510(k) Data Aggregation

    K Number
    K223651

    Validate with FDA (Live)

    Date Cleared
    2023-05-26

    (171 days)

    Product Code
    Regulation Number
    870.2700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K190205, K180881

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cerebral Adaptive Index (CAI) Algorithm is an informational index to help assess the level of coherence or lack thereof between Mean Arterial Pressure (MAP) and the Absolute Levels of Blood Oxygenation (StO2) in patient's cerebral tissue. MAP is acquired by the HemoSphere Pressure Cable and StO2 is acquired by the ForeSight Oximeter Cable.

    CAI is intended for use in patients over 18 years of age receiving advanced hemodynamic monitoring.

    CAI is not indicated to be used for treatment of any disease or condition and no therapeutic decisions should be made based solely on the Cerebral Adaptive Index (CAI) Algorithm.

    Device Description

    Cerebral Adaptive Index (CAI) Algorithm is a derived parameter that quantifies the dynamic relationship between two existing hemodynamic parameters, Mean Arterial Pressure (MAP) and the Absolute Levels of Blood Oxygenation Saturation (StO2) in the cerebral tissue. CAI is intended to show the level of coherence between MAP and cerebral StO2. The output will be represented as an index value and a trended graph.

    MAP is acquired from the HemoSphere Pressure Cable (initially cleared in K180881 on November 16, 2018). StO2 used for computing CAI is acquired from the ForeSight Oximeter Cable (cleared in K201446 on October 1, 2020).

    The CAI parameter can enhance clinician's understanding of the underlying hemodynamic changes behind cerebral desaturation events. It helps the clinician recognize/ identify possible causes of, for example, decrease in StO2 and clinical events related to StO2 decrease (e.g., hypotension as opposed to inadequate oxygen content).

    CAI will be continuously displayed at 20-second rate. The parameter will not have any alarm ranges and will only be represented as a number with a range between 0 to 100. A high CAI value (CAI ≥45) means that MAP and StO2 have a greater coherence and informs the clinician that alterations in MAP may result in concomitant changes in cerebral oxygen saturation Whereas a low CAI value (CAI < 45) means there is lesser coherence between the two parameters, and therefore alterations in MAP may not result in concomitant changes in cerebral oxygen saturation.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance Study for Cerebral Adaptive Index (CAI) Algorithm

    The Cerebral Adaptive Index (CAI) Algorithm is an informational index designed to assess the coherence between Mean Arterial Pressure (MAP) and Absolute Levels of Blood Oxygenation (StO2) in cerebral tissue.

    1. Acceptance Criteria and Reported Device Performance

    The performance goals for the CAI algorithm were established using a CAI threshold of 45.

    Acceptance CriteriaReported Device Performance at CAI Threshold of 45
    Sensitivity ≥ 80%0.84 [0.78, 0.88] (84%)
    Specificity ≥ 80%0.80 [0.76, 0.85] (80%)
    ROC AUC (summarizes overall performance)0.88 [0.85, 0.90]

    Conclusion: The device met all pre-defined acceptance criteria for sensitivity, specificity, and ROC AUC at the specified CAI threshold of 45.

    2. Sample Size and Data Provenance for Test Set

    • Sample Size: 145 subjects aged 18 or older.
    • Data Provenance: Retrospectively obtained from four different clinical sites within the US:
      • Northwestern University, Chicago
      • UC Davis, Sacramento
      • University of Minnesota, Minneapolis
      • Stanford University, Stanford
    • Patient Characteristics: Adult surgical patients (cardiac surgery, general surgery, and surgical ICU) whose StO2 (via Foresight Sensors) and MAP (via Flotrac Sensors) were being monitored. Patients were randomly selected.

    3. Number of Experts and Qualifications for Ground Truth

    The document does not explicitly state the number of experts or their qualifications used to establish the ground truth.

    4. Adjudication Method for Test Set

    The document does not explicitly state an adjudication method. The ground truth was based on the Pearson's Correlation Coefficient (Corr) between MAP and StO2 from the clinical data.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC comparative effectiveness study was not mentioned in the provided text. The study focused on the standalone performance of the algorithm.

    6. Standalone (Algorithm Only) Performance

    Yes, a standalone performance study was conducted. The reported sensitivity, specificity, and ROC AUC values are for the CAI Algorithm operating without human intervention, based on the retrospective clinical data.

    7. Type of Ground Truth Used

    The ground truth was established by classifying the relationship between MAP and StO2 using the Pearson's Correlation Coefficient (Corr) from the utilized time-series clinical data.

    • Weak/Moderate MAP-StO2 association: 0 < Corr < 0.7
    • Strong MAP-StO2 association: Corr ≥ 0.7

    8. Sample Size for Training Set

    The document does not explicitly state the sample size for the training set. It mentions "Validation using simulated time-series data" and "Validation using data retrospectively collected from animals" as part of the software verification. The clinical performance data summarized here appears to be solely for validation/testing, rather than training.

    9. How Ground Truth for Training Set Was Established

    The document does not explicitly state how ground truth for the training set (if any separate training occurred with ground truth) was established. It briefly mentions "Validation using simulated time-series data" and "Validation using data retrospectively collected from animals" for software verification, which might imply a component of algorithm development and initial testing, but specific ground truth establishment methods for a training set are not detailed in this section.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221704

    Validate with FDA (Live)

    Date Cleared
    2022-11-22

    (162 days)

    Product Code
    Regulation Number
    870.1425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K163381, K163381, K180881, K190205, K180003, K201446

