510(k) Data Aggregation

    K Number
    K213171

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-05-26

    (240 days)

    Product Code
    Regulation Number
    884.1690
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K181909

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AVETA SYSTEM:
    Hysteroscopy:
    The Aveta System is intended for intrauterine use by trained gynecologists to permit viewing of the cervical canal and the uterine cavity, provide liquid distension of the uterus and monitor the volume differential between the irrigation fluid flowing into and out of the uterus during diagnostic and surgical procedures to resect and remove tissue such as submucous myomas, endometrial polyps and retained products of conception.
    Cystoscopy:
    The Aveta System is intended for use in endoscopic access to and examination of the lower urinary tract, including the bladder. When combined with accessory instruments, the system allows the user to perform various diagnostic and therapeutic procedures.

    AVETA DISPOSABLE HYSTEROSCOPE (Pearl/Opal/Coral):
    The Aveta Disposable Hysteroscope (Pearl/Opal/Coral) is intended to permit viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures.

    AVETA DISPOSABLE CYSTOSCOPE (Coral):
    The Aveta Disposable Cystoscope (Coral) is intended for use in endoscopic access to and examination of the lower urinary tract, including the bladder. When combined with accessory instruments, the Cystoscope allows the user to perform various diagnostic and therapeutic procedures.

    Device Description

    The Aveta System is an integrated system which allows for visualization of the intended cavity for the purpose of performing diagnostic and operative procedures (hysteroscopy and cystoscopy). The Aveta System consists of the components listed in Table 1. The system includes a Controller with integrated fluid management which incorporates a dual peristaltic pump design to control the continuous inflow and outflow of saline to provide fluid distention of the cavity. The Controller provides continuous monitoring of the cavity pressure to the set pressure. For hysteroscopy, it also monitors the volume differential between saline inflow from the uterus (fluid deficit). The Controller connects to a sterile, single use disposable Scope (available in various configurations, see below) that allows visualization of the cavity and displays the images obtained from the Scope on a standard monitor. For operative hysteroscopy procedures, the Aveta System includes sterile, mechanical Disposable Resecting Device (available in various configurations, see below) powered by an integrated motor in the device handset or by use of an external Resecting Handset (available in two configurations, as a disposable device or a reusable device). The resecting device is inserted through the working channel of the sterile hysteroscope to resect the target tissue/pathology. For cystoscopy, when combined with accessory instruments the cystoscope is used for diagnostic and therapeutic procedures.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for the Meditrina Aveta System. This type of submission aims to demonstrate that a new device is substantially equivalent to a legally marketed predicate device, rather than proving its safety and effectiveness from scratch in the same way a PMA (Premarket Approval) would. As such, the information provided focuses on comparative performance and engineering tests, rather than clinical efficacy studies with specific acceptance criteria related to disease detection or treatment outcomes.

    Therefore, the document does not contain the information requested in points 1, 2, 3, 4, 5, 6, 7, 8, and 9 as it pertains to acceptance criteria for algorithmic performance, sample sizes for test sets, expert-established ground truth, MRMC studies, or multi-reader studies. The "performance data" section (Section VII) lists various engineering and bench tests, but these are not for the purpose of demonstrating diagnostic or clinical performance in a "study" as implied by the request.

    The "study that proves the device meets the acceptance criteria" in this context refers to the bench and engineering tests conducted to demonstrate that the modified Aveta System performs equivalently to its predicate and meets safety standards.

    Here's an analysis of the provided text in relation to your request:

    1. A table of acceptance criteria and the reported device performance:
    The document does not provide a table of acceptance criteria and reported device performance in the sense of accuracy metrics (e.g., sensitivity, specificity) for a diagnostic AI/ML device. The "performance data" section lists types of tests performed (e.g., "Pressure accuracy and control," "Optical testing," "Motor/Oscillation speed"), but it does not specify the numerical acceptance criteria for these tests nor the exact quantitative results achieved by the device against those criteria. It only states that the tests were performed "per approved test protocols."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
    No information about sample sizes for a clinical "test set" or data provenance (country, retrospective/prospective) is provided. The performance data listed (Section VII) refers to engineering and bench tests on the device's components and system function (e.g., integrity, dimensional inspection, functional testing). These are not data sets for an AI/ML algorithm.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
    This information is not provided. The study is a 510(k) submission, focused on engineering and functional equivalence rather than clinical diagnostic performance that would require expert-established ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    Not applicable/not provided. No clinical test set requiring adjudication is mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    No MRMC study or any study involving human readers with or without AI assistance is mentioned. The device described appears to be an endoscopic system, not an AI diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    Not applicable/not provided. The device is a physical system for visualization and surgical procedures, not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
    Not applicable/not provided. The "ground truth" in this context would be the design specifications and functional requirements met by the engineering tests, not a clinical ground truth for diagnostic accuracy.

