Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K232033

    Validate with FDA (Live)

    Date Cleared
    2023-11-29

    (145 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K072404, K143306, K231023, K190823

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HYBRIDknife flex is intended for:
    · monopolar cutting and coagulation.
    · needle-free injection and tissue-selective hydrodissection including lifting mucosal lesions by injection into the submucosa (soft tissue).
    The HYBRIDknife flex is used in endoscopic interventions.

    Device Description

    The HYBRIDknife flex is a flexible monopolar probe that combines the technologies of hydrosurgery and electrosurgery in one instrument.
    Each function can be activated without the need to change instruments. The HYBRIDknife flex is a sterile, single use device which is used with endoscopes with a minimal working channel diameter of 2.8mm. The hydrosurgical function is intended to deliver a pressurized fluid for tissue-selective hydrodissection and needle-free injection whereas the electrosurgical function is intended for cutting and coagulation of tissue. By means of needle-free injection a fluid cushion is formed in the submucosa, which provides a mechanical and thermal protection layer during cutting and coagulation of the target tissue. The elevation thus reduces the risk of perforation. All HYBRIDknife flex probes have a length of 2.3 meters and an outer diameter (OD) of 2.6mm. The only difference between the variants is the electrode type and length. The HYBRIDknife flex is available with a "T-type" electrode and an "I-type" electrode whereas both electrode types are available as a long (i.e. 2mm length) and short (i.e. 1.5mm) version. The protrusion of the T-type electrode gives the user the possibility to hook and move tissue. The instruments are designed for operation with the hydrosurgical unit ERBEJET 2 (K072404; K143306 & K231023) in combination with an Erbe Electrosurgical unit of the "VIO" series (e.g. VIO 3 K190823). HYBRIDknife flex is connected to the units via respective cables/tubings. The settings or adjustment of application parameters is performed via the units. Activation of the instrument is done by using a footswitch.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device called HYBRIDknife® flex. This document is a regulatory submission to the FDA, demonstrating that the device is substantially equivalent to legally marketed predicate devices.

    Crucially, this document is NOT a study report for an AI/ML medical device. It describes a physical electrosurgical cutting and coagulation device. Therefore, the questions about AI/ML specific acceptance criteria, test sets, ground truth establishment, expert adjudication, MRMC studies, and standalone algorithm performance are not applicable to this document. The term "device" in the context of this document refers to the physical surgical instrument, not an AI model.

    The document discusses non-clinical performance testing of the device, which aims to verify its safety and performance based on engineering principles and established standards for electrosurgical equipment. This is distinct from the type of performance study conducted for AI/ML devices, which involves evaluating the performance of an algorithm against a ground truth.

    Here's how to address the request based on the provided text, while clarifying the limitations:


    Acceptance Criteria and Device Performance (Based on the document for a physical medical device)

    The document primarily focuses on demonstrating substantial equivalence to predicate devices rather than defining explicit "acceptance criteria" in the way one might for an AI/ML model's performance metrics (e.g., specific sensitivity/specificity thresholds). Instead, the acceptance is based on demonstrating that the new device does not raise new questions of safety or effectiveness and performs comparably to the predicate devices for its intended use.

    The "performance" is verified through various non-clinical tests to ensure the device meets design specifications and relevant industry standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    As this is not an AI/ML device, a table of acceptance criteria for algorithm performance metrics (like sensitivity, specificity, AUC) is not provided in the document. Instead, the document discusses conformity to standards and comparability to predicate devices. The "performance" here refers to the device's functional integrity and safety.