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module: The HemoSphere Advanced Monitor when used with the HemoSphere Swan-Ganz Module and Edwards Swan-Ganz Catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of cardiac output [continuous (CO) and intermittent (iCO)] and derived hemodynamic parameters. It may also be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards Swan-Ganz catheter indications for use statement for information on target patient population specific to the catheter being used. Refer to the Intended Use statement below for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Advanced Monitor with HemoSphere Oximetry Cable: The HemoSphere Advanced Monitor when used with the HemoSphere Oximetry Cable and Edwards oximetry catheters is indicated for use in adult and pediatric crtical care patients requiring of venous oxygen saturation (SvO2 and Scv02) and derived hemodynamic parameters in a hospital environment. Refer to the Edwards oximetry catheter indications for use statement for information on target patient population specific to the catheter being used. Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Advanced Monitor with HemoSphere Pressure Cable: The HemoSphere Advanced Monitor when used with the HemoSphere Pressure Cable is indicated for use in critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac, Acumen IQ and TruWave DPT sensor indications for use statement for information on target patient population specific to the sensor being used. The Edwards Lifesciences Acumen Hypotension Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter. Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population. When used in combination with the Swan-Ganz Module connected to a compatible Swan-Ganz catheter, the Edward Lifesciences Right Ventricular Pressure (RVP) algorithm provides the clinician with physiological insight into the hemodynamic status of the right ventricle of the heart. The RVP algorithm is indicated for critically ill patients over 18 years of age receiving advanced hemodynamic monitoring in the operating room (OR) and intensive care unit (ICU). The RVP algorithm is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Right Ventricular Pressure (RVP) parameters. Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Advanced Monitor with HemoSphere Tissue Oximetry Module: The noninvasive FORE-SIGHT ELITE tissue oximeter module is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the Sensors in individuals at risk for reduced-flow or no-flow ischemic states. The FORE-SIGHT ELITE tissue oximeter module is intended to allow for the display of StO2 on the HemoSphere advanced monitor. · When used with large sensors is indicated for use on adults and transitional adolescents ≥40 kg. · When used with Medium Sensors, the FORE-SIGHT ELITE tissue oximeter module is indicated for use on pediatric subjects ≥3 kg. · When used with Small Sensors, the FORE-SIGHT ELITE tissue oximeter module is indicated for cerebral use on pediatric subjects <8 kg and non-cerebral use on pediatric subjects <5kg. Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Advanced Monitor with HemoSphere ClearSight Module: The HemoSphere advanced monitor when used with the HemoSphere ClearSight module, pressure controller and a compatible Edwards finger cuff are indicated for patients over 18 years of age in which the balance between cardiac function, fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the noninvasive system is indicated for use in patients with co-morbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere advanced monitor and compatible Edwards finger cuffs noninvasively measures blood pressure and associated hemodynamic parameters. The Edwards Lifesciences Acumen Hypotension Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter. Refer to the ClearSight finger cuff and Acumen IQ finger cuff indications for use statements for information on target patient population specific to the finger cuff being used.

    Device Description

    The HemoSphere Advanced Monitoring platform was designed to simplify the customer experience by providing one platform with modular solutions for their hemodynamic monitoring needs. The user can choose from the available optional sub-system modules or use multiple sub-system modules at the same time. This modular approach provides the customer with the choice of purchasing and/or using specific monitoring applications based on their needs. Users are not required to have all of the modules installed at the same time for the platform to function. HemoSphere Advanced Monitoring Platform consists of the HemoSphere Advanced Monitor that provides a means to interact with and visualize hemodynamic and volumetric data on a screen and five (5) optional external modules: the HemoSphere Swan-Ganz Module (K163381 Cleared, April 14, 2017), the HemoSphere Oximetry Cable (K163381 Cleared, April 14, 2017), HemoSphere Pressure Cable (K180881 Cleared, November 16, 2018), HemoSphere Technology Module (K190205 August 29, 2019). HemoSphere ForeSight Module (K180003, May 10, 2018), and the HemoSphere ClearSight Module (K201446 Cleared October 1, 2020).

    AI/ML Overview

    The provided FDA 510(k) summary (K221704) for the HemoSphere Advanced Monitoring Platform does not contain a table of acceptance criteria and reported device performance for the modifications made (specifically the Right Ventricular Pressure (RVP) algorithm). While it states that "All tests passed" and "demonstrated that the subject devices meet their predetermined design and performance specifications," specific numerical performance metrics and their corresponding acceptance criteria are not detailed in this document.

    However, based on the information provided, here's a breakdown of the other requested information regarding the study supporting the device:

    1. Table of Acceptance Criteria and Reported Device Performance

    Not available in the provided document. The document states that "all performance verification and validation activities demonstrated that the subject devices meet their predetermined design and performance specifications" and "All tests passed," but it does not specify the quantitative acceptance criteria or the numerical results achieved by the device against those criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states: "Clinical data (waveforms) were collected in support of the design and validation of the RVP algorithm."

    • Sample Size for Test Set: Not specified. The document does not provide the number of patients or waveforms used for the clinical data collection for the RVP algorithm validation.
    • Data Provenance: Not specified. The document does not mention the country of origin of the data or whether it was retrospective or prospective.

    3. Number of Experts Used to Establish Ground Truth and Their Qualifications

    Not applicable/Not specified. The document mentions the collection of "clinical data (waveforms)" for the RVP algorithm validation, but it does not describe a process involving human experts to establish ground truth from this data. The RVP algorithm likely derives its parameters directly from physiological waveform data obtained from the Swan-Ganz Module and Pressure Cable, rather than relying on expert interpretation for ground truth.

    4. Adjudication Method for the Test Set

    Not applicable/Not specified. As there is no mention of human expert-established ground truth, an adjudication method is not described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The document does not mention a Multi-Reader Multi-Case (MRMC) comparative effectiveness study, nor does it discuss human readers or AI assistance in a comparative context. The device focuses on monitoring physiological parameters rather than image interpretation or diagnostic tasks involving human readers.

    6. Standalone (Algorithm Only) Performance Study

    Yes, implicitly. The validation of the RVP algorithm described in the document is a standalone performance assessment. The statement "Clinical data (waveforms) were collected in support of the design and validation of the RVP algorithm" implies that the algorithm's performance was evaluated based on this collected data. The conclusion that the device "has successfully passed functional and performance testing, including software and algorithm verification and validation and bench studies" further supports that the algorithm's performance was assessed. However, specific standalone performance metrics are not provided.

