510(k) Data Aggregation

    K Number
    K220243

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-04-30

    (92 days)

    Product Code
    Regulation Number
    870.1425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K210458, K183099

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ultreon™ Software for Ultiri™ Measurement System is intended to be used only with compatible Ultiri™ Measurement Systems.

    The Ultiri™ Measurement System is intended for use in the catheterization and related cardiovascular specialty laboratories and will further compute and display various physiological parameters based on the output from one or more electrodes, transducers, or measuring devices. The physician may use the acquired physiological parameters, along with knowledge of patient history, medical expertise, and clinical judgment to determine if thervention is indicated.

    Indications for Use (capital equipment hardware)

    The Ultiri™ Measurement System is intended for use in the catheterization and related cardiovascular specialty laboratories and will further compute and display various physiological parameters based on the output from one or more electrodes, transducers, or measuring devices. The physician may use the acquired physiological parameters, along with knowledge of patient history, medical expertise, and clinical judgment to determine if thervention is indicated.

    Device Description

    The Ultiri™ Measurement System is a diagnostic computer designed to record, compute, display and store data received wirelessly from the PressureWire™ guidewire and Wi-box. The system's main component is an all-in-one personal computer (AiO-PC) operating Ultreon™ Software for Ultiri™ Measurement System. It is compatible with the PressureWire X Guidewire and Wi-Box AO Transmitter. The information is displayed as graphs as well as numerical values on the screen. Data includes, but is not limited to: systolic, diastolic and mean blood pressure, heart rate, Fractional Flow Reserve (FFR), and Resting Full-cycle Ratio (RFR). The physician may use FFR, Pd/Pa at rest and RFR parameters, along with knowledge of patient history, medical expertise and clinical judgment to determine if therapeutic intervention is indicated.

    The Ultreon™ Software provides guided instructions for the physician to setup a physiology procedure. The display of the user interface can be output to an external video monitor. Additional functions allow the user to import patient lists from the hospital's DICOM (Digital Imaging Communications server) system, export recorded measurement data to DICOM or to an external server location or save it to a USB memory stick. Ultiri™ Measurement System is not intended for patient contact and is a non-sterile device that can be draped for sterile interaction. Ultir™ Measurement System is distributed as a package with the following components: ● All-in-One PC (with Ultreon Software) DC power supply ● AC power cords ● Instructions for use ● . USB flash drive ● Video output adapter(s)

    AI/ML Overview

    The Ultiri™ Measurement System, which includes the Ultreon™ Software, is a diagnostic computer designed to record, compute, display, and store data received wirelessly from the PressureWire™ guidewire and Wi-box. The system measures and displays physiological parameters such as systolic, diastolic, and mean blood pressure, heart rate, Fractional Flow Reserve (FFR), and Resting Full-cycle Ratio (RFR). These parameters help assess the severity of coronary lesions by measuring the pressure drop across them.

    Here's an analysis of the acceptance criteria, study details, and ground truth establishment:

    1. Acceptance Criteria and Reported Device Performance:

    The document does not explicitly present a table of specific numerical acceptance criteria (e.g., sensitivity, specificity, accuracy thresholds) for the device's physiological measurements. Instead, it relies on demonstrating substantial equivalence to predicate devices through design characteristics and non-clinical bench performance testing.

    The "Comparison of Subject to Predicate Device" table on page 5 details the functional equivalences, which serve as the implicit acceptance criteria for technological characteristics.

    Feature / CriteriaPredicate Device (K210458)Predicate Device (K183099)Proposed Device (Ultiri™ Measurement System)Reported Device Performance
    Indications for Use (Physiology)IncludedPhysiology-onlyPhysiology-onlyEquivalent to predicates; uses FFR & RFR.
    Software / GUI / WorkflowUltreon™ SoftwareQuantien™ SoftwareUltreon™ SoftwareSame as primary predicate (OPTIS™ Next System).
    Operating SystemWindows 10Linux SoftwareWindows 10Same as primary predicate System.
    Measurement & Display Features (Physiology)Physiological parametersPhysiological parametersPhysiological parametersSame as secondary predicate and included in primary.
    Physiology Functionality (FFR, RFR, Pa, Pd, Wi-Box)FFR & RFR, Pa from external hemodynamic, Pd receiver Thermo View, Wi-Box transmitterFFR & RFR, Pa from external hemodynamic, Pd receiver Thermo View, Wi-Box transmitterFFR & RFR, Pa from external hemodynamic, Pd receiver Thermo View, Wi-Box transmitterUses same algorithms as predicates, same output receiver.
    Data ReceiverPressureWire X (Wireless & Corded)PressureWire X (Wireless & Corded)PressureWire X (Wireless only)Only wireless PressureWire X is supported.
    Compatible DevicesPressureWire X (Wireless), Wi-BoxPressureWire X (Wireless), Wi-BoxPressureWire X (Wireless), Wi-BoxNo change.
    Safety and EffectivenessAs safe and effective as predicateAs safe and effective as predicateTested and found as safe and effectiveMet design specifications through Abbott quality management system design control verification and validation processes.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size for Test Set: Not specified in the provided documents. The filing states that "non-clinical bench performance testing" was conducted. This typically involves a set of predefined test cases rather than patient data.
    • Data Provenance: Not applicable in the context of clinical data for the 510(k) submission. The testing was non-clinical bench testing. There is no mention of country of origin or whether it was retrospective or prospective, as it was not a clinical study involving patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