    Aspect of "Performance"Acceptance Criteria (Implied from testing)Reported Device Performance (Summary from text)
    Material SafetyBiocompatibility in compliance with ISO 10993."The different materials have no impact on safety as shown by biocompatibility testing in compliance with ISO 10993."
    SterilitySterile, single-use; SAL of 10-6; EO residual limits compliant with ISO 10993-7."Sterilization validation was performed in compliance with ISO 11135 and documentation was provided according to FDA Guidance... showing an SAL of 10-6. EO residual testing and limits are in compliance with ISO 10993-7."
    Packaging & Shelf-lifeCompliance with ISO 11607-1; successful accelerated aging."Packaging and shelf-life validation was performed in compliance with ISO 11607-1 and accelerated aged devices (ASTM F 1980)."
    Electrosurgical FunctionCutting and coagulation performance comparable to predicate."Tissue testing was performed... to validate cutting and coagulation performance... compared to the predicate device." "High Frequency (HF) Current with a maximum Electrical Capacity 4500 Vp (higher than predicate's 2500 Vp)."
    Waterjet FunctionEquivalent fluid cushion creation as predicate."As shown by respective tissue testing, the subject device HYBRIDknife flex creates equivalent fluid cushions as the primary predicate device." "Pressurized Sterile Normal Saline... with or without dye additives from Effect 1 to 60 which corresponds to a max. impact force between 0.04 – 0.063 N (14.5 to 870.2 psi)."
    Electrical Safety & EMCCompliance with IEC 60601-1, IEC 60601-2-2, IEC 60601-2-18, IEC 60601-1-2."EMC and Electrical safety of the subject device was tested in compliance with IEC 60601-1 Edition 3.2; IEC 60601-2-2 Edition 6.0; IEC 60601-2-18: Edition 3.0 and IEC 60601-1-2 Edition 4.0."
    Overall Performance & DesignPerforms as intended, meets design specifications, does not raise new safety/effectiveness questions."Functional testing and design controls... demonstrate that the differences do not raise any new issues of safety or effectiveness... The subject device has the same intended use, the same fundamental design, substantially equivalent performance characteristics, and the same energy source as the predicate devices."

    2. Sample size used for the test set and the data provenance
    Not applicable in the context of an AI/ML test set. The testing referenced are non-clinical (e.g., tissue testing, electrical testing, biocompatibility testing, sterilization validation) of a physical device. There isn't a "test set" of patient data in the AI/ML sense. Data provenance would refer to the characteristics of the materials or test conditions used in the lab testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    Not applicable. This pertains to AI/ML model validation using expert labels, which is not described here.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    Not applicable. This pertains to AI/ML model validation using expert labels, which is not described here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This pertains to AI/ML model validation. The device is a physical surgical instrument, not an AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not applicable. This pertains to AI/ML model validation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
    Not applicable in the AI/ML context. For this physical device, "ground truth" would be established by physical measurements, chemical analyses, biological assays (e.g., for biocompatibility), and direct observation of electrosurgical and hydrodissection effects on tissue models, as per standard engineering and medical device testing protocols. The "truth" is adherence to specifications and standards.

    8. The sample size for the training set
    Not applicable. This device is not an AI/ML model and does not have a "training set" in that sense.

    9. How the ground truth for the training set was established
    Not applicable. This device is not an AI/ML model and does not have a "training set" in that sense.


    Summary of the Study Proving Device Meets "Acceptance Criteria" (Non-Clinical Performance Testing)

    The "study" in this context refers to a series of non-clinical performance tests designed to demonstrate the safety and effectiveness of the HYBRIDknife® flex and its substantial equivalence to predicate devices. These tests were performed rather than a clinical trial or AI/ML validation study.

    The primary goal was to show that despite some technological differences (e.g., higher dielectric strength, slightly different dimensions, fixed electrode states vs. adjustable), the device performs its intended functions comparably to existing, legally marketed devices and adheres to relevant safety and performance standards.

    The non-clinical performance testing included:

    • Functional Testing and Design Controls: To verify overall safety and performance, ensuring the device performs as intended and meets design specifications (in compliance with 21 CFR 820.30).
    • Tissue Testing: To validate the cutting and coagulation performance and the waterjet function. This was done by comparing the HYBRIDknife® flex to the predicate device, specifically looking at the creation of equivalent fluid cushions. This would involve in-vitro or ex-vivo tissue models.
    • EMC and Electrical Safety Testing: To ensure compliance with international standards (IEC 60601-1, IEC 60601-2-2, IEC 60601-2-18, IEC 60601-1-2). This verifies that the device operates safely electrically and doesn't interfere with other medical equipment.
    • Biocompatibility Testing: To ensure the new materials used in the device are safe for contact with the body (in compliance with ISO 10993-1).
    • Sterilization Validation: To confirm the device can be consistently sterilized to a specific sterility assurance level (SAL of 10-6) and that ethylene oxide residuals are within safe limits (in compliance with ISO 11135 and ISO 10993-7).
    • Packaging and Shelf-life Validation: To ensure the device remains sterile and functional over its intended shelf-life, including accelerated aging tests (in compliance with ISO 11607-1 and ASTM F 1980).