    8. The sample size for the training set:
    Not applicable/not provided. This device is not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established:
    Not applicable/not provided. This device is not an AI/ML algorithm that requires a training set with established ground truth.

    In summary, the provided document details a 510(k) submission for a medical device (Aveta System, a hysteroscope/cystoscope system). It focuses on demonstrating substantial equivalence to a predicate device through engineering and bench testing, rather than reporting on the performance of an AI/ML model for diagnostic purposes as implied by the structure of your questions.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210512

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-06-30

    (127 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K181909, K190827, K19227

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Hysteroscopy: The LUMINELLE DTx System is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures. NOTE: Hysteroscopes are used as tools to access the uterine cavity and are not, in and of themselves, a method of surgery. When LUMINELLE DTx System is specifically used with the LUMINELLE Dx 360° Rotatable Disposable Sheath (Diagnostic), the system is limited to performing diagnostic procedures only. When the LUMINELLE DTx System is used with the LUMINELLE Bx 360° Rotatable Disposable Sheath (Biopsy), the system is limited to performing diagnostic procedures only. The Bx Sheath can be used for endometrial or lower urinary tract (including bladder) biopsy by aspiration with a syringe during diagnostic procedures. Generally recognized indications for diagnostic hysteroscopy include: Abnormal bleeding, Infertility and pregnancy wastage, Evaluation of abnormal hysterosalpingogram, Intrauterine foreign body, Amenorrhea, Pelvic pain. Generally recognized indications for use for operative hysteroscopy include: Directed endometrial biopsy, Polypectomy, Submucous myomectomy, Transection of intrauterine adhesions, Transection of intrauterine septa, Endometrial ablation. Cystoscopy: The LUMINELLE DTx System is intended for use in endoscopic access to and examination of the lower urinary tract, including the bladder. When combined with accessory instruments, the System allows the user to perform various diagnostic and therapeutic procedures. NOTE: When LUMINELLE DTx System is specifically used with the LUMINELLE Dx 360° Rotatable Disposable Sheath (Diagnostic), the system is limited to performing diagnostic procedures only. NOTE: When the LUMINELLE DTx System is used with the LUMINELLE Bx 360° Rotatable Disposable Sheath (Biopsy), the system is limited to performing diagnostic procedures only. The Bx Sheath can be used for endometrial or lower urinary tract (including bladder) biopsy by aspiration with a syringe during diagnostic procedures.

    Device Description

    The LUMINELLE DTx System, previously known as the LUMINELLE DTx Hysteroscope System, originally received 510(k) clearance in 2018 as a hysteroscopic system (K181909). In addition, LUMINELLE DTx Hysteroscopy System received 510(k) clearance in 2019 for the addition of a sheath component, LUMINELLE DTx 360° Rotatable Disposable Sheath Rigid to the system (K190827) and LUMINELLE Dx 360° Rotatable Disposable Sheath (Diagnostic) (K19227). The LUMINELLE DTx System permits viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures. The LUMINELLE DTx System is intended for use in endoscopic access to the examination of the lower urinary tract, including the bladder. When combined with accessory instruments, the LUMINELLE DTx System allows the user to perform various diagnostic and therapeutic procedures. The LUMINELLE DTx System is comprised of the following components: LUMINELLE DTx Scope; LUMINELLE Communication Cable; LUMINELLE Control Hub: USB cable: HDMI cable, and; power cord. UVision360, Inc. (hereafter UVision360) designed a new accessory, the LUMINELLE Bx 360° Rotatable Disposable Sheath (Biopsy) that is compatible with the LUMINELLE DTx System that permits direct viewing of the cervical canal and uterine cavity during biopsy procedures. LUMINELLE Bx 360° Rotatable Disposable Sheath (Biopsy) is a single use accessory and is made of the same materials as the previously cleared sheaths (K190827, K192278). The LUMINELLE DTx Scope is reprocessed after each use, depending on the procedure. High Level Disinfection (HLD) solutions can be used to reprocess the LUMINELLE DTx Scope when used with a cystoscopy procedure. 0.6% Ortho-Phthalaldehyde (OPA), a chemical disinfectant, has been previously cleared for HLD. However, UVision360 has verified and validated an additional chemical, Steris Revital-Ox™ RESERT® High Level Disinfectant, for HLD with the LUMINELLE DTx System. Steris Revital Ox™ RESERT® High Level Disinfectant has been tested in order to determine that it identically cleans the LUMINELLE DTx Scope as 0.6% OPA and does not adversely affect the functionality of the LUMINELLE DTx System. Before a hysteroscope procedure, the LUMINELLE DTx Scope must be sterilized to be reprocessed. In the previous submissions, Ethylene Oxide, qualified by Andersen Scientific, Inc., has been used to sterilize the LUMINELLE DTx Scope. An additional sterilization method, V-PRO® Low Temperature Sterilization, qualified by Steris located in Mentor, OH, has been validated and verified as an additional sterilization method to ensure that it properly sterilizes the LUMINELLE DTx Scope.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for the LUMINELLE DTx System. This type of document is a submission for a medical device to demonstrate it is substantially equivalent to a legally marketed predicate device. This typically involves performance and design testing rather than clinical study data regarding AI performance. Based on the provided text, the LUMINELLE DTx System and its new accessory (LUMINELLE Bx 360° Rotatable Disposable Sheath (Biopsy)) are a hysteroscopy and cystoscopy system, which are medical imaging devices for direct viewing of internal body cavities. There is no mention of Artificial Intelligence (AI) or an AI-powered component within the LUMINELLE DTx System.

