510(k) Data Aggregation

    K Number
    K231231

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-10-12

    (167 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K203011, K222194, K182307, K210869, K171715, K210390

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Monster® Screw System is indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, ligament fixation, fracture repair and fracture fixation, appropriate for the size of the device. Specific examples include: Fractures and Osteotomies, Hallux Valgus Correction, Arthrodesis/Deformity Correction, Fusion resulting from neuropathic osteoarthropathy (Charcot).

    The Baby Gorilla® / Gorilla® Bone Plates and Bone Screws of the Baby Gorilla® and Gorilla® Plating System are indicated for use in stabilization and fixation of fractures or osteotomies; intra and extra articular fractures, joint depression, and multi-fragmentary fractures; revision procedures, joint fusion and reconstruction of small bones of the toes, feet and ankles including the distal tibia, talus, and calcaneus, as well as the fingers, hands, and wrists. The system can be used in both adult and pediatric patients. Specific examples include: Forefoot, Mid/Hindfoot, Ankle, First metatarsal osteotomies for hallux valgus correction, Arthrodesis of the first metatarsophalangeal joint (MTP), Flatfoot, Charcot. In addition, the non-locking titanium screws and washers are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair and fracture fixation, appropriate for the size of the device.

    The Phantom® Small Bone Intramedullary Nail System is indicated for use in stabilization and fixation of the small bones of the feet and ankle for the treatment of fractures, osteotomies, nonunions, pseudarthroses and malunions by revision, joint fusion or reconstruction procedures.

    The Phantom® Hindfoot TTC/TC Nail System is indicated for tibiotalocalcaneal arthrodesis (fusion) and to provide stabilization of the hindfoot and ankle including the transverse tarsal joints coupling the mid-foot to the hindfoot. Examples of specific indications include: Post-traumatic or degenerative arthritis, Previously infected arthrosis, Revision of failed ankle arthrodesis, Revision of failed total ankle arthroplasty, Talar deficiency conditions such as avascular necrosis of the talus (requiring tibiocalcaneal arthrodesis), Neuromuscular deformity or other neuromuscular disease with severe deformity or instability of the ankle, Rheumatoid arthritis, Osteoarthritis, Nonunions or pseudarthrosis of hindfoot and distal tibia, Trauma (severe or malunited tibial pilon fracture), Charcot foot (neuroarthropathy), Severe end-stage degenerative arthritis, Instability and skeletal defects after tumor resection, Pantalar arthrodesis, Severe foot/ankle deformity.

    HammerTube™M System is indicated for fixation of The reconstruction and fusion of toes during correction procedures for hammertoe deformity, claw toe deformity, shortening osteotomies of the phalanges and mallet toe deformity as well as revision hammertoe procedures. The cannulated and solid HammerTube™ implants may be used without any other additional device. The cannulated implants may be used with K-Wires for delivery of implants or for the temporary stabilization of nearby joints, such as the metatarsophalangeal joint. The implantable K-Wires are indicated for use in the stabilization and fixation of small bones for use in bone reconstruction, osteotomy, arthrodesis, fracture repair and fixation, appropriate for the size of the joint. Additionally, the implantable K-Wires are indicated as guide pines for insertion of instruments and implants in the HammerTube™ System.

    The APEX 3D™ Total Ankle Replacement System is indicated as a total ankle replacement in primary surgery for patients with ankle joints damaged by severe rheumatoid, post-traumatic, or degenerative arthritis. Revision surgery for these patients is also indicated for patients with sufficient bone stock present. Components are intended for cemented use only.

    Device Description

    Monster® Screw System: The Monster® Screw System is comprised of threaded bone screws (Ti Alloy or Stainless Steel) offered in 2.0mm to 9.5mm diameters (in 0.5mm increments). In addition, a 2.7mm diameter is also part of the system. The overall screw length ranges from 8mm (for smaller diameters) thru 200mm (for larger diameters). The screws are available in a variety of designs including fully or partially threaded self-drilling or blunt, cannulated or solid, and headed or headless. Sized-matched washers are also available.

