Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K233302

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-11-16

    (48 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K203465, K193660, K213596, K221193

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OSSIOfiber® Compression Staple is indicated for fixation of arthrodesis, osteotomies and fractures in hand or foot surgery in the presence of appropriate brace and/or immobilization.

    The number and size of the OSSIOfiber® Compression Staples must be adapted to the indication.

    Device Description

    The OSSIOfiber® Compression Staple is a fixation implant made of degradable poly (L-lactide-co-D,Llactide) (PLDLA) reinforced with continuous mineral fibers. OSSIOfiber® implants have been shown to be biocompatible. The polymer content degrades by hydrolysis into alpha-hydroxy acids that are metabolized by the body. The fibers are made of minerals that are found in natural bone. As the OSSIOfiber® implants degrade, the load transfers to the surrounding anatomy throughout the healing period of the osteotomy. fusion, or fracture. Substantial decradation takes place within approximately 18 months as shown in pre-clinical studies, thus eliminating the requirement for future hardware removal surgery.

    The OSSIOfiber® Compression Staples are supplied sterile, for single patient use only, and nonpyrogenic. The additional devices included in this submission are: 9 x 10mm, and 11 x 10 mm.

    The OSSIOfiber® Compression Staples are designed to be used with commonly available orthopedic surgical tools such as ISO 9714 compatible instrumentations.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification summary for the OSSIOfiber® Compression Staple. It describes the device, its intended use, and the non-clinical data used to demonstrate substantial equivalence to predicate devices. However, this document does not contain information related to an AI/ML-driven medical device, digital health software, or any study involving human readers or ground truth established by experts.

    The primary focus of this submission is on the physical and mechanical properties of a medical implant (a compression staple) and its biocompatibility. The acceptance criteria discussed are related to the mechanical performance of the staple and its degradation profile.

    Therefore, I cannot provide the requested information regarding:

    • A table of acceptance criteria and reported device performance for an AI/ML device.
    • Sample size used for a test set or data provenance for an AI/ML device.
    • Number of experts used to establish ground truth or their qualifications.
    • Adjudication method for a test set.
    • Multi Reader Multi Case (MRMC) comparative effectiveness study or effect size.
    • Standalone (algorithm only) performance.
    • Type of ground truth used (expert consensus, pathology, outcomes data, etc.) for an AI/ML device.
    • Sample size for the training set or how ground truth was established for a training set (as this is not an AI/ML device).

    The document specifically mentions:

    • Non-Clinical Data: "Static bending, bending fatique and pull-out testing were performed to verify the strength and fixation properties of the OSSIOfiber® Compression Staple, and to compare them to those of the primary predicate device, the OS2®-VP Varisation Staple (K153770). Testing on the subject device were done initially and following in-vitro degradation. The in-vitro degradation profile was characterized."
    • Biocompatibility: "Biocompatibility for the implants was established primarily based on the referenced ISO 10993 data from the previously cleared predicate and reference devices (K212594, K203465) as well as a rationale."

    These are standard engineering and biocompatibility tests for physical medical devices, not performance metrics for AI/ML algorithms.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213415

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-03-28

    (160 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K181180, K190652, K201803, K193660, K203465

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OSSIOfiber® Suture Anchors are indicated for fixation of suture (soft tissue) to bone in the shoulder, footlankle, knee, hand/wrist, and elbow in the following procedures:

    • Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.
    • I Foot/Ankle: Lateral Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair and Bunionectomy.
    • I Knee: Anterior Cruciate Ligament Repair (4.75-5.5 Anchors Only), Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis and Quadriceps Tendon Repair. Secondary or adjunct fixation of ACL/PCL reconstruction or repair (4.75 - 5.5 Anchors only).
    • . Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction.
    • . Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis repair (Tennis Elbow).
    Device Description

