Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K251670

    Validate with FDA (Live)

    Date Cleared
    2026-01-26

    (241 days)

    Product Code
    Regulation Number
    870.2340
    Age Range
    0 - 21
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K173830, K210560

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indications for Use:
    The MAC 7 Resting ECG Analysis System is a non-invasive prescription device.

    • The device is indicated for use to acquire, analyze, display and print electrocardiograms.
    • The device is indicated for use to provide interpretation of the data for consideration by a physician.
    • The device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician. It is not intended as a sole means of diagnosis.
    • The interpretations of ECG offered by the device are only significant when used in conjunction with a physician over-read as well as consideration of all other relevant patient data.
    • The device is indicated for use on adult and pediatric (birth through 21 years of age) populations.

    Intended Use:
    The MAC 7 Resting ECG Analysis System is intended to acquire, analyze, display, and record electrocardiographic information from adult or pediatric populations. Basic system simultaneously acquires data from each lead. Once the data is acquired, it can be analyzed, reviewed, stored, printed or transmitted. Transmission and reception of ECG data and other clinical data to and from a central clinical information system is optional.

    The MAC 7 Resting ECG Analysis System is intended to be used under the direct supervision of a licensed healthcare practitioner, by trained operators in a hospital, medical professional's facility or wherever ECG testing is performed.

    Device Description

    The MAC 7 Resting ECG Analysis System is a mobile electrocardiograph designed to acquire, analyze, display, and record ECG signals from surface ECG electrodes.

    The device can capture 3, 6, 12 or 15 lead electrocardiograms, provide interpretive analysis, and print reports.

    The device can connect to a network, either through a wired LAN connection or via wireless WiFi access points. Once on the network, the device can optionally interface with cardiology information systems such as the GEHC MUSE® system to participate in a complete electrocardiology workflow.

    The device provides state-of-the-art information technology security features and a contemporary user interface. Mobility is provided via an optional trolley.

    AI/ML Overview

    The FDA 510(k) clearance letter for the MAC 7 Resting ECG Analysis System (K251670) does not contain a specific study proving the device meets acceptance criteria. Instead, it establishes substantial equivalence to predicate devices (K203786, K173830, K210560) based on similarities in intended use, indications for use, technology, and performance, along with compliance with voluntary standards and non-clinical testing.

    Therefore, the following information is extracted from the provided text to fulfill your request:

    1. Acceptance Criteria and Reported Device Performance

    The document describes the device's characteristics and compares them to predicate devices, demonstrating substantial equivalence rather than explicit acceptance criteria with numerical performance targets. The "Discussion of Differences" column often highlights that a change does not significantly affect substantial equivalence, implying that the performance remains acceptable.