    Baby Gorilla® / Gorilla® Plating System: The Baby Gorilla® / Gorilla® Plating System is comprised of bone plates, threaded bone screws, and washers. Gorilla® Plates are offered in "mini" and "standard" set sizes in a variety of shapes based upon the anatomical fixation required. Screws are also offered in "mini" and "standard" sets and, in addition, in locking and non-locking versions. Size-matched washers are available for use with the non0locking screws when the latter are used for fixation without the plates. Size-matched plate washers are also available for use with plate holes when there is no desire to use a screw. The Baby Gorilla® / Gorilla® Plating System implants are manufactured from medical grade titanium (per ASTM F67), stainless steel (per ASTM F138), and titanium alloy (per ASTM F136).

    Phantom® Small Bone Intramedullarv Nail System: The Phantom® Small Bone Intramedullary Nail System is comprised of small bone intramedullary nails, locking screws and threaded pegs. The Phantom® nails are offered in a variety of lengths to accommodate variations in patient anatomy. The Phantom® threaded pegs and locking screws insert through the intramedullary nail to secure the construct. These are offered in varving lengths to accommodate the anatomical fixation required. The system includes instruments for implantation.

    Phantom® Hindfoot TTC/TC Nail System: The Phantom® Hindfoot TTC/TC Nail System is comprised of intramedullary nails, screws and accessory components. The Phantom® nails are offered in a variety of sizes lengths, and configurations to accommodate variations in patient anatomy. The Phantom® screws insert through the intramedullary nail to secure the construct. These are offered in varying lengths to accommodate the anatomical fixation required.

    HammerTube™ System: The HammerTube™ System includes HammerTube™ and K-Wire implants. The HammerTube™ is a commercially pure titanium plasma sprayed PEEK cylinder in varying diameters, lengths, and angles. The K-Wires are stainless steel and double trocar pointed. The implants are sold sterile.

    APEX 3D Total Ankle Replacement System: The APEX 3D™ Total Ankle Replacement System is a cemented, fixed bearing device comprised of a tibial component, a talar component and an Ultra-High Molecular Weight Polyethylene (UHMWPE) component used for ankle joint replacement. Components are available in varying sizes and design configurations intended for both primary and revision applications.

    AI/ML Overview

    This FDA 510(k) summary (K231231) describes the MR Safety and Compatibility Testing and Labeling Updates for several Paragon 28 Orthopedic Fixation Devices. The document states that the purpose of this submission is to obtain clearance for Magnetic Resonance (MR) safety and compatibility testing and Magnetic Resonance Imaging (MRI) Safety Information updates to the labeling of previously cleared devices. It explicitly mentions that there are no other indications for use, design, material, processing, performance, or labeling modifications subject to the submission. This is crucial because it means the study described focuses solely on MR safety, not on clinical effectiveness of the orthopedic fixation devices themselves.

    Therefore, the study focuses on ensuring the devices meet MR Conditional labeling requirements based on established standards, rather than evaluating their clinical performance in human patients.

    Here's a breakdown of the requested information based on the provided text, with the caveat that the study is not a clinical effectiveness study of the orthopedic devices but rather a safety study for MRI compatibility:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategoryStandard/Test MethodDevice Performance (General Statement in document)
    Image ArtifactASTM F2119 Test Method for Evaluation of MR Image Artifacts from Passive ImplantsPerformance presented in the submission as recommended in the FDA Guidance. Based on results, each device will be labeled as "MR Conditional."
    Magnetically Induced Displacement ForceASTM F2052 Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance EnvironmentPerformance presented in the submission as recommended in the FDA Guidance. Based on results, each device will be labeled as "MR Conditional."
    Magnetically Induced TorqueASTM F2213 Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance EnvironmentPerformance presented in the submission as recommended in the FDA Guidance. Based on results, each device will be labeled as "MR Conditional."
    Radiofrequency (RF) Induced HeatingASTM F2182 Test Method for Measurement of Radio Frequency Induced Heating On or Near Passive Implants during Magnetic Resonance ImagingPerformance presented in the submission as recommended in the FDA Guidance. Based on results, each device will be labeled as "MR Conditional" with MRI Safety Information in the instructions for use as described in ASTM F2503.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not explicitly state the specific "sample size" of devices tested for MR safety. The testing was performed on the described orthopedic fixation devices (Monster® Screw System, Baby Gorilla® / Gorilla® Plating System, Phantom® Small Bone Intramedullary Nail System, Phantom® Hindfoot TTC/TC Nail System, HammerTube™ System, APEX 3D™ Total Ankle Replacement System).

