    Reference Devices :

    K180363

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ClearTip FNA and FNB Types are intended for Ultrasonically Guided Fine Needle Aspiration, (FNA) of submucosal and extraluminal lesions of the Gastrointestinal Tract, (e.g., lymph nodes, abnormal tissue in the mediastinum).

    Device Description

    The ClearTip is a manually operated endoscopic instrument intended to obtain tissue specimens of gastrointestinal tract (=digestive tract). The subject device mainly consists of a handle unit with an insertion part, a syringe, a stopcock, and a connector. ClearTip is offered in various needle gauges for insertion and aspiration. The available needle gauge sizes of 19, 22, and 25 gauge. The Needle which is dimpled for ultrasonic visualization is advanced into the target site for aspiration. The device allows for adjustment of the length of the coil sheath and the needle to enable the user to adjust for the working length of the endoscope and to control needle insertion depth. It is preloaded with a stylet to aid in inserting the Needle which is removed for injection and aspiration. This device passes through the working channel of endoscope, and the average contact time with mucosa of the human gastrointestinal tract is less than 1 hour. This device is supplied sterile for single-patient use and shall be not reused or re-sterilized.

    AI/ML Overview

    This FDA 510(k) clearance letter pertains to a medical device, and as such, it does not detail acceptance criteria or a study that would demonstrate the device meets those criteria in the way typically seen for AI/software-as-medical-device (SaMD) products. This document describes a traditional medical device (a biopsy instrument), not an AI system.

    The "acceptance criteria" and "device performance" mentioned in your request are usually quantifiable metrics (e.g., sensitivity, specificity, AUC) for SaMD products, often against a ground truth for diagnostic accuracy. For a physical device like the ClearTip FNA and FNB Types, acceptance criteria typically relate to physical properties, functionality, safety, and compatibility, which are assessed through non-clinical (engineering, bench) tests rather than clinical performance studies against a diagnostic "ground truth."

    Therefore, I will extract the information provided based on the context of a physical medical device submission, focusing on the nearest equivalents to your requested categories.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a physical device, the "acceptance criteria" are implied by the performance standards for the non-clinical tests conducted. The reported "device performance" is typically that the device met these standards. The document does not provide specific numerical acceptance criteria or performance metrics for each test, but rather lists the types of tests performed.

    Acceptance Criteria Category (Implied)Reported Device Performance (Implied)
    Sterility ValidationValidated (device is sterilizable by EO)
    Shelf-Life3 Years (device maintains integrity for this period)
    AppearanceMet (device's visual characteristics are acceptable)
    DimensionsMet (device dimensions are within specified limits)
    OperabilityMet (device functions as intended)
    ElasticityMet (device's elastic properties are acceptable)
    Bending StrengthMet (device withstands specified bending forces)
    Pull-outMet (device components withstand specified pull-out forces)
    Tensile ForceMet (device components withstand specified tensile forces)
    BiocompatibilityMet (device materials are biocompatible)
    Endoscope CompatibilityMet (device is compatible with specified endoscope channels)

    2. Sample Size Used for the Test Set and Data Provenance

    This document does not describe a "test set" in the context of clinical data for diagnostic performance. The "tests" mentioned are non-clinical (bench/engineering) tests conducted on the physical device itself. Therefore, information on sample size for a "test set" or "data provenance" (country of origin, retrospective/prospective) is not applicable or provided given the nature of the device and the submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. This is not an AI/SaMD product requiring expert-established ground truth for diagnostic accuracy. The "ground truth" for a physical device's performance often relates to engineering specifications and regulatory standards.

    4. Adjudication Method for the Test Set

    Not applicable. No "test set" requiring adjudication in a clinical context is described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. An MRMC study is typically for evaluating diagnostic performance of imaging or AI systems with human readers. This clearance is for a physical biopsy instrument.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No. This is not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's acceptance is based on engineering specifications, material standards, and regulatory requirements (e.g., for sterility, biocompatibility, mechanical strength). It is not pathology, outcomes data, or expert consensus in the diagnostic sense.

    8. The Sample Size for the Training Set

    Not applicable. This device does not involve a "training set" as it is not an AI/machine learning system.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. There is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231267

    Validate with FDA (Live)

    Device Name
    ClearTip
    Date Cleared
    2023-06-30

    (59 days)

    Product Code
    Regulation Number
    876.1075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K180363

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Clear Tip is intended for Ultrasonically Guided Fine Needle Aspiration, (FNA) of submucosal and extraluminal lesions of the Tracheobronchial Tree and Gastrointestinal Tract, (e.g., lymph nodes, abnormal tissue in the mediastinum).

    Device Description

    The ClearTip is a manually operated endoscopic instrument intended to obtain tissue specimens of gastrointestinal tract (=digestive tract) and tracheobronchial tree. The subject device mainly consists of a handle unit with an insertion part, a syringe, a stopcock, and a connector. This device passes through a working channel of an endoscope, and the average contact time with the submucosal or external lesions of a gastrointestinal tract or a tracheobronchial tree is less than 1 hour. This device is supplied sterile for single-patient use and shall be not reused or resterilized.

    AI/ML Overview

    The provided document does not contain information regarding the acceptance criteria of a study that proves the device meets specific performance metrics. It's a 510(k) summary for a medical device (ClearTip) seeking substantial equivalence to predicate devices, not a study report demonstrating specific performance against acceptance criteria.

    The document discusses substantial equivalence based on:

    • Indications for Use: The ClearTip is intended for Ultrasonically Guided Fine Needle Aspiration (FNA) of submucosal and extraluminal lesions of the Tracheobronchial Tree and Gastrointestinal Tract.
    • Comparison to Predicate Devices: The document provides a detailed comparison table between the subject device (ClearTip) and two predicate devices (SonoTip Pro and Pro Flex EBUS-TBNA Needle System, and Clear-Tip EUS-FNA) regarding various characteristics like device name, classification, product code, regulation number, indications for use, needle gauge, material, length, sheath material, accessory channel diameter, endoscope compatibility, use with syringe/stopcock, principle of operation, shelf life, sterility, and single use.
    • Non-clinical Tests: The following tests were performed on the subject device to support substantial equivalence: Sterilization Validation Test, Shelf-Life Test, Appearance, Dimensions, Operability, Elasticity, Bending Strength, Pull-out, Tensile Force, and Biocompatibility Tests. The document states that the test results support the device's substantial equivalence but does not provide specific acceptance criteria or reported performance values for these tests.

    Therefore, the requested information for a table of acceptance criteria and reported device performance, sample sizes, ground truth establishment, adjudication methods, or MRMC/standalone study results cannot be extracted from this document. The document focuses on demonstrating substantial equivalence to existing legally marketed devices, rather than establishing primary performance metrics through a clinical or non-clinical study with defined acceptance criteria.