    SpecificationPredicate Product: MAC 7 Resting ECG Analysis System (K203786)Proposed Product: MAC 7 Resting ECG Analysis SystemReported Device Performance (as implied by "Discussion of Differences")
    Intended UseAs described in the predicateAs described in the proposed productEquivalent: "The change in the intended use statement reflects the flexibility of the system without impacting the core functionality or safety profile." and "The change in the intended use statement doesn't alter the substantial equivalence of the device."
    Indications for UseAs described in the predicateAs described in the proposed productEquivalent: "The updated language to include healthcare practitioner broadens the description to reflect current clinical practices without altering the device's safety or performance."
    ContraindicationsAs described in the predicateAs described in the proposed productIdentical
    Patient PopulationAdult and pediatric (birth through 21 years of age), with ACS interpretation exception < 16 years.Adult and pediatric (birth through 21 years of age), with Lead reversal detection exception ≤ 15 years and ACS interpretation exception < 16 years.Substantial Equivalent: "The subject device maintains full alignment with the adult and pediatric patient population indications of the predicate and reference devices." and "Age-related limitations for Lead Reversal Detection and ACS interpretation are consistent with those of the predicate/reference devices and do not alter the overall patient population equivalence."
    Environment of UseAs described in the predicateAs described in the proposed productIdentical
    Patient Acquisition CircuitryIntegrated in the device, digitalizing functions provided by the device.Integrated in the device, digitalizing functions provided by the device.Identical (for comparison with MAC 7 predicate); Equivalent: "The proposed device uses the same acquisition module as the reference device. The only difference is that the proposed product uses a standard USB port. These differences do not affect the substantial equivalence of the device." (for comparison with MAC VU360 reference)
    Interpretive ECG AnalysisYesYesIdentical
    Critical ValuesIdentified, indicated via dialog box and printed report; user acknowledgement required.Identified, indicated via dialog box and printed report; user acknowledgement required.Identical
    ECG Pacemaker Detection and HD PaceDigital detection, separate printable/viewable channel; 12SL disabled for acquisition module detections.Digital detection, separate printable/viewable channel (configurable on/off, default enabled); 12SL disabled for acquisition module detections.Substantial Equivalent: "The change involves adding the ability to configure the separate pacemaker pulses channel to enable or disable detection, with the default setting being enabled. This modification provides additional flexibility without affecting the device's core functionality or safety."
    Frequency Response0.04 to 150Hz0.04 to 300Hz (default 0.04 to 150Hz)Equivalent: "The proposed product expands bandwidth support from 150 to 300Hz as included in the K221321. There was no change in measurements or accuracy... The difference does not significantly affect substantial equivalence."
    Prior ECGNot supportedDownload, review, print recent previous ECG from same patient.Equivalent: "The proposed MAC 7 can download the most recent previous ECG from the ECG management server for the same patient. This change has been verified to not significantly impact substantial equivalence." (for MAC 7 predicate); Equivalent: "The design of the Prior ECG feature uses current patient identification to query historical ECG data from the management system, enabling the retrieval and comparison of the most recent previous ECG with the current one from the same patient. The output of this process is consistent between the proposed and reference device. The only notable difference is in the labelling of the printed report, where MAC 7 V2 displays "Prior ECG" while TC30 uses "Previous ECG"... The labelling difference does not alter the functionality, or performance of the feature. Therefore, the addition of this feature does not impact the substantial equivalence of the proposed device." (for TC30 predicate)
    Display type, size, resolution, and information10 inch diagonal LCD, 1280 x 800, displays patient name, lead label, patient I.D., heart rate, date/time.10 inch diagonal LCD, 1280 x 800, displays patient name, lead label, patient I.D., heart rate, date/time.Identical
    Battery OperationRechargeable and user replaceableRechargeable and user replaceableIdentical
    Recorder MethodThermal dot arrayThermal dot arrayIdentical
    Number of ChannelsSelectable 3, 6, or 12 channels + pace annotationSelectable 3, 6, 12 or 15 channels + pace annotationEquivalent: "The proposed device supports up to 15 channels due to the addition of three more electrodes and their corresponding signal acquisition."
    Thermal Paper sizeA4 or Letter format, thermal paper Z-foldA4 or Letter format, thermal paper Z-foldIdentical
    Network Printer OptionNot supportedSupport to print report via network printer"The contents of the network printer reports are the same as thermal printer reports." (Implies acceptable performance by producing identical reports)
    eDeliveryNot supportedSupport for self-registration, activation, and software update notifications."This is a service feature which make it easier to deliver new software version to customer, it does not affect substantial equivalence." (Implies acceptable performance as it doesn't impact core function)
    RSvPNot supportedSupport to upload service snapshot to remote server."This is a service feature which make it easier to get device data for trouble shooting, it does not affect substantial equivalence." (Implies acceptable performance as it doesn't impact core function)
    Interpretation Statements12SL™ analysis algorithm (v23.1) for 10 seconds ECG.12SL™ analysis algorithm (v24) for 10 seconds ECG.Equivalent: "Interpretive Statements are provided by 12SL (v24) ECG Analysis Program which was previously cleared under K221321... The difference does not significantly affect substantial equivalence." (Implies performance is acceptable as per prior clearance)
    Lead Reversal DetectionLimb lead reversal detectionLimb lead and chest lead reversal detection.Equivalent: "The primary change involves an algorithm update from 12SL (v23.1) to 12SL (v24), which allows the detection of additional lead reversals without altering the core substantial equivalence of the device. The MAC 7 interface presents the new detection capability, without compromising the device's safety and performance."
    Acute Coronary Syndrome (ACS)Provides interpretation statement for ACS.Provides interpretation statement for ACS.Identical
    Dimensions and Weight40 x 32 x 21 cm, 5.2 Kg40 x 32 x 21 cm, 5.2 KgIdentical

    2. Sample Size Used for the Test Set and Data Provenance

    The document states: "Summary of Clinical Tests: The subject of this premarket submission, MAC 7 Resting ECG Analysis System, did not require clinical studies to support substantial equivalence." This indicates that no specific test set data from clinical studies was used for performance evaluation in this submission. The "acceptance" is based on the device's technical characteristics aligning with or improving upon those of legally marketed predicate devices, supported by non-clinical testing and previous clearances for core components (like the 12SL™ analysis algorithm v24).

