Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K232618

    Validate with FDA (Live)

    Date Cleared
    2023-10-27

    (59 days)

    Product Code
    Regulation Number
    882.1480
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K191861

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aurora Surgiscope System is intended for use in neurosurgery and endoscopic neurosurgery and pure neuroendoscopy (i.e. ventriculoscopy) for visualization, diagnostic and/or therapeutic procedures such as ventriculostomies, biopsies and removal of cysts, tumors and other obstructions.

    Device Description

    The Aurora Surgiscope System consists of two components: (1) a sterile, single use, sheath with integrated illumination LEDs and camera, with an obturator, and (2) a non-sterile, reusable control unit, Image Control Box (ICB).

    The sheath is intended to provide access to the surgical site by acting as the insertable portion of the device, as well as the instrument channel to accommodate other surgical tools. Depth markers are present along the length of the sheath for user reference.

    At the proximal end of the sheath is the imager, which comprises the following components: LEDs (light emitting diodes), camera (and optical components), and focus knob.

    • The LEDs provide illumination to the surgical field by directing light down the sheath, along . the instrument channel.
    • The camera captures video image of the surgical field. ●

    The proximal end of the sheath also contains a tab, which may be used to manually hold the device. To facilitate insertion of the surgical site, an obturator is provided with the device. During insertion, the obturator is fully inserted into the sheath, and the entire unit is advanced to the desired location. The distal end of the obturator is conical in shape to minimize tissue damage. In addition, the proximal handle of the obturator is designed to accommodate various stereotactic instruments for neuronavigation. Once inserted, the obturator is removed. Two sterile, single use accessories optional for use are provided with the Aurora Surgiscope System: an Irrigation Device and 12 French Suction Device.

    The ICB is a non-sterile device that provides three main functions in the Aurora Surgiscope System:

    • To power the Surgiscope LEDs and camera
    • . To relay the video feed captured by the Surgiscope camera to a display monitor for real-time image visualization
    • . To allow the user to make adjustments to the displayed video feed (e.g., contrast, brightness), as well as vary the LED light output.

    The user interface is a membrane keypad with buttons located on the ICB that can be depressed for image adjustment, such as zoom, contrast, brightness, and orientation. The ICB is supplied with two cables: A power cable for connection to an AC wall outlet, and a display cable for connection to a high definition surgical monitor.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the Aurora Surgiscope System do not contain information typically found in a clinical study report or performance evaluation for an AI/software device. The document focuses on demonstrating substantial equivalence to a predicate device, which means proving that the new device is as safe and effective as a legally marketed device, rather than rigorously quantifying performance against defined acceptance criteria in a study setting.

    Specifically, the document does not include:

    • A table of acceptance criteria and reported device performance related to a diagnostic or AI function.
    • Sample sizes for test sets or data provenance for AI model validation.
    • Details about expert readers, ground truth establishment, or adjudication methods for AI performance.
    • Information on multi-reader multi-case (MRMC) comparative effectiveness studies.
    • Standalone algorithm performance data.
    • Training set details for an AI model.

    The "testing" mentioned in the document pertains to traditional medical device testing for hardware, biocompatibility, electrical safety, and mechanical aspects. While it states "Software verification and validation testing" was conducted and "documentation provided as recommended by the FDA Guidance Content of Premarket Submissions for Device Software Functions," it does not provide any specific performance metrics or acceptance criteria for software functionality that would typically be associated with an AI/ML-driven device's diagnostic performance. The "Image Control Box" software mentioned focuses on image adjustment and display, not diagnostic interpretation.

    Therefore, based solely on the provided text, it is not possible to describe the acceptance criteria and the study proving the device meets those criteria from an AI/ML perspective. The device, as described, appears to be a neurological endoscope system for visualization, diagnostic, and therapeutic procedures, with software for image display and adjustment, not an AI-powered diagnostic tool.

    If this were an AI-powered device, the information requested would be crucial for its evaluation. Without it, I cannot fulfill the request for AI-related performance criteria.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1