    • Number of Experts: Not applicable. Since no clinical testing was performed or relied upon, there was no ground truth established by experts on patient data for the purpose of demonstrating device performance in this 510(k). The "ground truth" for bench testing would be defined by engineering specifications and expected output values in controlled environments, verified by engineers.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set:

    • Adjudication Method: Not applicable. No clinical test set requiring expert adjudication was utilized for this 510(k) submission. Bench testing results would be compared against expected outputs and engineering specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No. The document explicitly states: "No clinical testing was completed, nor relied upon, in support of this Traditional 510(k)." Therefore, no MRMC study, or any clinical study involving human readers and AI assistance, was performed or presented.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Standalone Performance: The core of the device's functionality is a standalone algorithm that computes and displays physiological parameters. The non-clinical bench performance testing would have validated the accuracy of these computations in a standalone manner, generating outputs that are then interpreted by a physician. While not explicitly termed "standalone performance" in the document, the "Measurement & Display Features" and "Physiology Functionality" sections imply that the algorithms for FFR, RFR, and other parameters were validated for their accuracy in processing input data.

    7. The Type of Ground Truth Used:

    • Type of Ground Truth: For the non-clinical bench performance testing, the ground truth would be based on:
      • Engineering specifications and controlled inputs: Known pressure/flow values (simulated or generated) fed into the system.
      • Mathematical accuracy: Verification that the algorithms correctly calculate FFR, RFR, and other parameters based on the known inputs and established physiological models used in the predicate devices.
      • Comparison to predicate device outputs: Direct comparison of the Ultiri™ Measurement System's output to the known outputs or established performance of the legally marketed predicate devices under identical simulated conditions.

    8. The Sample Size for the Training Set:

    • Sample Size for Training Set: Not applicable. This device is a diagnostic computer that computes physiological parameters using established algorithms. It does not appear to employ machine learning or AI that requires a "training set" in the conventional sense for image recognition or predictive modeling. The algorithms are based on scientific principles and, as stated, are the "Same algorithms as used in the predicates."

    9. How the Ground Truth for the Training Set was Established:

    • Ground Truth for Training Set Establishment: Not applicable, as there is no mention of a training set for machine learning/AI. The algorithms themselves, derived from scientific understanding of physiology and validated in predicate devices, form the basis of the device's functionality.
    Ask a Question

    Ask a specific question about this device

    K Number
    K201881

    Validate with FDA (Live)

    Date Cleared
    2020-11-13

    (129 days)

    Product Code
    Regulation Number
    870.1425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K183099, K092105

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CoroFlow is indicated to provide hemodynamic information for use in the diagnosis of patients with cardiovascular diseases.

    CoroFlow is intended for use in catheterization and related cardiovascular specialty laboratories to compute and display various physiological parameters based on the output from one or more measuring devices.

    Device Description

    CoroFlow Cardiovascular system is used to calculate, display and store physiological parameters based on pressure and temperature measurements from Abbott Medical's PressureWire and Wi-box. Calculated parameters include physiological indices to assess coronary lesion severity (FFR, Pd/Pa, RFR) and indices to assess coronary micro-circulation (IMR, CFR). The system also provides novel indices based on the same raw pressure and temperature measurements (IMR_Corr, RRR, Absolute Flow/Resistance, dP/dt, Tau). CoroFlow™ is installed on a personal computer and receives measurement data wirelessly via the CoroHub™ Receiver. Information is displayed on the computer screen which can optionally be slaved to a monitor inside the coronary cathlab. Data can be stored on a local storage unit or transferred to a network location.

    AI/ML Overview

    The provided document does not contain details about specific acceptance criteria and a study demonstrating the device meets those criteria. Instead, it describes the CoroFlow™ Cardiovascular System and asserts its substantial equivalence to predicate devices (QUANTIENT™ Measurement System (K183099) and RadiAnalyzer® Xpress (K092105)) based on non-clinical testing and clinical evaluation of Software as a Medical Device (SaMD) principles, rather than a direct clinical study to prove performance against pre-defined acceptance criteria.

    The document explicitly states: "No clinical study was performed as a part of either the product development or in support of the substantial equivalence of CoroFlow as the intended use/indications for use and technological characteristics are equivalent to the predicate devices."

    Therefore, I cannot provide concrete information for some of your requested points, as a direct clinical study with acceptance criteria and reported performance was not conducted for this specific device clearance.

    However, I can extract information related to the device's technical specifications and the general approach to demonstrating substantial equivalence, which serves as an indirect form of meeting performance expectations established by the predicate devices.

