510(k) Data Aggregation

    K Number
    K212070

    Validate with FDA (Live)

    Date Cleared
    2021-08-30

    (59 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K181655, K171657, K113561, K201605

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KMTI S141 Lumbar Interbody Fusion System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1.

    Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). KMTI S141 System implants are to be used with autogenous bone graft. Patients should be skeletally mature and have at least six months of non-operative treatment. The KMTI S141 System must be used with supplemental fixation cleared by FDA for use in the lumbar spine.

    Device Description

    The Kyocera Medical Technologies, Inc. (KMTI) KMTI S141 Lumbar Interbody Fusion System (also called S141 LIF System) consists of cages which can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion. The S141 LIF System includes the Tesera-P, Tesera-T, PEEK Straight TLIF, and PEEK TLIF families. The implants are offered in additively manufactured Titanium Alloy (Titanium-6Aluminum-4Vanadium) per ASTM F2924 and PEEK (Polyetheretherketone) per ASTM F2026.This submission updates the subject titanium alloy material specifications and standards, modifies the additive printing process, and expands the for the previously cleared S141 Lumbar Interbody Fusion System.

    The KMTI S141 Lumbar Interbody Fusion System is comprised of a variety of implant sizes to accommodate various patient anatomy and pathology, and associated instrumentation. The implants may be inserted via an open or minimally invasive approach. The Tesera P and PEEK PLIF use a posterior approach. The PEEK Straight TLIF, Tesera ST, and Tesera T use an oblique approach. The hollow geometry of the implants allows them to be packed with autogenous bone graft. The superior and inferior surfaces of the Titanium Alloy implant consist of Tesera porous titanium structure to facilitate osseous integration. The superior and inferior surfaces of the S141 PEEK devices have "teeth" to help prevent the device from migration after surgically positioned. Additionally, the S141 PEEK devices contain tantalum markers (per ASTM F560) to assist the surgeon with proper placement of the device. The implants are provided terminally sterilized via gamma irradiation prior to end-user receipt.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the KMTI S141 Lumbar Interbody Fusion System, which is a medical device for spinal fusion.

    Here's the breakdown of the acceptance criteria and the study performed, based only on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Inferred from Predicate Equivalence)Reported Device Performance
    Adequate and substantially equivalent mechanical strength for intended use.Mechanical testing demonstrated adequate and substantially equivalent mechanical strength for its intended use.
    Similar intended uses and indications.Identical to primary predicate device.
    Similar technological characteristics.Nearly identical to predicate devices; minor differences do not raise new safety/effectiveness issues.
    Similar principles of operation.Similar to predicate devices.

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: Not explicitly stated. The text mentions "mechanical testing" was conducted, but does not provide details on the number of devices or iterations tested.
    • Data Provenance: The testing was conducted by Kyocera Medical Technologies, Inc. within the context of a 510(k) submission to the FDA. No information is provided regarding the country of origin of the data or whether it was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This question is not applicable as the study presented is a mechanical performance study, not a clinical study involving ground truth established by experts.

    4. Adjudication method for the test set

    • This question is not applicable as the study presented is a mechanical performance study, not a clinical study requiring adjudication of expert opinions.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This question is not applicable. The device is an intervertebral body fusion system, a physical implant, not an AI-powered diagnostic or assistive tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This question is not applicable as the device is a physical implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The "ground truth" for the mechanical testing would be defined by engineering specifications and established performance standards for spinal implants, as derived from the predicate devices and relevant ASTM standards (e.g., ASTM F136, F2924, F2026, F560 mentioned for materials). The device's mechanical strength was compared to these standards and to the performance of predicate devices to establish substantial equivalence.

    8. The sample size for the training set

    • This question is not applicable This is a mechanical device, not an AI/machine learning system that requires a training set.

