Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K190833

    Validate with FDA (Live)

    Date Cleared
    2019-12-15

    (258 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K172167 & K181354

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AXS Vecta Intermediate Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The AXS Vecta Intermediate Catheter is also indicated for use as a conduit for retrieval devices.

    Device Description

    The AXS Vecta Intermediate Catheter is a single lumen, flexible stiffness catheter. It has a radiopaque marker band on the distal end and a Luer hub at the proximal end. The AXS Vecta Intermediate Catheter shaft has a lubricious hydrophilic coating at the distal end (distal 25cm) to reduce friction during use. It is packaged with one Scout Introducer, one hemostasis valve, and two peel-away introducers.

    The Scout Introducer may be used in conjunction with the AXS Vecta Intermediate Catheter to facilitate in the introduction of the AXS Vecta Intermediate Catheter into distal vasculature and aid in navigation to distal anatomy. The Scout Introducer has a lubricious hydrophilic coating at the distal end to reduce friction during use. The inner lumen of the AXS Vecta Intermediate Catheter is compatible with the Scout Introducer, guide wires and microcatheters. The inner lumen of the Scout Introducer is compatible wires and microcatheters of an outer diameter of less than 0.044in.

    AI/ML Overview

    The AXS Vecta Intermediate Catheter's acceptance criteria and the studies proving it meets them are described below. The device is not an AI/ML device, so certain categories will be marked as "Not Applicable".

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance CriteriaReported Device Performance
    Design Verification Testing
    Particulate CharacterizationParticulate generated during simulated use must be acceptable.Particulate generation was acceptable.
    Design Validation Testing
    Simulated Use (Torque, ID, Product Compatibility)Test catheters must meet applicable user needs: introduction of interventional devices into the peripheral and neuro vasculature.All test samples met the applicable user needs and design specifications.
    Shelf Life TestingMust meet established acceptance criteria (specific criteria not detailed in the provided text).Results met the established acceptance criteria.
    Sterilization TestingMust meet established acceptance criteria (specific criteria not detailed in the provided text) and achieve a sterility assurance level (SAL) of 10-6.Results met the established acceptance criteria. A SAL of 10-6 has been demonstrated.
    BiocompatibilityMust be in accordance with EN ISO 10993-1 for an externally communicating medical device with circulating blood contact for less than 24 hours.Testing for the identical catheter component in the Reference device (K172167 and K181354) supports the biocompatibility of the subject device.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document primarily refers to "test samples" for bench testing and does not explicitly state the numerical sample size for each bench test. For the "Simulated Use" test, it mentions "All test samples," implying N > 1, but the exact number isn't provided.

    • Particulate Characterization: Not explicitly stated.
    • Simulated Use: Not explicitly stated.
    • Shelf Life Testing: Not explicitly stated.
    • Sterilization Testing: Not explicitly stated.
    • Biocompatibility: Not explicitly stated for specific tests, but based on the Reference device.

    Data Provenance: The studies were conducted by Stryker Neurovascular ("Stryker Neurovascular has demonstrated...") and are retrospective for most tests as they leverage previous device submissions (K172167 and K181354). The animal study, for instance, was also performed to support previously cleared devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    Not applicable. This device is a physical medical catheter, and the evaluation involves engineering and performance testing, not analysis of medical images or data requiring expert interpretation for ground truth.

    4. Adjudication Method:

    Not applicable. See point 3.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No. This is a physical medical device (catheter), not an AI/ML system that would involve human readers.

    6. Standalone (Algorithm Only) Performance:

    No. This is a physical medical device.

    7. Type of Ground Truth Used:

    For the performance tests, the "ground truth" is defined by compliance with established engineering design specifications, international standards (e.g., EN ISO 10993-1), and acceptable performance in simulated use models. For example:

    • Particulate Characterization: Acceptable levels of particulates.
    • Simulated Use: Meeting applicable user needs and design specifications in an in vitro neurovascular model.
    • Shelf Life/Sterilization: Meeting established acceptance criteria for shelf life and achieving a specific sterility assurance level.
    • Biocompatibility: Compliance with EN ISO 10993-1.