    Data provenance is not directly about country of origin in this context; it's about the nature of the testing. The testing was prospective in relation to the submission, performed specifically to generate the data for the MR safety labeling update. The tests are in vitro (laboratory-based) rather than in vivo (on human subjects).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This question is not applicable to the described study. The "ground truth" here is compliance with established engineering standards for MR safety, not a medical diagnosis or outcome requiring expert medical consensus. The testing involved standardized methods (ASTM standards) and engineering evaluations, not interpretation by medical experts for a test set of patient data.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This question is not applicable. Adjudication methods like 2+1 or 3+1 typically refer to the process of resolving discrepancies among multiple human readers (e.g., radiologists) in a clinical diagnostic study. This study is an engineering safety performance study of medical devices, not a diagnostic study.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This study is not about AI assistance or human reader performance. It is a safety study evaluating the compatibility of orthopedic implants with MRI environments.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No, this question is not applicable. This study involves the physical testing of orthopedic implants in an MRI environment to determine their safety and compatibility characteristics, not the performance of an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this study is compliance with established, recognized engineering standards for MR safety as defined by ASTM (American Society for Testing and Materials) and the FDA's guidance document "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment." These standards provide quantitative metrics and methodologies for evaluating magnetic force, torque, heating, and image artifact.

    8. The sample size for the training set

    This question is not applicable. The described study is not a machine learning or AI study, so there is no concept of a "training set."

    9. How the ground truth for the training set was established

    This question is not applicable, as there is no training set in this type of device safety study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223184

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-12-16

    (65 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K131722, K182307

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Phantom Metatarsal Shortening System is indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair and fracture fixation, appropriate for the size of the device. Specific examples include:
    Metatarsal and phalangeal osteotomies
    Metatarsal deformity correction
    Hammertoe
    Revision hammertoe
    Claw toe
    Mallet toe
    Proximal Interphalangeal Joint Arthrodesis
    Distal Interphalangeal Joint Arthrodesis

    Device Description

    The Phantom Metatarsal Shortening System includes a series of titanium (Ti-6Al-4V ELI per ASTM F136) intramedullary implants used for the correction of small bones in the foot. The implants are designed to provide stability and fixation of bone fragments to ultimately achieve fusion.

    AI/ML Overview

    The provided document is a 510(k) summary for the Phantom Metatarsal Shortening System. It describes the device, its intended use, and substantial equivalence to predicate devices, supported by performance testing. However, it explicitly states that clinical data are not needed to support the safety and effectiveness of the subject device. Therefore, the document does not contain information about a study involving human readers or ground truth determination as described in your request.

    The "device" in this context is a medical implant (titanium intramedullary implants), not an AI algorithm or diagnostic tool. The performance testing refers to mechanical and material tests.

