Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    DEN220014

    Validate with FDA (Live)

    Date Cleared
    2023-02-01

    (343 days)

    Product Code
    Regulation Number
    892.5060
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K190978

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RefleXion Medical Radiotherapy System (RMRS) is indicated for treatment planning and precise delivery of image-guided radiation therapy, stereotactic radiotherapy, or stereotactic radiosurgery for tumors or other targeted tissues anywhere in the body when radiation treatment is indicated. while minimizing the delivery of radiation to vital healthy tissue. The megavoltage X-ray radiation is delivered in a rotational, modulated, image-guided format in accordance with the physician approved plan.

    The RefleXion Medical Radiotherapy System is also indicated for FDG-guided treatment which includes modeling, planning and precise delivery of FDG-guided radiation therapy, a type of Biology-guided Radiotherapy (BgRT), in five or fewer fractions for adults. It is indicated for tumor volumes in lung and bone subject to potential motion and positional uncertainty that have each been assessed with on-board PET/CT prior to delivery for adequate localization, sufficient FDG metabolic activity, local contrast and consistent biodistribution to meet the RMRS requirements, while minimizing the delivery of radiation to vital healthy tissue. BgRT involves the detection of signals from F18 during active beam delivery as a guide to deliver megavoltage X-ray radiotherapy in a rotational, modulated format in accordance with a physician approved treatment plan.

    For complete fludeoxyglucose F18 prescribing information, refer both to the current medical imaging agent labeling and to this device labeling under "FDG Medical Imaging Agent Information".

    Device Description

    The RMRS is a linear accelerator capable of delivering intensity-modulated radiation therapy, stereotactic body radiation therapy, stereotactic radiotherapy, stereotactic radiosurgery, and biology-guided radiotherapy (BgRT).

    The RMRS is comprised of six major subsystems as shown in Figure 1.

    The major subsystems of the RefleXion Medical Radiotherapy System are: 1) Compact linear accelerator, 2) primary collimation, 4) PET scanner, 5) kVCT, and 6) MV x-ray detectors.

    The linear accelerator portion of the device was previously cleared under a 510(k) premarket notification (K190978). The subject device in K190978 included the PET hardware, but the PET hardware and associated software was inactive and outside the scope of K190978. The subject of the current submission is the addition of the BgRT functionality and the activation of the PET hardware.

    Biology-guided Radiation Therapy

    BgRT is a type of volumetric modulated arc therapy that allows radiation dose delivery in a slice-by-slice fashion using a multi-pass couch motion based on the collection and processing of PET data from the radiotracer fludeoxyglucose F18 (FDG). BgRT also allows the dose delivery to be adjusted in real-time to account for target motion in the lung and bone.

    The PET system is comprised of two PET detector arcs mounted on the ring gantry that rotate with the system to generate complete tomographic samples of radiotracer distribution required for the BgRT algorithm.

    Rapidly acquired "limited time sampled" (LTS) PET imaging data are used to form accumulated lines of response (projection data). These data are then used to guide the beam using the mapping calculated during BgRT planning, maintaining the ability to track the target in real-time as it moves within a pre-defined volume called the biology-tracking zone (BTZ).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study proving the device meets them, based on the provided text, structured as requested:

    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text details several performance tests and their acceptance criteria, primarily within the "SUMMARY OF NONCLINICAL/BENCH STUDIES" and "SUMMARY OF CLINICAL INFORMATION" sections. Due to the diverse nature of these tests, a single comprehensive table is challenging. Instead, I've organized them by the type of test (Bench Studies and Clinical Studies) and extracted the relevant acceptance criteria and results.