    Therefore, many of the requested items (e.g., acceptance criteria for AI performance, sample size for AI test sets, expert ground truth establishment for AI, MRMC studies, standalone AI performance) are not applicable to this device submission.

    However, I can extract the general acceptance criteria and study information related to the device's performance, safety, and effectiveness as a medical device, which are based on "bench testing" (engineering tests) and compliance with relevant standards.

    Here's the breakdown of the information that is available in the document:

    1. Table of acceptance criteria and the reported device performance (based on the device's technical specifications and safety/performance testing):

    Acceptance Criteria/CharacteristicReported Device Performance (LUMINELLE DTx System)Comparison to Predicate (LUMINELLE DTx Hysteroscopy System K192278)
    Indications for UseHysteroscopy: Viewing cervical canal and uterine cavity for diagnostic/surgical procedures. Cystoscopy: Endoscopic access/examination of lower urinary tract (bladder) for diagnostic/therapeutic procedures. Bx Sheath for endometrial/lower urinary tract biopsy by aspiration.Identical, with the addition of biopsy aspiration capability for the new Bx Sheath. This difference does not raise new questions of safety and effectiveness.
    Intended UseSame as Indications for Use.Identical.
    Product CodeHIH (Hysteroscope and Accessories), FAJ (Cystoscope and Accessories), HFF (Aspirator, Endometrial), FCK (Instrument, Biopsy, Suction)Primary codes identical. The addition of HFF and FCK product codes is due to the new Bx Sheath's biopsy aspiration capability.
    BiocompatibilityISO 10993 Compliant. Results confirmed materials are not cytotoxic, sensitizing, or irritating.Identical compliance.
    ComponentsLUMINELLE DTx Scope, Communication Cable, Control Hub, USB/HDMI cables, power cord. New accessory: LUMINELLE Bx 360° Rotatable Disposable Sheath (Biopsy).Main components identical. The new Bx Sheath was validated/verified to not affect safety/effectiveness.
    Rigid/Flexible SheathBoth flexible (with PEEK hypotube) and rigid (with 304 stainless steel inner tube) sheaths available.Identical.
    Sheath ChannelsDTx & DTx Rigid Sheaths: 4 channels (scope, 2 fluid, 1 operative). Dx & Bx Sheaths: 2 channels (scope, 1 fluid) merging into single channel.Bx Sheath channels identical to Dx Sheath channels of predicate.
    Syringes during Biopsy10 mL Luer Lock syringe attached to the sheath (for biopsy procedures).Not applicable for predicate specifically, but difference (Bx adapter) validated to not affect safety/effectiveness.
    ImagesDevice can take and transmit images.Identical.
    Scope Working Length240 mmIdentical.
    Image Processing & DisplayDigital Video Processor; Standard HD Monitor/TV.Identical.
    Objective LensFocal Length: 5-50 mm; Field of View: 120° in air; Direction of View: Forward (0°).Identical.
    Image ResolutionCMOS chip is 400 x 400 pixels; USAF 1951 bar code Group-Element: 1-5.Identical.
    Power Requirements120V ACIdentical.
    Power SupplyLUMINELLE Control Hub (converts 120V AC to 12V DC, image converter, visualization connection).Identical.
    CablesLUMINELLE Communication Cable (powers scope, transmits image), HDMI, USB.Identical.
    Mode of OperationContinuousIdentical.
    Electrical SafetyCompliant with IEC 60601-1, IEC 60601-2-18.Identical compliance.
    Thermal SafetyCompliant with IEC 60601-1, IEC 60601-2-18.Identical compliance.
    Electromagnetic CompatibilityCompliant with IEC 60601-1-2 4th edition.Identical compliance.
    Degree of Protection Against Electrical ShockType BFIdentical.
    Degree of Protection Against Invasion of LiquidsIPX7Identical.
    Site of UseHospitals and Physician offices.Identical.
    Reprocessing (Sheaths)Single-use, disposable, provided sterile. No reprocessing required.Identical.
    Sterilization Cycles (Scope)Andersen Scientific Standard Ethylene Oxide (EO) Flexible Bag Process. New: Steris V-PRO® Low Temperature Sterilization.EO cycle identical. V-PRO added and validated to ensure no effect on safety/effectiveness.
    High Level Disinfectants (Scope for Cystoscopy)0.6% ortho-Phthalaldehyde (OPA). New: Steris Revital-Ox™ RESERT® High Level Disinfectant.OPA identical. Steris Revital-Ox™ added and validated to ensure no effect on safety/effectiveness.

