    Intended Use

    The Synergy Gold (SG) Portable (Models MM011450 and MM011460) is a powered breast pump to be used by lactating women to express and collect milk from their breast. The Synergy Gold (SG) Portable (Models MM011450 and MM011460) breast pump is intended for home use by a single user.

    Device Description

    The Synergy Gold (SG) Portable is a powered breast pump intended to express and collect milk from the breasts of lactating women. Pumping can be performed on one breast (single pumping) or both breasts (double pumping) at the same time. The user employs buttons to select one or both sides for pumping, to switch from massage mode to expression mode and to control the vacuum levels within those modes. Massage mode consists of 5 suction levels, while expression mode has 12 suction levels. The SG Portable breast pump is capable of providing vacuum levels from 50-270 mmHz with cycling rate or 100 cycles per minute in massage mode and 26 cycles per minute in expression mode. The SG Portable breast pump is powered by a 5V DC adaptor or an internal rechargeable lithium-ion polymer battery.

    The SG Portable breast pump is sold in two configurations (with and without breast shield and bottle sets). Model MM011450 includes the SG Portable breast pump, double collection kit (MM12302-C; breast shield set - medium 24 mm), power adapter, and two collection bottles. Model MM011460 includes the SG Portable breast pump and power adaptor: this model is to be used with separately purchased compatible breast shields and bottle sets.

    The SG Portable breast pump and associated breast pump kits are intended for a single user.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for a powered breast pump (Synergy Gold (SG) Portable). This type of document focuses on establishing substantial equivalence to a predicate device based on non-clinical testing, rather than extensive clinical studies or acceptance criteria for an AI/ML device.

    Therefore, the document does not contain the information requested regarding acceptance criteria for an AI/ML device, a study proving it meets those criteria, sample sizes for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details, or ground truth establishment methods for an AI/ML context.

    The document primarily details:

    • Device Name: Synergy Gold (SG) Portable (Models MM011450 and MM011460)
    • Regulation Number: 21 CFR 884.5160
    • Regulation Name: Powered Breast Pump
    • Product Code: HGX
    • Regulatory Class: II
    • Predicate Device: K191208 Spectra Cashmere Breast Pump
    • Intended Use: For lactating women to express and collect milk from their breast, intended for home use by a single user.
    • Non-Clinical Testing Summarized: Electrical Safety (ANSI/AAMI ES60601-1:2005+A1:2012, IEC 60601-1-11:2015, IEC 62133:2012), Risk Analysis (ISO 14971:2007), Electromagnetic Compatibility (IEC 60601-1-2:2014), Biocompatibility (by reference to predicate), Software Validation ("Moderate" level of concern per FDA Guidance 2005), Bench Testing (vacuum, cycles, AC/battery power, various breast shield sizes, backflow protection), Battery and pump use-life testing.

    The "acceptance criteria" discussed in this document refer to the device meeting its performance specifications through bench testing and compliance with electrical safety and EMC standards, a standard process for a physical medical device. It does not relate to the performance metrics of an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210759

    Validate with FDA (Live)

    Device Name
    Solo, Swing Maxi
    Manufacturer
    Date Cleared
    2021-08-11

    (149 days)

    Product Code
    Regulation Number
    884.5160
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K200508, K181937, K020518, K151632

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Solo™ / Swing Maxi™ breast pumps are powered breast pumps to be used by lactating women to express and collect milk from their breasts.

    The Solo™ / Swing Maxi™ breast pumps are intended for a single user.

    The breast pumps are intended to be used in a home environment.

    Device Description

    Solo™ and Swing Maxi™ are breast pump systems intended to be used in a home environment (or similar such as an office). Solo™ and Swing Maxi™ comprise a pump unit, power adapter, and one (Solo™) or two (Swing Maxi™) PersonalFit Flex™ connectors that include the connector body, membrane (diaphragm), and connector back cap. The breast pump systems also include breast shields (21, 24, 27, and 30 mm), bottle lid, bottle stand, and tubing. The device is provided non-sterile.

    The breast pumps can be used on one breast (single pumping; Solo™ and Swing Maxi™)) or on both breasts (double pumping) at the same time (Swing Maxi 101).

    The Solo"M and Swing Maxi™ breast pumps feature 2-Phase Expression® technology, which runs pumping in two phases (Stimulation and Expression) by applying a cyclic negative pressure to mimic a baby's natural nursing rhythm. A DC motor is used to drive a membrane aggregate. This membrane aggregate creates the negative pressure (suction) required to extract the breast milk. The pump unit includes the following features:

    • user-adjustable controls: "On/Pause/Off" for powering on/off or pausing the device, "Letdown" for switching between pumping modes, and "Increase vacuum"/ "Decrease vacuum" for controlling vacuum intensity levels;
    • a port for connection of the tubing that channels the vacuum for breast pumping;
    • a port for connection of the power supply;
    • a central LED light as status indicator; and
    • a textile lanyard as an interface to the user's clothes/body.

    The Solo™ and Swing Maxi™ breast pump systems allow the user to adjust the vacuum levels in both phases (stimulation and expression). The powered breast pumps are preprogrammed with variable vacuum levels and cvcle rates. The powered breast pumps are capable of providing vacuum levels from -45 to -140 mmHg with cycle speeds up to 111 cycles per minute for stimulation and from -45 to -245 mmHg with cycle speeds from 75 to 45 cycles per minute for expression.

    Two different product configurations are available for the Swing Maxi™ breast pump: without and with Bluetooth®. These two product configurations share the same mechanical and electrical design components; however, the Swing Maxi111 breast pump with Bluetooth® additionally features always-on enabled Bluetooth® connectivity. This allows monodirectional wireless data transmission from the breast pump to a Bluetooth®-enabled personal mobile device, such as a smartphone or tablet computer. When the Medela FamilyTM smart application is installed on a compatible mobile device and the device is paired via Bluetooth® with the breast pump unit, the user can: automatically record pumping data (session length, phases, and vacuum intensity levels), manually input the amount of milk expressed, and get notified when the battery is low as indicated in provided instructional material and in-app information.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the Medela Solo™ / Swing Maxi™ breast pumps. It describes the device, its intended use, and compares it to a predicate device. However, it does not contain information regarding acceptance criteria or a study that proves the device meets specific acceptance criteria in the context of an AI/ML medical device.

    The document states that the breast pumps comply with voluntary standards for electrical safety, electromagnetic compatibility, use in the home healthcare environment, and usability. It also lists performance data provided to support the substantial equivalence determination. These performance data are:

    • Electrical Safety testing: In accordance with ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012, and IEC 60601-11:2015.
    • Risk analysis: In accordance with ISO 14971:2007.
    • Electromagnetic compatibility testing: In accordance with IEC 60601-1-2:2014.
    • Biocompatibility evaluation: Completed according to FDA guidance "Use of International Standard ISO 10993-1," concluding no new testing was required as patient-contacting materials are identical to the predicate device.
    • Software validation: In accordance with FDA Guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (May 11, 2005). The software was considered a "Moderate" level of concern.
    • Bench testing: Conducted to determine minimum and maximum vacuum levels and cycle rates compared to specifications.
    • Battery and pump use life testing: Conducted to demonstrate device maintenance of specifications throughout its use life.

    Therefore, I cannot provide the requested information about acceptance criteria and a study proving those criteria are met for an AI/ML device because the provided text is for a traditional medical device (breast pump) and does not describe an AI/ML component or associated performance studies against specific AI/ML metrics.

    If you have a document related to an AI/ML medical device, please provide that, and I will do my best to extract the information you need.