    Bench Studies: Dosimetric Accuracy and Motion Tracking

    TestPurposeAcceptance CriteriaReported Device Performance
    Moving target with 3-D respiratory and 3-D non-respiratory motion; organs at risk (OARs) independent motionDemonstrate BgRT dose delivery accuracy to target(s) and measure dose to OARs with respiratory or non-respiratory continuous motion pattern; evaluate BgRT dose delivery accuracy to target and nearby normal tissues (OARs) for varying degrees of clinical motion patterns; compare with stereotactic body radiation therapy (SBRT) treatment delivery.Coverage margin loss: Maximum of (CMP-CMD*) in a given plane ≤ 3mm. Dosimetric coverage: Dose at all points measured on film within the CTV ≥ 97% of prescription dose, but ≤ 130% of maximum planned dose. OAR point dose: Calibrated ion chamber measurement ≤ expected plan dose. OAR maximum dose: Maximum dose measured on radiochromic film inserted into OAR structure < maximum plan dose."Both dosimetric coverage criteria as well as OAR dose criteria were met for all BgRT test cases. Under the same test circumstances, SBRT plans with an ITV approach for the target were not able to successfully meet the same criteria as the BgRT plans."
    Target or OAR with a single step shift - (positional uncertainty)Demonstrate BgRT dose delivery accuracy to target(s) and OARs in the case of a single step shift of the target; evaluate effectiveness of BgRT treatment providing conformal dose to the target while sparing nearby normal tissues (OARs); evaluate BgRT dose delivery accuracy; compare with SBRT treatment delivery.Same acceptance criteria as "Moving target with 3-D respiratory and 3-D non-respiratory motion.""Both dosimetric coverage criteria as well as OAR dose criteria were met for all BgRT test cases. Under the same test circumstances, SBRT plans with an ITV approach for the target were not able to successfully meet the same criteria as the BgRT plans."
    Test clinical variations of targets (stationary)Demonstrate BgRT dose delivery accuracy for various clinical situations: various target-to-background (T:B) contrast, different FDG biodistribution patterns between planning and delivery while keeping the same anatomic shape of the target, and homogeneous or heterogeneous backgrounds around the target.PET evaluation results: AC/NTS/bDVH. Bounded DVH: 95% of points in the calculated dose DVHs of all static structures must be within the bounds of the bDVH from the treatment plan. Gamma Index: 90% of comparison points (above 10% of prescription dose) within either 3% in dose or closer than 3mm in 3D space (DTA) to a point with the same dose. Overall Acceptance: PET Evaluation must pass AND either post-delivery DVH ≥ 95% points within bounds OR Gamma index > 90%."For all tests where the acceptance criteria required both the PET evaluation to pass and either the post-treatment dose evaluation against the plan and/or the Gamma index to pass met the dose accuracy criteria. For cases where there was a significant change in biodistribution in the target, low contrast between the target and background, or a change in the background signal, the PET evaluation did not pass, and the resulting dose accuracy showed mixed results."
    Tests evaluating special edge cases (stationary)Demonstrate robustness of the BgRT system under challenging situations: variations between planning and delivery such as changes in PET biodistribution patterns; PET-avid OAR in the BTZ prior to delivery; dose delivery accuracy to target in the presence of nearby PET-avid OAR; PET Evaluation step challenge with large changes between planning and delivery.Same acceptance criteria as "Test clinical variations of targets (stationary).""For cases where there was a significant change in biodistribution between planning and delivery, or a situation when an FDG-avid target moves into the BTZ, the PET evaluation did not pass."
    BgRT ON/OFF with Motion ON/OFFDemonstrate the benefit of BgRT in the same test setup by running with BgRT ON compared to BgRT OFF; evaluate BgRT dose delivery accuracy to target and measure dose to OAR with respiratory continuous motion pattern for the target; evaluate BgRT dose delivery accuracy to target and measure dose to OAR in the case of a single shift of the target.Coverage margin loss: Maximum of (CMP-CMD*) in a given plane ≤ 3mm. Dosimetric coverage: Dose at all points measured on film within the CTV ≥ 97% of prescription dose, but ≤ 130% of maximum planned dose. OAR maximum dose: Maximum dose measured on radiochromic film over all valid dose points < (103% of bDVH maximum dose from plan)."BgRT ON Motion ON - The tracked dose delivery was successfully provided where the CTV was fully covered (100%) in all cases. In cases for which the PET evaluation did not pass, some degradation to margin loss was observed. BgRT OFF Motion ON - In all cases dose accuracy failed the criteria. The CTV dose coverage was less than 100% and margin loss higher than 3.0mm. BgRT ON Motion OFF - All test cases met the dose accuracy criteria. BgRT OFF Motion OFF - All test cases met the dose accuracy criteria."
    Multi-target testingDetermine accuracy of delivery with cases of lower target to background contrast.Coverage margin loss: Maximum of (CMP-CMD*) in a given plane ≤ 3mm. Dosimetric coverage: Dose at all points measured on film within the CTV ≥ 97% of prescription dose, but ≤ 130% of maximum planned dose."For all 3-target, moving tests, including with reduced target:background for the second and third targets, all dosimetric criteria (CTV dose coverage and Margin Loss) were met."

