510(k) Data Aggregation

    K Number
    K250373

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-04-09

    (58 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K190488, K223430, K212520

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When used as an interbody fusion device, the PROLIFT® Expandable Spacer System is intended for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). It is to be used in patients who have had at least six months of non-operative treatment. Patients with previous non-fusion spinal surgery at involved level(s) may be treated with the device. This device is intended to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone and a supplemental internal spinal fixation system (e.g., pedicle screw or anterolateral plating system) that is cleared for use in the lumbosacral spine.

    Device Description

    The PROLIFT® Expandable Spacer System is available in a range of sizes and footprints and can expand to the desired height (8mm to 16mm) to suit the individual pathology and anatomical conditions of the patient. It is fabricated and manufactured from titanium alloy (Ti-6Al-4V ELI) as described by ASTM F136. The implant allows packing of autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone to help promote fusion. The superior and inferior surfaces have teeth to assist in the interface with the vertebral bodies to prevent rotation and/or migration.

    All implants are provided sterile and intended for SINGLE USE ONLY and should not be reused under any circumstances. Do not use any of the PROLIFT® Expandable Spacer System components with components from any other system or manufacturer. The PROLIFT® Expandable Spacer System components should never be reused under any circumstances.

    AI/ML Overview

    The provided FDA 510(k) clearance letter for the ProLift Expandable Spacer System does not describe the specific acceptance criteria and study proving the device meets those criteria in the level of detail requested. This document primarily focuses on the regulatory aspects of substantial equivalence for a medical device (an intervertebral body fusion device), not a software or AI-based diagnostic tool, which typically entails the type of performance evaluation outlined in your request.

    Therefore, many of the requested details about acceptance criteria, specific performance metrics, sample sizes for test/training sets, ground truth establishment, expert qualifications, and MRMC studies are not applicable or not present in this type of regulatory submission for a physical implant.

    However, I can extract the information that is present and explain why other information is not available in this document.


    Analysis of the Provided Document for Acceptance Criteria and Study Details

    The provided document is an FDA 510(k) clearance letter for a medical device, specifically an intervertebral body fusion device (ProLift Expandable Spacer System). This type of clearance is based on demonstrating "substantial equivalence" to a legally marketed predicate device, rather than a de novo approval that often requires extensive clinical performance data with specific acceptance criteria as you've requested for an AI/software device.

    For a physical device like this, performance data typically refers to mechanical testing, biocompatibility, and sterilization validation, not diagnostic accuracy metrics.

    Here's what can be gathered from the document, and what cannot:

    1. Table of Acceptance Criteria and Reported Device Performance

    This document does not contain a table of acceptance criteria and reported device performance in the context of diagnostic accuracy (e.g., sensitivity, specificity, AUC). For this physical device, "performance" refers to mechanical properties and material characteristics.

    What is present regarding "performance":

    • Performance Data Section (Page 6): "Through the Comparison Analysis it is determined that the mechanical performance has not change as the worst case is still the same and therefore is substantially equivalent to the Life Spine ProLift Expandable Spacer System (K190488) and ProLift Micro Expandable Spacer System (K223430 & K212520)."

    This indicates that the key "acceptance criterion" related to performance for this 510(k) submission was demonstrating that the mechanical performance of the new device is substantially equivalent to its predicate devices, meaning it performs similarly in relevant mechanical tests (e.g., strength, durability under load). No specific numerical acceptance values for mechanical performance are provided in this regulatory letter itself, as they would typically be detailed in the full 510(k) submission document.

    2. Sample sizes used for the test set and the data provenance

    • Not applicable/Not provided in this document. For a physical mechanical device, "test set" in the context of AI/software performance is not relevant. Mechanical testing would involve a certain number of device samples, but this is not typically referred to as a "test set" in the same way. The document does not specify the number of devices or test conditions used for mechanical performance comparison. Data provenance (country of origin, retrospective/prospective) is also not applicable here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable/Not provided in this document. "Ground truth" established by experts (like radiologists for imaging devices) is for diagnostic devices where human interpretation is being evaluated or augmented. For an intervertebral body fusion device, the "ground truth" of its mechanical properties would be established through engineering tests following industry standards, not expert consensus on interpretations.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable/Not provided in this document. Adjudication methods are relevant for resolving discrepancies in human expert interpretations, which is not the type of evaluation described for this device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable/Not provided in this document. MRMC studies are specific to evaluating diagnostic accuracy, especially for AI-assisted image interpretation. This device is a physical implant, not a diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable/Not provided in this document. This applies to algorithms. The device is a physical implant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For this device, the "ground truth" for its performance would implicitly be based on engineering standards, material specifications (e.g., ASTM F136 for Ti-6Al-4V ELI), and biomechanical testing standards to ensure it meets safety and functional requirements equivalent to established devices. The document explicitly mentions "mechanical performance" and "material."

