Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K183043

    Validate with FDA (Live)

    Date Cleared
    2019-04-17

    (166 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K180115

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MantaRay Reperfusion Catheters, with the Imperative Care Aspiration Tubing Set and Imperative Care Aspiration Pump (or equivalent vacuum pump), are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    The Imperative Care Aspiration Tubing Set is intended to connect the Imperative Care Reperfusion Catheter to the canister of the Imperative Care Aspiration Pump and to allow the user to control the fluid flow.

    Device Description

    The MantaRay Reperfusion Catheter is a single lumen, braid and coil reinforced, variable stiffness catheter that facilitates removal of thrombus/clot from the neurovasculature when connected to a vacuum source generated from the Imperative Care Aspiration Pump using the Imperative Care Aspiration Tubing. The MantaRay Reperfusion Catheter will be offered with various working lengths and nominal inner diameters (ID) and outer diameters (OD) as shown in Table 1 below.

    The MantaRay Reperfusion Catheter is comprised of a hollow cylindrical tube which is bonded to a standard luer fitting. The wall of the tube is constructed using a combination of metal coils/braids and medical grade polymers.

    The distal section of the MantaRay Reperfusion Catheter has a hydrophilic coating to enhance tracking through the vasculature. The beveled distal tip allows for atraumatic tracking past vessel branches during insertion. A radiopaque marker provides the user with visual confirmation of the distal tip location under fluoroscopy.

    The MantaRay Reperfusion Catheter is packaged with an accessory Rotating Hemostasis Valve (RHV). The RHV is designed to be attached to the proximal luer of the catheter and helps the user maintain hemostasis.

    AI/ML Overview

    This document describes the premarket notification (510(k) summary) for the MantaRay Reperfusion Catheters and Imperative Care Aspiration Tubing Set. The study primarily relies on a "substantial equivalence" argument by comparing the subject device to a predicate device (Penumbra System Reperfusion Catheter) through various bench, lab, and animal tests, rather than a standalone clinical study demonstrating a specific clinical outcome or a multi-reader, multi-case study for human-AI improvement.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document provides a comprehensive table (Table 4) detailing various test attributes and their corresponding specifications, which serve as acceptance criteria. The "Results" column in this context indicates that all tests passed or met these specifications, demonstrating the device's performance.

