510(k) Data Aggregation

    K Number
    K210711

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-12-22

    (287 days)

    Product Code
    Regulation Number
    872.4120
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K200805, K191363, K191605, K173402, K182776, K202100

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The NGS is intended for use in partially edentulous adult patients who qualify for dental implants.

    When Yomi Plan software is used for preplanning on third party PCs, it is intended to perform the planning (pre-operative) phase of dental implantation surgery. Yomi Plan provides pre-operative planning for dental implantation procedures. The output of Yomi Plan is to be used with the Neocis Guidance System (NGS).

    Device Description

    The purpose of subject device is for modification of the Neocis Guidance System (K202264) to allow for Wi-Fi to be continuously active in the Yomi Plan v2.0.1 while it is powered on. All other software and hardware features/functions remain identical to the predicate. In terms of FDA regulations, the Neocis Guidance System (NGS) is a dental stereotaxic instrument (Product Code PLV) and a powered surgical device for bone cutting (21 CFR 872.4120).

    The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The NGS is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants.

    Major components include: Monitor, Planning Station Laptop PC, Lift Column, Base Cart, Robotic Guide Arm, and Patient Tracker.

    The system allows the user to plan the surgery virtually in Yomi Plan (K191363-cleared for use alone on third party PCs for preplanning). The operative plan is based on a cone beam computed tomography (CBCT) scan of the patient, which is used to create a 3-D model of the patient anatomy in our planning software. The plan is used by a guidance system to provide physical, visual, and audible feedback to the surgeon during the implant site preparation. The NGS robotic arm holds and guides a standard FDAcleared third party powered bone cutting instrument (K191605).

    The patient tracking portion of the NGS is comprised of linkages from the patient to the NGS, which include the Chairside Patient Splint (CPS) (K173402) or Edentulous Patient Splint (EPS) (K200805), the End Effector (EE) and the Patient Tracker (PT).

    AI/ML Overview

    The request is to describe the acceptance criteria and the study that proves the device meets the acceptance criteria for the Neocis Guidance System (NGS) with Yomi Plan v2.0.1.

    Based on the provided text, the submission does not contain a study proving that the device meets specific performance acceptance criteria for its clinical function. Instead, it modifies an existing, previously cleared device (NGS with Yomi Plan v2.0) to allow continuous Wi-Fi activity and describes the testing conducted to ensure this modification maintains safety and effectiveness, primarily through software verification, wireless coexistence testing, and EMC testing.

    Therefore, the following information is extracted and presented in relation to the modification and associated testing, rather than a clinical performance study with acceptance criteria for the device's primary function of dental implantation guidance.


    Acceptance Criteria and Study Proving Device Meets Criteria

    The Neocis Guidance System (NGS) with Yomi Plan v2.0.1 is a modification of a previously cleared device (Neocis Guidance System (NGS) with Yomi Plan v2.0, K202264). The primary change in this submission is to allow Wi-Fi to be continuously active in the Yomi Plan v2.0.1 while it is powered on. Therefore, the "acceptance criteria" and "study" described below relate to the safety and effectiveness of this change and the overall system's compliance with relevant standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    Category / TestAcceptance CriteriaReported Device Performance
    Software VerificationCompliance with relevant software and risk management standards.Fully executed according to: - ANSI AAMI ISO 14971: 2019 (Risk Management) - ANSI AAMI IEC 62304:2006/A1:2016 (Software Life Cycle Processes) - FDA Guidance for Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005) - FDA Guidance for Cybersecurity in Medical Devices (Oct 2, 2014; Dec 28, 2016) - Cybersecurity for Networked Medical Devices (Jan 14, 2005) - AAMI TIR57: 2016 (Medical Device Security)
    Wireless CoexistenceSafe and effective operation with continuous Wi-Fi in the presence of other radio-frequency devices.Tested according to: - AAMI TIR69: 2017 (Risk management of radio-frequency wireless coexistence for medical devices and systems) - IEEE ANSI C63.27-2017 (American National Standard for Evaluation of Wireless Coexistence)
    Electromagnetic Compatibility (EMC)Compliance with EMC standards for medical electrical equipment.Tested according to: - IEC 60601-1-2 Edition 4.1 2020-09 (Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests)
    Clinical Performance (Primary Function)Not explicitly detailed in this 510(k) summary for the subject device. This submission is for a modification to a previously cleared device, assuming its fundamental clinical performance is already established."Our performance testing demonstrates substantially equivalent performance of the subject device as compared to the predicate." (This statement refers to equivalence, not necessarily a new clinical performance study for the modified device).