    7. Type of Ground Truth Used for the Test Set

    The ground truth for the RVP algorithm's validation would be the physiological waveform data itself, specifically from the Swan-Ganz Module and Pressure Cable. The algorithm processes this raw physiological data to derive parameters like SYSRVP, DIARVP, MRVP, PRRVP, RV dp/dt, and RVEDP. The validation would involve comparing the algorithm's derived parameters against established methods or calculations from the same direct physiological measurements (e.g., from the Swan-Ganz catheter and pressure sensors).

    8. Sample Size for the Training Set

    Not specified. The document mentions "clinical data (waveforms) were collected in support of the design and validation of the RVP algorithm," but it does not differentiate between data used for design/training and data used specifically for validation (test set), nor does it specify the sample size for any such training.

    9. How the Ground Truth for the Training Set Was Established

    Not specified/Not explicitly described. Given that the RVP algorithm processes physiological signals from existing, cleared hardware, the "ground truth" for any potential training would inherently be the raw physiological signals themselves, as measured by the Swan-Ganz Module and Pressure Cable. The algorithm's development would likely be based on established physiological principles and signal processing techniques to derive the mentioned RVP parameters. The document does not detail a specific "training set" or a separate process for establishing ground truth for training data beyond the intrinsic nature of the physiological measurements.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221833

    Validate with FDA (Live)

    Date Cleared
    2022-11-07

    (137 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K163381, K180881, K190205, K201446

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module:

    The HemoSphere Advanced Monitor when used with the HemoSphere Swan-Ganz Module and Edwards Swan-Ganz Catheters is indicated for use in adult and pediatric critical care patients requiring of cardiac output [continuous (CO) and intermittent (iCO)] and derived hemodynamic parameters. It may also be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards Swan-Ganz catheter indications for use statement for information on target patient population specific to the catheter being used.

    Refer to the Intended Use statement below for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Advanced Monitor with HemoSphere Oximetry Cable:

    The HemoSphere Advanced Monitor when used with the HemoSphere Oximetry cable and Edwards is indicated for use in adult and pediatric crtical care patients requring of venous oxygen saturation (SvO2 and ScvO2) and derived hemodynamic parameters in a hospital environment. Refer to the Edwards oximetry catheter indications for use statement for information on target patient population specific to the catheter being used.

    Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Advanced Monitor with HemoSphere Pressure Cable:

    The HemoSphere Advanced Monitor when used with the HemoSphere Pressure Cable is indicated for use in critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac, Acumen IQ and TruWave DPT sensor indications for use statement for information on target patient population specific to the sensor being used.

    The Edwards Lifesciences Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical and nonsurgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.

    Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Advanced Monitor with HemoSphere Tissue Oximetry Module:

    The noninvasive FORE-SIGHT ELITE tissue oximeter module is intended for use as an adjunct monitor for absolute regional hemoglobin oxygen saturation of blood under the Sensors in individuals at risk for reduced-flow or no-flow ischemic states. The FORE-SIGHT ELITE tissue oximeter module is intended to allow for the display of StO2 on the HemoSphere advanced monitor.

    · When used with large sensorsis indicated for use on adults and transitional adolescents ≥40 kg.

    · When used with Medium Sensors, the FORE-SIGHT ELITE tissue oximeter module is indicated for use on pediatric subjects ≥3 kg.

    · When used with Small Sensors, the FORE-SIGHT ELITE tissue oximeter module is indicated for cerebral use on pediatric subjects <8 kg and non-cerebral use on pediatric subjects <5kg.

    Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Advanced Monitor with HemoSphere ClearSight Module:

    The HemoSphere Advanced Monitor when used with the HemoSphere ClearSight module, pressure controller and a compatible Edwards finger cuff are indicated for patients over 18 years of age in which the balance between cardiac function, fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the non-invasive system is indicated for use in patients with co-morbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere advanced monitor and compatible Edwards finger cuffs noninvasively measures blood pressure and associated hemodynamic parameters.

    Refer to the ClearSight finger cuff indications for use statements for information on target patient population specific to the finger cuff being used

    Device Description

    HemoSphere Advanced Monitoring Platform consists of the HemoSphere Advanced Monitor that provides a means to interact with and visualize hemodynamic and volumetric data on a screen and five (5) optional external modules: the HemoSphere Swan-Ganz Module (K163381 Cleared, April 14, 2017), the HemoSphere Oximetry Cable (K163381 Cleared, April 14, 2017), HemoSphere Pressure Cable (K180881 Cleared, November 16, 2018), HemoSphere Tissue Oximetry Module (K190205 August 29, 2019), and the HemoSphere ClearSight Module (K201446 Cleared October 1, 2020).

    AI/ML Overview

    Acceptance Criteria and Device Performance for Edwards HemoSphere ClearSight Module

    Based on the provided text, the Edwards HemoSphere ClearSight Module has undergone a modification to its existing APCO algorithm. The acceptance criteria and performance evaluation are related to ensuring this modification did not adversely affect the safety and effectiveness of the device, particularly concerning Cardiac Output accuracy.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Predicate Device Specifications: The modified algorithm's performance (specifically Cardiac Output accuracy) should meet the specifications cleared for the predicate device (Edwards HemoSphere Advanced Monitoring Platform, K201446)."All testing passed without exception." The retrospective analysis of clinical data demonstrated that the modification did not adversely affect the safety and effectiveness of the subject device, and "All tests passed."
    No Adverse Effect on Safety and Effectiveness: The modification should not negatively impact other aspects of the device's safety and effectiveness."System verification activities confirmed that the modification to the device did not adversely affect the safety and effectiveness of the subject device."
    Software Verification: The software modification should comply with FDA guidance for software in medical devices, including design, development, and traceability."Software verification was performed per FDA's Guidance for Industry and FDA Staff... All tests passed."
    System Verification: The algorithm change should be integrated without issues."the change in the algorithm was integrated without any concern and all integration passed with no exceptions."
    Unchanged Design, Materials, Energy Source, User Interface, Measurement Principle, and Performance Specifications: These aspects of the HemoSphere ClearSight Module should remain the same."Design, materials, energy source, user interface, measurement principle, and all performance specifications of the modified HemoSphere ClearSight Module remain unchanged."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The text states that "retrospective analysis of clinical data from multiple independent datasets, comprised of data from patients over the age of 18 years" was used. However, a specific numerical sample size (e.g., number of patients or data points) is not provided in the document.
    • Data Provenance: The data was "retrospective analysis of clinical data from multiple independent datasets." The country of origin of the data is not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The text does not describe how the "ground truth" for the clinical data used in the retrospective analysis was established, nor does it mention the use of experts for this purpose.