    Test AttributeAcceptance Criteria (Specification)Reported Device Performance (Results from document)
    Delivery, Compatibility, and Retraction (Trackability)The catheter shall be able to be delivered, deployed, and retracted per the IFU within a simulated neurological model without incurring any damage to the catheter.Passed (Demonstrated)
    Clot RetrievalThe catheter will be able to be delivered in a clinically relevant model and used to effectively aspirate clots with the aspiration pump and aspiration tubing set when placed at appropriate locations based on the size of the catheter.Passed (Demonstrated)
    Flexibility and Kink ResistanceThere shall be no kinking of shaft (permanent deformation) after simulated use.Passed (No kinking)
    Compatibility with other devices (external)The catheters shall be able to be delivered through the minimum introducer sheath or guide catheter size indicated in the product labeling.Passed (Demonstrated)
    Guidewire compatibilityThe catheters shall be able to be delivered over the guidewire size indicated in the product labeling.Passed (Demonstrated)
    Interventional device compatibility (internal)The catheters shall be able to accommodate other interventional devices (e.g., support catheter, diagnostic catheter) up to the maximum size indicated in the product labeling.Passed (Demonstrated)
    Luer compatibilityDevices and accessories shall be compatible with standard syringe luer fittings per ISO 80369-7.Passed (Compatible)
    Accessory compatibilityDevices shall be compatible with the accessory RHV.Passed (Compatible)
    VisualFree of kinks, breaks, separation or particulate (greater than 0.25mm²). No exposed metal.Passed (Free of defects)
    DimensionalAll defined catheter dimensions are within the specified tolerances.Passed (Within tolerances)
    Catheter Bond StrengthThe catheter shall have sufficient bond strengths to remain intact throughout a procedure.Passed (Sufficient bond strength)
    Flowrate - positive (forward) pressureThe catheter lumen shall allow for a minimum flowrate comparable to competitive products.Passed (Comparable flowrate)
    Flowrate – vacuum pressureThe flowrate under a vacuum shall be similar to or greater than competitive devices.Passed (Similar or greater flowrate)
    Freedom from Leakage – positive pressureNo liquid leakage from the hub or catheter shaft at 46psi for 30 seconds.Passed (No leakage)
    Freedom from Leakage – negative pressureNo air leakage into a 20cc syringe when vacuum pulled for 15 seconds.Passed (No leakage)
    Burst PressureCatheter does not burst under pressures that could be seen when performing contrast injections with a standard 10cc syringe.Passed (Did not burst)
    Catheter Torque StrengthNo separation of any portion of the catheter when rotated at least two (2) full rotations (720 degrees).Passed (No separation)
    Lumen IntegrityThe catheter lumen shall not collapse under vacuum.Passed (Did not collapse)
    Kink ResistanceThere shall be no kinking of the catheter shaft (permanent deformation) after wrapping around anatomically relevant bend radii.Passed (No kinking)
    FlexibilityThe flexibility of the catheter tip shall be comparable to competitive products and allow for easily tracking the device to the desired target anatomy.Passed (Comparable flexibility)
    Coating - ParticulateThe amount of particulate matter that comes off the hydrophilic-coated shaft during simulated use testing shall be determined and compared to competitive products and techniques.Passed (Acceptable particulate)
    Coating - lubricityCoating must be lubricious and maintain a minimum lubricity over 15 test cycles.Passed (Lubricious, maintained lubricity)
    Corrosion ResistanceNo visible corrosion present on devices after saline immersion followed by 30 minutes in boiling water followed by 48 hours in 37°C water bath.Passed (No visible corrosion)
    RadiopacityThe radiopaque marker on the catheter can be seen under fluoroscopy during use.Passed (Visible under fluoroscopy)
    Vacuum Force at Catheter TipThe vacuum force at the tip of the catheter should be comparable to the vacuum force at the tip of the predicate devices.Passed (Comparable vacuum force)
    Aspiration VolumeThe volume of fluid aspirated when using the catheter should be comparable to the predicate devices.Passed (Comparable aspiration volume)

    Biocompatibility Test Summary (Table 5)