    Here's a breakdown of the information that can be extracted or inferred from the provided text, and where the requested AI/human study details are missing:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Mechanical Performance (Implied from testing types)All performance testing conducted for the Phantom Metatarsal Shortening System met the predetermined acceptance criteria.
    Torque to failure per ASTM F543Met acceptance criteria.
    Insertion and removal torque per ASTM F543Met acceptance criteria.
    Static pullout of threads per ASTM F543Met acceptance criteria.
    Static pullout of the prongsMet acceptance criteria.
    Static cantilever bendingMet acceptance criteria.
    Dynamic cantilever bending testingMet acceptance criteria.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size: Not specified. The document only mentions "All necessary testing has been performed on representative Phantom Metatarsal Shortening System components."
    • Data provenance: Not applicable. This was mechanical/material testing of physical devices, not clinical data from patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. Ground truth in the context of expert review for AI performance is not relevant to this type of device submission. The "ground truth" for this device would be the physical properties and performance characteristics measured against established ASTM standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. Adjudication methods are used in clinical studies with human readers, which were not part of this submission. The "adjudication" (if one could even call it that) of the mechanical tests would be the measurement against the ASTM standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. An MRMC comparative effectiveness study was not done. This is a submission for a physical medical implant, not an AI-powered diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No. A standalone performance study of an algorithm was not done. This device is a physical implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The "ground truth" for this device's performance is established by validated mechanical testing standards (e.g., ASTM F543) and the inherent material properties of titanium (Ti-6Al-4V ELI per ASTM F136). The acceptance criteria are derived from these engineering and material standards.

    8. The sample size for the training set

    • Not applicable. There is no "training set" as this is a physical medical device, not an AI algorithm.

    9. How the ground truth for the training set was established

    • Not applicable. There is no "training set" or corresponding ground truth establishment process for a physical implant. The design and manufacturing process would be subject to quality control and engineering principles.
    Ask a Question

    Ask a specific question about this device

    K Number
    K192163

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-01-31

    (175 days)

    Product Code
    Regulation Number
    888.3020
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K182307, K171376, K120419

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TTC Phantom® Intramedullary Nail system is intended for tibiotalocalcaneal arthrodesis (fusion) and to provide stabilization of the hindfoot and ankle including the transverse tarsal joints coupling the mid-foot to the hindfoot. Examples of specific indications include:

    • · Post-traumatic or degenerative arthritis
    • · Previously infected arthrosis
    • Revision of failed ankle arthrodesis
    • · Revision of failed total ankle arthroplasty
    • · Talar deficiency conditions such as avascular necrosis of the talus (requiring tibiocalcaneal arthrodesis)
    • · Neuromuscular deformity or other neuromuscular disease with severe deformity or instability of the ankle
    • · Rheumatoid arthritis
    • · Osteoarthritis
    • Nonunions or pseudarthrosis of hindfoot and distal tibia
    • · Trauma (severe or malunited tibial pilon fracture)
    • · Charcot foot (neuroarthropathy)
    • · Severe end-stage degenerative arthritis
    • · Instability and skeletal defects after tumor resection
    • · Pantalar arthrodesis
    • · Severe foot/ankle deformity
    Device Description

    The Paragon 28® TTC Phantom® Intramedullary Nail System is comprised of intramedullary nails, screws and accessory components. The Phantom® nails are offered in a variety of sizes lengths, and configurations to accommodate variations in patient anatomy. The Phantom® screws insert through the intramedullary nail to secure the construct. These are offered in varying lengths to accommodate the anatomical fixation required.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the TTC Phantom® Intramedullary Nail System, which is a medical device for tibiotalocalcaneal arthrodesis. It does NOT describe an AI/ML powered device, and therefore the requested information regarding acceptance criteria, study design for AI/ML performance, ground truth establishment, sample sizes for training/test sets, expert involvement, and reader studies is not applicable to this submission.

    The document focuses on demonstrating substantial equivalence to predicate devices through:

    1. Indications for Use: Listing the specific conditions for which the device is intended.
    2. Device Description: Detailing the components of the system.
    3. Preclinical Testing: Describing mechanical testing (static and dynamic per ASTM F1264-14), wear testing, and biocompatibility evaluation (per ISO 10993-1 and CDRH's 2016 Biocompatibility Guidance). Additional testing mentioned includes bacterial endotoxins and sterilization.
    4. Conclusion: Stating that the device is substantially equivalent to predicates based on intended use, technological characteristics, and preclinical testing.

    Therefore, I cannot provide details on AI/ML acceptance criteria or studies from the provided text.