    4. Adjudication Method for the Test Set

    This information is not provided in the document.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study is not mentioned or described in the document. This study focuses on an algorithm modification for a medical device (HemoSphere ClearSight Module) measuring physiological parameters, not on human reader performance with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone performance evaluation of the algorithm was done. The "Algorithm Verification (Clinical Performance Data)" section specifically states: "Algorithm performance was tested using clinical data." This indicates an evaluation of the algorithm's performance independent of real-time human interaction.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The document refers to the evaluation of "Cardiac Output accuracy." For a device that measures physiological parameters like Cardiac Output, the "ground truth" would typically refer to a gold standard measurement technique for that parameter. However, the exact gold standard method used to establish the ground truth for Cardiac Output in the clinical data is not explicitly stated in the provided text. It implies the use of "clinical data" which would have reference measurements for comparison but does not detail the nature of these reference measurements.

    8. The Sample Size for the Training Set

    The document only mentions "retrospective analysis of clinical data" for testing the algorithm modification. It does not provide any information regarding a "training set" or its sample size. This suggests that the modification might have been made to an existing algorithm, and the focus of this submission is on verifying the impact of that modification using a test set, rather than developing a new algorithm from scratch requiring a separate training set.

    9. How the Ground Truth for the Training Set Was Established

    Since no training set is mentioned, this information is not provided in the document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223127

    Validate with FDA (Live)

    Date Cleared
    2022-11-07

    (35 days)

    Product Code
    Regulation Number
    870.1425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K163381, K180881, K190205, K203687

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indications for Use: HemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module The HemoSphere Advanced Monitor when used with the HemoSphere Swan-Ganz Module and Edwards Swan-Ganz Catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of cardiac output [continuous (CO) and intermittent (iCO)] and derived hemodynamic parameters. Pulmonary artery blood temperature monitoring is used to compute continuous and intermittent CO with thermodilution technologies. It may also be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards Swan-Gatheter indications for use statement for information on target patient population specific to the catheter being used.

    Refer to the Intended Use statement below for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Advanced Monitor with HemoSphere Oximetry Cable

    The HemoSphere Advanced Monitor when used with the HemoSphere Oximetry Cable and Edwards is indicated for use in adult and pediatric critical care patients requiring of venous oxygen saturation (SvO2 and Scv02) and derived hemodynamic parameters in a hospital environment. Refer to the Edwards oximetry catheter indications for use statement for information on target patient population specific to the catheter being used.

    Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Advanced Monitor with HemoSphere Pressure Cable

    The HemoSphere Advanced Monitor when used with the HemoSphere Pressure Cable is indicated for use in critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac, Acumen IQ, and TruWave DPT sensor indications for use statement for information on target patient population specific to the sensor being used.

    The Edwards Lifesciences Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical or nonsurgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.

    Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Advanced Monitor with Acumen Assisted Fluid Management Feature and Acumen IQ Sensor: The Acumen Assisted Fluid Management (AFM) software feature provides the clinician with physiological insight into a patient's estimated response to fluid therapy and the associated hemodynamics. The Acumen AFM software feature is intended for use in surgical patients ≥18 years of age, that require advanced hemodynamic monitoring. The Acumen AFM software feature offers suggestions regarding the patient's physiological condition and estimated response to fluid therapy.

    Acumen AFM fluid administration suggestions are offered to the clinician; the decision to administer a fluid bolus is made by the clinician, based upon review of the patient's hemodynamics. No therapeutic decisions should be made based solely on the Assisted Fluid Management suggestions.

    HemoSphere Advanced Monitor with HemoSphere Technology Module and ForeSight Oximeter Cable The noninvasive ForeSight Oximeter Cable is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced-flow or no-flow ischemic states. The ForeSight Oximeter Cable is also intended to monitor relative changes of total hemoglobin of blood under the sensors. The ForeSight Oximeter Cable is intended to allow for the display of StO2 and relative change in total hemoglobin on the HemoSphere advanced monitor.

    • When used with large sensors, the ForeSight Oximeter Cable is indicated for use on adults and transitional adolescents >40 kg.

    · When used with medium sensors, the ForeSight Oximeter Cable is indicated for use on pediatric subjects >3 kg.

    · When used with small sensors, the ForeSight Oximeter Cable is indicated for cerebral use on pediatric subjects < 8 kg and non-cerebral use on pediatric subjects <5kg.

    Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Advanced Monitor with HemoSphere ClearSight Module

    The HemoSphere Advanced Monitor when used with the HemoSphere ClearSight module, pressure controller and a compatible Edwards finger cuff are indicated for patients over 18 years of age in which the balance between cardiac function. fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the noninvasive system is indicated for use in patients with co-morbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere advanced monitor and compatible Edwards' finger cuffs noninvasively measures blood pressure and associated hemodynamic parameters.

    The Edwards Lifesciences Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.

    Refer to the ClearSight finger cuff and Acumen IQ finger cuff indications for use statements for information on target patient population specific to the finger cuff being used.

    Refer to the ClearSight finger cuff indications for information on target patient population specific to the finger cuff being used.

    Intended Use of HemoSphere Advanced Monitor: The HemoSphere Advanced Monitoring Platform is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. The Viewfinder Remote mobile application can be used for supplemental near real-time remote display of monitored hemodynamic parameter data as well as Faults, Alerts and Notifications generated by the HemoSphere Advanced Monitoring Platform.

    The HemoSphere Advanced Monitoring Platform is intended for use with compatible Edwards Swan-Ganz and Oximetry Catheters, FloTrac sensors, Acumen IQ sensors, TruWave DPT sensors, ForeSight sensors, and ClearSight/Acumen IQ finger cuffs.