    8. The sample size for the training set

    • Not applicable/Not provided in this document. "Training set" is a concept for machine learning models.

    9. How the ground truth for the training set was established

    • Not applicable/Not provided in this document.

    In summary, the provided FDA 510(k) clearance letter for the ProLift Expandable Spacer System is for a physical medical implant, not an AI or software-based diagnostic device. Therefore, the detailed performance study information typically associated with AI/software (acceptance criteria tables, expert adjudication, MRMC studies, ground truth establishment for training/test sets) is not present and generally not part of a 510(k) submission for this class of device.

    The "study" that proves the device meets the (unspecified but implied mechanical) acceptance criteria is the "Comparison Analysis" mentioned on Page 6, which determined that the mechanical performance of the new device is substantially equivalent to the predicate devices. The specifics of this comparison (e.g., number of tests, exact parameters) would be in the full 510(k) submission, not typically in the public clearance letter itself.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243668

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-01-23

    (57 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ProLift Pivot Expandable System is intended for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiquous levels (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). It is to be used in patients who have had at least six months of non-operative treatment. Patients with previous non-fusion spinal surgery at involved level(s) may be treated with the device. This device is intended to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone and a supplemental internal spinal fixation system (e.g., pedicle screw or anterolateral plating system) that is cleared for use in the lumbosacral spine.

    Device Description

    The ProLift® Pivot Expandable System is available in a range of sizes and footprints and can expand to the desired height (8mm to 15mm) to suit the individual pathology and anatomical conditions of the patient. It is fabricated and manufactured from titanium alloy (Ti-6AI-4V ELI) and Cobalt Chrome (CO- 26Cr-6MO) as described by ASTM F136and ASTM F1537. The implant allows packing of autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone to help promote fusion. The superior and inferior surfaces have teeth to assist in the interface with the vertebral bodies to prevent rotation and/or migration.

    The ProLift® Pivot Expandable Spacer System is available in a range of sizes and footprints and can expand to the desired height (8mm to 15mm) to suit the individual pathology and anatomical conditions of the patient.

    • The ProLift implants are fabricated and manufactured from titanium alloy (Ti-6Al-. 4V ELI) as described by ASTM F136.
      . The ProLift Pivot-U implants are fabricated and manufactured from titanium alloy (Ti-6Al-4V ELI) and Cobalt Chrome (CO-26Cr-6MO) as described by ASTM F136 & ASTM F1537.
      The implant allows packing of autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone to help promote fusion. The superior and inferior surfaces have teeth to assist in the interface with the vertebral bodies to prevent rotation and/or migration.
      All ProLift Pivot implants are provided sterile and intended for SINGLE USE ONLY and should not be reused under any circumstances. Do not use any of the ProLift® Pivot Expandable Spacer System components with components from any other system or manufacturer.
    AI/ML Overview

    This document pertains to the clearance of a medical device, the ProLift Pivot Expandable Spacer System, not an AI/ML powered device. As such, the requested information regarding acceptance criteria, study details, expert involvement, and ground truth for AI models is not applicable to this submission.

    The document discusses the substantial equivalence of the ProLift Pivot Expandable Spacer System to predicate devices based on design, materials, indications for use, sizing, and mechanical performance.

    Here's a breakdown of the relevant information provided:

    1. A table of acceptance criteria and the reported device performance:
      The document does not present a formal table of acceptance criteria with reported performance values in the context of an AI/ML study. Instead, it states that "Performance Data: Static compression shear, Static Axial Compression, Dynamic Compression Shear and Dynamic Axial Compression according to ASTM F2077 was presented to demonstrate the substantial equivalency of the Life Spine ProLift Expandable Spacer System (K190488)." This implies that the acceptance criteria are adherence to the mechanical testing standards outlined in ASTM F2077, and the device performance met these standards to demonstrate substantial equivalence to the predicate device (K190488).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable, as this is a medical device clearance based on mechanical testing, not an AI/ML study involving a test dataset of patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable. "Ground truth" in the context of this device refers to the mechanical properties and performance standards defined by ASTM F2077 and assessed through engineering tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI-assisted diagnostic or therapeutic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): The ground truth for this device's performance is established by the mechanical testing standards outlined in ASTM F2077. This standard defines the methodologies and expected performance for intervertebral body fusion devices under various loading conditions (static compression shear, static axial compression, dynamic compression shear, dynamic axial compression).