    The OSSIOfiber® Suture Anchor consists of an eyelet and anchor body preloaded on an inserter. The anchor body and eyelet are made from poly (L-lactide-co-D,L-lactide) (PLDLA) reinforced with continuous mineral fibers. OSSIOfiber® implants have been shown to be biocompatible. The polymer content degrades by hydrolysis into alpha-hydroxy acids that are metabolized by the fibers are made from minerals that are found in natural bone. As the OSSIOfiber® implants degrade, the load transfers to the surrounding anatomy throughout the healing period of the bone. Substantial degradation takes place within approximately 18 months as shown in pre-clinical studies, thus eliminating the requirement for future hardware removal surgery. Sutures, needles and suture snare may also be provided with the device depending on configuration.

    The OSSIOfiber® Suture Anchors are sterile, single-use, and non-pyrogenic.

    AI/ML Overview

    The provided text is a 510(k) summary for the OSSIOfiber® Suture Anchor, a medical device. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving a device (like an AI model) meets acceptance criteria based on performance metrics such as sensitivity, specificity, or reader improvement.

    Therefore, I cannot provide the information requested in your prompt based on the provided text, as the document describes a mechanical medical device (suture anchor) and its non-clinical testing for substantial equivalence, not a study involving AI or human reader performance.

    The request asks for details like:

    • A table of acceptance criteria and reported device performance (in terms of AI metrics)
    • Sample size and data provenance for a test set
    • Number and qualifications of experts for ground truth
    • Adjudication methods
    • MRMC comparative effectiveness study results (AI vs. human)
    • Standalone algorithm performance
    • Type of ground truth used (e.g., pathology, outcomes)
    • Training set sample size and ground truth establishment

    None of this information is present in the provided 510(k) summary for the OSSIOfiber® Suture Anchor, which is a physical implant, not a diagnostic AI device. The "non-clinical data" section pertains to mechanical tests (static pull-out, cyclic pull-out, torsional strength, driving torque, insertion testing) and biocompatibility, which are relevant for a suture anchor but not for the type of AI-driven diagnostic device your prompt describes.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212594

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-01-21

    (158 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K181180, K190652, K193660, K201803

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OSSIOfiber® Staple is indicated for fixation of arthrodesis, osteotomies in hand or foot surgery in the presence of appropriate brace and/or immobilization.

    The number and size of the OSSIOfiber® Staples must be adapted to the indication.

    Device Description

    The OSSIOfiber® Staple is a fixation implant made of degradable poly (L-lactide) (PLDLA) reinforced with continuous mineral fibers. OSSIOfiber® implants have been shown to be biocompatible. The polymer content degrades by hydrolysis into alpha-hydroxy acids that are metabolized by the body. The fibers are made of minerals that are found in natural bone. As the OSSIOfiber® implants degrade, the load transfers to the surrounding anatomy throughout the healing period of the osteotomy, fusion, or fracture. Substantial degradation takes place within approximately 18 months as shown in pre-clinical studies, thus eliminating the requirement for future hardware removal surgery.

    The OSSIOfiber® Staples are supplied sterile, for single patient use only, and non-pyrogenic. They are available in several sizes: 11-25 mm bridge lengths, and 15-22 mm leg lengths.

    The OSSIOfiber® Staples are designed to be used with commonly available orthopedic surgical tools such as ISO 9714 compatible instrumentations.

    AI/ML Overview

    This document, K212594, is a 510(k) Premarket Notification for the OSSIOfiber® Staple. It outlines the FDA's determination of substantial equivalence to a legally marketed predicate device. This document focuses on the mechanical and material properties of the absorbable staple, not on an AI/ML-driven device or study involving human readers analyzing images. Therefore, many of the requested points regarding AI/ML study design, reader performance, ground truth establishment for AI training sets, etc., are not applicable to the content of this document.

    However, I can extract information related to the device's performance based on non-clinical data and the rationale for its acceptance.