    Data Provenance and "Test Set" Details (for a physical device):

    The data would originate from laboratory and bench testing conditions. This is typically controlled and prospective, conducted specifically to gather data for regulatory submission. There isn't a "country of origin of the data" in the sense of patient data, but rather the location where the testing labs are.

    In conclusion, the provided document is a regulatory submission for a physical medical device, not an AI/ML model. Therefore, many of the specific questions about AI/ML acceptance criteria and study design are not applicable. The "acceptance criteria" for this device are demonstrated through adherence to design specifications, relevant industry standards, and demonstrated substantial equivalence to predicate devices via rigorous non-clinical testing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230427

    Validate with FDA (Live)

    Date Cleared
    2023-07-11

    (144 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    12 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K191601, K181389, K083452, K190823, K143161

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CUSA® Clarity Ultrasonic Surgical Aspirator System is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft and hard (e.g. bone) tissue is desirable.

    The CUSA Clarity Ultrasonic Surgical Aspirator is indicated for use in:
    Plastic and Reconstructive surgery, Orthopedic Surgery, Gynecological Surgery and Thoracic Surgery and the following specific uses:
    Neurosurgery - including removal of primary and secondary malignant and benign brain and spinal tumors, including but not limited to meningiomas and gliomas
    Gastrointestinal and Affiliated Organ Surgery - including removal of benign or malignant tumors or other unwanted tissue, including hepatic parenchyma, in open or laparoscopic procedures, hepatic resection, tumor resection, lobectomy or trisegmentectomy, or removal of tissue during liver allotransplantation and donor hepatectomy
    Urological surgery- including removal of renal parenchyma during nephrectomy or partial nephrectomy
    General Surgery - including removal of benign or malignant tumors or other unwanted soft or hard tissue in open or minimally invasive general surgical procedures
    Laparoscopic Surgery - including removal of hepatic parenchyma in laparoscopic hepatic resection, lobectomy or trisegmentectomy, in laparoscopic donor hepatectomy or laparoscopic cholecystectomy or laparoscopic pancreatic jejunostomy, or pancreatectomy, or laparoscopic appendectomy, laparoscopic colon resection or laparoscopic partial gastrectomy

    Device Description

    The device within the scope of this premarket notification is the optional CUSA® Electrosurgery Module (CEM) accessory that is intended to be used with the 23 kHz components of the CUSA® Clarity Ultrasonic Surgical Aspirator System.

    The purpose of this submission is to modify the CEM nosecone accessory currently offered with CUSA Clarity to allow for connection with additional electrosurgical generators, to continue to provide electrosurgical capabilities to the user. The additional electrosurgical generators that the modified CEM Nosecone may be used with include the Medtronic FT10 (K191601), Medtronic FX8 (K181389), Erbe VIO 300D (K083452), and Erbe VIO 3 (K190823). Compatibility with the Medtronic Force FX (K143161) will be maintained as well.

    The CUSA Clarity 23kHz Expanded CEM Nosecone has the same intended use and technological characteristics as the predicate CUSA Clarity 23 kHz CEM Nosecone (K190180). The subject CEM nosecone will continue to allow the surgeon to apply immediate electrosurgical coagulation to bleeding tissue at the surgical site, with the same handpiece assembly that is removing unwanted tissue.

    AI/ML Overview

    The provided text describes the CUSA Clarity Ultrasonic Surgical Aspirator System, specifically focusing on a modified CEM nosecone accessory for the 23 kHz components. This submission is a 510(k) premarket notification claiming substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present acceptance criteria in a quantitative table format with corresponding device performance values. Instead, it describes various non-clinical tests undertaken to ensure the safety and efficacy of the device and its substantial equivalence to the predicate.

    The reported device performance is broadly stated as:

    • "Testing was determined successful and supports the conclusion that all product specifications and design inputs have been met."
    • "The results of the non-clinical testing indicate that the intended use of the device, fundamental scientific technology, and performance of the CUSA Clarity 23 kHz Expanded CEM Nosecone is substantially equivalent to the predicate device."

