Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K202957

    Validate with FDA (Live)

    Date Cleared
    2020-10-29

    (29 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K182186

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Flexible Video Cysto-Urethroscope C-VIEW is used to provide visualization and operative access during diagnostic and therapeutic endoscopic procedures of urinary tract including the urethra, bladder, ureters, and kidneys.

    Device Description

    The Flexible Video Cysto-Urethroscope C-VIEW (Part Number: 11272VUE) is intended to be used with multiple compatible CCUs: C-MAC (8403ZX, cleared via K182186) and C-HUB II (20290301, cleared via K182186). Identical to the predicates, the scope cannot be operated on its own because it is a videoscope whose image data output is provided in the form of video signals and sent to the CCU for decoding and display. Therefore, when the scope is connected to the compatible CCUs, it becomes the Flexible Video Cysto-Urethroscope C-VIEW System, which provides visualization and operative access in the urinary tract.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the Flexible Video Cysto-Urethroscope (C-view). This type of submission aims to demonstrate substantial equivalence to a legally marketed predicate device, rather than proving efficacy through extensive clinical trials as would be required for a PMA. As such, the information typically provided to meet acceptance criteria, specifically for software-based medical devices or AI/ML-driven devices, is not present here.

    Based on the provided document, the "device" in question is a hardware endoscope, not an AI/ML-driven device or a device that typically has "acceptance criteria" for software performance. The document focuses on demonstrating substantial equivalence to predicate devices through technical specifications, physical characteristics, and compliance with recognized standards.

    Therefore, many of the requested items (e.g., sample size for test/training sets, experts for ground truth, MRMC study, standalone algorithm performance, ground truth type) are not applicable to this submission as they pertain to the evaluation of AI/ML software or diagnostic image processing, which is not the primary function of this endoscope.

    However, I can extract and interpret the available information relevant to device performance and the study conducted.

    Here's an analysis of the provided information in relation to your request:

    Acceptance Criteria and Device Performance (Interpreted)

    Since this is a hardware device submission focused on substantial equivalence, the "acceptance criteria" are implicitly defined by the technical specifications and performance characteristics compared to the predicate devices. The "study" proving it meets these involves non-clinical bench testing and adherence to recognized standards.

    Table 1: Acceptance Criteria (Implied) and Reported Device Performance

    Feature/MetricAcceptance Criteria (Implied by Predicate/Standards)Reported Device Performance (Subject Device)
    Device TypeFlexible Video Cysto-UrethroscopeFlexible Video Cysto-Urethroscope
    Insertion Shaft DiameterSimilar to predicates (5.5 mm)5.2 mm
    Insertion Shaft LengthSame as primary predicate (37 cm)37 cm
    Working Channel DiameterSimilar to primary predicate (2.3 mm)2.3 mm
    Suction Port/ChannelPresent (via working channel)Via working channel
    Deflection (°) - UpSimilar to predicates (210°)210°
    Deflection (°) - DownSimilar to predicates (140°)140°
    Type of ImagerCMOSCMOS
    Field of ViewSimilar to predicates (100°)100°
    Direction of ViewSimilar to predicates (0°)
    Depth of FieldSimilar to predicates (5-50 mm)5-50 mm
    On-axis Resolution (minimal)Comparable to predicates (e.g., 2.5 lp/mm @ 3mm, 40 lp/mm @ 50mm for primary predicate)1.8 lp/mm @ 5 mm, 16 lp/mm @ 50 mm
    Light SourceInternal LEDInternal LED
    Compatible CCUsAbility to connect to compatible CCUsC-MAC (8403ZX), C-HUB II (P/N20290301)
    Electrical SafetyCompliance with IEC 60601-1, IEC 60601-1-2, etc.Compliant
    EMCCompliance with IEC 60601-1-2Compliant
    BiocompatibilityCompliance with ISO 10993Compliant
    Performance (Optical)Color Reproduction & Color Contrast, Illumination Detection Uniformity, Spatial Resolution & Depth of Field, Latency, Distortion, Signal-to-Noise Ratio (SNR) & SensitivityTested and met design specifications
    Reprocessing ValidationCompliance with FDA Guidance DocumentCompliant

    Note on "Acceptance Criteria" for this specific device: For a 510(k) of a hardware device like an endoscope, acceptance criteria are generally not explicitly stated as performance thresholds in the same way they would be for an AI/ML algorithm (e.g., "sensitivity must be >X%"). Instead, the acceptance criteria are implicitly met by demonstrating that the device is as safe and effective as the predicate devices and meets applicable consensus standards. The "study" mainly involves engineering tests and comparisons.

    Study Details

    1. Sample Size Used for the Test Set and Data Provenance:

      • Not applicable in the context of a typical "test set" for an AI/ML algorithm. The "test set" for this hardware device primarily refers to the physical units subjected to bench testing and measurements. No specific sample size (e.g., N=X images or patient cases) is provided as it's not a software performance study.
      • Data Provenance: The "data" comes from engineering measurements and tests performed on the device itself, not from patient studies or image datasets.
    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

      • Not applicable. "Ground truth" in the sense of expert annotation for images or clinical outcomes is not relevant here. The "ground truth" for the device's physical and optical properties is established by objective engineering measurements and validated against industry standards and the predicate device's specifications.
    3. Adjudication Method for the Test Set:

      • Not applicable. Adjudication methods (e.g., 2+1, 3+1) are used for resolving disagreements among human readers/annotators in clinical studies, particularly for creating ground truth for AI algorithms. This is a hardware validation.
    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. An MRMC study is a clinical study design used to evaluate the diagnostic performance of a system (often AI-assisted) by comparing multiple readers' interpretations of cases. This device is an endoscope providing visualization; it's not an AI-based diagnostic tool. Therefore, an MRMC study was not performed, nor is it relevant to demonstrate substantial equivalence for this type of device.
    5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

      • Not applicable. This is a hardware device (an endoscope). There is no "algorithm only" performance to be evaluated. Its function is to provide real-time visualization for a human operator.
    6. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

      • For this hardware device, the "ground truth" for its performance characteristics (e.g., resolution, field of view, deflection angles) is established by objective engineering measurements and adherence to recognized industry standards (e.g., IEC, ISO standards for electrical safety, biocompatibility, optical performance). The comparison to the predicate devices also serves as a benchmark for "ground truth" in terms of what is "safe and effective."
    7. The Sample Size for the Training Set:

      • Not applicable. There is no AI/ML algorithm training involved with this hardware device.
    8. How the Ground Truth for the Training Set Was Established:

      • Not applicable. As there is no training set, this question is not relevant.

    In summary: The provided document is a 510(k) submission for a conventional medical device (an endoscope). The "acceptance criteria" and "study" described are focused on engineering performance, safety, and substantial equivalence to existing predicate devices, rather than the rigorous statistical evaluation and ground truth establishment typically required for AI/ML-driven diagnostic software.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1