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not specify sample sizes for test sets related to clinical performance. The testing described relates to technical compliance (software, wireless, EMC).

    • Data Provenance: Not applicable in the context of clinical data for this specific 510(k) submission, as it doesn't describe a clinical study for performance. The testing pertains to engineering and regulatory standards for the device modification.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    • Not applicable. This submission focuses on engineering testing (software, wireless, EMC) rather than a clinical study requiring expert-established ground truth for device performance validation in a patient setting.

    4. Adjudication Method for the Test Set

    • Not applicable. This submission focuses on engineering testing, not a study requiring adjudication of clinical outcomes or interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done as described in the provided text. The device is a computerized navigational system for dental implantation, not an AI-assisted diagnostic or interpretation tool for human readers. This submission focuses on a software modification (continuous Wi-Fi) and its impact on technical safety and performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not explicitly detailed. The described testing pertains to the integrated system's technical compliance and software behavior (Yomi Plan v2.0.1) in the context of continuous Wi-Fi. While individual software components would undergo standalone verification, the submission does not describe a standalone clinical performance study. The device is described as providing "navigational guidance of the surgical instruments" and "haptic feedback to the surgeon by constraining the motion of the bone cutting instrument to the plan," implying human-in-the-loop operation.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    • Not applicable. For the technical testing (software verification, wireless coexistence, EMC), the "ground truth" is adherence to established engineering standards, cybersecurity protocols, and functional specifications, rather than clinical ground truth (e.g., pathology, expert consensus on images).

    8. The Sample Size for the Training Set

    • Not applicable. The submission does not describe a machine learning or AI model that requires a training set. The software is a planning and guidance system not explicitly described as employing AI in a way that requires a training set for model development.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. As no training set is described for an AI/ML model.
    Ask a Question

    Ask a specific question about this device

    K Number
    K203401

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-12-15

    (26 days)

    Product Code
    Regulation Number
    872.4120
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K161399, K173402, K202100, K182776, K190059, K014263

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The NGS is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants.

    Device Description

    The Neocis Guidance System (NGS) (K161399) is a dental stereotaxic instrument (Product Code PLV) and a powered surgical device for bone cutting (21 CFR 872.4120). The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides precise and accurate navigational guidance of surgical instruments, with regard to planning in dental implantation procedures. The system allows the user to plan the surgery virtually in software using a cone beam computed tomography (CBCT) scan of the patient, and the plan is used by a guidance system to provide physical, visual, and audible feedback to the surgeon during the implant site preparation. The holds and guides a standard FDA-cleared powered bone cutting instrument.

    The implant process occurs in two phases. First, the dental surgeon plans the surgical procedure with the planning software. A virtual implant is placed at the desired location in the CT scan, allowing the dental surgeon to avoid interfering with critical anatomical structures during implant surgery. Second, when the implant plan is optimally positioned, the NGS provides accurate guidance of the dental surgical instruments according to the pre-operative plan. The NGS provides haptic feedback to the surgeon by constraining the motion of the bone cutting instrument to the plan. This allows the surgeon to feel resistance to attempts at motions that may deviate from the plan.