    Clinical Study: Cohort I (RRFD and PET Imaging Performance)

    Endpoint #DescriptionAcceptance CriteriaReported Device Performance
    Primary Objective: RRFDFDG Dose: Enables Activity Concentration necessary for BgRT functioning (i.e., 5 kBq/ml or higher).All 6 evaluable cases met the threshold (> 5 kBq/ml), with values ranging from 7.46 to 21.05 kBq/ml.
    Secondary Objective: BgRT PET Imaging-only Session Performance
    1Percent of cases where there was an agreement between a site investigator (SI) and agreement standard (AS) for the BgRT PET imaging-only session localization decision (overall percent agreement).Not explicitly stated as a pass/fail criterion, but a high percentage indicates agreement.Overall % agreement = 83.3% (5/6 eval. cases). Positive % agreement = 83.3%, Negative % agreement = N/A.
    2Percent of cases where there was concordance of the positive "plan proceed" decision between the BgRT imaging-only session PET and a cleared, third-party diagnostic PET/CT (positive percent agreement).Not explicitly stated as a pass/fail criterion, but high concordance is desirable.Positive % agreement = 100% (6/6 eval. cases). Overall % agreement = 83.3%*, Negative % agreement = 0.
    3Percent of cases where RefleXion PET data could be used to generate an acceptable BgRT plan such that relevant dosimetric parameters for the target and the nearby normal anatomy were met based on Investigator assessment.Not explicitly stated as a pass/fail criterion, but a high percentage indicates acceptability.67% (4/6 evaluable cases).
    4Percent of cases where the intended dose distribution of the BgRT plan was achieved in a physical phantom, defined as meeting a standard gamma index for external beam radiotherapy quality assurance, i.e., whether 90% of pixels meet the 3 mm/3% deviation standard.90% of pixels meet 3 mm/3% deviation standard.67% (4/6 evaluable cases). "100% (4/4) for all cases where an acceptable plan was created."

    Clinical Study: Cohort II (BgRT Delivery Consistency and Workflow)

    Endpoint #DescriptionAcceptance CriteriaReported Device Performance
    Primary Objective: BgRT Dose Distribution Consistency
    Percent of radiotherapy fractions where the emulated BgRT dose distribution in silico was shown to be consistent with the approved BgRT treatment plan.95% of DVH Delivered points for the BTZ and OAR fell within bounded DVH of the approved BgRT plan.16/18 fractions (88.9%) met the 95% bDVH criterion. One lung case failed the first fraction (94.52%) and the last fraction "was not evaluable as the PET Evaluation did not pass." Another lung case resulted in a calculated volume reduction of -0.41%, indicating a slightly larger PTV than SBRT for that specific case. Overall, 88.9% success rate for this primary endpoint.
    Secondary Objective: Workflow and Deliverability
    1Percent of fractions where there was concordance between the physical and digital phantoms of emulated BgRT delivery derived from human participant PET emissions.90% of pixels meet 3 mm/3% deviation standard (gamma index).100% (18/18)
    2Percent of cases where there was agreement between a SI and the AS for the BgRT PET PreScan localization decision (overall percent agreement).Not explicitly stated as a pass/fail criterion, but a high percentage indicates agreement.Overall % agreement = 72.2% (13/18). Positive % agreement = 80%, Negative % agreement = 33%.
    3Percent of cases where there was concordance of a positive localization decision between the short-duration PET PreScan and a third-party diagnostic PET/CT scan (positive percent agreement).Not explicitly stated as a pass/fail criterion, but high concordance is desirable.Positive % agreement = 100% (7/7). Overall % agreement = 100%, Negative % agreement = N/A.
    5aWorkflow: Percent of PET imaging sessions at RRFD that met the AC threshold for BgRT (5 kBq/mL).5 kBq/mL.100% (18/18 fractions)
    5bWorkflow: Percent of PET imaging sessions which led to acceptable BgRT plans.Acceptable BgRT plans based on user-defined coverage/avoidance goals.100% (9/9 cases)
    5cWorkflow: Percent of approved BgRT plans that went on to pass physics quality assurance.90% of pixels meet 3 mm/3% deviation standard (gamma index).100% (9/9 cases)
    5dWorkflow: Percent of PET valuations on the day of fraction delivery that elicited a "Pass" signal."Pass" signal.94% (17/18 fractions)

    2. Sample Size Used for the Test Set and Data Provenance

    • Bench Studies: The sample sizes for the bench studies are not explicitly stated in terms of a numerical count of "test sets" or cases. Instead, they describe the methodology (e.g., "large anthropomorphic phantom," "ArcCHECK Model") and the types of configurations tested (e.g., respiratory/non-respiratory motion, single step shifts, various target-to-background contrasts, edge cases, multi-target). The nature of these tests suggests a comprehensive exploration of the device's performance under simulated conditions.

      • Provenance: In silico (simulated) and phantom studies. No specific country of origin is mentioned for the bench testing.
    • Clinical Studies:

      • Cohort I: 8 patients were enrolled, with 6 having evaluable images for the secondary endpoints.
      • Cohort II: 9 patients were enrolled (4 with bone tumors, 5 with lung tumors).
      • Provenance: Prospective clinical trial conducted at two sites in the U.S.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Bench Studies: Ground truth for bench studies was established through the controlled and measurable nature of the phantom setups and physical dosimetric measurements (film, ion chambers). There's no mention of human experts establishing ground truth in terms of image review or clinical assessment for the bench tests.