    2. Sample sized used for the test set and the data provenance:

    • Test Sets (Bench Testing): The document describes various bench tests:
      • Performance - Bench: "useful life testing following multiple cycles of high-level cleaning and disinfection," "performance testing after accelerated aging," "durability of the reusable handle markings."
      • Design Validation: "usability testing in accordance with IEC 62366-1 and ANSVAAMI HE75 in simulated uterine and bladder models." "in vivo models" (for LUMINELLE Bx sheath).
      • Shelf-Life: "an established 100% Ethylene Oxide (EO) sterilization process in accordance with ISO 1135:2014, with a Sterility Assurance Level (SAL) of 10-6." "leak testing in accordance with ASTM F2096-11 and seal strength testing in accordance with ASTM F88-15 following accelerated aging in accordance with ASTM F1980-16." "simulated distribution testing in accordance with ASTM D4169-16, Distribution Cycle 13."
      • Reprocessing: Validation per "FDA guidance document 'Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling'"
      • Biocompatibility: In accordance with "ISO 10993-1," "ISO 10993-10:2016," "ISO 10993-5:2009."
      • Electrical Safety/EMC: Compliant with "IEC 60601-1, IEC 60601-1-6, IEC 60601-2-18 and IEC 60601-1-2."
    • Data Provenance: The studies are described as "Design verification and validation" and "Bench testing." These are laboratory-based engineering and performance tests, not clinical studies with patient data. The provenance is internal company testing to meet regulatory standards. No country of origin for patient data (as there isn't any) or retrospective/prospective status (as it's not a human study) is relevant here.
    • Sample Size: Specific numerical sample sizes for these bench tests are not provided in the summary. The document states "multiple cycles," "accelerated aging," and refers to compliance with various ASTM and ISO standards, which would imply defined sample sizes within those methodologies, but the specific numbers are not disclosed in this summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • For the "usability testing," it states "Instructions for Use were also evaluated by the physicians during the design validation testing." and "The reprocessing instructions were validated for use by Sterile Processing Department technicians."
    • For the "in vivo models" with the LUMINELLE Bx sheath, it aimed to "confirm that [...] the user is able to operate the system as intended, and the sheath conforms to user needs."
    • The document implies that qualified professionals (physicians, Sterile Processing Department technicians) provided input on usability and operational aspects, serving as "experts" for this kind of "ground truth" (i.e., whether the device is usable and meets user needs). However, no specific number of experts or detailed qualifications (e.g., 10 years of experience) are provided.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This is a medical device submission based on bench and performance testing, not a study evaluating human reader performance or image interpretation requiring adjudication of discrepancies.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done:

    • No. This device is not an AI-powered diagnostic tool, and the submission does not involve comparative effectiveness studies of human readers with versus without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • No. There is no algorithm or AI component mentioned in the LUMINELLE DTx System.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for this device's validation is primarily based on engineering specifications, compliance with international standards (ISO, IEC, ASTM), and user/usability feedback. For biocompatibility, it's lab test results showing lack of cytotoxicity, sensitization, or irritation. For performance, it's successful operation within defined parameters, durability under testing conditions, and maintaining sterile barriers. For usability, it's feedback from physicians and technicians. There is no mention of pathology or outcomes data as "ground truth" in this 510(k) summary for the device itself.