    Therefore, details on sample size, country of origin, or retrospective/prospective nature of a clinical test set are not available in this document.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    As no clinical studies were required for this submission to support substantial equivalence, there is no information provided regarding experts establishing ground truth for a test set.

    4. Adjudication Method for the Test Set

    Given that no clinical studies were performed, there is no adjudication method described for a test set.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The document does not mention any MRMC comparative effectiveness study. The focus is on establishing substantial equivalence to existing devices, not on demonstrating improved human reader performance with AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    The document states that the "Interpretive Statements are provided by 12SL (v24) ECG Analysis Program which was previously cleared under K221321." This implies that the performance of the 12SL™ algorithm itself (a standalone interpretation algorithm) would have been assessed during its prior clearance (K221321). However, the details of that standalone performance study are not included in this K251670 submission.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    Since no new clinical studies were performed for this submission, there is no mention of the type of ground truth used for a test set. For the 12SL™ analysis algorithm (v24) which provides interpretation statements, the ground truth would have been established during its prior clearance (K221321), but those details are not provided here.

    8. The Sample Size for the Training Set

    The document does not provide information on the sample size for any training set. As noted, the approval is based on substantial equivalence and non-clinical testing rather than specific training data for a new algorithm.

    9. How the Ground Truth for the Training Set was Established

    The document does not provide information on how ground truth was established for any training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231010

    Validate with FDA (Live)

    Device Name
    Corvair
    Manufacturer
    Date Cleared
    2024-06-07

    (427 days)

    Product Code
    Regulation Number
    870.1025
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K081437, K110266, K173830, K232035

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AliveCor's Corvair ECG analysis system assists the healthcare professional (HCP) in measuring and interpreting resting diagnostic ECGs for rhythm and morphological information by providing an initial automated interpretation. The interpretation by the analysis program may then be confirmed, edited, or deleted by the HCP. The analysis program is intended for use in the general population ranging from healthy subjects to patients with cardiac abnormalities. Corvair is intended for use by healthcare professionals, or trained personnel in healthcare facilities (e.g. the doctor's office or hospital) and in acute settings.

    Corvair analyses should be used only as an adjunct to clinical history, symptoms, and the results of other non-invasive and/or invasive tests. Corvair analyses are considered unconfirmed and must be reviewed by a qualified physician. The provisional automated ECG analysis should not be used for clinical action if it has not been reviewed by a qualified healthcare professional capable of independently interpreting the ECG signal.

    Device Description

    Corvair is Software as a Medical Device (SaMD) intended for use by healthcare professionals to analyze a diagnostic-bandwidth ECG. Corvair analyzes a 10-second ECG and provides rhythm analysis, morphological analysis, and ECG interval estimation. Corvair provides 35 separate determinations with 14 rhythm and 21 morphology determinations. Rhythm determinations include Normal Sinus Rhythm, Atrial fibrillation, Atrial flutter, Paced Rhythm, Junctional Rhythm, and Bigeminy, with the modifiers of 1* Degree AV Block, Higher Degree AV Block (including 2nd and 3rd degree AV blocks), Sinus Arrhythmia, Marked Sinus Arrhythmia, Marked Bradycardia, Sinus Tachycardia, and PVCs. Morphology determinations include Intraventricular block (RBBB, LBBB, and Other Intraventricular Block), Hypertrophy (LVH, and RVH), Atrial Enlargement (LAE and RAE), Acute Myocardial Infarction (Anterior MI, Inferior MI, Lateral MI), Old/Previous Myocardial Infarction (Anterior Old MI, Inferior Old MI, Lateral Old MI), Ischemia (Anterior, Inferior, Lateral), Prolonged QT, Paced ECG, Other Morphological Defects (Early Repolarization, Wolff-Parkinson-White Syndrome (WPW)), and Normal or Otherwise Normal. Rhythm and morphology determinations are overlapping. i.e., an ECG could receive multiple rhythm and morphology determinations (e.g., Sinus Rhythm, Acute MI). The device also provides global ECG measurements (PR, ORS, OT, OTcB, OTcF, and Heart Rate). No beat-level analysis is provided by the device. Corvair may fail to detect or misidentify conduction system pacing and demand pacing. Corvair does not detect sinus pause. While Corvair provides PR interval estimation and does detect WPW, it does not have a separate determination of abnormally short PR intervals.

    This SaMD provides these capabilities in the form of an Application Program Interface (API) library. Any software or device ("target device") can incorporate the Corvair API library into its device software to provide users with resting ECG analytics. The input ECG is provided by the target device to Corvair, to which the various Corvair algorithms are applied, and outputs generated accordingly. Corvair has a C++ interface and a distributed binary (library), which is used by the target device to statically link to Corvair. Viewing of Corvair's ECG analysis is handled by the target device.

