510(k) Data Aggregation

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VersaTap™ Suture Anchor is intended to be used for soft tissue fixation during general orthopedic surgery. The VersaTap™ Suture Anchor is intended for use in arthroscopic or open surgical approaches for fixation of soft tissue and ligaments to bone/healthy tissue during tendon and ligament repairs, during procedures such as:
    • Shoulder: Rotator cuff repair, biceps tenodesis, SLAP repair, Bankart repair
    • Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus
    • Elbow: Tennis elbow repair
    • Knee: Medial and lateral collateral ligament repair.
    • Wrist: Scapholunate ligament reconstruction.

    The VersaPEEK™ Suture Anchor is intended to be used for soft tissue fixation during general orthopedic surgery. The VersaPEEK™ Suture Anchor is intended for use in arthroscopic or open surgical approaches for fixation of soft tissue and ligaments to bone/healthy tissue during tendon and ligament repairs, during procedures such as:
    • Shoulder: Rotator cuff repair, biceps tenodesis, SLAP repair, Bankart repair
    • Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus
    • Elbow: Tennis elbow repair
    • Knee: Medial and lateral collateral ligament repair.
    • Wrist: Scapholunate ligament reconstruction.

    The MicroTi™ Suture Anchor is intended to be used for soft tissue fixation during general orthopedic surgery. The MicroTi™ Suture Anchor is intended for use in arthroscopic or open surgical approaches for fixation of soft tissue and ligaments to bone/healthy tissue during tendon and ligament repairs, during procedures such as:
    • Elbow: Ulnar/Medial Collateral Ligament Repair
    • Foot/Ankle: Achilles Tendon Repair, Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Mid-foot Reconstruction, Hallux Valgus Reconstruction, Metatarsal Ligament Repair.
    • Hand/Wrist: Scapholunate Ligament Reconstruction and Ulnar/ Radial Collateral Ligament Reconstruction.

    The VersaLat™ Ti Suture Anchor is intended to use for fixation of soft tissue and ligaments to bone/ healthy tissue during tendon and ligament repairs, during procedures such as:
    • Shoulder: Rotator Cuff Repair, Biceps Tenodesis
    • Elbow: Biceps Tendon Reattachment, Ulnar/Radial Collateral Ligament Reconstruction.
    • Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, and Iliotibial Band Tenodesis
    • Foot/Ankle: Lateral stabilization, medial stabilization, Achilles tendon repair, mid-foot reconstruction, hallux valgus reconstruction, metatarsal ligament repair, metatarsal tendon repair, bunionectomy, and digital tendon transfers.

    The VersaLat™ Suture Anchor is intended to use for fixation of soft tissue and ligaments to bone/ healthy tissue during tendon and ligament repairs, during procedures such as:
    • Shoulder: Rotator Cuff Repair, Biceps Tenodesis
    • Elbow: Biceps Tendon Reattachment, Ulnar/Radial Collateral Ligament Reconstruction.
    • Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, and Illiotibial Band Tenodesis.
    • Foot/Ankle: Lateral stabilization, medial stabilization, Achilles tendon repair, mid-foot reconstruction, hallux valgus reconstruction, metatarsal ligament repair, metatarsal tendon repair, bunionectomy, and digital tendon transfers.

    Device Description

    The VersaTap™ Suture Anchor is a self-tapping suture anchor. This anchor is designed to combine the advantages of both PEEK and Titanium. The titanium tip makes the product self-tapping, whilst the majority of the anchor is manufactured from PEEK to minimize post-operative imaging effects.

    The VersaPEEK™ Suture Anchor is manufactured completely from PEEK. This material composition minimizes post-operative imaging effects. The suture anchor is designed for ultimate mechanical properties (pullout strength, tensile strength, etc). The VersaPEEK™ Suture Anchor also has a variety of suture/suture tape configurations.

    The MicroTiTM Suture Anchor is a small screw-in suture anchor used in a variety of small-joint applications. Despite its small diameter, the specially engineered thread combines with cortical bone to provide tremendous pull-out strength.

    The VersaLat™ Ti Suture Anchor is a knotless-/suture anchor. This titanium screw-in anchor can be combined with the VersaTap™ Suture Anchor in double rotator cuff repair surgery. Flexibility is what differentiates this anchor from its competitors with surgeons being able to fixate any number of sutures/tapes (K150438) by adjusting the size of the characteristic front loop. The extra suture provided by this anchor can also be used for the fixation of the biceps tendon or any ligament fragments.

