510(k) Data Aggregation

    K Number
    K203772

    Validate with FDA (Live)

    Date Cleared
    2022-01-27

    (400 days)

    Product Code
    Regulation Number
    862.1358
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K180045, K192659

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Omnipod 5 SmartBolus Calculator is software intended for the management of diabetes in persons aged 6 and older requiring rapid-acting U-100 insulin. The Omnipod 5 SmartBolus Calculator calculates a suggested bolus dose based on user-entered carbohydrates, most recent sensor glucose reading (or blood glucose reading if using fingerstick), rate of change of the sensor glucose (if applicable), insulin on board (IOB), and programmable correction factor, insulin to carbohydrate ratio, and target glucose value. The Omnipod 5 SmartBolus Calculator is intended for single patient, home use and requires a prescription.

    Device Description

    The Omnipod 5 SmartBolus Calculator is a software device that is a component of the Omnipod 5 Automated Insulin Delivery System. The SmartBolus Calculator exists on the Omnipod 5 App portion of the Omnipod 5 ACE Pump and relies on the user interface of the App.

    The Omnipod 5 SmartBolus Calculator receives input parameters and settings from the other components of the system and calculates a suggested bolus amount of insulin to correct an elevated glucose level (a correction bolus) and/or to cover carbohydrates from a meal (meal bolus). The Omnipod 5 SmartBolus Calculator allows users to have the option of populating the current estimated glucose value and trend, which is communicated by the connected iCGM. Users may also manually enter the estimated glucose value or a blood glucose (BG) reading from a blood glucose meter. In addition to glucose, the suggested bolus dose is calculated based on the following parameters: user-entered carbohydrates, rate of change of the sensor glucose (if using a CGM), correction factor, insulin to carbohydrate ratio, target glucose value, and insulin on board (IOB). Once the calculation is complete, the user has the option of delivering the suggested dose of insulin, modifying the amount, or canceling.

    The Omnipod 5 SmartBolus Calculator can be used in the Omnipod 5 Automated Insulin Delivery System with both Manual Mode and Automated Mode. When the Omnipod 5 SmartBolus Calculator is used with manually-entered BG readings, it suggests a bolus dose based on the same calculations as the currently cleared Omnipod DASH Insulin Management System (K180045, most recently cleared in K192659).

    AI/ML Overview

    The document describes the Omnipod 5 SmartBolus Calculator, a software device for diabetes management. Here's a breakdown of the acceptance criteria and the study proving it meets these criteria:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied through the statement that the device meets "Insulin Therapy Adjustment Device special controls and to be safe and effective." The clinical study focused on glycemic control within 4 hours of bolusing. While explicit acceptance criteria for these percentages are not given as strict pass/fail thresholds, the performance observed is presented as evidence of safety and effectiveness, particularly regarding hypoglycemia.

    Glycemic Measure (as measured by CGM)Acceptance Criteria (Implied: Safety & Effectiveness)Reported Performance (CGM-Informed SmartBolus Calculator)
    Time in range (70-180 mg/dL)Expected to maintain time in range within acceptable levels given bolus calculations63.8% (SD 15.7)
    Time spent < 70 mg/dL (Hypoglycemia)Expected to minimize hypoglycemia, especially post-bolus2.1% (SD 2.0) Lower than Standard SmartBolus Calculator
    Time spent < 54 mg/dL (Severe Hypoglycemia)Expected to minimize severe hypoglycemia0.3% (SD 0.7)
    Time spent > 180 mg/dL (Hyperglycemia)Expected to manage hyperglycemia34.0% (SD 16.0)
    Time spent ≥ 250 mg/dLExpected to manage high hyperglycemia9.7% (SD 10.3)
    Time spent ≥ 300 mg/dLExpected to manage very high hyperglycemia2.6% (SD 3.7)

    Note: The primary analysis focused on comparing the CGM-informed SmartBolus Calculator with the standard one, and a statistically significant difference was observed for time spent < 70 mg/dL, indicating improvement in minimizing hypoglycemia.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 25 participants
    • Data Provenance: The document does not explicitly state the country of origin. The study appears to be a prospective clinical study, as participants actively used the device in two distinct phases of device usage.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The "ground truth" for the clinical study was measured directly by CGM data. There is no mention of experts establishing ground truth for the bolus calculations themselves or for the CGM data.