    9. How the ground truth for the training set was established

    • This question is not applicable This is a mechanical device, not an AI/machine learning system that requires a training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K193320

    Validate with FDA (Live)

    Date Cleared
    2020-02-04

    (64 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K181655

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KMTI Tesera SA Anterior Lumbar Interbody Fusion (ALIF) System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). KMTI Tesera SA ALIF System implants are to be used with autogenous bone graft. Patients should be skeletally mature and have at least six months of non-operative treatment. The KMTI Tesera SA ALIF System is a standalone device and is intended to be used with the cover plate and screws provided and requires no additional supplementary fixation. The anterior cover plate must be utilized whenever the device is implanted using the bone screws provided. Should the physician choose to use fewer than the four screws provided, additional supplemental fixation cleared by the FDA for use in the lumbar spine must be used. Supplemental fixation, cleared by the FDA for use in the lumbosacral spine, must be used with implants ≥20°.

    Device Description

    The Kyocera Medical Technologies, Inc. Tesera SA Anterior Lumbar Interbody Fusion (ALIF) System is a change to the FDA cleared Kyocera S128 Anterior Lumbar Interbody Fusion (ALIF) System (Ti6Al4V and PEEK cages, Cover Plate, screws and instruments; K131122. K140106. and K180502). Only the Ti6Al4V cages are relevant to this application which describes a change in manufacturing and material standard. There are no other changes in cage dimensions (L, W, H and lordosis), sterilization, packaging or shelf-life. The new system will be marketed as the Tesera SA Anterior Lumbar Interbody Fusion (ALIF) System ("Tesera SA ALIF System") and will include Ti6Al4V cages only.

    The Tesera SA ALIF System includes cages of a variety of lengths, widths, and lordosis to suit the individual pathology and anatomical conditions of the patient. The different shape of the footprint allows for different surgical approaches for insertion. Tesera SA ALIF System cages are standalone when implanted with the Cover Plate and screws with the exception of implants ≥20° which require supplemental fixation.

    The Ti6Al4V is compliant with ASTM F2924 Standard Specification for Additive Manufacturing Titanium-6 Aluminum-4 Vanadium with Powder Bed Fusion
    The Tesera SA ALIF System cages that are the subject of this application are gamma sterilized.

    The system also includes the previously FDA cleared Cover Plate and screws; previously FDA cleared and new instruments to allow for implant determination, trialing and disc preparation. New instruments are manufactured from stainless steels compliant with ASTM A564 Standard Specification for Hot-Rolled and Cold-Finished Age-Hardening Stainless Steel Bars and Shapes.

    AI/ML Overview

    This document describes the KMTI Tesera SA Anterior Lumbar Interbody Fusion (ALIF) System, which is a medical device. The information provided is primarily for a 510(k) premarket notification, indicating the device is substantially equivalent to previously cleared devices. Therefore, the "acceptance criteria" and "device performance" are primarily benchmarks against predicate devices and adherence to relevant ASTM standards, rather than clinical trial results with specific performance metrics like sensitivity or specificity for AI.

    Here's an analysis based on the provided text, focused on aspects relevant to device performance and study methodology, even if not directly an "AI" study:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Bench Testing)Reported Device Performance
    Compliance with ASTM F2077-18 Test Methods For Intervertebral Body Fusion Devices (Dynamic Shear Compression)Successful dynamic shear compression testing conducted.
    Compliance with ASTM E8/E8M-16 Standard Test Methods for Tension Testing of Metallic Materials(Not explicitly stated as a test conducted for this submission, but implies material properties meet this standard)
    Compliance with ASTM F2924 Standard Specification for Additive Manufacturing Titanium-6 Aluminum-4 Vanadium with Powder Bed Fusion (for Ti6Al4V material)Cages are manufactured to ASTM F2924.
    Manufacturing change validationManufacturing change has been validated.
    Substantial equivalence to predicate devices (K131122, K140106, K180502) in terms of indications for Use, dimensions, sizes, packaging, and sterilization.The Tesera SA ALIF System has the same Indications for Use as the predicate devices. The cages are additively manufactured; have the same dimensions and sizes, are packaged and gamma sterilized the same as the predicate cages.
    Absence of new or different issues of safety or effectiveness associated with the new cages.The change in manufacturing and material standard do not result in new or different issues of safety or effectiveness associated with the new cages.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not describe a clinical study with a "test set" in the context of an AI device. The performance evaluation is based on bench testing of the device itself and comparison to predicate devices. Therefore, a sample size for a test set of patient data, data provenance, or retrospective/prospective nature is not applicable here. The "sample" in this context refers to the manufactured devices tested. The document does not specify the exact number of devices or test specimens used for the bench testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This is not an AI study involving human expert ground truth for interpretation of medical images or data. The "ground truth" for this device's performance is established by engineering standards and direct physical testing of the device.