    Here's a breakdown of the requested information based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a direct table of numerical "acceptance criteria" against "reported performance" for the device in the typical AI/ML context (e.g., sensitivity, specificity thresholds). Instead, it states that the device's performance was evaluated through non-clinical mechanical testing and compared to a predicate device. The acceptance criterion is "at least equivalent performance" to the predicate.

    Performance MetricAcceptance Criteria (Compared to Predicate/Reference)Reported Device Performance
    Static Bending StrengthAt least equivalent to OS2®-C Compression Staple (K153395)Demonstrated "at least equivalent performance" (initially and after in vitro degradation)
    Static Bending StiffnessAt least equivalent to OS2®-C Compression Staple (K153395)Demonstrated "at least equivalent performance" (initially and after in vitro degradation)
    Bending Fatigue StrengthAt least equivalent to OS2®-C Compression Staple (K153395)Demonstrated "at least equivalent performance" (initially and after in vitro degradation)
    Pull-out Fixation PropertiesAt least equivalent to OSSIO® Pin Product Family (K181180)Demonstrated "at least equivalent performance" (initially and after in vitro degradation)
    BiocompatibilityEstablished based on referenced ISO 10993 data from previously cleared devices and a rationaleEstablished via referenced ISO 10993 data and rationale.
    MR SafetyEvidence to support MR safe labelingRationale provided to support MR safe labeling.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This document details non-clinical laboratory testing of mechanical properties, not a clinical study on human subjects or an AI/ML test set. Therefore, concepts like "test set sample size" (in terms of patient data) and "data provenance" (country of origin, retrospective/prospective) are not applicable. The testing was conducted in a laboratory setting. The specific number of samples for each mechanical test (e.g., number of staples tested for bending) is not detailed in this summary, but standards like ASTM F564-17 typically specify minimum sample sizes for such tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as the device is an implant, and the testing involves mechanical and material properties, not interpretation of medical images or data by experts to establish a "ground truth" for an AI model. "Ground truth" here refers to the measured physical properties of the device.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not an AI/ML study involving human readers and adjudicated interpretations. Mechanical tests have objective outputs (e.g., force, displacement, cycles to failure).

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document describes a medical device (an absorbable staple), not an AI-driven diagnostic or assistive tool, and therefore no MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is not an AI algorithm. The "standalone performance" refers to the intrinsic mechanical properties of the device itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the mechanical testing, the "ground truth" is defined by the measured physical properties (e.g., ultimate bending strength, fatigue life, pull-out force) determined through standardized laboratory tests (e.g., ASTM F564-17). For biocompatibility, the ground truth is established by compliance with ISO 10993 standards and a rationale.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device that requires training data.

    9. How the ground truth for the training set was established

    Not applicable. As above, this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203465

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-01-06

    (43 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K190652, K193660

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OSSIOfiber® Trimmable Fixation Nails are indicated for maintenance of alignment and fixation of bone fractures, osteotomies, arthrodesis, and bone grafts in the presence of appropriate additional immobilization implants, cast, brace).

    Device Description

    OSSIOfiber® Cannulated Trimmable Fixation Nails are indicated for maintenance of alignment and fixation of bone fractures, osteotomies, arthrodesis, and bone grafts in the presence of appropriate additional immobilization (e.g. rigid fixation implants, cast, brace).

    The OSSIOfiber® Cannulated Trimmable Fixation Nail is a fixation device made of degradable poly (L-lactide-co-D.L-lactide) (PLDLA) reinforced with continuous mineral fibers. The polymer content degrades by hydrolysis into alpha-hydroxy acids that are metabolized by the fibers are made from materials that are found in natural bone. As the OSSIOfiber® implants degrade, the load transfers to the surrounding anatomy throughout the healing period of the osteotomy, fusion, or fracture. Substantial degradation takes place within approximately 18 months as shown in pre-clinical studies, thus eliminating the requirement for future hardware removal surgery.