    The VersaLat™ Suture Anchor is a knotless-/suture anchor. This titanium screw-in anchor can be combined with the VersaTap™ Suture Anchor in double rotator cuff repair surgery. Flexibility is what differentiates this anchor from its competitors with surgeons being able to fixate any number of sutures/tapes (K150438) by adjusting the size of the characteristic front loop. The extra suture provided by this anchor can also be used for the fixation of the biceps tendon or any ligament fragments.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for several suture anchor devices (VersaTap™, VersaPEEK™, MicroTi™, VersaLat™ Ti, and VersaLat™ Suture Anchors). This document focuses on demonstrating substantial equivalence to previously cleared predicate devices through non-clinical testing. It does not involve AI/ML technology, imaging, or human reader studies.

    Therefore, many of the requested criteria, such as "acceptance criteria and reported device performance" in the context of an AI/ML study, "sample size for the test set," "number of experts used to establish ground truth," "adjudication method," "MRMC comparative effectiveness study," "standalone performance," and "training set details," are not applicable to this type of submission.

    The study described here is a mechanical and biocompatibility assessment of medical devices, not an AI/ML performance study.

    Here's an adaptation of the requested information based on the provided document:


    Acceptance Criteria and Study for Suture Anchors (Non-AI/ML Medical Device)

    The provided submission, K242296, for the VersaTap™, VersaPEEK™, MicroTi™, VersaLat™ Ti, and VersaLat™ Suture Anchors, is a 510(k) Premarket Notification. The purpose of this submission is to demonstrate substantial equivalence to legally marketed predicate devices, not to prove performance of an AI/ML algorithm or derive clinical effectiveness from human reader studies.

    The acceptance criteria and supporting studies are based on non-clinical testing to ensure the new devices are as safe and effective as their predicates.

    1. Table of Acceptance Criteria (for Non-Clinical Testing) and Reported Device Performance

    Acceptance Criteria CategorySpecific Test/RequirementAcceptance Criteria (Implicit)Reported Device Performance and Conclusion
    Mechanical PerformanceInsertion Torque TestingPerformance comparable to or better than the predicate device to ensure proper implantation and stability.Substantial equivalence supported by the results of mechanical testing, including insertion torque. The devices support safety and effectiveness. (Details not explicitly quantified in this summary, but determined to be acceptable).
    Static Pullout TestingPerformance comparable to or better than the predicate device to ensure adequate fixation strength.Substantial equivalence supported by the results of mechanical testing, including static pullout. The devices support safety and effectiveness. (Details not explicitly quantified in this summary, but determined to be acceptable).
    Dynamic Pullout (Fatigue)Justification for not performing dynamic pullout testing explicitly, demonstrating it's unnecessary to prove equivalence or safety.Justification for dynamic pullout (fatigue) testing provided according to FDA guidance.
    Component InterconnectionAssurance of proper and secure connection of device components.Justification for component interconnection provided.
    Corrosion TestingMaterials and design ensure resistance to corrosion in the body environment.Justification for corrosion testing provided.
    BiocompatibilityMaterial CompositionMaterials must be biocompatible and safe for implantation, consistent with previously cleared devices. Performance must be non-pyrogenic.Subject devices manufactured using the same approved suppliers, manufacturing facilities, processes, chemicals, materials, and cleaning processes as predicate devices. All materials and processes have been used in previously cleared devices. Tested to be non-pyrogenic. Biocompatibility is substantially equivalent.
    SterilitySterilityDevices must achieve and maintain sterility for their shelf-life.Methods, thresholds, and criteria for sterility are identical to the predicate device.
    Shelf-LifeShelf-LifeDevices must maintain their properties and safety over their intended shelf-life.Methods, thresholds, and criteria for shelf-life are identical to the predicate device.
    PackagingPackaging IntegrityPackaging must protect the device and maintain sterility until use.Methods, thresholds, and criteria for packaging are identical to the predicate device.
    Quality ControlManufacturing QAManufacturing processes adhere to established quality systems (ISO 13485, ISO 9001).All manufacturing is conducted according to processes defined by ISO 13485 and ISO 9001. Methods, thresholds, and criteria for quality control measures are identical to the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    This is not an AI/ML study with a "test set" of data in the traditional sense. It's a medical device submission based on non-clinical bench testing and comparison to predicates.