    8. The sample size for the training set:

    • Not applicable. There is no AI component, so no training set.

    9. How the ground truth for the training set was established:

    • Not applicable. There is no AI component, so no training set ground truth.
    Ask a Question

    Ask a specific question about this device

    K Number
    K192278

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-11-01

    (71 days)

    Product Code
    Regulation Number
    884.1690
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K181909

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Hysteroscopy: The LUMINELLE DTx Hysteroscopy System is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures.

    NOTE: Hysteroscopes are used as tools to access the uterine cavity and are not, in and of themselves, a method of surgery. When LUMINELLE DTx Hysteroscopy System is specifically used with the LUMINELLE Dx 360 Rotatable Disposable Sheath (Diagnostic), the system is limited to performing diagnostic procedures only.

    Generally recognized indications for diagnostic hysteroscopy include:

    • Abnormal bleeding
    • Infertility and pregnancy wastage
    • Evaluation of abnormal hysterosalpingogram
    • Intrauterine foreign body
    • Amenorrhea
    • Pelvic Pain

    Generally recognized indications for use for operative hysteroscopy include:

    • Directed endometrial biopsy
    • Polypectomy
    • Submucous myomectomy
    • Transection of intrauterine adhesions
    • Transection of intrauterine septa
    • Endometrial ablation

    Cystoscopy: The LUMINELLE DTx Hysteroscopy System is intended for use in endoscopic access to and examination of the lower urinary tract, including the blader. When combined with accessory instruments, the user to perform various diagnostic and therapeutic procedures.

    NOTE: When LUMINELLE DTx Hysteroscopy System is specifically used with the LUMINELLE Dx 360 Rotatable Disposable Sheath (Diagnostic), the system is limited to performing diagnostic procedures only.

    Device Description

    The LUMINELLE DTx Hysteroscopy System originally received 510(k) clearance in 2018 as a hvsteroscopic and cvstoscopic svstem (K181909). In addition, LUMINELLE DTx Hysteroscopy System received 510(k) clearance in 2019 for the addition of a sheath component, LUMINELLE DTx 360° Rotatable Disposable Sheath Rigid to the system (K190827).

    Based on user feedback for a thinner option for the disposable sheath for only diagnostic use, UVision360, Inc. (hereafter UVision) proposes the addition of a new accessory, the LUMINELLE Dx 360° Rotatable Disposable Sheath (Diagnostic), as part of an engineering performance change.

    The LUMINELLE DTx Hysteroscopy System indications for use is identical to the previously cleared indications for use for the system. However, when the system is used with the LUMINELLE Dx 360° Rotatable Disposable Sheath (Diagnostic), the indications for use are limited to diagnostic use only.

    No changes are proposed for the previously cleared components of the LUMINELLE DTx Hysteroscopy System. The new LUMINELLE Dx 360° Rotatable Disposable Sheath (Diagnostic) accessory is a modification of the LUMINELLE DTx 360° Rotatable Disposable Sheath Rigid component of the previously cleared LUMINELLE DTx Hysteroscopy System (K190827) with the following main changes:

    • . The new accessory has one fluid line rather than the two fluid lines of the previously cleared component.
    • . The new accessory has a combined channel for the scope and the fluid line rather than the separate channels of the previously cleared component.
    • The new accessory has no operative channel available unlike the previously cleared . component, due to the fact that it is for diagnostic use only.
    AI/ML Overview

    The provided text describes a 510(k) submission for the LUMINELLE DTx Hysteroscopy System, which includes the addition of a new accessory, the LUMINELLE Dx 360° Rotatable Disposable Sheath (Diagnostic).

    The study aims to demonstrate substantial equivalence to a previously cleared predicate device (LUMINELLE DTx Hysteroscopy System, K190827).

    Summary of Acceptance Criteria and Device Performance:

    The document primarily focuses on verifying the safety and effectiveness of the new accessory by demonstrating that the modified device remains as safe and effective as the predicate. It does not provide specific quantitative acceptance criteria for diagnostic performance outcomes (e.g., sensitivity, specificity for detecting specific conditions) for the new diagnostic sheath, as the intent of this 510(k) is an engineering performance change and limitation of use for the new accessory rather than a new diagnostic claim. Instead, the "acceptance criteria" are implied through the successful completion of various engineering, biocompatibility, and usability tests.

