Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K251402

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-12-19

    (227 days)

    Product Code
    Regulation Number
    874.4680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K221340, K181193

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LIA-1 Catheter is intended for use through a compatible bronchoscope in the tracheobronchial tree for Optical Coherence Tomography (OCT) guided Fine Needle Aspiration (FNA) and Fine Needle Biopsy (FNB) of endobronchial lesions, peripheral lung nodules, and lung masses. The device collects real-time two-dimensional, cross-sectional depth images which are visualized for evaluation of human tissue microstructure on the LIA Console.

    Do not use this device for any purpose other than its intended use. This device is intended to provide an image of tissue microstructure in the tracheobronchial tree and parenchyma. The safety and effectiveness of this device for diagnostic analysis (i.e., differentiating normal versus specific abnormalities) in any tissue microstructure or specific disease has not been evaluated.

    Device Description

    LIA-1 Catheter is a sterile, single-use, Optical Coherence Tomography (OCT) image-guided biopsy needle which is an assembly consisting of a Handle, Outer Sheath, Needle, and Imaging Stylet (iStylet). The Outer Sheath and Needle are attached to Actuation Handle and the iStylet is located within the Needle.

    Outer Sheath: The LIA-1 Catheter Outer Sheath consists of a clear tube attached to the Handle.

    Handle: The LIA-1 Catheter Handle allows the user to actuate the needle from the distal tip of the Outer Sheath, and to actuate the iStylet into or out of the Needle tip, or to be retracted for biopsy collection, as necessary.

    Needle: The 21G stainless-steel Needle has a beveled, sharpened tip for puncture and tissue collection.

    Imaging Stylet: The iStylet consists of a shaft and imaging core serving as an optical imaging probe. It provides real-time imaging for its intended use.

    The LIA-1 Catheter is intended to be used only in conjunction with the LIA Console in order to function as intended.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1