    Therefore, a table of quantitative acceptance criteria and specific reported device performance cannot be generated from the provided text. The information indicates that all tests were passed and specifications met, implying the device performed within acceptable limits.

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes non-clinical testing but does not specify a "test set" in terms of subject or patient data. The tests are focused on device characteristics rather than clinical outcomes with a patient population.

    • Sample size for test set: Not applicable in the context of device performance testing described. The tests are on the device itself (e.g., handpiece life, functionality, mechanical properties).
    • Data provenance: Not applicable in the context of patient data. The provenance is from internal testing conducted by Integra LifeSciences Corporation. There is no mention of country of origin for data related to clinical or patient studies, as none were performed. The tests are prospective as they were conducted as part of the submission process.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of experts: Not applicable. The "ground truth" for the non-clinical tests is established by engineering specifications, recognized standards (FDA guidance documents, ISO standards for biocompatibility, EMC, and electrical safety), and comparison to the predicate device's established performance.
    • Qualifications of experts: Not specified as a separate set of experts for ground truth. However, the development and testing would have been overseen by Integra LifeSciences Corporation's engineering and regulatory teams.

    4. Adjudication Method for the Test Set

    • Adjudication method: Not applicable. The testing described is objective device performance (e.g., sterilization, biocompatibility, electrical safety, mechanical, thermal effects). Success or failure is determined by meeting predefined engineering specifications and regulatory standards, not by an adjudication process between human experts on a specific outcome.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • MRMC study: No. The document explicitly states: "No clinical studies were performed or required as all conducted performance tests appropriately support a determination of substantial equivalence compared with the predicate device."
    • Effect size of human readers with/without AI assistance: Not applicable, as no MRMC study or study involving human readers with or without AI assistance was conducted. The device is a surgical aspirator system, not an AI diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Standalone study: Not applicable. The CUSA Clarity Ultrasonic Surgical Aspirator System is a physical medical device, not an algorithm. Therefore, "standalone" performance in the context of an algorithm is not relevant.

    7. The Type of Ground Truth Used

    The "ground truth" for the device's acceptable performance is based on:

    • Established engineering specifications and design inputs: Ensuring the device functions as intended.
    • Compliance with FDA guidance documents and recognized standards: Such as those for sterilization, biocompatibility, EMC, electrical safety, thermal effects, and capacitive coupling.
    • Substantial equivalence to the legally marketed predicate device (CUSA® Clarity Ultrasonic Surgical Aspirator System K190180): This implies that the predicate's established safety and effectiveness profile serves as a benchmark for the new component.

    8. The Sample Size for the Training Set

    • Sample size for training set: Not applicable. This device is not an AI/ML algorithm that requires a training set. The performance is assessed through traditional engineering and regulatory compliance testing.

    9. How the Ground Truth for the Training Set Was Established

    • How ground truth for training set was established: Not applicable, as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Erbe APC 3 with accessories is intended to deliver argon gas for argon plasma coagulation and ablation of tissue as well as argon-assisted cutting of tissue when used in conjunction with a compatible Erbe VIO Electrosurgical Generator (ESU) and applicators or probes.

    Device Description

    The Erbe Argon Plasma Coagulation Unit Model APC 3 is an APC unit for use in conjunction with an Erbe Electrosurgical Unit (ESU) compatible Model (currently Model VIO® 3). The APC 3 (as well as VIO 3) can be mounted/secured to a cart or on a ceiling mount using fastening sets, i.e. connecting cables and hardware. The APC 3 is connected to an argon gas source via a pressure reducer and connected to the VIO 3 which powers and controls the APC unit. The ESU further provides a touch-screen monitor (i.e., interactive display) with an onscreen tutorial, settings and operational information as well as safety features (i.e., audio and visual error system). When using the APC 3 together with the VIO 3, the user is offered various APC Modes as well as argon-supported Modes which can be adjusted by the physician via effect levels in order to achieve the desired tissue effect. The ESU VIO 3 provides the high-frequency (HF) current required for thermal coagulation. For argon plasma coagulation and ablation, the HF current is applied non-contact to the tissue to be coagulated through ionized and hence electrically conductive argon gas (argon plasma). The argon gas is ionized in the high-frequency electrical field between the electrode of the applicator and the tissue. For argon-assisted cutting, the ESU provides the regular cutting/coagulation Mode while the argon gas provided by the APC 3 circulates the active electrode. This application is with contact to tissue.