    Intended Use of Viewfinder Remote: Viewfinder Remote is a mobile application, which provides supplemental remote near real-time display of hemodynamic data measured by a connected HemoSphere advanced monitoring platform. Viewfinder Remote allows clinicians to view continuous monitoring data and alarms/alerts remotely for multiple patients. All displayed data is generated by connected HemoSphere advanced monitoring platforms, and not by Viewfinder Remote. Viewfinder Remote is intended for use by clinicians as a supportive visual aid, and not as a replacement for in-person patient monitoring with connected HemoSphere advanced monitoring platforms.

    Device Description

    The HemoSphere Advanced Monitoring platform was designed to simplify the customer experience by providing one platform with modular solutions for their hemodynamic monitoring needs. The user can choose from the available optional sub-system modules or use multiple sub-system modules at the same time. This modular approach provides the customer with the choice of purchasing and/or using specific monitoring applications based on their needs. Users are not required to have all of the modules installed at the same time for the platform to function.

    HemoSphere Advanced Monitoring Platform consists of the HemoSphere Advanced Monitor that provides a means to interact with and visualize hemodynamic and volumetric data on a screen and five (5) optional external modules: the HemoSphere Swan-Ganz Module (K163381 Cleared, April 14, 2017), the HemoSphere Oximetry Cable (K163381 Cleared, April 14, 2017), HemoSphere Pressure Cable (K180881 Cleared, November 16, 2018), HemoSphere Tissue Oximeter Module (K190205 August 29, 2019), HemoSphere ForeSight Oximeter Cable (K213682 cleared June 22, 2022). and the HemoSphere ClearSight Module (K203687 cleared May 28, 2021).

    AI/ML Overview

    The regulatory submission K223127 for the HemoSphere Advanced Monitoring Platform indicates a modification to the existing StO2 algorithm of the HemoSphere ForeSight Oximeter Cable. The submission claims substantial equivalence to a predicate device (K213682 cleared June 22, 2022) and an additional predicate (Fore-Sight Elite Module Tissue Oximeter, K143675 cleared April 10, 2015) for the StO2 algorithm.

    Here's an analysis based on the provided text, fulfilling the requested information points:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission states that "All testing passed without exception" and the "modification did not adversely affect the safety and effectiveness of the subject device." However, specific numerical acceptance criteria (e.g., accuracy +/- X%, bias Y, precision Z) and the corresponding reported performance values for the StO2 algorithm are not explicitly provided in the given text.

    The text vaguely indicates that:

    • "Algorithm performance was tested using the same method and criteria as previously used in the predicate device."
    • "The same methods, protocols and acceptance criteria as the predicate device (K213682) were used to evaluate the modification."
    • "Design, materials, energy source, user interface, measurement principle and all performance specifications of the modified HemoSphere ForeSight Oximeter cable remain unchanged."

    Without the actual specific criteria and reported values from the predicate device's clearance, a detailed table with numerical data cannot be generated from this document.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample size used for the test set. It mentions "Algorithm Verification" and "System Verification" without detailing the number of cases or patients included in these tests. The data provenance (e.g., country of origin, retrospective or prospective) is also not provided.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    The document does not specify the number of experts used or their qualifications for establishing ground truth for the test set. Given that the modified device measures "absolute regional hemoglobin oxygen saturation (StO2)," the ground truth would typically be established through a reference method (e.g., co-oximetry of arterial and venous blood samples, or another validated oximetry technique) rather than expert consensus on images.

    4. Adjudication Method for the Test Set

    The document does not mention any adjudication method for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    The document does not indicate that an MRMC comparative effectiveness study was performed. The device is a monitoring platform providing quantitative physiological parameters, not an imaging device requiring human reader interpretation in the same way.

    6. Standalone (Algorithm Only) Performance Study

    Yes, a standalone performance study of the algorithm was done. The submission explicitly states:

    • "Algorithm performance was tested using the same method and criteria as previously used in the predicate device."
    • "The results establish that the modification did not adversely affect the safety and effectiveness of the subject device."

    This "Algorithm Verification" section refers to directly testing the algorithm's performance.

    7. Type of Ground Truth Used

    The document does not explicitly state the type of ground truth used but implies it would be a comparison to existing specifications from the predicate devices. For a tissue oximeter measuring StO2, the ground truth would typically involve comparison to a validated reference method for oxygen saturation, possibly through in-vivo or in-vitro testing. It is not expert consensus on images or pathology in this context.

    8. Sample Size for the Training Set

    The document does not provide the sample size for the training set. This is a modification to an existing algorithm, so the original algorithm would have been developed and trained, but details about that original training are not in this submission.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide details on how the ground truth for the training set (of the original algorithm) was established. It only refers to the modification of an "existing StO2 algorithm."

    Ask a Question

    Ask a specific question about this device

    K Number
    K211465

    Validate with FDA (Live)

    Date Cleared
    2021-07-08

    (58 days)

    Product Code
    Regulation Number
    870.1425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K163381, K180881, K190205, K180003