    The OSSIOfiber® Cannulated Trimmable Fixation Nails are cannulated, supplied sterile, for single patient use only, and non-pyrogenic. They are available in several sizes: 10-70 mm long, and 2.80-4.60 mm nominal diameter (core diameter of 2.4-4.00 mm).

    The OSSIOfiber® Cannulated Trimmable Fixation Nails are designed to be used with commonly available orthopedic surgical tools such as ISO 9714 compatible instrumentations.

    AI/ML Overview

    This document describes a medical device clearance rather than a study evaluating device performance against acceptance criteria for an AI/ML powered device. Therefore, much of the requested information regarding AI/ML studies (such as sample sizes for test and training sets, number of experts, adjudication methods, MRMC studies, or standalone performance results) is not applicable or present in this provided text.

    Based on the provided text, the device in question is the OSSIOfiber® Cannulated Trimmable Fixation Nail. This device is a bone fixation fastener, not an AI/ML powered medical device. The document is a 510(k) summary, which outlines the substantial equivalence of this new device to legally marketed predicate devices.

    Here's the relevant information that can be extracted from the provided text:

    1. A table of acceptance criteria and the reported device performance:

    Since this is a 510(k) submission for a physical medical device and not an AI/ML software, "acceptance criteria" here refers to the performance of the device in mechanical testing compared to a predicate device, rather than diagnostic accuracy metrics typically seen with AI. Specific numeric acceptance criteria are not explicitly stated in the summary, but the general criterion is "at least equivalent performance" to the predicate.

    Acceptance Criterion (Implicit)Reported Device Performance
    Mechanical Performance: Equivalent or superior flexural bending strength compared to predicate.Mechanical testing demonstrated at least equivalent performance both initially and after in vitro degradation compared to the predicate device (OSSIO® Pin Product Family, K181180).
    Mechanical Performance: Equivalent or superior pull-out strength compared to predicate.Mechanical testing demonstrated at least equivalent performance both initially and after in vitro degradation compared to the predicate device (OSSIO® Pin Product Family, K181180).
    Mechanical Performance: Equivalent or superior shear strength compared to predicate.Mechanical testing demonstrated at least equivalent performance both initially and after in vitro degradation compared to the predicate device (OSSIO® Pin Product Family, K181180).
    Biocompatibility: Meet ISO 10993 standards and be rationally justified.Biocompatibility was established primarily based on referenced ISO 10993 data from previously cleared predicate and reference devices (K181180, K190652) and a rationale.
    Degradation Profile: Characterized in-vitro degradation similar to what is expected for the material.The in-vitro degradation profile (i.e., change in material properties) was characterized. Substantial degradation takes place within approximately 18 months as shown in pre-clinical studies.
    Safety and Effectiveness: Demonstrate substantially equivalent safety and effectiveness to the predicate.Non-clinical testing data demonstrate that the OSSIOfiber® Cannulated Trimmable Fixation Nail has a substantially equivalent safety and effectiveness profile as the predicate device.

    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not specified for the mechanical testing mentioned. The data provenance is pre-clinical studies and bench testing. The country of origin for the studies is not specified in this document.
    • Data Provenance: Pre-clinical studies and in-vitro mechanical testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable as this is a physical device being evaluated through mechanical and pre-clinical testing, not an AI/ML diagnostic device requiring expert ground truth for image interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC study was not done as this is not an AI/ML diagnostic device with human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable as this is not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the mechanical tests, the "ground truth" would be the measured physical properties (e.g., strength, stiffness) as determined by standardized laboratory methods, compared against the properties of the predicate device.
    • For biocompatibility, the ground truth is established through adherence to ISO 10993 standards and a rationale.

    8. The sample size for the training set

    • Not applicable as this is not an AI/ML device.

    9. How the ground truth for the training set was established

    • Not applicable as this is not an AI/ML device.

    In summary, the provided text details the FDA clearance of a physical medical device based on its substantial equivalence to an existing predicate device, primarily through mechanical and biocompatibility testing, rather than a study for an AI/ML product.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1