    8. Sample Size for the Training Set:

    Not applicable. This is a physical medical device, not an AI/ML system requiring a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191768

    Validate with FDA (Live)

    Date Cleared
    2019-07-30

    (29 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K172167, K181354

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AXS Vecta Aspiration Catheter, as part of the AXS Vecta Aspiration System, is indicated in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - MI and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment.

    Device Description

    The Stryker AXS Vecta Aspiration System includes an aspiration catheter and its accessories, including the Scout Introducer, the Peel-Away Introducers, the Hemostasis Valve, the AXS Universal Aspiration Tubing, the Medela Dominant Flex Pump, and the AXS Universal Liner Set.

    AI/ML Overview

    This document is a 510(k) summary for the AXS Vecta Aspiration System, which is a percutaneous catheter used for revascularization in patients with acute ischemic stroke. The submission seeks to demonstrate substantial equivalence to a previously cleared predicate device (K190212).

    Here's an analysis of the provided information regarding acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document provides a "Performance Testing Summary" table (Table 2) which includes the "Test" conducted and the "Conclusion". While it states "All units met the acceptance criteria and passed," the specific acceptance criteria values for each test are not explicitly listed in this document. The table implicitly states that the device met the unlisted criteria.

    TestReported Device Performance (Conclusion)Explicit Acceptance Criteria
    Visual Inspection (Packaging: Pouch Visual)All units met the acceptance criteria and passed Packaging Visual Inspection.Not explicitly stated
    Visual Inspection (Packaging: Undamaged Product)All units met the acceptance criteria and passed Packaging Visual Inspection.Not explicitly stated
    Tensile StrengthAll units met the acceptance criteria and passed Tensile Strength testing.Not explicitly stated
    PTFE DelaminationAll units met the acceptance criteria and passed PTFE Delamination testing.Not explicitly stated
    Torque StrengthAll samples met acceptance criteria and passed Torque Strength testing.Not explicitly stated
    Catheter BurstAll samples met acceptance criteria and passed Catheter Burst testing.Not explicitly stated
    Leak (Liquid)Leak (Air)All samples met acceptance criteria and passed the Air and Liquid Leakage testing.Not explicitly stated
    Dimensional (ID, OD, & Working Length)All samples met acceptance criteria and passed Dimensional testing.Not explicitly stated
    Kink ResistanceAll samples met acceptance criteria and passed Kink Resistance testing.Not explicitly stated
    Visual Inspection (Transition & Tip)All samples met acceptance criteria and passed both the transition and tip visual inspections.Not explicitly stated
    Lumen PatencyAll samples met acceptance criteria and passed Lumen Patency testing.Not explicitly stated
    Vacuum DropAll samples met acceptance criteria and passed Vacuum Drop testing.Not explicitly stated
    Tip FlexibilityAll samples met acceptance criteria and passed Tip Flexibility testing.Not explicitly stated
    Friction ForceAll samples met acceptance criteria and passed Friction Force testing.Not explicitly stated

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document states, "All units met the acceptance criteria" or "All samples met acceptance criteria" for each test, implying that the acceptance criteria were met by all tested units/samples. However, the specific number of samples used for each non-clinical bench test is not explicitly mentioned.

    No animal or clinical studies were performed for the current submission. The document states that animal testing previously conducted for the AXS Vecta Aspiration System was leveraged (from K190212, K172167, and K181354). The provenance of this leveraged animal data (country of origin, retrospective/prospective) is not specified in this 510(k) summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This 510(k) summary primarily relies on non-clinical bench testing. Therefore, there are no human experts establishing ground truth in the context of diagnostic performance or clinical outcomes. The "ground truth" for bench tests is defined by engineering specifications and test protocols.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Since no clinical or human-reader studies were conducted for this submission, an adjudication method for a test set is not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done as this is a medical device (catheter) and not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to an algorithm's performance without human interaction. This is not applicable as the device is a physical catheter, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical bench tests (e.g., tensile strength, burst pressure, dimensions), the ground truth is based on engineering specifications and established test methods that define what constitutes acceptable performance for a medical device of this type.

    8. The sample size for the training set

    This is not applicable as the device is a physical medical device (catheter) and not a machine learning or AI model that requires a training set.

    9. How the ground truth for the training set was established

    This is not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1