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module
    The HemoSphere Advanced Monitor when used with the HemoSphere Swan-Ganz Module and Edwards Swan-Ganz Catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of cardiac output [continuous (CO) and intermittent (iCO)] and derived hemodynamic parameters. It may also be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards Swan-Ganz catheter indications for use statement for information on target patient population specific to the catheter being used.
    Refer to the Intended Use statement below for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Advanced Monitor with HemoSphere Oximetry Cable
    The HemoSphere Advanced Monitor when used with the HemoSphere Oximetry cable and Edwards is indicated for use in adult and pediatric critical care patients requiring of venous oxygen saturation (SvO2 and SevO2) and derived hemodynamic parameters in a hospital environment. Refer to the Edwards oximetry catheter indications for use statement for information on target patient population specific to the catheter being used.
    Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Advanced Monitor with HemoSphere Pressure Cable
    The HemoSphere Advanced Monitor when used with the HemoSphere Pressure Cable is indicated for use in critical care patients in which the balance between cardiac function. fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac, Acumen IQ and TruWave DPT sensor indications for use statement for information on target patient population specific to the sensor being used.
    The Edwards Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Acumen Hypotension Prediction Index (HPI) parameter.
    Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Advanced Monitor with HemoSphere Tissue Oximetry Module and FORE-SIGHT ELITE Tissue Oximeter Module
    The non-invasive Fore-Sight Elite tissue oximeter module is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced-flow or no-flow ischemic states. The Fore-Sight Elite tissue oximeter module is intended to allow for the display of StO2 on the HemoSphere advanced monitor.
    · When used with large sensors, the Fore-Sight Elite tissue oximeter module is indicated for use on adults and transitional adolescents ≥40 kg.
    · When used with medium sensors, the Fore-Sight Elite tissue oximeter module is indicated for use on pediatric subjects >3 kg.
    · When used with small sensors, the Fore-Sight Elite tissue oximeter module is indicated for cerebral use on pediatric subjects <8 kg and non-cerebral use on pediatric subjects <5kg.
    Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Advanced Monitor with HemoSphere ClearSight Module
    The HemoSphere advanced monitor when used with the HemoSphere ClearSight module, pressure controller and a compatible Edwards finger cuff are indicated for patients over 18 years of age in which the balance between cardiac function, fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the non-invasive system is indicated for use in patients with co-morbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere advanced monitor and compatible Edwards finger cuffs noninvasively measures blood pressure and associated hemodynamic parameters. Refer to the ClearSight finger cuff indications for use statements for information on target patient population specific to the finger cuff being used.
    Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    Intended use - Viewfinder Remote:
    Viewfinder Remote is a mobile application which provides supplemental remote near real-time display of non-invasive hemodynamic parameters measured by a connected HemoSphere advanced monitoring platform. Viewfinder Remote allows clinicians to view continuous monitoring data and alarms/alerts remotely for multiple patients. All displayed data is generated by connected HemoSphere advanced monitoring platforms, and not by Viewfinder Remote. Viewfinder Remote is intended for use by clinicians as a supportive visual aid, and not as a replacement for in-person patient monitoring with connected HemoSphere advanced monitoring platforms.

    Device Description

    The HemoSphere Advanced Monitoring platform was designed to simplify the customer experience by providing one platform with modular solutions for their hemodynamic monitoring needs. The user can choose from the available optional sub-system modules or use multiple sub-system modules at the same time. This modular approach provides the customer with the choice of purchasing and/or using specific monitoring applications based on their needs. Users are not required to have all of the modules installed at the same time for the platform to function.
    HemoSphere Advanced Monitoring Platform consists of the HemoSphere Advanced Monitor that provides a means to interact with and visualize hemodynamic and volumetric data on a screen and five (5) optional external modules: the HemoSphere Swan-Ganz Module (K163381 Cleared, April 14, 2017), the HemoSphere Oximetry Cable (K163381 Cleared, April 14, 2017), HemoSphere Pressure Cable (K180881 Cleared, November 16, 2018), HemoSphere Tissue Oximetry Module (K190205 August 29, 2019), HemoSphere ForeSight Module (K180003, May 10, 2018), and the HemoSphere ClearSight Module (K201446 cleared October 1, 2020).
    The HemoSphere Advanced Monitor (K201446 most recently cleared October 1, 2020), subject of this submission, is being modified to enable connectivity to a Viewfinder Remote mobile application via a software-based Viewfinder Hub and Viewfinder Cloud.
    The Viewfinder Remote mobile application provides clinicians with a supplemental near-real time display of the patient hemodynamic data from the connected HemoSphere Advanced Monitoring Platform. The Viewfinder Remote application is part of the Edwards Viewfinder network, which includes Viewfinder Hub and Viewfinder Cloud. The Viewfinder Remote mobile application functions as a supportive visual aid for patient status communication between clinicians and allows them to view multiple patient monitoring sessions at once from their mobile device. The near-real time updates to patient monitoring sessions includes non-invasive hemodynamic parameter data and the associated physiological alarm notifications, historical trend data and parameter waveform data.

    AI/ML Overview

    The provided text describes modifications to an existing device, the HemoSphere Advanced Monitoring Platform, to include a new feature: the Viewfinder Remote mobile application. The primary predicate device is the HemoSphere Advanced Monitoring Platform itself (K201446). The submission focuses on the addition of the Viewfinder Remote and changes to the HemoSphere Advanced Monitor to support this remote viewing functionality.

    The acceptance criteria and study information provided are primarily related to the safety and functionality of the new remote viewing feature and the integrated system, rather than the clinical performance of the underlying hemodynamic monitoring parameters. No clinical performance data is deemed necessary for the modifications.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative acceptance criteria for parameters like accuracy, sensitivity, or specificity for the Viewfinder Remote mobile application. Instead, it states that "All tests passed" for various verification activities, implying that the device met internal performance specifications for these tests. The focus is on ensuring data integrity, safety, and usability of the remote viewing function.

    Acceptance Criteria CategorySpecific Test/DescriptionReported Device Performance
    System VerificationEnd-to-end system verification for data integrity and accuracy (Monitor to Remote View mobile application)All tests passed.
    Electrical Safety & EMCIEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 62304, IEC 62366, IEC 60601-2-34 and IEC 60601-2-49 standardsAll tests passed.
    Wireless CoexistenceANSI C63.27/D1.0, bench and simulated environment testing on the entire platform, including modules and the Viewfinder RemoteAll tests passed.
    Software VerificationFDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"All tests passed. (HemoSphere Advanced Monitor: Major Level of Concern; Viewfinder Remote: Moderate Level of Concern)
    Usability StudyFDA's guidance, "Applying Human Factors and Usability Engineering to Medical Devices"All tests passed.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify sample sizes for any of the verification tests (system, wireless, software, usability). It does not mention clinical data proving the device meets the acceptance criteria, as clinical data was not required for the modifications. The tests appear to be non-clinical (bench and simulated environment).