    The patient tracking portion of the NGS is comprised of linkages from the NGS, which for partially edentulous patients include the Chairside Patient Splint (CPS) (K173402) or the Clamped Chairside Patient Splint (CCPS) (K202100), the End Effector (EE) and the Patient Tracker (PT). The CPS or CCPS is attached to the contralateral side of the patient's mouth over stable teeth. The CPS is placed on the patient using on-label dental materials (K182776) prior to the presurgical CBCT scan. A Fiducial Array (FA) with radio-opaque fiducial markers is placed on the splint prior to the CBCT scan so the virtual plan can be related to the physical space of the system using the markers. The PT is an electromechanical feedback system that is connected to the splint on the patient, which relays information to the control software in order to track patient movement. If patient movement occurs during the surgical procedure, the system will respond by altering the prescribed surgical cutting angle, position, and depth to accommodate the patient movement, which will maintain the accuracy of the osteotomy.

    The subject of this submission is a design change to the sleeves in our Edentulous Patient Splint (EPS) (K200805). The EPS enables use of the NGS in fully edentulous patients. It is affixed to the anterior mandible or maxilla using standard bone screws. Like the CPS and CCPS, the EPS serves as rigid connection to the patient for robotic tracking of the patient during the procedure. The EPS is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants.

    AI/ML Overview

    The provided text discusses the Neocis Guidance System (NGS) with Edentulous Patient Splint (EPS) and a design change to its sleeves. However, it does not contain a detailed study proving the device meets acceptance criteria for performance, especially not in the context of diagnostic accuracy (e.g., sensitivity, specificity, AUC).

    Instead, the document focuses on demonstrating substantial equivalence to a predicate device (Neocis Guidance System (NGS) with Patient Splints, K200805) after a design change to the EPS sleeves. The "Performance Testing" section lists various tests conducted, primarily related to the physical and biological aspects of the device, rather than a clinical performance study.

    Therefore, many of the requested elements (like sample size for test/training sets, data provenance, number of experts for ground truth, adjudication method, MRMC studies, standalone performance, type of ground truth for training) are not available in the provided text for a clinical performance study.

    Here's what can be extracted and inferred from the text, focusing on the design change and the tests mentioned:


    Acceptance Criteria and Device Performance (Design Change Validation)

    The document describes a design change to the sleeves within the Edentulous Patient Splint (EPS) component of the Neocis Guidance System (NGS). The acceptance criteria are implicitly related to ensuring this design change does not negatively impact the safety and effectiveness of the device, and that it remains substantially equivalent to the predicate.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific Test/RequirementReported Device Performance/Conclusion
    Mechanical Performance (Splint)EPS Weighted Deflection Test with Optical Tracking in Sawbones®Implied to have met predefined limits for deflection, ensuring stability and accuracy. The document states "The new design is functionally the same as the predicate device."
    System AccuracyTotal System AccuracyImplied to have met accuracy requirements (likely related to navigational guidance precision), ensuring the design change does not degrade the overall system's ability to guide surgical instruments accurately.
    Risk ManagementANSI AAMI ISO 14971:2019 Medical devices Applications of risk management to medical devicesRisk analysis performed and controls implemented to mitigate risks associated with the design change. Concludes no new questions of safety or effectiveness.
    SterilizationANSI AAMI ISO 17665-1:2006/(R)2013 Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical deviceSterilization process validated for the new material/design.
    BiocompatibilityANSI AAMI ISO 10993-1:2009/(R)2013 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management processBiological evaluation indicating the new materials (Neocis design titanium for sleeves, removal of stainless steel from patient-contacting materials) are safe for patient contact.
    ANSI AAMI ISO 10993-5:2009/(R)2014 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityNo cytotoxicity detected.
    ANSI AAMI ISO 10993-10:2010/(R)2014 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitizationNo irritation or skin sensitization detected.
    ISO 10993-11 Third edition 2017-09 Biological evaluation of medical devices - Part 11: Tests for systemic toxicityNo systemic toxicity detected.
    ANSI AAMI ISO 10993-12: 2012 Biological Evaluation of Medical Devices - Part 12: Sample Preparation and Reference MaterialsSample preparation and reference materials used were appropriate for the biocompatibility testing.
    Conclusion of Substantial Equivalence (Overall System)Overall, the design changes to the EPS sleeves have been verified using well-established methods. The new design is functionally the same as the predicate device. The subject device different questions of safety and effectiveness.The subject device is substantially equivalent to the predicate, meaning it is as safe and effective as the previously cleared device.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated for performance tests like "EPS Weighted Deflection Test" or "Total System Accuracy". These are typically engineering verification tests, and the "sample size" would refer to the number of units tested.
    • Data Provenance: Not specified for these engineering tests. "Sawbones®" is mentioned, indicating laboratory testing on synthetic bone models. This is not clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable for the reported tests. The tests are engineering verification tests, not diagnostic accuracy studies requiring expert-established ground truth.