    Corvair is intended to be used with standard diagnostic-bandwidth, resting ECG recordings collected using 'wet' Ag/AgCl electrodes with conductive gel/paste. Corvair only requires 4 ECG leads for analysis, specifically, either Leads {I, II, V2, and V4}, or Leads {I, II, V1, and V4}. Compatible devices include resting ECGs from GE Medical Systems® (e.g., K081437, MAC 1600, K110266, MAC 5500, K173830, MAC VU360, etc.), and AliveCor's Impala (K232035). Regardless of the lead configuration, Corvair provides the same set of rhythm, morphological, and interval determinations. Corvair has two modes of operation, Symptomatic Mode, which is used when the pre-test probability for a specific rhythm or morphology is high, and Asymptomatic Mode, which optimizes the PPV, by optimizing the specificity, to detect the various rhythms and morphologies. The target device can choose which lead set and which mode of determinations to utilize based on the target clinical application and the patient's clinical presentation.

    Corvair utilizes several deep neural networks (DNNs) for its analysis. These DNNs were trained on a dataset of approximately 1 million 12-Lead ECGs acquired from about 400K clinical patients at the Emory University Hospital over several decades between 1985 and 2010. Each ECG has a physician overread confirmed diagnosis with multiple diagnostic codes. The dataset had a 52%/48% ratio of ECGs from male and female patients, respectively. The average age of the patient was 61.3 ± 16. The dataset included 56% white, 33% African American, 2.2% Asian, 9% other races/ethnicities.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document states that Corvair was evaluated against a large set of ECGs and compared its analysis output against a known reference using standard ECG performance metrics. These outputs were evaluated against clinically relevant acceptance criteria. However, the specific numerical acceptance criteria for sensitivity, specificity, PPV, and error margins for interval outputs are not explicitly stated in the provided text. The text only mentions that acceptable performance was demonstrated.

    Table of Acceptance Criteria and Reported Device Performance (Summary based on text):

    Performance MetricAcceptance Criteria (Not explicitly stated numerically)Reported Device Performance
    Interpretive Outputs (Rhythm & Morphology)Clinically relevant acceptance criteria for sensitivity, specificity, and PPV.Performance demonstrated as effective and substantially equivalent to predicate.
    Interval Outputs (PR, QRS, QT)Clinically relevant acceptance criteria for mean error and standard deviation of error.Performance demonstrated as effective and substantially equivalent to predicate using CSEDB and AliveCor proprietary datasets.
    Heart Rate AccuracyClinically relevant acceptance criteria for mean absolute error.Performance demonstrated as effective and substantially equivalent to predicate using AliveCor proprietary dataset.

    2. Sample size used for the test set and the data provenance

    • Test Set Size: The document states that Corvair was evaluated "against a large set of ECGs" and mentions "additional large validation datasets" created from sites independent of the training data. For PR, QRS, QT interval estimation, the Common Standards for Quantitative Electrocardiography Standard Database (CSEDB) and an AliveCor proprietary dataset developed from ECGs collected in a clinical study at the Mayo Clinic's Genetic Heart Rhythm Clinic were used. For Heart Rate and QTcF validation, the AliveCor proprietary dataset was also used. The exact numerical sample sizes for these test sets are not explicitly provided.
    • Data Provenance:
      • Training Data: Approximately 1 million 12-Lead ECGs acquired from about 400,000 clinical patients at the Emory University Hospital over several decades (1985-2010).
      • Test Data:
        • CSEDB (Common Standards for Quantitative Electrocardiography Standard Database) - an established public database.
        • AliveCor proprietary dataset from clinical study at Mayo Clinic's Genetic Heart Rhythm Clinic.
        • Additional large validation datasets from sites independent of the training data.
      • Retrospective/Prospective: The Emory University Hospital data (1985-2010), used for training, is retrospective. The Mayo Clinic data used for the proprietary dataset likely has a prospective component if collected specifically for this study, but the text doesn't explicitly state its collection method. The "additional large validation datasets" are not detailed regarding their collection method.
      • Country of Origin: Emory University Hospital and Mayo Clinic are in the United States.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document mentions that the training dataset ECGs "each has a physician overread confirmed diagnosis with multiple diagnostic codes." For the test sets, the ground truth is against a "known reference." While it refers to "physician overread confirmed diagnosis" for the training set, it does not explicitly state the number or specific qualifications of experts who established the ground truth for the test set. For CSEDB, the ground truth is part of the established database, which typically involves expert consensus. For the AliveCor proprietary dataset, the text implies a clinical study setting, but details on ground truth establishment by experts are missing.