    Table of Acceptance Criteria (Implied) and Reported Device Performance:

    Acceptance Criteria (Implied)Reported Device Performance
    Sterilization adoption for the new sheath in accordance with AAMI TIR28:2016.Product adoption and Process Equivalence for Ethylene Oxide Sterilization was conducted. Results demonstrated that the LUMINELLE Dx 360° Rotatable Disposable Sheath sterilization may be adopted into the existing validation protocol after microorganism characterization and EO residuals testing.
    13-month shelf life for the new sheath, confirmed by accelerated aging.An accelerated aging shelf life study was performed for packaged LUMINELLE Dx 360° Rotatable Disposable Sheath (Diagnostic) in accordance with ASTM F1980-16. Visual inspection and functional design verification confirmed that all samples met specifications after aging.
    Packaging maintaining a sterile barrier throughout normal transportation and distribution.Distribution simulation testing in accordance with ASTM D4169-16 (Distribution Cycle 13: Assurance Level I) was performed. This included seal strength testing (ASTM F88-15) and bubble leak testing (ASTM F2096-11). Packages were confirmed to maintain integrity following simulated distribution.
    New sheath performing according to its intended use in in vivo models; user able to operate the system as intended; product conforming to user needs.Usability and design validation testing was repeated for the LUMINELLE DTx Hysteroscopy System with the LUMINELLE Dx 360° Rotatable Disposable Sheath (Diagnostic). This testing was conducted "in in vivo models." Devices met predefined acceptance criteria.
    New sheath not inducing sensitization or irritation.Sensitization and irritation testing was performed for the final finished LUMINELLE Dx 360° Rotatable Disposable Sheath (Diagnostic) in accordance with ISO 10993-10:2016. Results demonstrate the device is not irritating or sensitizing.
    New sheath not inducing cytotoxicity.Cytotoxicity testing was performed for the final finished LUMINELLE Dx 360° Rotatable Disposable Sheath (Diagnostic) in accordance with ISO 10993-5:2009. Results demonstrate the device is not cytotoxic.

    Detailed Information about the Study:

    1. A table of acceptance criteria and the reported device performance: Refer to the table above.

      • Note: The document does not specify quantitative diagnostic performance metrics because the submission is for an engineering change and a limitation of use for an accessory, not a new diagnostic claim that would require clinical performance study. The "acceptance criteria" are primarily related to safety, sterility, shelf life, and basic functionality as a medical device to ensure equivalence to the predicate.
    2. Sample sizes used for the test set and the data provenance:

      • Test Set Sample Size: The document does not explicitly state specific sample sizes for the various tests (e.g., how many sheaths were tested for accelerated aging, how many "in vivo models" were used for usability). It refers to "all samples met specifications" or "packages were confirmed."
      • Data Provenance: The information provided is retrospective in nature, as it's a summary of testing already performed for a 510(k) submission. The country of origin for the data is implied to be within the scope of UVision360 Inc.'s operations (Raleigh, NC, USA), but not explicitly stated for the testing sites.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable for this type of submission. The "ground truth" for these engineering, biocompatibility, and usability tests is established by adherence to recognized standards (e.g., AAMI TIR28, ASTM F1980-16, ASTM D4169-16, ASTM F88-15, ASTM F2096-11, ISO 10993-10:2016, ISO 10993-5:2009) and predefined acceptance criteria set during the device's development and validation. Experts in relevant engineering, toxicology, and quality assurance fields would have designed and overseen these tests, but their specific number and qualifications are not detailed in this regulatory summary.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. This document describes technical device performance testing (e.g., sterility, shelf life, biocompatibility, mechanical function), not clinical diagnostic performance requiring expert adjudication of patient cases.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML-enabled device and no MRMC study was performed or needed for this type of submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI/ML-enabled device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The "ground truth" for the tests described is the adherence to specified technical and safety standards (e.g., ISO, ASTM standards for biocompatibility, packaging integrity, and accelerated aging) and predefined functional specifications for the device. For the usability and design validation, it implicitly relies on objective performance measures and user feedback in "in vivo models" compared against "predefined acceptance criteria."
    8. The sample size for the training set: Not applicable. This is not an AI/ML device, so there is no "training set."

    9. How the ground truth for the training set was established: Not applicable.