    4. Adjudication method for the test set:

    • Not applicable. See point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was mentioned or conducted. The device is a surgical guidance system, not a diagnostic AI tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not explicitly described as a standalone algorithm performance study. The "Total System Accuracy" test would assess the device's accuracy in guiding the surgical instrument, which is its primary function (albeit with a human surgeon operating the instrument under guidance). The text focuses on the mechanical and system accuracy of the guidance mechanism itself, not a diagnostic algorithm.

    7. The type of ground truth used:

    • For mechanical tests: Engineering specifications, precision measurements, or established physical benchmarks are the "ground truth."
    • For biological tests: Standards (e.g., ISO 10993) and established laboratory protocols define the "ground truth" for material properties and effects.

    8. The sample size for the training set:

    • Not applicable. The document does not describe the development or training of an AI algorithm in the context of a "training set" for diagnostic performance.

    9. How the ground truth for the training set was established:

    • Not applicable. See point 8.

    Summary of Device Performance (from the document's conclusion):
    The primary conclusion is that "The design changes to the EPS sleeves have been verified using well established methods. The new design is functionally the same as the predicate device. The subject device different questions of safety and effectiveness. Therefore, the subject device is substantially equivalent to the predicate." This implies that all the listed performance tests were successfully passed, ensuring that the modified device remains as safe and effective as its predecessor.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200805

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-07-28

    (123 days)

    Product Code
    Regulation Number
    872.4120
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K173402, K190059, K014263

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The NGS is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants.

    Device Description

    The Neocis Guidance System (NGS) (K161399) is a dental stereotaxic instrument (Product Code PLV) and a powered surgical device for bone cutting (21 CFR 872.4120). The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides precise and accurate navigational guidance of surgical instruments, with regard to planning in dental implantation procedures. The system allows the user to plan the surgery virtually in software using a cone beam computed tomography (CBCT) scan of the patient, and the plan is used by a guidance system to provide physical, visual, and audible feedback to the surgeon during the implant site preparation. The holds and guides a standard FDA-cleared powered bone cutting instrument.

    The implant process occurs in two phases. First, the dental surgical procedure with the planning software. A virtual implant is placed at the desired location in the CT scan, allowing the dental surgeon to avoid interfering with critical anatomical structures during implant surgery. Second, when the implant plan is optimally positioned, the NGS provides accurate guidance of the dental surgical instruments according to the pre-operative plan. The NGS can be used for flapless dental implant procedures, which is a type of minimally invasive surgical approach. The NGS provides haptic feedback to the surgeon by constraining the bone cutting instrument to the plan. This allows the surgeon to feel resistance to attempts at motions that may deviate from the plan.