    • Clinical Studies:

      • Ground Truth for Planning and Treatment decisions:

        • Site Investigator (SI): Involved in the "localization decision" and "investigator assessment" for acceptable BgRT plans. Their specific qualifications are not detailed beyond being a "site investigator."
        • Agreement Standard (AS): Used in conjunction with the SI for localization decisions. The nature or composition of the AS is not explicitly defined (e.g., whether it's a consensus of experts, a reference standard from a central review).
        • Radiation Oncologist: Approved the detailed treatment plan and decided whether radiotherapy delivery with PET guidance was warranted based on metabolic characteristics (AC and NTS).
        • Medical Physicist or Dosimetrist: Developed the detailed treatment plan and conducted patient-specific quality assurance.
        • Clinical Expertise: Decisions on BgRT plan acceptability and patient positioning were made by clinical staff.
      • "Cleared, third-party diagnostic PET/CT": This served as a comparative ground truth for evaluating PET imaging performance, implying that its established diagnostic accuracy serves as a reference.

    4. Adjudication Method for the Test Set

    • Bench Studies: No specific "adjudication method" involving human experts is applicable, as the ground truth was based on physical measurements and controlled simulations.

    • Clinical Studies:

      • For endpoint 1 (Cohort I) and 2 (Cohort II) (localization decisions), agreement was assessed between a "Site Investigator (SI)" and an "Agreement Standard (AS)." The method for establishing the "Agreement Standard" (e.g., consensus, expert panel) is not specified. It appears to be a direct comparison against a predefined reference, rather than a multi-reader, multi-case adjudication process.
      • For endpoint 2 (Cohort I) and 3 (Cohort II) (concordance with diagnostic PET/CT), the comparison was against a "cleared, third-party diagnostic PET/CT," which serves as the established reference without explicit "adjudication" in the sense of multiple human readers resolving disagreements.
      • No formal 2+1 or 3+1 adjudication method involving multiple independent expert readers is described for the clinical study data in the provided text. Decisions and assessments appear to be made by the "Investigator" or clinical staff.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and what was the effect size of how much human readers improve with AI versus without AI assistance

    • No MRMC comparative effectiveness study involving human readers assisting with or being assisted by the RMRS (which incorporates AI/algorithmic guidance) is described in the provided text.
    • The clinical studies focus on the RMRS's performance against predefined criteria and existing diagnostic modalities, not on evaluating human reader improvement with AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • The "SUMMARY OF NONCLINICAL/BENCH STUDIES" section, particularly the detailed performance metrics for motion tracking, dosimetric coverage, and gamma index evaluations, represent the standalone algorithm performance in a controlled, simulated environment. These tests involved the device delivering radiation based on its algorithms and PET data, without human real-time intervention in the delivery process, and assessing the physical outcome.
    • In Cohort II, the "emulated BgRT dose distribution in silico" and the concordance between "physical and digital phantoms of emulated BgRT delivery" also reflect the standalone performance of the algorithm in generating and confirming deliverable fluence based on human patient PET emissions. While human input is involved in initial planning and approval, the core evaluation of dose delivery consistency and accuracy is algorithmic.

    7. The Type of Ground Truth Used

    • Bench Studies: Ground truth was established through physical measurements (e.g., film dosimetry, ion chamber measurements) in highly controlled phantom environments, along with the precise knowledge of programmed motion and simulated targets. This constitutes a manufactured or precisely controlled ground truth.
    • Clinical Studies:
      • Comparative Reference: For PET imaging performance, a "cleared, third-party diagnostic PET/CT" served as a comparative ground truth for localization decisions and "plan proceed" decisions.
      • Clinical Assessment/Investigator Opinion: For whether RefleXion PET data could be used to generate an "acceptable BgRT plan," the ground truth was based on Investigator assessment.
      • Dosimetric QA Standards: For the achievement of intended dose distribution and plan quality assurance, the ground truth was defined by established radiotherapeutic quality assurance standards (e.g., gamma index meeting 3 mm/3% deviation standard, DVH metrics).
      • In Silico Emulation: For the primary endpoint in Cohort II, the consistency of the emulated dose distribution was compared against the "approved BgRT plan." This is a form of design-specific ground truth derived from the planning system's output.

    8. The Sample Size for the Training Set

    • The provided text describes performance testing and clinical studies for validation. It does not mention a separate training set or its sample size for the development of any AI/algorithm components within the RMRS. The information focuses solely on the validation data. It's possible the device's algorithms were developed internally or on proprietary datasets not detailed in this submission summary.

    9. How the Ground Truth for the Training Set Was Established

    • Since no training set and its use in algorithm development are explicitly described, there is no information provided on how its ground truth was established.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1