    • Test Set Sample Size: Not specified.
    • Data Provenance: Non-clinical (bench and simulated environment testing).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided. The verification activities mentioned are technical and regulatory compliance tests, not typically requiring expert ground truth in the way a diagnostic algorithm's performance would. For the usability study, "clinicians" are mentioned as users, but their number and specific qualifications are not detailed.

    4. Adjudication Method for the Test Set

    This information is not provided.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No such study was mentioned. The device (Viewfinder Remote) is a "supportive visual aid" for remote display of parameters, not an AI-assisted diagnostic tool that would improve human reader performance in interpreting complex medical images or data.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The Viewfinder Remote itself is a standalone application in the sense that it displays data from the HemoSphere monitor, but its "performance" is based on accurately reflecting the data generated by the primary monitoring platform. The verification activities focus on the integrity and accuracy of this data transfer and display, not on the Viewfinder Remote generating independent algorithmic outputs. The "Acumen Hypotension Prediction Index Feature" is mentioned as a component of the HemoSphere system, providing "additional quantitative information," but the document doesn't detail standalone performance studies for this specific AI feature outside of its integration with the HemoSphere platform and the context of the Viewfinder Remote submission.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the HemoSphere Advanced Monitoring Platform itself, the ground truth for its core hemodynamic parameters would typically be established through validation against gold-standard invasive measurements or other well-established methods. However, for the modifications related to the Viewfinder Remote, the "ground truth" for the verification tests involved comparing the displayed data on the remote application against the data displayed on the primary HemoSphere monitor, and ensuring it passed technical and regulatory standards.

    • For the Viewfinder Remote modifications: The ground truth for data integrity and accuracy would be the parameters and alarms generated and displayed on the HemoSphere Advanced Monitor itself.
    • For the overall HemoSphere Advanced Monitor (implied from its existing clearance K201446 and general device type): Ground truth for hemodynamic parameters like cardiac output, pressure, and oxygen saturation would typically rely on established clinical measurement techniques (e.g., thermodilution for CO, direct arterial line measurement for BP, co-oximetry for SvO2).

    8. The Sample Size for the Training Set

    This information is not provided. Given that clinical data was not required for these specific modifications, and the focus is on a remote display application rather than a new diagnostic algorithm requiring extensive training data, a training set as typically understood for AI models is not explicitly mentioned. If the Acumen HPI feature involves AI, its training data details are not provided in this document.

    9. How the Ground Truth for the Training Set Was Established

    This information is not provided. (See point 8).

    Ask a Question

    Ask a specific question about this device

    K Number
    K203687

    Validate with FDA (Live)

    Date Cleared
    2021-05-28

    (162 days)

    Product Code
    Regulation Number
    870.1425
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K163381, K190205

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module:
    The HemoSphere Advanced Monitor when used with the HemoSphere Swan-Ganz Module and Edwards Swan-Ganz Catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of cardiac output [continuous (CO) and intermittent (iCO)] and derived hemodynamic parameters. It may also be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards Swan-Ganz catheter indications for use statement for information on target patient population specific to the catheter being used.
    Refer to the Intended Use statement below for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Advanced Monitor with HemoSphere Oximetry Cable:
    The HemoSphere Advanced Monitor when used with the HemoSphere Oximetry cable and Edwards is indicated for use in adult and pediativ critical care patients requiring of venous oxygen saturation (Sv02 and ScvO2) and derived hemodynamic parameters in a hospital environment. Refer to the Edwards oximetry catheter indications for use statement for information on target patient population specific to the catheter being used.
    Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Advanced Monitor with HemoSphere Pressure Cable:
    The HemoSphere Advanced Monitor when used with the HemoSphere Pressure Cable is indicated for use in critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac, Acumen IQ and TruWave DPT sensor indications for use statement for information on target patient population specific to the sensor being used.
    The Edwards Lifesciences Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.
    Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Advanced Monitor with HemoSphere Tissue Oximetry Module:
    The noninvasive FORE-SIGHT ELITE tissue oximeter module is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the Sensors in individuals at risk for reduced-flow or no-flow ischemic states. The FORE-SIGHT ELITE tissue oximeter module is intended to allow for the display of StO2 on the HemoSphere advanced monitor.
    · When used with large sensorsis indicated for use on adults and transitional adolescents ≥40 kg.
    · When used with Medium Sensors, the FORE-SIGHT ELITE tissue oximeter module is indicated for use on pediatric subjects ≥3 kg.
    · When used with Small Sensors, the FORE-SIGHT ELITE tissue oximeter module is indicated for cerebral use on pediatric subjects <8 kg and non-cerebral use on pediatric subjects <5kg.
    Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Advanced Monitor with HemoSphere ClearSight Module:
    The HemoSphere advanced monitor when used with the HemoSphere ClearSight module, pressure controller and a compatible Edwards finger cuff are indicated for patients over 18 years of age in which the balance between cardiac function, fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the noninvasive system is indicated for use in patients with co-morbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere advanced monitor and compatible Edwards finger cuffs noninvasively measures blood pressure and associated hemodynamic parameters.
    The Edwards Lifesciences Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.
    Refer to the ClearSight finger cuff and Acumen IO finger cuff indications for information on target patient population specific to the finger cuff being used.
    Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    Device Description