    4. Adjudication Method for the Test Set

    This information is not provided in the document. Given that the clinical study involved direct physiological measurements (CGM data), it is unlikely that an adjudication method for "ground truth" labels was necessary in the traditional sense.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, What was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    An MRMC study was not conducted, and therefore, there is no information on the effect size of human readers improving with AI assistance. This device is an automated bolus calculator, not an AI assisting human interpretation of images or data in an MRMC setting.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The presented clinical study is a comparison of two versions of the SmartBolus Calculator (standard vs. CGM-informed) in a human-in-the-loop setting, as users still had the option to deliver, modify, or cancel the suggested dose. While the algorithm calculates the dose standalone, the clinical study evaluates its performance within the user's decision-making process. The document also mentions "Software Validation" as a non-clinical performance test, which would likely include standalone algorithm validation, but specific details of such a study are not provided in terms of performance metrics or acceptance criteria.

    7. The Type of Ground Truth Used

    The "ground truth" for the clinical performance data was continuous glucose monitoring (CGM) data. This is objective physiological measurement data collected from the participants.

    8. The Sample Size for the Training Set

    This information is not provided in the document. The document describes a clinical study as "Summary of Clinical Performance Data" which would be considered the test set for validating the device's clinical impact. There is no mention of a separate training set for the algorithm "training" itself within the scope of this regulatory submission summary. The bolus calculator is based on established algorithms for insulin dose calculation.

    9. How the Ground Truth for the Training Set was Established

    This information is not provided in the document. As mentioned in point 8, the document describes a clinical validation study rather than a de novo algorithm training process. The bolus calculation algorithm is based on programmable settings (correction factor, insulin to carbohydrate ratio, target glucose value) and real-time data (glucose readings, carbs, IOB), which are clinical parameters established through diabetes management principles rather than a "ground truth" derived from a specific training dataset in the AI/ML sense.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191679

    Validate with FDA (Live)

    Date Cleared
    2019-09-20

    (88 days)

    Product Code
    Regulation Number
    880.5730
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K182630, K180045

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Omnipod DASH Insulin Management System (the Pump) with interoperable technology is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin.

    The Pump is able to reliably and securely compatible, digitally connected devices, including automated insulin dosing software, to receive, execute and confirm commands from these devices.

    The Pump is intended for single patient, home use and requires a prescription. The Pump is indicated for use with NovoLog, Humalog, Admelog, or Apidra U-100 insulin.

    Device Description

    The Omnipod DASH™ Insulin Management System with interoperable technology provides for the management of insulin therapy by patients with diabetes mellitus. It is comprised of two primary components: the disposable insulin infusion pump (Pod), and an associated Bluetooth Low Energy (BLE) enabled remote controller. The Omnipod DASH System with interoperable technology is provided with the DASH Personal Diabetes Manager (PDM), but future alternate controllers may be established. The DASH PDM incorporates a suggested bolus calculator which aids the user in determining the insulin bolus dosage needed based on carbohydrates ingested, most recent blood glucose reading, programmable correction factor, insulin to carbohydrate ratio, target blood glucose value and Insulin on Board (IoB).

    The Pod is a body-wearable insulin pump that affixes to the user on the back of the arm, the lower back or abdomen, the thigh area, or any site that has a layer of fatty tissue available. It is held in place by an adhesive pad and provides up to three days of insulin before it is removed and replaced with a new Pod. The DASH PDM is a handheld device that controls the Pod. The user interfaces with the device system through the DASH PDM using a touch screen, similar to a smartphone, where they control basal and bolus delivery and various insulin program settings and calculations. The DASH PDM also has a food library to assist with carbohydrate calculations, and it maintains several variables in a history log for the viewer to track their diabetes therapy. The device system is for prescription use only.

    The remote control design of the Pod inherently enables connectivity to other interoperable controllers with new functionality. Capabilities can be built into compatible controllers to add functionality such as Automated Insulin Delivery (AID) systems. In this design, a controller may contain an algorithm and connect to an iCGM system. In such an integrated system, the AID controller would be responsible for coordinating the interoperable devices (Omnipod DASH and iCGM) in order to automate delivery. It would read the Pod for insulin delivery status, read the iCGM for the sensor value, compute an automated delivery amount and then command the Pod to deliver the required insulin amount. For this automated delivery to occur, the Controller is required to be in range of the Pod. The Pod is designed to the programmed basal rate in the case of extended loss of communication.

    AI/ML Overview

    This document describes the Omnipod DASH™ Insulin Management System with interoperable technology. It is a 510(k) premarket notification for a Class II medical device, an alternate controller enabled infusion pump.

