Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K203156

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-11-20

    (29 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K183676, K163122

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DTX Studio Clinic is a software program for the acquisition, management, transfer and analysis of dental and craniomaxillofacial image information, and can be used to provide design input for dental restorative solutions. It displays and enhances digital images from various sources to support the diagnostic process and treatment planning. It stores and provides these images within the system or across computer systems at different locations.

    Device Description

    DTX Studio Clinic is a software interface for dental/medical practitioners used to analyze 2D and 3D imaging data, in a timely fashion, for the treatment of dental, craniomaxillofacial and related conditions. DTX Studio Clinic displays and processes imaging data from different devices (i.e. Intraoral and extraoral X-rays, (CB)CT scanners, intraoral scanners, intraoral and extraoral cameras).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the DTX Studio Clinic device, based on the provided text:

    Important Note: The provided text is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device, not necessarily a comprehensive clinical study report. Therefore, some information requested (like specific sample sizes for test sets, the number and qualifications of experts for ground truth, adjudication methods, MRMC study effect sizes, and detailed information about training sets) is not explicitly stated in this document. The focus here is on software validation and verification.


    Acceptance Criteria and Reported Device Performance

    The document does not explicitly state numerical "acceptance criteria" in the format of a table with specific metrics (e.g., sensitivity, specificity, accuracy thresholds). Instead, the acceptance is based on demonstrating that the DTX Studio Clinic software performs its intended functions reliably and safely, analogous to the predicate and reference devices, as verified through software validation and engineering testing.

    The "reported device performance" is primarily described through the software's functionality and its successful verification and validation.

    Feature/CriterionReported Device Performance (DTX Studio Clinic)Comments (Based on 510(k) Summary)
    Clinical UseSupports diagnostic and treatment planning for craniomaxillofacial anatomical area."Primarily the same" as the predicate device CliniView (K162799). Differences in wording do not alter therapeutic use.
    Image Data Import & AcquisitionAcquires/imports DICOM, 2D/3D images (CBCT, OPG/panorex, intra-oral X-ray, cephalometric, clinical pictures). Also imports STL, NXA, PLY files from intraoral/optical scanners. Directly acquires images from supported modalities or allows manual import. Imports from 3rd party PMS systems via VDDS or OPP protocol.Similar to CliniView, with additional capabilities (STL, NXA, PLY, broader PMS integration). Subject device does not control imaging modalities directly for acquisition settings, distinguishing it from CliniView.
    Data Visualization & ManagementDisplays and enhances digital images. Provides image filters, annotations, distance/angular measurements, volume and surface area measurements (for segmentation). Stores data locally or in DTX Studio Core database. Comparison of 3D images and 2D intraoral images in the same workspace.Core functionality is similar to CliniView. Enhanced features include volume/surface area measurements and comparison of different image types within the same workspace.
    Airway Volume SegmentationAllows volume segmentation of indicated airway, volume measurements, and constriction point determinations.Similar to reference device DentiqAir (K183676), but specifically limited to airway (unlike DentiqAir's broader anatomical segmentation).
    Automatic Image Sorting (IOR)Algorithm for automatic sorting of acquired or imported intra-oral X-ray images to an FMX template. Detects tooth numbers (FDI or Universal).This is a workflow improvement feature, not for diagnosis or image enhancement.
    Intraoral Scanner Module (ioscan)Dedicated intraoral scanner workspace for acquisition of 3D intraoral models (STL, NXA, PLY). Supports dental optical impression systems.Classified as NOF, 872.3661 (510(k) exempt). Does not impact substantial equivalence.
    Alignment of Intra-oral/Cast Scans with (CB)CT DataImports 3D intraoral models or dental cast scans (STL/PLY) and aligns them with imported CB(CT) data for accurate implant planning.Similar to reference device DTX Studio Implant (K163122).
    Implant PlanningFunctionality for implant planning treatment. Adds dental implant shapes to imported 3D data, allowing user definition of position, orientation, type, and dimensions.Similar to reference device DTX Studio Implant (K163122), which also adds implants and computes surgical templates.
    Virtual Tooth SetupCalculates and visualizes a 3D tooth shape for a missing tooth position based on indicated landmarks and loaded intra-oral scan. Used for prosthetic visualization and input for implant position.A new feature not explicitly present in the predicate devices but supported by the overall diagnostic and planning workflow.
    Software Validation & VerificationDesigned and manufactured under Quality System Regulations (21 CFR § 820, ISO 13485:2016). Conforms to EN IEC 62304:2006. Risk management (ISO 14971:2012), verification testing performed. Software V&V testing conducted as per FDA guidance for "Moderate Level of Concern." Requirements for features have been met.Demonstrated through extensive software engineering and quality assurance processes, not clinical performance metrics.

    Study Information

    1. Sample sizes used for the test set and the data provenance:

      • Not explicitly stated in the provided text. The document mentions "verification testing" and "validation testing" but does not detail the specific sample sizes of images or patient cases used for these tests.
      • Data Provenance: The document does not specify the country of origin of the data or whether it was retrospective or prospective. It focuses on the software's functionality and its comparison to predicate devices, rather than the performance on specific clinical datasets.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not explicitly stated in the provided text. The 510(k) summary primarily addresses software functionality verification and validation, not a diagnostic accuracy study involving expert ground truth.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not explicitly stated in the provided text.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done or reported. The document explicitly states: "No clinical data was used to support the decision of substantial equivalence." This type of study would involve clinical data and human readers.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, in spirit. The software validation and verification described are for the algorithm and software functionalities operating independently. While the device does not make autonomous diagnoses (it "supports the diagnostic process and treatment planning"), its individual features (like airway segmentation, image sorting, virtual tooth setup) are tested in a standalone manner in terms of their computational correctness and adherence to specifications. However, this is distinct from standalone clinical performance (e.g., an AI algorithm making a diagnosis without human input). The document focuses on the technical performance of the software.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the software verification and validation, the implicit "ground truth" would be the software's functional specifications and requirements. For features like measurements or segmentation, this would likely involve mathematical correctness checks or comparison to pre-defined anatomical models or manually delineated reference segmentations. It is not based on expert consensus, pathology, or outcomes data in a clinical diagnostic sense, as no clinical data was used for substantial equivalence.
    7. The sample size for the training set:

      • Not explicitly stated in the provided text. The document describes a medical device software for image management and analysis, not a machine learning model that typically requires a large 'training set' in the deep learning sense. If any features (like the automatic image sorting or virtual tooth setup) utilize machine learning, the details of their training (including sample size) are not provided in this 510(k) summary.
    8. How the ground truth for the training set was established:

      • Not explicitly stated in the provided text. As mentioned above, details about training sets are absent. If machine learning is involved in certain features, the ground truth would typically be established by expert annotation or curated datasets, but this is not detailed here.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1