    Intended Use

    FibriCheck is indicated for self-testing by patients who have been diagnosed with, or are susceptible to developing, atrial fibrillation and who would like to monitor and record their heart rhythms on an intermittent basis.

    Device Description

    FibriCheck is a medical device software that can be used to determine heart rhythm conditions, with a primary focus on the detection of atrial fibrillation.

    The FibriCheck Mobile Application is a smartphone application for patients intended to record, display, store and transmit photoplethysmography (PG) data. Heart rhythm measurements are performed by placing the fingertip on the camera of a smartphone device. This way, volumetric changes can be detected.

    The FibriCheck Portal is a web application for healthcare providers intended to display data that was captured by the FibriCheck Mobile Application. Healthcare providers can use the issue FibriCheck prescriptions, to manage their patients and to consult measurement data.

    AI/ML Overview

    This document is a 510(k) Pre-market Notification summary for the FibriCheck device. It outlines the device details, its intended use, comparison to a predicate device, and the non-clinical and clinical tests conducted to demonstrate safety and effectiveness.

    Here’s a breakdown of the information relevant to acceptance criteria and the study proving the device meets them:

    1. Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with specific numerical thresholds for performance metrics. Instead, it states that "The performance metrics are in line with those reported for the predicate device, and did not raise concerns regarding the safety or effectiveness of the subject device." This implies the acceptance criterion was "performance in line with the predicate device," which is K173872, also named FibriCheck.

    The document reports the following general performance statement for the clinical validation study:

    Acceptance Criteria (Implied)Reported Device Performance
    Performance in line with the predicate device (K173872 FibriCheck)Capable of irregular rhythm detection accuracy in line with the predicate device. Performance metrics did not raise concerns regarding safety or effectiveness.

    2. Sample Size and Data Provenance

    • Sample Size for Test Set: 252 cardiology patients.
    • Data Provenance: The patients were "recruited in-clinic or consulting the cardiac outpatient clinic," implying the data was collected prospectively in a clinical setting. The country of origin is not explicitly stated, but given the applicant's address in Belgium, it's highly probable the study was conducted there.

    3. Number of Experts and Qualifications

    The document states, "The participants were instructed to record PPG measurements using the subject device which were compared against the gold-standard reference diagnosis." It does not explicitly specify the number of experts used to establish ground truth or their qualifications. However, given the "gold-standard reference diagnosis" in cardiology and the mention of "cardiology patients," it can be inferred that diagnoses were made by qualified medical professionals, likely cardiologists.

    4. Adjudication Method

    The document does not describe an adjudication method for the test set. It mentions comparison against a "gold-standard reference diagnosis," suggesting a single, definitive ground truth rather than a process requiring adjudication among multiple readers.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    The document does not mention an MRMC study or any comparative effectiveness study involving human readers with and without AI assistance. The study described focuses on the device's standalone performance against a gold standard.

    6. Standalone (Algorithm Only) Performance

    Yes, a standalone performance study was done. The clinical validation study directly compared the PPG measurements from the FibriCheck device (acting as the algorithm) against a "gold-standard reference diagnosis." The statement "The participants were instructed to record PPG measurements using the subject device which were compared against the gold-standard reference diagnosis" highlights the device's algorithmic performance.

    7. Type of Ground Truth Used

    The ground truth used was a "gold-standard reference diagnosis." While the specific nature of this diagnosis (e.g., expert consensus based on ECG, long-term monitoring data, etc.) is not detailed, in cardiology, a "gold-standard reference diagnosis" for atrial fibrillation typically involves a 12-lead electrocardiogram (ECG) or continuous ECG monitoring interpreted by a cardiologist.

    8. Sample Size for Training Set

    The document does not provide information on the sample size used for the training set. It only describes the clinical validation study (test set).

    9. How Ground Truth for Training Set Was Established

    The document does not provide information on how the ground truth for the training set was established, as it does not discuss the training process or dataset.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192415

    Validate with FDA (Live)

    Date Cleared
    2020-01-17

    (135 days)

    Product Code
    Regulation Number
    870.2920
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K182456, K173872

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Study Watch with Irregular Pulse Monitor is indicated for use by adult patients (22 years and older) who have been diagnosed with, or are susceptible to developing, atrial fibrillation enabling them to monitor and record their heart rhythms. Study Watch is also intended to record, store, transfer, and display single-channel electrocardiogram (ECG) rhythms. The Irregular Pulse Monitor is indicated for use in professional healthcare facility environments, while worn as a wrist watch.