    TestAcceptance CriteriaResults
    Cytotoxicity: ISO MEM ElutionSample extracts must yield cell lysis grade 2 or lower.Pass, Non-cytotoxic
    Cytotoxicity: ISO MTT AssayThe percentage of cells exhibiting lysis should be similar for all test devices.Pass, No Significant Differences
    Sensitization: Magnusson-Kligman MethodTest Group shall yield Grade < 1 score on Magnusson and Kligman scale (provided control Grade < 1).Pass, Non-Sensitizing
    Irritation: ISO Intracutaneous IrritationThe difference in the mean test article and mean control score must be grade 1.0 or lower.Pass, Non-Irritant
    Systemic Toxicity: ISO Materials MediatedSample Extracts must not cause a total rise in body temperature of $\ge$ 0.5°C.Pass, Non-pyrogenic
    Pyrogen Systemic Toxicity: ISO AcuteSample extracts must not cause the following: > 10% weight loss in 3 or more test animals; Mortality of 2 or more test animals; Toxic signs such as convulsions and prostration in 2 or more test animals.Pass, Non-Toxic
    Systemic Injection Test HemocompatibilityThe concentrations of C3a and SC5b-9 in the test devices are statistically similar to the predicate device (Penumbra Neuron Max 6F Catheter) and statistically lower than the positive control for all exposure times.Pass, No Significant Differences. The test device had statistically similar or lower concentrations than the predicate and negative controls.
    Complement Activation HemocompatibilitySample extracts must be nonhemolytic ( $\le$ 2% hemolytic index).Pass, Non-hemolytic
    Hemolysis (Direct Contact Method)Sample extracts must be nonhemolytic ( $\le$ 2% hemolytic index).Pass, Non-hemolytic
    Hemolysis (Extract Method)The device must have similar or lesser thrombogenic potential after 4 hours in vivo when compared to a control device (Penumbra Neuron Max 6F Catheter).Pass, Minimal Thrombogenic Potential. The test device had less thrombogenic potential than the control device.
    Standard In vivo ThrombogenicityThe device must have similar or lesser thrombogenic potential after 4 hours in vivo when compared to a control device (Penumbra Neuron Max 6F Catheter).Pass, Minimal Thrombogenic Potential. The test device had less thrombogenic potential than the control device.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Description: The document focuses on bench, lab (in-vitro), and animal testing for substantial equivalence. There isn't a "test set" in the traditional clinical trial sense with human patient data.
    • Sample Size:
      • Bench and Lab Testing: The largest and smallest diameter catheter sizes for both the subject and predicate devices were tested for the various performance attributes. The exact number of units per test is not specified, but the testing was conducted per "company approved protocols and test methods based primarily on catheter performance standard ISO 10555-1."
      • Animal Studies: Performed in a porcine model. The number of animals used is not specified directly for the main animal study, but the thrombogenicity test was "Performed in duplicate."
      • Biocompatibility Testing: Guinea pigs and rabbits were used for sensitization and irritation tests, respectively. Mice were used for acute systemic injection tests. The number of animals per test is not explicitly stated in the summary table.
      • Physician Validation Study: Active US physicians experienced in interventional neuroradiology and neurovascular surgery were recruited. The number of physicians is not specified.
    • Data Provenance: Not explicitly stated as country of origin of the data. All studies were conducted to support a US FDA 510(k) submission, implying they meet US regulatory standards. The animal and lab tests are pre-clinical. The physician validation study involved "US physicians." The nature of these studies is prospective as they were designed specifically to evaluate the device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Bench and Lab Testing: The "ground truth" for these tests are the defined engineering specifications (acceptance criteria) and physical measurements. These are established by engineering standards (e.g., ISO 10555-1) and internal company protocols. No external "experts" in the sense of clinical specialists are mentioned for establishing ground truth for these quantifiable measures.
    • Animal Studies: The evaluation of vascular response and pathological findings was "histologically comparable" to the predicate device. This implies assessment by experts (likely veterinary pathologists), but their number and specific qualifications are not detailed.
    • Physician Validation Study: "Active US physicians experienced in interventional neuroradiology and neurovascular surgery" participated. They performed clot retrieval procedures, and their performance/feedback constituted the evaluation. Their number and years of experience are not specified.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • There is no mention of an adjudication method in the context of expert consensus or dispute resolution for a "test set" of clinical cases. The various performance tests are against defined specifications or compared to a predicate device in controlled environments.
    • For the physician validation study, the "results... were analyzed and compared," implying a statistical comparison rather than an expert adjudication process for ground truth.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done in the context of AI assistance. This document describes a medical device (catheter and aspiration system), not an AI device for diagnostic imaging or interpretation.
    • The Physician Validation Study (Section L) involved physicians performing procedures with both the subject and predicate devices, demonstrating usability and comparable performance, but it was not a comparative effectiveness study designed to measure the improvement of human readers with AI assistance. It was a usability/performance study for a physical medical device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This question is not applicable. The device described, the MantaRay Reperfusion Catheter and Aspiration Tubing Set, is a physical medical device (catheter system) for mechanical thrombectomy, not an algorithm, AI, or software-only device. Its performance is inherently tied to human use (human-in-the-loop) and the mechanical action of aspiration.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Bench and Lab Testing: Engineering specifications, physical measurements, and performance against industry standards (e.g., ISO 10555-1).
    • Animal Studies: Histological analysis of vascular response and pathological findings; direct observation of successful revascularization, absence of perforation, dissection, or thrombosis. Comparison to the predicate device served as the control.
    • Biocompatibility Testing: Standardized biological test methods (e.g., ISO 10993-x series) with defined pass/fail criteria.
    • Physician Validation Study: Direct observation of procedural success in a simulated model (clot retrieval), and comparison of performance metrics between the subject and predicate devices.

    8. The sample size for the training set

    • This question is not applicable as the studies described are for regulatory clearance of a physical medical device. There is no mention of machine learning or an "AI" component requiring a training set in the document.

    9. How the ground truth for the training set was established

    • This question is not applicable, as there is no "training set" for an AI or machine learning model in this submission.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1