510(k) Data Aggregation

    K Number
    K220360

    Validate with FDA (Live)

    Date Cleared
    2022-11-10

    (275 days)

    Product Code
    Regulation Number
    872.4760
    Panel
    Dental
    Age Range
    18 - 65
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K182758, K171904, K953806

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tandry CMF System is a Maxillofacial plate and screw system intended for osteotomy, stabilization and rigid fixation of maxillofacial/midface fractures and reconstruction including infraorbital area and maxillofacial, and mandibular and maxillary correction surgery.

    Each plate is indicated to be used in the following anatomic regions:

    • · 1.2mm upper-face plates Orbital floor region, zygomatic region, and dento-alveolar region
    • · 1.5mm mid-face plates Zygomatic region, orbital floor region, and maxilla
    • · 2.0 mm mini plates Orthognathics midface/mandible
    • · 2.0/2.3mm mandible plates Mandibular
    Device Description

    The Tandry CMF Plate System consists of various sized plates and screws. The plates and screws are fabricated from pure titanium and Ti-6Al-4V. The plates are designed to distribute for respective anatomical and indicated areas, besides, they can be used with self-tapping, self-drilling, emergency self-tapping and locking screws.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from Microware Precision Co., Ltd. for their Tandry CMF Plate System. It describes the device, its intended use, and compares it to predicate devices to demonstrate substantial equivalence.

    Based on the provided text, here's an analysis of the acceptance criteria and study information:

    Key Takeaway: The submission explicitly states that clinical studies were NOT required to support substantial equivalence for this device. Therefore, information related to clinical performance, human reader studies, and specific AI acceptance criteria (which would typically be evaluated in such studies) is not present in this document. The "study" here refers to non-clinical tests.


    1. A table of acceptance criteria and the reported device performance

    The document does not present a formal table of acceptance criteria in the sense of a numerical performance target (e.g., sensitivity, specificity, accuracy) combined with reported device performance from a clinical study. This is because, as stated, clinical studies were not required.

    Instead, the acceptance criteria are implicitly met by demonstrating substantial equivalence to predicate devices through non-clinical testing (biomechanical tests, biocompatibility, reprocessing, and sterilization).

    Here's a breakdown of the non-clinical test results:

    Acceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance (Non-Clinical)
    Biomechanical Performance- ASTM F543 and F382 (bone plate and screw performance) were performed.- The results support substantial equivalence to predicate devices.
    Biocompatibility- Sample preparation per ISO 10993-12.- Biological Safety Assessment guided by ISO 10993-1.- Cytotoxicity testing per ISO 10993-5.- The previous testing results for the predicate device (Tandry Locking Plate System K171904) are utilized, asserting the proposed device meets validation requirements.
    Reprocessing and Sterilization- Previous testing results for the predicate device (Tandry Locking Plate System K171904) are utilized, as the material, manufacturing process, geometry, physico-chemical, body/fluid contact, and sterilization process are the same.- Validated steam sterilization instructions to a sterility assurance level (SAL) of 10-6 using the biological indicator overkill method (in accordance with ANSI/AAMI/ISO 17665-1 and ANSI/AAMI/ISO 14937).
    Design Characteristics- Dimensions, shape, and sizes are similar to predicate devices (as detailed in the comparative table on page 6).
    Material Composition- Fabricated from pure titanium (ASTM F67) and Ti-6Al-4V (ASTM F136), which are standard materials used in predicate devices.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    This information is not applicable as no clinical test set or human data was used for a performance study. The "test set" for this submission consists of device samples evaluated in non-clinical lab settings (biomechanical and biocompatibility tests). No specific sample sizes for these lab tests are provided in the summary, other than "representative samples."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not applicable as no clinical study with expert interpretation was conducted.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not applicable as no clinical study with human readers requiring adjudication was conducted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done or required. The device is a traditional medical implant (plate and screw system), not an AI-powered diagnostic or assistive technology.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable as the device is not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical tests, the "ground truth" is defined by established engineering standards and biological safety guidelines (e.g., ASTM standards for biomechanics, ISO series for biocompatibility and sterilization). There is no clinical "ground truth" (like pathology or outcomes data) established or reported because no clinical study was performed.

    8. The sample size for the training set

    This information is not applicable as the device is not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable as the device is not an AI/ML algorithm that requires a training set.


    In summary, the document details a 510(k) submission for a traditional mechanical medical device, where substantial equivalence is demonstrated through non-clinical bench testing and comparison to predicate devices, rather than through clinical trials or AI performance evaluations.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212391

    Validate with FDA (Live)

    Date Cleared
    2022-10-21

    (445 days)

    Product Code
    Regulation Number
    882.5320
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K971297, K161821, K141452, K121624, K182758, K944565

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MCI - Neuro Fixation System is indicated for use in selective trauma of the cranial skeleton, cranial surgery and reconstructive procedure.

    Device Description

    The MCI – Neuro Fixation System is composed of plates (including burr hole covers), screws, and meshes. The bone plates and meshes are made from commercially pure titanium (ASTM F67) and the bone screws are manufactured from titanium alloy - Ti-6Al-4V (ASTM F136) and are available in different sizes and shapes, according to the site of the implantation and the extension of the fracture. The surface of plates, meshes and screws are colored-anodized.

    MCI - Neuro Fixation System devices are for single use. The devices are provided non-sterile and must be properly sterilized before use, according to the recommendations provided in the Instructions for Use.

    The devices must only be used by qualified surgeons mastering the surgical technique, having been trained and qualified in cranial surgery.

    AI/ML Overview

    The provided text describes information about the MCI-Neuro Fixation System, a cranial fixation device. However, it does not contain the kind of detailed information about acceptance criteria, clinical study design, sample sizes, expert qualifications, or comparative effectiveness studies that your request asks for, especially concerning an AI/ML-driven device.

    The document discusses:

    • Regulatory Clearance (K212391): It's a 510(k) premarket notification for a Class II medical device.
    • Device Description: The system consists of plates, screws, and meshes made from titanium.
    • Predicate Device: K141911 - OPTIMUS NEURO SYSTEM.
    • Performance Data: This section focuses on bench testing for physical and material properties (metallographic tests, cyclic polarization, cycle bend testing of plates, screw torsion yield test, pull-out testing, driving torque testing).
    • Biocompatibility: Stated as supported by a reference device (K182758).
    • Lack of Clinical Data: Explicitly states, "No clinical data were included in this submission."

    Since this device is a physical fixation system and not an AI/ML driven diagnostic or prognostic tool, the concepts of human readers, AI assistance, ground truth establishment (beyond material specifications), training sets, and test sets in the context of diagnostic accuracy are not applicable to this document.

    Therefore, I cannot provide a table of acceptance criteria and device performance, or details about clinical study design, sample sizes, expert qualifications, or adjudication methods as requested, because this information is not present in the provided text. The document focuses on demonstrating substantial equivalence through material and mechanical bench testing.

    Ask a Question

    Ask a specific question about this device