    4. Adjudication Method for the Test Set

    Not applicable. There is no human adjudication process described for the bench testing. Compliance with standards and successful performance are determined by the test results themselves.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This is not an AI device, and therefore no MRMC study or assessment of human reader improvement with AI assistance was conducted.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    Not applicable. This is not an algorithm or AI device. The device itself (an interbody fusion system) is "standalone" in the sense that it doesn't require "additional supplementary fixation" in most cases, but this is a physical characteristic, not an algorithmic performance metric.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the device's technical performance is based on:

    • Established engineering standards: ASTM F2077-18, ASTM E8/E8M-16, ASTM F2924.
    • Physical properties and performance characteristics measured during bench testing (e.g., dynamic shear compression).
    • Comparison to previously cleared predicate devices based on design, materials, manufacturing processes, and intended use.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI device, so there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there is no training set, there is no ground truth established for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182007

    Validate with FDA (Live)

    Date Cleared
    2018-12-07

    (133 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K181655

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Renovis Tesera SC Stand-alone ACF System is indicated for intervertebral body fusion procedures in skeletally mature patients with cervical degenerative disc disease at one level from C2-T1. Renovis Tesera SC Stand-alone ACF System implants are to be used with autogenous bone graft. Patients should be skeletally mature and have at least six weeks of non-operative treatment prior to implantation.

    The Renovis Tesera SC Stand-alone ACF System is a stand-alone system when used with the cover plate and screws provided, and requires no additional supplemental fixation. When used as a stand-alone system, the cages require the use of three (3) screws and the cover plate assembly. When used without the cover plate and three screws the Renovis Tesera SC Stand-alone ACF System is a non-stand-alone system and requires additional supplemental fixation cleared by the FDA for use in the cervical spine to augment stability.

    The Renovis Tesera C ACF System is indicated for intervertebral body fusion procedures in skeletally mature patients with cervical degenerative disc disease at one level from C2-T1. Renovis Tesera C ACF System implants are to be used with autogenous bone graft. Patients should be skeletally mature and have at least six weeks of non-operative treatment prior to implantation.

    The Renovis Tesera C ACF System is a non-stand-alone system and requires additional supplemental fixation cleared by the FDA for use in the cervical spine to augment stability.

    Device Description

    Tesera (or T3) describes Renovis' proprietary Tesera Trabecular Technology™ Porous Structure for the manufacture of the porous titanium on the superior and inferior surfaces of the implant.

    The Renovis Tesera C/Tesera SC Anterior Cervical Fusion (ACF) System is a single, cervical implant System that includes two (2) "subsystems", the:

    • Renovis Tesera C Anterior Cervical Fusion (ACF) System
    • Renovis Tesera SC Stand-alone Anterior Cervical Fusion (ACF) System .

    The Renovis Tesera C/Tesera SC Anterior Cervical Fusion (ACF) System cages are additively manufactured from Ti-6Al-4V and are available in a variety of sizes (widths, heights, and lordotic angles) to suit the individual pathology and anatomical conditions of the patient.

    The Renovis Tesera SC Stand-alone Anterior Cervical Fusion (ACF) System includes cages, and existing bone screws (self-drilling and self-tapping options) and anterior cover plate assembly. When used with the cover plate and screws, the Tesera SC ACF System cages require no supplementary fixation systems. The screws protrude through the interbody portion of the device into the adjacent vertebral bodies. The cages have windows in the side for visualization of the bone graft. When used without the cover plate and three screws the Renovis Tesera SC Stand-alone ACF cages require supplemental fixation.

    The Renovis Tesera C Anterior Cervical Fusion (ACF) System is not stand alone and requires supplemental fixation.

    The Renovis Tesera C/Tesera SC ACF System cages comply with ASTM F2924-14 Standard Specification for Additive Manufacturing Titanium-6 Aluminum-4 Vanadium with Powder Bed Fusion.