    Device Description

    The Study Watch with Irregular Pulse Monitor is a miniaturized physiological data monitoring device that is intended to record, store, transfer and display single-channel electrocardiogram (ECG) rhythms. The Study Watch with Irregular Pulse Monitor is intended to notify the user in the event of irregular pulse, such as atrial fibrillation (AF), and recommend acquisition of an ECG. The device utilizes a band and proprietary watch that incorporates two dedicated sensing electrodes to obtain a single-channel ECG measurement. In practice, a healthcare professional (HCP) may prescribe the Study Watch with Irregular Pulse Monitor to a patient and recommend the capture and transmission of one or more ECG rhythms on a daily basis for analysis by the HCP. The patient uses prompts provided on the graphical user interface (GUI) of the watch to collect a realtime 60-second ECG measurement. The patient is able to capture an ECG rhythm on-demand by navigation to the ECG menu, in response to a daily ECG reminder, or upon receipt of an irregular pulse notification. While an animated ECG waveform is briefly displayed on the GUI of the watch, the patient will not have direct access to the collected wave form. Using the provided charging dock (Cradle) and Verily's Study Hub device, the ECG measurements are securely transferred to Verily's cloud server and are viewed by the HCP using a web portal. The web portal is solely intended for use by the HCP to view time-stamped ECG waveforms collected by the patient and does not include any analysis features. Collectively, the Study Watch consists of the wearable watch and band, the Study Hub (used to transfer data from the watch to the cloud), a Cradle, and the web portal.

    AI/ML Overview

    The document describes the Study Watch with Irregular Pulse Monitor, focusing on its ability to detect irregular pulses, particularly atrial fibrillation (AF). The performance data presented primarily relates to the Irregular Pulse Monitor algorithm, which uses photoplethysmogram (PPG) data as input.

    Here's a breakdown of the acceptance criteria and study findings:

    1. Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaRequirement (Implicit)Reported Device Performance (PPG-based Irregular Pulse Monitor algorithm)
    Sensitivity (Per-interval)Not explicitly stated as a numerical acceptance criterion, but the text states the algorithm "exceeds the pre-specified per-interval performance thresholds for sensitivity."0.85 (85%) with a confidence interval of (0.79-0.90)
    Specificity (Per-interval)Not explicitly stated as a numerical acceptance criterion, but the text states the algorithm "exceeds the pre-specified per-interval performance thresholds for specificity."0.96 (96%) with a confidence interval of (0.93-0.99)

    Note: The document does not provide explicit numerical acceptance criteria values, only that the reported performance exceeds them. These reported values are for the PPG-based irregular pulse monitor algorithm, not necessarily the ECG functionality of the watch which is for recording, storing, transfer, and display.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not explicitly state a numerical sample size for the test set of the clinical study. It mentions the study "collected Study Watch PPG waveform (Test) data in clinic from patients with a history of AF."
    • Data Provenance: The document does not specify the country of origin. It indicates the data was collected "in clinic," implying a prospective clinical study specifically for this device's performance validation.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • The document states that "contemporaneous Holter ECG (Reference) data" was used as "ground truth." It does not mention the use of human experts to establish this ground truth for the test set. Holter ECG data is considered an objective gold standard for rhythm analysis.

    4. Adjudication Method for the Test Set

    • The document implies that the ground truth was established by "contemporaneous Holter ECG (Reference) data," which would not require a human adjudication method in the traditional sense (e.g., 2+1, 3+1). The Holter ECG itself serves as the reference.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • The document does not describe an MRMC comparative effectiveness study where human readers improve with AI vs. without AI assistance. The study focuses purely on the standalone performance of the irregular pulse monitor algorithm against a clinical reference standard (Holter ECG).

    6. Standalone (Algorithm Only) Performance

    • Yes, a standalone performance evaluation was done. The reported sensitivity and specificity values of 85% and 96%, respectively, are for the "PPG-based Irregular Pulse Monitor algorithm" operating on the Study Watch PPG data, with Holter ECG as the ground truth. This indicates algorithm-only performance.

    7. Type of Ground Truth Used

    • The ground truth used for the clinical study was contemporaneous Holter ECG data. This is a clinical gold standard for continuous heart rhythm monitoring.

    8. Sample Size for the Training Set

    • The document does not provide information regarding the sample size used for the training set of the algorithm.

    9. How the Ground Truth for the Training Set Was Established

    • The document does not provide information on how the ground truth for the training set was established. Given the nature of a machine learning algorithm for pulse monitoring, it's highly probable that ground truth for training data would also have been established using reference ECGs or similar clinical gold standards.