510(k) Data Aggregation

    K Number
    K220107

    Validate with FDA (Live)

    Date Cleared
    2023-06-16

    (520 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K193228,K180098,K103211,K142679,K201957

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SV600, SV800 Ventilators are intensive care situations for long-term or during transport within a professional healthcare facility. The SV600, SV800 Ventilators are intended to provide ventilation assistance and breathing support for adult and pediatric patients with a minimum body weight of 10 kg (all pediatric subgroups except newborns (neonates)). The SV600, SV800 Ventilators should be operated by properly-trained and authorized medical personnel. This equipment is not suitable for use in an MRI environment.

    Device Description

    The SV600 and SV800 Ventilators are pneumatically-driven and electronically-controlled ventilators. The Ventilators consists of a main unit (including pneumatic circuit, electronic system, mechanical structure, display, CO2 module, SpO2 module), trolley and support arm.

    AI/ML Overview

    This document outlines the substantial equivalence of the Mindray SV600 and SV800 Ventilators to a predicate device (Evita XL Ventilator K083050) and reference devices. The submission focuses on technical comparisons and various types of performance testing to demonstrate safety and effectiveness.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly provide a table of acceptance criteria with corresponding device performance for specific thresholds. Instead, it states that "The functional and system level testing showed that the devices continue to meet specifications and the performance of the device is equivalent to the predicate." It also notes that the device "meets its accuracy specification and is substantially equivalent to the predicate device."

    The closest to "acceptance criteria" are the technical characteristics compared to the predicate and reference devices, and the standards the device was tested against. The "reported device performance" is implied by the statement of compliance with these standards and equivalency to the predicate.

    Here's an example of how one might infer "acceptance criteria" from the technical characteristics table:

    Technical Characteristic (Acceptance Criteria - inferred from predicate/reference)Reported Device Performance (Implied by "Same" or "Similar" comparison and compliance statement)
    Ventilation Mode:Compliance:
    V-A/C: YesSame as predicate
    P-A/C: YesSame as predicate
    PRVC: YesSame as predicate
    V-SIMV: YesSame as predicate
    P-SIMV: YesSame as reference device Servo-U (K180098)
    ... (and so on for all listed modes)
    Specifications - Ventilator setting parameter:
    TV range (Adult): 100 to 2000 ml (Predicate)Adult: 100 to 4000 ml (Similar, wider range)
    TV range (Ped.): 20 to 300 ml (Predicate)Ped: 20 to 300 ml (Same)
    O2% range: 21 to 100 vol.% (Predicate)21 to 100 vol.% (Same)
    ... (and so on for all listed parameters)
    Biocompatibility:Compliance:
    Adherence to ISO 10993-1, ISO 18562-1, -2, -3Found to comply
    Software Verification & Validation:Compliance:
    Adherence to FDA Guidance for Premarket Submissions for SoftwareWas conducted and documentation provided
    Electromagnetic Compatibility & Electrical Safety:Compliance:
    Adherence to ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, ISO 80601-2-12, ISO 80601-2-55, ISO 80601-2-61, AIM Standard 7351731Found to comply
    Bench Testing:Compliance:
    Meets accuracy specification compared to predicateMeets accuracy specification and is substantially equivalent
    Human Factors Testing:Compliance:
    Demonstrates device is safe and effective for intended users, uses and use environmentsWas conducted