    The patient tracking portion of the NGS is comprised of linkages from the NGS, which include the Chairside Patient Splint (CPS) (K173402), the End Effector (EE) and the Patient Tracker (PT). The Patient Splint is attached to the contralateral side of the patient's mouth over stable teeth. The CPS is placed on the patient using on-label dental materials (K182776) prior to the presurgical CBCT scan. A Fiducial Array (FA) with radio-opaque fiducial markers is placed on the CPS prior to the CBCT scan so the virtual plan can be related to the physical space of the system using the markers. The PT is an electromechanical feedback system that is connected to the CPS on the patient, which relays information to the control software in order to track patient movement. If patient movement occurs during the surgical procedure, the system will respond by altering the prescribed surgical cutting angle, position, and depth to accommodate the patient movement, which will maintain the accuracy of the osteotomy.

    The subject of this submission is our Edentulous Patient Splint (EPS). The EPS is affixed to the anterior mandible or maxilla using standard bone screws. Like the CPS, the EPS serves as rigid connection to the patient for robotic tracking of the patient during the procedure. The EPS is intended for use in partially edentulous and fully edentulous adult patients.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the Neocis Guidance System (NGS) with Patient Splints, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document primarily focuses on demonstrating substantial equivalence to a predicate device (X-Guide Surgical Navigation System, K150222) rather than defining absolute acceptance criteria for novel performance claims. However, specific accuracy metrics are listed and compared.

    Acceptance Criteria (from Predicate/Prior Clearance)Reported Device Performance (NGS with Patient Splints)
    System Lateral Accuracy: RMS < 1 mmRMS < 1 mm
    System Depth Accuracy: RMS < 1 mmRMS < 1 mm
    System Angular Accuracy: RMS < 6.0°RMS < 6.0°
    CT Scan Quality Requirement: Not specified for predicate, but NGS has tighter tolerances0.3 mm Voxel, 0.3 mm Slice Thickness, Matrix 512 x 512, Full 13 cm 21 sec. Multi 2 DICOM format.
    F/T Sensor Force Measurement Range: N/A for predicate+/- 30 N
    F/T Sensor Torque Measurement Range: N/A for predicate+/- 2 Nm
    F/T Sensor Single Axis Force Overload Limit: N/A for predicate200 N
    F/T Sensor Single Axis Torque Overload Limit: N/A for predicate20 Nm
    Upper limit specification for Guidance Arm Translation Speed: N/A for predicate1.25 m/s
    Biocompatibility: Yes (ISO 10993-1, -5, -10, -12)Yes (ISO 10993-1, -5, -10, -12)
    Sterilization: Steam (ISO 17665-1)Steam (ISO 17665-1)
    Software Level of Concern: ModerateModerate

    2. Sample Size Used for the Test Set and Data Provenance

    • EPS Cadaver Testing:
      • Sample Size: 2 human cadaver heads.
      • Data Provenance: Not specified, but likely from a US-based facility where cadaver studies are conducted. This is a prospective test.
    • IDE Study G190282 (Clinical Validation):
      • Sample Size: 10 adult patients (5 per site), resulting in 67 dental implants placed.
      • Data Provenance: Prospective, two-center study conducted in private practices (presumably in the US, as it's an FDA submission).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • EPS Cadaver Testing: One surgeon experienced with the NGS performed the procedures and answered qualitative performance questions. Qualifications: "Surgeon experienced with the NGS."
    • IDE Study G190282 (Clinical Validation): Two clinical investigators (surgeons), one per site. Qualifications: "Fully licensed to practice dental implant surgery and were trained on use of the study device prior to starting the study."