    The system also consists of reusable instrumentation. The instruments comply with ASTM A564/A64M-13e1 Standard Specification for Hot-Rolled and Cold-Finished Age-Hardening Stainless Steel Bars and Shapes or ASTM F136-13 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Renovis Tesera C/Tesera SC Anterior Cervical Fusion (ACF) System. This submission focuses on establishing substantial equivalence to a predicate device through mechanical testing and does not involve a study evaluating AI performance. Therefore, many of the requested sections related to AI acceptance criteria, ground truth, expert involvement, and reader studies are not applicable.

    Here's an analysis of the provided information, focusing on what is available:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Standard)Reported Device Performance
    Mechanical Testing:
    ASTM F2077 (Static Compression, Dynamic Compression, Static Shear Compression, Dynamic Shear Compression, Static Torsion, Dynamic Torsion Testing)Conducted and found to be acceptable
    Expulsion testingConducted and found to be acceptable
    Subsidence testing (with and without screws) per ASTM F2267Conducted and found to be acceptable
    Sterilization Validation:Completed
    AAMI/ANSI/ISO 10993-1 (Biological Evaluation)(Implied acceptable, no specific result mentioned)
    AAMI/ANSI ST72 (Bacterial Endotoxins)(Implied acceptable, no specific result mentioned)
    AAMI/ANSI/ISO 11137-2 (Radiation Sterilization Dose)(Implied acceptable, no specific result mentioned)
    AAMI/ANSI/ISO 11737-1 (Microbiological Methods)(Implied acceptable, no specific result mentioned)
    ISO 11607-2 (Packaging for Sterilized Devices)(Implied acceptable, no specific result mentioned)
    ISO 17665-1 (Moist Heat Sterilization)(Implied acceptable, no specific result mentioned)
    Materials/Manufacturing:
    ASTM F2924-14 (Additive Manufacturing Titanium)Complies
    ASTM A564/A64M-13e1 (Stainless Steel for Instruments)Complies
    ASTM F136-13 (Titanium for Instruments)Complies
    Porous Structure:
    T3 porous shell evaluationSuccessfully tested
    Other Standards:
    ASTM F1854-15 (Stereological Evaluation of Porous Coatings)(Implied acceptable, no specific result mentioned)
    ASTM F1147-05 (Tension Testing of Coatings)(Implied acceptable, no specific result mentioned)
    ASTM F1044-05 (Shear Testing of Coatings)(Implied acceptable, no specific result mentioned)
    ASTM F1160-14 (Shear and Bending Fatigue)(Implied acceptable, no specific result mentioned)
    ASTM F1978-17 (Abrasion Resistance of Coatings)(Implied acceptable, no specific result mentioned)
    ISTA 2A:2011 (Packaged-products weighing 150 lbs.)(Implied acceptable, no specific result mentioned)
    ASTM D4169:2016 (Performance Testing of Shipping Containers)(Implied acceptable, no specific result mentioned)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable as the submission describes a medical device, not an AI/software device that would typically use a "test set" of data in the common sense. The testing described is bench testing of the physical implant.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable. Ground truth, expert consensus, and expert qualifications are relevant for studies evaluating human or AI performance on diagnostic tasks. This submission is for a physical implant, where ground truth is established through standardized engineering and biological tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. Adjudication methods are relevant for resolving discrepancies in expert interpretations during diagnostic studies.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. MRMC studies are used to evaluate the impact of an AI system on human reader performance. This submission is for a physical medical implant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. "Standalone" performance refers to an AI algorithm operating independently, which is not relevant to a physical medical implant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the device's performance is established through adherence to recognized ASTM, AAMI/ANSI, and ISO standards for mechanical properties, biocompatibility, and sterilization. For example, the mechanical strength of the implant components and assemblies is evaluated against the parameters outlined in ASTM F2077. The porous structure's characteristics were "successfully tested," implying adherence to relevant specifications.

    8. The sample size for the training set

    This information is not applicable. A "training set" is used for machine learning. This submission is for a physical implant.

    9. How the ground truth for the training set was established

    This information is not applicable. This is relevant for machine learning models.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1