    HemoSphere Advanced Monitoring Platform consists of the HemoSphere Advanced Monitor that provides a means to interact with and visualize hemodynamic and volumetric data on a screen and five (5) optional external modules: the HemoSphere Swan-Ganz Module (K163381 Cleared, April 14, 2017), the HemoSphere Oximetry Cable (K163381 Cleared, April 14, 2017), HemoSphere Pressure Cable (K180881 Cleared, November 16, 2018), HemoSphere Tissue Oximetry Module (K190205 August 29, 2019), and the HemoSphere ClearSight Module (K201446 Cleared October 1, 2020).
    The HemoSphere Advanced Monitor with HemoSphere ClearSight module is a non-invasive monitoring platform intended to continuously and noninvasively measure blood pressure and associated hemodynamic parameters.
    The HemoSphere Advanced Monitoring Platform when used with ClearSight Module uses the same technology as the predicate device. The volume clamp method of Peñáz is the measurement method, and the pressure waveform reconstruction is based on the generalized transfer function and level of correction of Gizdulich.
    The platform also includes the Acumen Hypotension Prediction Index (HPI) feature for the ClearSight (non-invasive) technology. Currently, the HemoSphere Pressure Cable enables the HPI feature when connected to an Acumen IQ sensor. This feature has been updated to also enable the HPI feature when connected to the Acumen IQ finger cuff (ClearSight finger cuff).
    The modified ClearSight System receives the pressure signal from the finger cuff and reconstructs it to a radial arterial pressure representation of the signal. This radial reconstructed signal is then used to calculate the previously available key hemodynamic parameters; PR, MAP, SYS, DIA, PPV and SVV. Additionally, the radial reconstructed signal is also used to calculate the parameters associated with the Hypotension Prediction Index feature (HPI, Eadyn, and dP/dt).
    The cardiac output (CO) and other measurements derived from cardiac output, such as stroke volume (SV), cardiac output index (CI), stroke volume index (SVI) remain unchanged and will continue to use previously cleared ClearSight algorithm to reconstruct the finger pressure waveform into a brachial arterial pressure waveform.

    AI/ML Overview

    The provided text describes the HemoSphere Advanced Monitoring Platform, which is a medical device. The submission focuses on modifications to the ClearSight algorithm and the addition of the Acumen Hypotension Prediction Index (HPI) feature for non-invasive technology.

    Here's an analysis of the acceptance criteria and the study that proves the device meets those criteria, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria with specific numerical targets. Instead, it states that "Completion of all verification and validation activities demonstrated that the subject devices meet their predetermined design and performance specifications." and "All tests passed." for various types of testing. The primary performance claim is related to the substantial equivalence to the predicate device, particularly regarding safety and effectiveness.

    For the Acumen HPI feature, the indication states: "The Acumen HPI feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics." The acceptance is implied by the successful clinical performance study demonstrating that the usage of the HPI algorithm with non-invasive technology did not adversely affect the safety and effectiveness.

    Acceptance Criteria CategoryReported Device Performance
    Overall PerformanceSubstantially equivalent to predicate device (K201446). Does not raise any concerns of safety and effectiveness.
    UsabilityIntended users can perform primary operating functions and critical tasks without any usability issues that may lead to patient or user harm. (Passed usability study).
    System VerificationMeets predetermined design and performance specifications. Differences in design/materials did not adversely affect safety/effectiveness. Measured and derived parameters tested using bench simulation. Individual modules and integrated system tested for safety and effectiveness. (All tests passed).
    Electrical Safety & EMCComplies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 62304, IEC 62366-1, IEC 60601-2-34, IEC 60601-2-57, IEC 60601-2-49 and IEC 80601-2-49. (All tests passed).
    Wireless CoexistencePassed bench and simulated environment testing on the entire HemoSphere Advanced Monitoring Platform. (All tests passed).
    Software VerificationPerformed per FDA guidance. Software on individual modules tested at a sub-system level to ensure device safety. (All tests passed).
    Algorithm VerificationThe usage of the HPI algorithm with non-invasive technology did not adversely affect the safety and effectiveness of the subject device.
    Clinical PerformanceDemonstrates safety and effectiveness profile is substantially equivalent to the predicate device. The changes regarding the Acumen HPI feature and ClearSight algorithm did not adversely affect safety and effectiveness.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: The document mentions "multiple independent datasets" for the clinical performance study. It states these datasets were comprised of "patients over the age of 18 years undergoing surgical procedures with both minimally invasive and non-invasive monitoring." However, a specific numerical sample size for the test set is not provided.
    • Data Provenance: The data was from "Prospective analyses of retrospective clinical data." The country of origin is not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. The document states that the HPI feature defines a hypotensive event as "mean arterial pressure < 65 mmHg for at least one minute in duration," which is an objective physiological measure, not one requiring expert interpretation for ground truth. For other parameters, substantial equivalence to the predicate device is claimed, implying the established methods for those measurements serve as ground truth or reference.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the document. Given the nature of the device (physiological monitoring), traditional adjudication methods involving expert review of images or diagnoses are unlikely to be directly applicable in the same way as, for example, diagnostic imaging AI. The ground truth for hypotension is defined objectively by MAP thresholds.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study involving human readers and AI assistance is mentioned or implied. The HemoSphere Advanced Monitoring Platform and its HPI feature are presented as tools for clinicians to gain physiological insight, not as AI that directly 'assists' a human reader in interpreting complex visual data. The HPI feature is described as providing "additional quantitative information... for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter." There is no discussion of human reader improvement or effect size.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance of the algorithm was done as part of the "Algorithm Verification (Clinical Performance Data)" and "Clinical Performance" sections. The algorithms (ClearSight algorithm update and Acumen HPI feature) were tested using clinical data to establish that their usage did not adversely affect safety and effectiveness. This implies evaluation of the algorithm's output against clinical ground truth without direct human interpretation of the algorithm's internal workings or intermediate outputs.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The primary ground truth for the Acumen HPI feature, as implied by its definition, is physiological outcomes data: "mean arterial pressure < 65 mmHg for at least one minute in duration." For other measured and derived hemodynamic parameters, the ground truth would typically be established by validated reference methods or the performance against the predicate device which itself has established accuracy.

    8. The sample size for the training set

    The sample size for the training set is not explicitly provided. The document states that "No modifications to the core predictive HPI algorithm have been made from the granted version in DEN160044 (granted March 16, 2018)." This suggests the original HPI algorithm was trained prior to this submission and its training data is not detailed here. The current submission focuses on verifying the algorithm's performance with non-invasive technology and changes to the ClearSight algorithm, not on re-training.

    9. How the ground truth for the training set was established

    This information is not provided in the document for the HPI algorithm specifically. Since the core HPI algorithm was not modified, its original training ground truth would have been established during its initial development and clearance (DEN160044). For physiological monitoring devices, training ground truth typically comes from simultaneously acquired data from a highly accurate, often invasive, reference standard.