    4. Adjudication Method for the Test Set

    • Cadaver Testing: Qualitative; likely based on the single surgeon's assessment. No formal adjudication method like 2+1 or 3+1 is mentioned.
    • Clinical Validation (IDE Study): Implant location accuracy was examined using a "before and after analysis of CT data showing the location of the implant in the preop plan versus postop CT." It's not explicitly stated if independent experts or an adjudication panel reviewed these "before and after" CTs. The statement "All implants met system specifications for accuracy" suggests an evaluation against a predetermined quantitative threshold rather than a consensus-based adjudication process.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    No MRMC comparative effectiveness study is mentioned in the provided text. The study focuses on the performance and usability of the device for dental implant surgery, with the device providing robotic guidance, rather than an AI assistance tool for human readers interpreting images.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    For the specific accuracy metrics (Lateral, Depth, Angular Accuracy), it's stated "The full system is used to perform a simulated clinical procedure on a typodont to measure the system accuracy," and also "The Patient Tracker was evaluated for accuracy per ASTM F2554," and "The positional accuracy of the Guidance Arm was evaluated by collecting 27 data points in spaces within two work volumes (54 total points) against a calibrated CMM." While these involve the system, the core device described is a robotic guidance system, not an AI algorithm that operates standalone. The accuracy values likely represent the standalone performance of the robotic system in guiding surgical instruments.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    • System Accuracy (Lateral, Depth, Angular): Ground truth was established by precise measurements against a "calibrated CMM" (Coordinate Measuring Machine) for the guidance arm's positional accuracy and a "typodont" (dental model) for overall system accuracy. For clinical validation, post-operative CT data was compared to the pre-operative plan.
    • Clinical Validation (IDE Study): The ground truth for effectiveness was direct comparison of post-operative implant location (via CT) against the pre-operative plan. Safety was assessed via follow-up for wound site healing. Usability was assessed via qualitative feedback from surgeons.

    8. The Sample Size for the Training Set

    The document does not specify a training set sample size for any machine learning or AI component. The Neocis Guidance System is described as a "computerized navigational system" providing "robotic guidance," and while it uses software and control systems, the text doesn't indicate a machine learning model that would typically have a "training set" in the context of image interpretation or diagnostic aid. The "planning software" is described as a tool for virtual implant placement.

    9. How the Ground Truth for the Training Set Was Established

    As no training set for a machine learning component is described, this question is not applicable based on the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191605

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-09-11

    (86 days)

    Product Code
    Regulation Number
    872.4120
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K030163, K070084, K161399, K173402

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments.

    Device Description

    The Neocis Guidance System (NGS) is a stereotaxic medical device that guides surgeons during dental implant surgery. The system allows the user to plan the surgery virtually in software using a CT scan of the patient, and the plan is used by a quidance system to provide physical, visual, and audible feedback to the surgeon during the implant site preparation.

    The implant process occurs in two phases. First, the dental surgeon plans the surgical procedure with the planning software. A virtual implant is placed at the desired location in the CT scan, allowing the dental surgeon to avoid interfering with critical anatomical structures during implant surgery. Second, when the implant plan is optimally positioned, the NGS provides accurate quidance of the dental surgical instruments according to the pre-operative plan.

    Physical quidance is provided via the Guidance Arm. The Guidance Arm grips a standard dental drill from the back end, allowing the surgeon to grip the drill as normal. The Guidance Arm does not move unless the surgeon applies a manual force to the drill. The Guidance Arm will constrain the surgeon to drill according to the prescribed surgical plan, preventing deviation. The surgeon is constantly in control of the drilling.

    Visual guidance is provided by 3D graphics and 2D cross sections that indicate the position and orientation of the drill in relation to the pre-operative plan and scan. The visual feedback is updated in real-time so any relative motion between the dental handpiece and the patient properly update the visualization.

    The patient tracking portion of the NGS is comprised of the Patient Splint and the Patient Tracker. The Patient Splint is attached to the contralateral side of the patient's mouth. The Patient Splint is placed on the patient prior to the CT scan. A fiducial array with fiducial markers is placed on the Patient Splint prior to the CT scan so the virtual plan can be related to the physical space of the system. The Patient Tracker is a mechanical feedback system that is connected to the Patient Splint on the patient, which relays information to the control software in order to track patient movement. If patient movement occurs during the surgical procedure, the system will respond by altering the prescribed surgical cutting angle and position to accommodate the patient movement, which will maintain the accuracy of the drill placement.