    8. The sample size for the training set: Not applicable, as this is not an AI/ML device.

    9. How the ground truth for the training set was established: Not applicable, as this is not an AI/ML device.

    In summary, the document describes the clearance of a physical medical implant (ProLift Pivot Expandable Spacer System) based on its substantial equivalence to previously cleared devices, supported by mechanical performance data against recognized ASTM standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242826

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-10-08

    (20 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K203361, K190488, K153627/K171633

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When used as an interbody fusion device, the PROLIFT® Wedge System is intended for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). It is to be used in patients who have had at least six months of non-operative treatment. Patients with previous non-fusion spinal surgery at involved level(s) may be treated with the device is intended to be used with autograft bone graft and/or allogeners bone graft composed of cancellous and/or corticocancellous bone and a supplemental internal spinal fixation system (e.g., pedicle screw or anterolateral plating system) that is cleared for use in the lumbosacral spine.

    Device Description

    The PROLIFT® Wedge Expandable Spacer System is available in a range of sizes and footprints and can expand to the desired lordosis (0-15°) to suit the individual pathology and anatomical conditions of the patient. It is fabricated and manufactured from titanium alloy (Ti-6AI-4V ELI) as described by ASTM F136. The implant allows packing of autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone to help promote fusion. The superior and inferior surfaces have teeth to assist in the interface with the vertebral bodies to prevent rotation and/or migration.

    All implants are provided sterile and intended for SINGLE USE ONLY and should not be reused under any circumstances. Do not use any of the PROLIFT® Wedge Expandable Spacer System components with components from any other system or manufacturer. The PROLIFT® Wedge Expandable Spacer System components should never be reused under any circumstances.

    AI/ML Overview

    The provided text describes a medical device called the "ProLift Wedge Expandable Spacer System" and its 510(k) submission to the FDA. This document is a regulatory clearance letter and summary, not a study report that would typically include detailed acceptance criteria and performance data for an AI/ML medical device.

    Therefore, the information required to populate the table and answer your specific questions about acceptance criteria, study design, sample sizes, ground truth establishment, expert qualifications, and MRMC studies for an AI/ML algorithm is not present in the provided text.

    The document states:

    • "Performance Data: Static compression, dynamic compression, static shear and dynamic shear testing according to ASTM F2077 & load induced subsidence and static expulsion according to F2267, was presented to demonstrate the substantial equivalency of the Life Spine ProLift Wedge Expandable Spacer System (K203361)." This indicates mechanical testing was performed for a physical intervertebral body fusion device, not a software algorithm.

    Based on the information provided, I cannot answer the questions as they pertain to the evaluation of an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223430

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-05-04

    (171 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K212520, K190488, K153400, K113447

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When used as an interbody fusion device, the ProLift® Micro Expandable System is intended for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radios. DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). It is to be used in patients who have had at least six months of non-operative treatment. Patients with previous non-fusion spinal surgery at involved level(s) may be treated with the device is intended to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone and a supplemental internal spinal fixation system (e.g., pedicle screw or anterolateral plating system) that is cleared for use in the lumbosacral spine.

    Device Description

    The ProLift® Micro Expandable System is available in a range of sizes and footprints and can expand to the desired height (8mm to 16mm) to suit the individual pathology and anatomical conditions of the patient. It is fabricated and manufactured from titanium alloy (Ti-6Al-4V ELI) as described by ASTM F136. The implant allows packing of autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone to help promote fusion. The superior and inferior surfaces have teeth to assist in the interface with the vertebral bodies to prevent rotation and/or migration.

    All implants are provided sterile and intended for SINGLE USE ONLY and should not be reused under any circumstances. Do not use any of the ProLift® Micro Expandable System components with components from any other system or manufacturer. The ProLift® Micro Expandable System components should never be reused under any circumstances.

    AI/ML Overview

    The provided text describes a medical device, the ProLift® Micro Expandable Spacer System, and its clearance through a 510(k) premarket notification. This type of submission relies on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting new clinical performance data to establish specific acceptance criteria for device performance in a clinical setting.

