of its intended use and evaluation.

    7. Type of Ground Truth Used

    • GLP Animal Study: Fluoroscopic CT scans were used to objectively verify pilot hole depth and location, establishing the ground truth for bone anatomy and breach detection.
    • Cadaver Study: Likely anatomical dissection or expert visualization for the human cadaver component, though not explicitly detailed.

    8. Sample Size for the Training Set

    • Not Applicable / Not Provided. The document describes a medical device, not an AI/ML algorithm that undergoes a distinct training phase with a specific dataset. The "device" in this context refers to the physical instrument and its integrated electronics/software, which are likely designed and calibrated based on engineering principles and preclinical testing, rather than "training" on a dataset like a machine learning model.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable / Not Provided. As explained above, there isn't a "training set" in the common AI/ML sense for this device. The principles of operation (spectral reflectance for color changes) are built into the device design rather than learned from labeled data.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1