    4. Adjudication method for the test set

    The document does not explicitly state the adjudication method used for establishing the ground truth of the test set (e.g., 2+1, 3+1, none). It refers to "physician overread confirmed diagnosis" for the training data and "known reference" for the test data, implying an established ground truth, but the process of its establishment is not detailed.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and if so, what was the effect size of how much human readers improve with AI vs without AI assistance.

    The document explicitly states: "No clinical testing was required or conducted to support a determination of substantial equivalence." This indicates that an MRMC comparative effectiveness study was not performed for this submission. The device is intended to assist healthcare professionals, providing an "initial automated interpretation" that "may then be confirmed, edited, or deleted by the HCP," but its impact on human performance was not part of this specific submission's evidence.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done.

    Yes, a standalone performance evaluation was done. The nonclinical performance testing sections describe evaluating Corvair's analysis output against a known reference using standard ECG performance metrics (sensitivity, specificity, PPV, mean error, standard deviation of error, mean absolute error). This directly assesses the algorithm's performance without a human in the loop. The device provides its analysis as an API library, which integrates into other software, implying a standalone analysis capability.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the training set was based on "physician overread confirmed diagnosis with multiple diagnostic codes." For the test sets, it was against a "known reference." For the CSEDB, it's generally accepted expert consensus reference data. For the proprietary dataset from Mayo Clinic, it likely involves clinical diagnoses and expert review, but the specific process (e.g., expert consensus vs. single physician overread) is not detailed.

    8. The sample size for the training set

    The training set consisted of approximately 1 million 12-Lead ECGs acquired from about 400,000 clinical patients.

    9. How the ground truth for the training set was established

    The ground truth for the training set was established through "a physician overread confirmed diagnosis with multiple diagnostic codes" for each ECG. This indicates that medical professionals reviewed and confirmed the diagnoses, which were then used as the labels for training the deep neural networks.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203786

    Validate with FDA (Live)

    Date Cleared
    2021-04-30

    (123 days)

    Product Code
    Regulation Number
    870.2340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    MAC VU360 Resting ECG Analysis System (K173830)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MAC 7 Resting ECG Analysis System is a non-invasive prescription device.
    • The device is indicated for use to acquire, analyze, display and print electrocardiograms.
    • The device is indicated for use to provide interpretation of the data for consideration by a physician.
    • The device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician. It is not intended as a sole means of diagnosis
    • The interpretations of ECG offered by the device are only significant when used in conjunction with a physician over-read as well as consideration of all other relevant patient data.

    Device Description

    The MAC 7 Resting ECG Analysis System is a mobile electrocardiograph designed to acquire, analyze, display, and record ECG signals from surface ECG electrodes.
    The MAC 7 can capture 3, 6 or 12 lead electrocardiograms, provide interpretive analysis, and print reports.
    The device can connect to a network, either through a wired LAN connection or via wireless WiFi access points. Once on the network, the device can optionally interface with the cardiology information systems such as the GEHC MUSE®
    The device provides state-of-the-art information technology security features and a contemporary user interface. Mobility is provided via an optional trolley.

    AI/ML Overview

    The provided text doesn't contain specific acceptance criteria or details about a study proving the device meets those criteria. The 510(k) summary indicates that no clinical studies were required to support substantial equivalence for the MAC 7 Resting ECG Analysis System.

    However, it does state that the device underwent non-clinical tests to demonstrate compliance with applicable voluntary standards and quality system regulations.

    Therefore, for your request, the following information is not available in the provided document:

    1. A table of acceptance criteria and the reported device performance.
    2. Sample size used for the test set and the data provenance.
    3. Number of experts used to establish the ground truth for the test set and their qualifications.
    4. Adjudication method for the test set.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, or the effect size of human improvement with AI vs without AI assistance.
    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.) for the test set.
    8. The sample size for the training set.
    9. How the ground truth for the training set was established.

    The document explicitly states:
    "Summary of Clinical Tests: The subject of this premarket submission, MAC 7 Resting ECG Analysis System, did not require clinical studies to support substantial equivalence."

    This indicates that the FDA's clearance was based on non-clinical testing and substantial equivalence to a predicate device (MAC VU360 Resting ECG Analysis System K173830), rather than on a clinical study with specific acceptance criteria for diagnostic performance. The substantial equivalence argument relies on the new device employing "the same fundamental scientific technology, basic design, construction, materials, energy source, control mechanism, operating principles as the predicate device MAC VU360".

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1