    2. Sample sized used for the test set and the data provenance

    The document mentions "functional and system level testing" and "bench testing" but does not specify the sample sizes for these tests. It also does not specify the data provenance (e.g., country of origin of the data, retrospective or prospective) for any of the testing. The nature of these tests (bench, functional, system level) generally implies laboratory or simulated environments, rather than patient data from a specific country or clinical setting.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The testing described (biocompatibility, software, EMC, electrical safety, bench testing) does not typically involve "experts establishing ground truth" in the way a clinical study would. These are generally engineering and performance evaluations against defined specifications and standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided. Adjudication methods are typically relevant in clinical studies (e.g., for image interpretation or diagnosis) to resolve discrepancies between multiple evaluators. The types of testing described here (functional, system, bench, safety, EMC) do not involve such adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no mention of a multi-reader multi-case (MRMC) comparative effectiveness study. This document describes a ventilator (hardware and software for life support), not an AI-assisted diagnostic or interpretative device that would typically involve human "readers" or "AI assistance" in that context.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document does not explicitly use the term "standalone" in the context of an algorithm. However, the majority of the testing described (software verification and validation, electromagnetic compatibility, electrical safety, bench testing) are inherently evaluations of the device's performance without continuous human intervention during the test itself, focusing on the device's adherence to specifications and standards. The "human factors testing" mentioned relates to the usability and user interface, ensuring safe and effective interaction for the intended users (medical personnel), rather than a "human-in-the-loop" performance study of an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The concept of "ground truth" as typically understood in clinical or AI performance studies (e.g., pathology for a biopsy, expert consensus for an image diagnosis) is not directly applicable or explicitly stated for the types of testing described. For this device (ventilator), the "ground truth" for its performance is its adherence to:

    • Engineering specifications (e.g., flow range, pressure range accuracy)
    • Consensus standards (e.g., ISO, IEC for safety, EMC, performance characteristics)
    • Functional requirements (e.g., does it deliver the correct ventilation modes)

    These are verified through various physical and software tests against known engineering principles and standardized test methods.

    8. The sample size for the training set

    This information is not applicable/not provided. Ventilators are not typically "trained" using data sets in the way AI/ML algorithms are. This is a traditional medical device whose functionality is based on pre-programmed logic, control algorithms, and pneumatic/electronic design.

    9. How the ground truth for the training set was established

    This information is not applicable/not provided for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193056

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-11-20

    (385 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K180098

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Puritan Bennett 980 Series Ventilator System is designed for use on patient population sizes from Neonatal (NICU) through Adult who require respiratory support or mechanical ventilation and weigh a minimum of 0.3kg (0.66lb). It is suitable for service in hospital (institutions) and intra-hospital transport to provide continuous positive pressure ventilatory support using medical oxygen and compressed medical air from either an internal air sources to deliver oxygen concentrations of 21% to 100%. Ventilatory support can be delivered invasively to patients who require the following types of ventilator support

    • Positive Pressure Ventilation, delivered invasively (via endotracheal tube or trachectorny tube) or non-invasively (via mask or nasal prongs)

    • Assist/Control, SIMV, or Spontaneous modes of ventilation

    Device Description

    The Puritan Bennett 980 Series Ventilator System is a dual-microprocessor-based, touch-screen controlled; critical care ventilator intended to provide continuous ventilation for neonate to adult patients who require either invasive ventilation or non-invasive ventilation. It can be used in hospitals and institutions and for intra -hospital transport applications with access to the appropriate services.

    The ventilator system offers features for patient comfort while delivering sensitive, precise breaths to critically ill patients. The product ventilates Neonatal, Pediatric, and Adult patients with predicted body weights from 0.3 kg, and with tidal volumes for mandatory volume-controlled breaths from 2 mL to 2500 mL.

    The following options are being made to the Subject Device:

    • Initial release -
      • Integrated Nebulizer option ●
      • High Flow Oxygen Therapy (HFO2T) option ●
      • NIV+ Software option ●
      • New Monitored Parameter Porive ●
    • Modification to previously cleared feature -
      • Updated IE Sync Algorithm
    AI/ML Overview

    This document is an FDA 510(k) Premarket Notification for the Puritan Bennett 980 Series Ventilator System. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving a device meets specific acceptance criteria in the context of an AI/ML medical device.

    Therefore, none of the requested information regarding acceptance criteria, device performance, sample sizes, expert involvement, adjudication, MRMC studies, standalone performance, ground truth types, or training set details can be extracted from the provided text. The document states that "Clinical evidence was not necessary to show substantial equivalence," and the non-clinical evidence focuses on verification and validation against standards and software/system tests, not a performance study as would be expected for a new or modified algorithm with performance metrics.