    Therefore, the document does not contain information typically found in a study that proves a device meets acceptance criteria related to AI/algorithm performance, human reader improvement, or detailed clinical efficacy. It focuses on the mechanical and material equivalence of the new device to existing ones.

    Based on the provided text, I can extract information relevant to the device's technical specifications and the basis for its substantial equivalence, but not the specific clinical performance criteria, study design, or AI-related metrics requested in the prompt.

    Here's an analysis of the information not present in the provided text, related to your specific questions:

    • Acceptance Criteria and Reported Device Performance (Table): The document does not describe clinical acceptance criteria (e.g., sensitivity, specificity, accuracy) or report specific clinical performance metrics. It focuses on mechanical testing for substantial equivalence.
    • Sample Size for Test Set and Data Provenance: No clinical test set or patient data is mentioned as part of a performance study.
    • Number of Experts and Qualifications: No experts were used to establish ground truth for a clinical test set because no such study is described.
    • Adjudication Method: Not applicable as no clinical test set with ground truth is discussed.
    • Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: No such study is mentioned. The document is about a physical medical implant, not an AI or diagnostic tool that would directly assist human readers.
    • Standalone Performance (Algorithm Only): Not applicable, as this is a physical implant, not an algorithm.
    • Type of Ground Truth Used: Not applicable for a performance study in the context of your questions. The "ground truth" for this device relates to its mechanical integrity and biocompatibility, which are verified through material specifications and mechanical testing rather than clinical outcomes established by experts or pathology.
    • Sample Size for Training Set & How Ground Truth for Training Set was Established: Not applicable as this is not an AI/machine learning device.

    Information that is extractable from the document, though not directly answering your specific questions about AI/clinical performance studies:

    The document describes the mechanical performance data used to demonstrate substantial equivalence, which is a different kind of "acceptance criterion" based on engineering standards:

    • Acceptance Criteria/Performance (Mechanical):
      • Type of Testing: Static Axial Compression, Dynamic Axial Compression, Static Shear, and Dynamic Shear testing.
      • Standard: According to ASTM F2077.
      • Purpose: To demonstrate substantial equivalency to the predicate device (Life Spine ProLift Micro Expandable System K212520).
      • Reported Performance: The document states that this testing "was presented to demonstrate the substantial equivalency," implying the device met the requirements of the standard consistent with the predicate. Specific numerical results are not provided in this summary.

    In summary, the provided FDA 510(k) document is for a physical medical device (an expandable spinal spacer) and therefore does not contain the type of information requested about AI/algorithm performance studies, human-in-the-loop studies, or clinical ground truth establishment.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201721

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-09-04

    (73 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K190488

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When used as an interbody fusion device, the TRULIFT® Expandable Spacer System is intended for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). It is to be used in patients who have had at least six months of non-operative treatment. Patients with previous non-fusion spinal surgery at involved level(s) may be treated with the device. This device is intended to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone and a supplemental internal spinal fixation system (e.g., pedicle screw or anterolateral plating system) that is cleared for use in the lumbosacral spine.

    Device Description

    The TRULIFT® Expandable Spacer System is available in a range of sizes and footprints and can expand to the desired height (8mm to 16mm) to suit the individual pathology and anatomical conditions of the patient. It is fabricated and manufactured from titanium alloy (Ti-6AI-4V ELI) as described by ASTM F136. The implant allows packing of autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone to help promote fusion. The superior and inferior surfaces have teeth to assist in the interface with the vertebral bodies to prevent rotation and/or migration.

    All implants are provided sterile and intended for SINGLE USE ONLY and should not be reused under any circumstances. Do not use any of the TRULIFT® Expandable Spacer System components with components from any other system or manufacturer. The TRULIFT® Expandable Spacer System components should never be reused under any circumstances.

    AI/ML Overview

    I apologize, but the provided text from the FDA 510(k) clearance letter for the "TruLift Expandable Spacer System" deals with a spinal implant (intervertebral body fusion device), not an AI/ML-enabled medical device for which the requested information regarding acceptance criteria, study findings, sample sizes, expert ground truth, and human-in-the-loop performance would be relevant.

    The document discusses the mechanical and material equivalence of the TruLift system to predicate spinal implants, based on ASTM performance standards for devices of its type (e.g., static axial compression, dynamic compressive shear). It does not involve any form of AI, image analysis, or diagnostic/prognostic capabilities that would require the kind of clinical validation study you've described.

    Therefore, I cannot extract the information you've requested about acceptance criteria and a study proving an AI device meets them, because this document pertains to a traditional, mechanical Class II medical device.