    • Sample Size:
      • For mechanical testing (insertion torque, static pullout), specific sample sizes would have been used per test standards (e.g., ASTM, ISO). These exact numbers are not detailed in the summary document but are implied to be sufficient for demonstrating equivalence.
      • For biocompatibility, the assessment relies on the prior clearance of the same materials and processes in predicate devices. New testing (e.g., non-pyrogenic) would involve a sample size relevant to that specific test.
    • Data Provenance: The data provenance is from bench laboratory testing conducted by Ortho-Design (Pty) Ltd in South Africa. It is prospective testing of the new devices and comparison to the known performance and characteristics of the already cleared predicate devices.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • This is not applicable as this is not an AI/ML study involving human interpretation of data where "ground truth" is established by experts in an annotation or diagnostic capacity.
    • The "ground truth" in this context is established by engineering and biocompatibility standards and the documented performance of legally marketed predicate devices, as determined by mechanical and chemical testing.

    4. Adjudication Method for the Test Set

    • Not applicable. There is no human interpretation or diagnostic "test set" requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • No, an MRMC comparative effectiveness study was not done. This type of study is specifically relevant to AI/ML software where the aim is to evaluate how AI assistance impacts human reader performance (e.g., in medical imaging diagnosis). This submission pertains to physical medical devices (suture anchors) and their mechanical/biocompatibility characteristics.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    • Not applicable. This is not an AI/ML algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for demonstrating substantial equivalence for these suture anchors is established through:

    • Engineering Standards: Adherence to recognized national and international standards for mechanical testing of bone anchors (e.g., those referenced in the FDA Guidance: 'Bone Anchors - Premarket Notification (510(k)) Submissions').
    • Predicate Device Performance: The established safety and effectiveness profile of the legally marketed predicate devices (cleared in K212381), which serve as the benchmark for equivalence.
    • Material Biocompatibility Data: Established safety profiles of the materials (PEEK, Titanium) as used in previously cleared medical devices, and specific biocompatibility testing (e.g., pyrogenicity) performed on the subject devices.

    8. The Sample Size for the Training Set

    • Not applicable. This is not an AI/ML study involving a "training set" of data.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. As there is no "training set," there is no ground truth established for it.
    Ask a Question

    Ask a specific question about this device

    K Number
    K192532

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-10-19

    (33 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K173788, K180594, K082810, K101679

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Arthrex SwiveLock Anchor is intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures:
    • Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
    · Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus reconstruction, Mid-foot reconstruction, Metatarsal Ligament Repair, Tendon Repair, Bunionectomy
    • Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis

    • · Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction
    • · Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondy.itis Repair
      · Hip: Capsular Repair, Acetabular labral repair
    Device Description

    The Arthrex SwiveLock Anchor is a sterile two-component suture anchor comprised of an eyelet and a hollow anchor body. The Arthrex SwiveLock Anchor is pre-mounted on a driver with the anchor body and eyelet physically separated on the driver shaft. Arthrex 510(k) cleared suture may also be provided with the device.

    AI/ML Overview

    This document describes the FDA clearance for the Arthrex SwiveLock Anchor. As such, the information provided focuses on the regulatory review process and the basis for substantial equivalence to predicate devices, rather than a clinical study evaluating the performance of a medical device against specific acceptance criteria.

    Therefore, the specific information requested in your prompt regarding acceptance criteria and a study to prove the device meets these criteria (including sample sizes, expert involvement for ground truth, adjudication methods, MRMC studies, standalone performance, training sets, etc.) is not present in the provided text.

    The document primarily states:

    • Performance Data: "Mechanical testing demonstrated that the pull-out (tensile) strength of the proposed Arthrex SwiveLock Anchor met the criteria established by the predicate device and that the anchor was successfully inserted into the bone substitute. Geometric analysis and Load vs Degradation Time comparison was conducted to demonstrate that the expected degradation and pull out strength of the proposed anchor is not a new worst case over the predicate and referenced predicate devices. Bacterial endotoxin per USP <161>, USP <85>, FDA Guidance for Industry Pyrogen & Endotoxin Testing and EP 2.6.14 was conducted to demonstrate that the device meets pyrogen limit specifications."

    This indicates that mechanical and biocompatibility tests were performed, and the results were compared against established criteria or predicate device performance. However, it does not provide the quantitative acceptance criteria, the specific methodology of these tests (e.g., number of samples tested, blinding, etc.), or details of a study meeting the criteria you've outlined for clinical performance or an AI/human-in-the-loop system.

    In summary, the provided text from the FDA 510(k) clearance letter and summary does not contain the detailed information necessary to answer the questions about acceptance criteria and a study proving the device meets those criteria in the context of an AI-driven or diagnostic device. This document pertains to a mechanical surgical implant, and the assessment for such a device is typically based on mechanical properties, biocompatibility, and substantial equivalence to existing devices, rather than the types of performance metrics and study designs common for AI/human reader studies.