    Key takeaway: This documentation confirms that the Omnipod DASH system is a medical device for insulin delivery, and the submission is for expanding its indications for use to include interoperable technology, allowing it to communicate with compatible digital devices. The FDA has determined it is substantially equivalent to existing devices.

    Here's an analysis based on the provided text, focusing on the requested criteria for an AI/ML device, even though this document is for an infusion pump. It's important to note that the provided text is for an insulin pump, not an AI/ML diagnostic or assistive device. Therefore, many of the requested criteria (like ground truth, MRMC studies, expert adjudication) are not applicable in their typical sense for this type of medical device.

    However, I will interpret the acceptance criteria in the context of a medical device submission, particularly for an infusion pump, and extract information where parallels can be drawn.


    Acceptance Criteria for Omnipod DASH™ Insulin Management System with Interoperable Technology

    The acceptance criteria for this device, as outlined in the "Special Control" section, predominantly focus on the safety and performance of the insulin pump itself, especially given its new interoperable functionality. The performance data presented demonstrates how the device meets these controls, often by stating that the subject device is identical to previous cleared versions (K182630 and K180045) and that previous testing is applicable, with some additional characterization.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria Category (Special Control)Reported Device Performance and Evidence Provided
    Infusion Delivery Accuracy: Device infusion delivery accuracy conforms to defined user needs and intended uses and is validated to support safe use under actual use conditions (includes specifications for ambient temp, pressure, fluid properties; testing met design input requirements for programmable range of rates/volumes, statistically valid number of devices).All performance testing from K182630 and K180045 is applicable. Further characterization conducted and submitted in this 510(k). Delivery accuracy was specifically conducted for this submission (though specific results are not detailed in this summary).
    Hazard Detection: Ability of the pump to detect relevant hazards (occlusions, air in line) within a clinically relevant timeframe across the range of programmable drug delivery rates and volumes. Appropriate performance validation under conditions of use.The subject device is identical to DASH system in K182630/K180045. All testing (except delivery accuracy) was completed to demonstrate hazard detection in K182630 and K180045. Traceability of hazards to risk controls and verification evidence is included in the System Hazard Analysis.
    Drug Compatibility: Compatibility with specified drugs, including assessment of drug stability under reasonably foreseeable use conditions (temp, light, etc.).The subject device is identical to DASH system in K182630/K180045. All validation testing demonstrating compatibility with insulin was provided in K180045 and K182630. Device is only intended for U-100 insulin delivery.
    Biocompatibility: Device parts directly or indirectly contacting the patient must be demonstrated to be biocompatible (chemical, particulate characterization, risk of harm from residues).The subject device is identical to DASH system in K182630/K180045. Biocompatibility of patient-contacting parts (adhesive pad) demonstrated with cytotoxicity, sensitization, skin irritation studies (all passing). Biocompatibility of fluid-path parts demonstrated with cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, material-mediated pyrogenicity, subacute/subchronic toxicity, genotoxicity, implantation, and hemocompatibility (all passing).
    Reliability over Use Life: Evidence verifying and validating that the device is reliable over the ACE pump use life in terms of all device functions and pump performance.The subject device is identical to DASH system in K182630/K180045. No additional shelf life testing was completed for Omnipod DASH with interoperable technology.
    Electrical Safety, EMC, RF Wireless Safety: Designed and tested for electrical safety, electromagnetic compatibility, and radio frequency wireless safety and availability consistent with patient safety requirements.The subject device is identical to DASH system in K182630/K180045. No additional electrical safety and electromagnetic compatibility testing was completed for Omnipod DASH with interoperable technology.
    Cross-channeling (for multi-drug devices): Risk of cross-channeling drugs must be adequately mitigated.Omnipod DASH is only intended for U-100 insulin delivery. (Therefore, cross-channeling is not applicable).
    Reprocessing (for multi-patient use): Validation of reprocessing procedures if intended for multiple patient use, demonstrating all functional and performance requirements are met after reprocessing.Omnipod DASH with interoperable technology is intended only for single-patient use. The Pod is a single-use disposable. The PDM is a non-sterile single-use patient component. (Therefore, reprocessing is not applicable).
    Interoperable Interface Specifications: Validated interface specifications for digitally connected devices (secure authentication, secure/accurate/reliable data transmission, sharing necessary state information, safe operation with out-of-bounds data, detailed process for sharing/validating protocol).The interface between the Omnipod DASH controller (PDM) and pump (DASH Pod) has been specified and validated. Insulet has a detailed process for sharing the pump interface specification and validating correct implementation.
    Critical Event Logging: Record of critical events stored and accessible for auditing of communications and sharing information with responsible parties. Minimum events: drug delivery, commands/confirmations, malfunctions, alarms/alerts, connectivity events.Insulin delivery commands are logged in the PDM's memory. Execution of commands stored in Pod's memory (basal/bolus pulses) and confirmed with PDM. Pod/PDM logs allow auditing of failures (includes all special controls requirements). PDM stores 90 days of alarms/alerts/acknowledgements. PDM stores short duration (~3 days) of communication connectivity. Insulet ensures logging requirements met/verified for external controllers.
    Human Factors Study: Design verification and validation includes human factors study demonstrating safe use by intended user.Not explicitly detailed in this summary, but implied by the statement "No additional shelf life testing was completed for Omnipod DASH with interoperable technology" and the general nature of device clearance. The larger context of a 510(k) submission would typically include this or refer to previous submissions where it was performed.
    Device Labeling: Labeling must include: compatible drugs/concentration, min/max basal/bolus rates/increments, pump accuracy description (bolus/basal at min/intermediate/max), hazard alarm perf (occlusion), wireless QoS, reprocessing instructions (if applicable).Omnipod DASH User Guide updated to contain information from special controls. The performance testing provided in K182630 and K180045 are applicable to the subject device.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes for tests. It generally refers to previous clearances (K182630, K180045) for most testing, stating that the subject device is "identical" or that previous testing is "applicable," with "further characterization" or "delivery accuracy" specifically conducted for this submission.