    Applicators or probes connected to the APC 3 for coagulating tissue via argon plasma and/or for cutting tissue supported by argon gas may be activated via footswitches connected to the ESU or by use of applicators that have an activation finger switch.

    The APC unit and its accessories (i.e. pressure reducer and fastening sets) are supplied non-sterile and are reusable. Cleaning and disinfections are provided in the APC user manual or notes on use of the involved accessory. The compatible applicators and probes are provided sterile as well as are single-use.

    AI/ML Overview

    This document describes the Erbe Argon Plasma Coagulation Unit APC 3 with Accessories, and its substantial equivalence to a predicate device. It primarily focuses on the device's technological characteristics and safety standards rather than clinical performance or AI-driven improvements in human reading. Therefore, much of the requested information regarding AI acceptance criteria, clinical study specifics (sample size, expert involvement, MRMC studies, ground truth), and training set details for AI cannot be extracted from this document because it is about an electrosurgical device, not an AI/ML-driven diagnostic device.

    Here's a breakdown of what can be extracted, and where limitations exist based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document doesn't provide specific quantitative acceptance criteria or detailed performance metrics in a table format for clinical efficacy. Instead, it focuses on general safety, performance, and substantial equivalence to a predicate device based on technical specifications and adherence to recognized standards.

    Acceptance Criteria (Implied from the document):

    • Safety: Adherence to electrical safety standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2).
    • Performance: Ability to deliver argon gas for argon plasma coagulation, ablation, and argon-assisted cutting as intended. This is assessed through "Output/Mode Comparison," "Tissue Testing," and "Capacitive Coupling" as listed in the "Performance Evaluations/Testing" section.
    • Usability: Conformance to usability engineering standards (e.g., IEC 62366, IEC 60601-1-6).
    • Compatibility: Operation with compatible Erbe VIO Electrosurgical Generators (ESU) and specific applicators/probes.
    • Substantial Equivalence: Demonstrating that the new device has "essentially the same principles of operation, technological characteristics, as well as performance characteristics as the predicate APC unit."

    Reported Device Performance:
    The document states that "Validation and verification activities in design control that includes testing/certification to designated standards (including electrical safety standards) and performance testing of the proposed devices has demonstrated substantial equivalence to the predicate." However, specific numerical performance data (e.g., coagulation efficacy rates, ablation depth measurements, cutting speeds) are not provided. The "Output/Mode Comparison" and "Tissue Testing" are mentioned as performance evaluations, but their results are summarized qualitatively as proving substantial equivalence.

    2. Sample size used for the test set and the data provenance

    This document describes a medical device (an electrosurgical unit), not an AI/ML diagnostic algorithm tested on a dataset of patient data. Therefore, the concepts of "test set sample size" and "data provenance" in the context of clinical/imaging data are not applicable here. The testing involves engineering and performance validation of the device itself, rather than testing on a retrospective or prospective clinical dataset.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is not an AI/ML diagnostic device requiring expert annotation of medical images or data to establish ground truth for a test set. The validation is technical and engineering-focused.

    4. Adjudication method for the test set

    Not applicable. As this is not a diagnostic device involving human interpretation of data, there is no need for adjudication of readings.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is not an AI-assisted diagnostic tool, so an MRMC study and
    assessment of human reader improvement with AI assistance is irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm. It is a hardware medical device.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance evaluation is based on engineering standards, physical measurements, and comparison to the predicate device's established performance characteristics. It relies on:

    • Compliance with recognized consensus standards: e.g., electrical safety, electromagnetic compatibility, usability (IEC, ISO standards listed).
    • Physical (in-vitro/bench) testing: "Output/Mode Comparison," "Tissue Testing," "Capacitive Coupling." These tests would compare the device's output and tissue effects to specifications and the predicate.
    • Functional verification: Ensuring the device operates as intended when connected to compatible components.

    It does not involve "expert consensus," "pathology," or "outcomes data" in the sense of clinical diagnostic accuracy.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML system that requires a "training set" of data.

    9. How the ground truth for the training set was established

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1