    Here's an analysis based on the information available:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not present a table of acceptance criteria for specific performance metrics of the CoroFlow™ system, nor does it report specific performance outcomes from a study designed to meet such criteria.
    Instead, it highlights that the signal performance specifications (pressure and temperature) of the CoroFlow™ are identical to the predicate devices. This implies that the device is expected to perform at the same level as the predicate devices in these aspects.

    Metric (Implied "Acceptance Criteria" based on Predicate)Reported Device Performance (CoroFlow™)Notes
    Pressure Signal PerformanceIdentical to Predicate
    Range-30 to +300mmHgThe CoroFlow™ device is designed to measure pressure within this range, matching the predicate device.
    Accuracy± 1 mmHg plus ± 1% of readings (-30 to 50mmHg) ± 3% of reading (50 to 300mmHg)The CoroFlow™ device is designed to achieve this accuracy, matching the predicate device.
    Frequency response0 – 25HzThe CoroFlow™ device is designed to have this frequency response, matching the predicate device.
    Temperature Signal PerformanceIdentical to Predicate
    Range15 - 42℃The CoroFlow™ device is designed to measure temperature within this range, matching the predicate device.
    Accuracy0.05℃ or 10% ΔT (whichever greatest)The CoroFlow™ device is designed to achieve this accuracy, matching the predicate device.
    Calculated Indices (FFR, Pd/Pa, CFR, IMR, RFR, RRR, IMR_Corr, Absolute Flow/Resistance, dP/dt, Tau, PB-CFR)Yes (calculates these)The software calculates these indices. For the indices not present in both predicate devices (e.g., IMR_Corr, Absolute Flow, RRR, PB-CFR, Tau), the document references "See clinical testing section below," implying their clinical validity was established through the SaMD clinical evaluation approach.

    2. Sample size used for the test set and the data provenance:

    • Test set sample size: Not applicable for a traditional clinical study with a test set. The substantial equivalence was demonstrated through comparison to predicate devices and a SaMD clinical evaluation. The document does not specify a "test set" in the context of a prospective study with patients.
    • Data provenance: Not explicitly stated as a separate test set. The clinical evaluation followed FDA guidance for Software as a Medical Device (SaMD), which typically relies on existing clinical data (retrospective) or scientific literature to establish clinical association and validation. The document implies a broad clinical evaluation, not a focused test set from a specific country. However, since the device received clearance in the US, the supporting data would have been reviewed by the FDA.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not provided. As no traditional clinical study with a "test set" requiring ground truth establishment by experts was conducted, these details are not relevant to the clearance pathway described.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • This information is not provided. Again, this is typically relevant for studies involving human readers or expert consensus on a test set, which was not the primary method for this clearance.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The device, CoroFlow™, is a "Programmable Diagnostic Computer" for calculating and displaying physiological parameters, not an AI-assisted diagnostic imaging interpretation tool in the context of MRMC studies. It assists human clinicians by computing parameters from measurement devices, rather than directly improving human reader performance in interpreting images or clinical data.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, implicitly. The "clinical evaluation" described for this SaMD involved establishing "clinical association, analytical validation and clinical validation of the SaMD" as per FDA guidance. This typically involves demonstrating that the algorithm (CoroFlow™ calculating parameters from raw pressure/temperature data) accurately computes these parameters correctly and that these parameters are clinically meaningful, similar to how predicate devices compute and display them. The device's "raison d'être" is the standalone calculation and display of these physiological parameters.

    7. The type of ground truth used:

    • The document states that the substantial equivalence was evaluated by following FDA guidance for Software as a Medical Device (SaMD): "the performed clinical evaluation showed a valid clinical association, analytical validation and clinical validation of the SaMD."
      • Analytical Validation: Likely involved comparing the device's computed parameters against established reference methods or highly accurate physical models, ensuring the calculations are correct.
      • Clinical Validation: Involved demonstrating that the calculated parameters (e.g., FFR, CFR, IMR) are clinically meaningful and correlate with patient outcomes or established diagnostic thresholds as widely accepted in cardiology. This would rely on existing clinical data and scientific literature that supports the utility of these physiological indices, rather than generating new ground truth from a dedicated prospective study.
      • For indices not present in the primary predicate (e.g., IMR_Corr, Absolute Flow, RRR, PB-CFR, Tau), the "clinical testing section below" (which refers back to the general SaMD clinical evaluation statement) would have addressed their clinical validation by leveraging scientific literature and existing clinical evidence for these emerging parameters.

    8. The sample size for the training set:

    • This information is not provided. The CoroFlow™ device is described as a software for calculating physiological parameters based on direct measurements, implying rule-based algorithms rather than a machine learning model that requires a "training set" in the conventional sense of AI/ML. The analytical and clinical validation would be sufficient for such a device.

    9. How the ground truth for the training set was established:

    • Not applicable, as there's no explicitly mentioned "training set" in the context of machine learning. The "ground truth" for the device's functionality would be the scientifically established physiological formulas for the calculated indices and the accurate measurements from the input sensing devices (Abbott PressureWire and Wi-box).