    • Sample Size: Not explicitly stated for specific tests in this summary. It mentions "statistically valid number of devices" as a requirement for delivery accuracy testing in the special control.
    • Data Provenance: The data comes from the manufacturer's (Insulet Corporation) internal testing and previous FDA submissions (K182630, K180045). The country of origin for the data is not specified but is implied to be within Insulet's testing facilities (Acton, MA, USA is their address). The data would be prospective in nature, generated specifically for these regulatory submissions.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    This information is not applicable in the context of this device. The Omnipod DASH Insulin Management System is an infusion pump, not a diagnostic device that relies on human expert interpretation of data (like medical images) to establish ground truth. The "ground truth" for this device relates to its physical and functional performance (e.g., accurate insulin delivery, proper communication), which is established through engineering and performance testing against predefined specifications.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods like 2+1 or 3+1 are used in studies where human readers are interpreting data or making diagnoses, and their initial assessments need to be reconciled to establish a consensus ground truth. For an infusion pump, performance is measured objectively via test equipment and quantitative outputs.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC study was not done or applicable. MRMC studies compare the performance of human readers, sometimes with and without AI assistance, especially in diagnostic imaging or similar fields. This device is an insulin pump, not a diagnostic tool where human interpretation is being augmented.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    This concept is partially applicable but in a different context. The "algorithm" here is the pump's internal software controlling insulin delivery and communication. The "standalone" performance would be the pump's ability to deliver insulin accurately and communicate reliably based on pre-programmed settings or commands, independent of a human directly monitoring every millisecond. The performance tests (delivery accuracy, hazard detection, interoperability) are essentially evaluations of the pump's "standalone" (or automated) functional performance. The human (user) interacts with the PDM to initiate commands, but the core function (insulin delivery) is automated by the pump's internal algorithms.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is based on engineering specifications, established physical measurement methods, and regulatory standards for insulin pump function and safety.

    Examples:

    • For "infusion delivery accuracy," the ground truth is the precisely measured volume of fluid delivered compared to the commanded volume, determined using calibrated equipment.
    • For "hazard detection," the ground truth is a simulated hazard (e.g., occlusion) and the objective measurement of the time taken for the pump to detect and alarm, compared against predefined safety limits.
    • For "biocompatibility," the ground truth is the absence of adverse biological reactions, confirmed by standardized in-vitro and in-vivo tests according to ISO standards.

    8. The Sample Size for the Training Set

    This is not applicable in the AI/ML sense. This device is a traditional medical device (an infusion pump) with embedded software. It does not learn or get "trained" from data in the way a machine learning algorithm does. Its operation is based on deterministic algorithms and hardware.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable as there is no "training set" for an AI/ML model for this device. The ground truth for its development and validation (e.g., engineering specifications, performance requirements) is established through established medical device development processes, risk analysis, and compliance with national and international standards (e.g., ISO, FDA guidance).