    Several steps are required for calibration and measurement during the procedure. The drill is calibrated using the Calibration Drill Bit inserted into a precise position on the Patient Tracker. During the surgery, each drill bit must be measured with the Depth Gauge to determine the proper length of the bit. These measurements complete the loop so the entire NGS is accurate to the tip of the drill.

    The NGS is a supporting device, providing additional information and guidance to the decision- making process during the surgical procedure. It is not intended to replace the surgeon's judqment. The final clinical decisions are the sole responsibility of the surgeon. The surgeon can at any time during the surgical procedure modify the planned implant positions. Under no circumstances does the device relieve the surgeon of his or her ultimate clinical responsibility.

    The subject device is the same as the NGS cleared under K182776 (the predicate device), except for a change to the dental drill supplier and dental drill collar design. The dental handpiece and motor have received previous 510(k) clearance under K070084 and K030163.

    The splint is a key component for patient tracking for the NGS. The patient tracking portion of the NGS is comprised of the Chairside Splint and the Patient Tracker. The Chairside Splint is attached to the contralateral side of the patient's mouth. The Chairside Splint is affixed to the patient's teeth using dental materials specified in the labeling. The Chairside Splint is placed on the patient prior to the CT scan. A fiducial array with fiducial markers is placed on the Chairside Splint prior to the CT scan so the virtual plan can be related to the physical space of the system. The Patient Tracker is a mechanical feedback system that is connected to the Patient Chairside on the patient, which relays information to the control software in order to track patient movement. If patient movement occurs during the surgical procedure, the system will respond by altering the prescribed surgical cutting angle and position to accommodate the patient movement, which will maintain the accuracy of the drill placement.

    AI/ML Overview

    The provided text describes information about the Neocis Guidance System (NGS), a computerized navigational system for dental implantation surgery. However, the document (a 510(k) premarket notification) primarily focuses on demonstrating substantial equivalence to a predicate device due to a change in dental drill supplier and collar design, rather than proving the device meets new acceptance criteria established for this specific submission.

    Therefore, the information regarding acceptance criteria and study details is largely drawn from previous clearances (K173402 and K161399) as the current submission leverages prior performance testing.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document doesn't explicitly define "acceptance criteria" presented as a single, consolidated list with corresponding performance for the current submission (K191605). Instead, it refers to the performance characteristics of the device, which serve as criteria for substantial equivalence to the predicate. The "Prior Performance Testing" sections (K173402 and K161399) detail various verification and validation activities.

    For this submission, the comparison table (Table 1) between the subject device and the predicate device outlines several technical characteristics. The implicit acceptance criterion for these is "no difference" compared to the predicate, as highlighted in the "Comments" column.

    Acceptance Criteria (Derived from Predicate Device Performance)Reported Device Performance (Subject Device)Comments (from Table 1)
    System Lateral Accuracy: RMS < 1 mmRMS < 1 mmNo difference
    System Depth Accuracy: RMS < 1 mmRMS < 1 mmNo difference
    System Angular Accuracy: RMS < 6.0°RMS < 6.0°No difference
    CT Scan Quality Requirements: 0.3 mm Voxel, 0.3 mm Slice Thickness, Matrix 512x512, Full 13cm 21 sec, Multi 2 DICOM format.0.3 mm Voxel, 0.3 mm Slice Thickness, Matrix 512x512, Full 13cm 21 sec, Multi 2 DICOM format.No difference
    F/T Sensor Force Measurement Range: +/- 30 N+/- 30 NNo difference
    F/T Sensor Torque Measurement Range: +/- 2 Nm+/- 2 NmNo difference
    F/T Sensor Single Axis Force Overload Limit: 200 N200 NNo difference
    F/T Sensor Single Axis Torque Overload Limit: 20 Nm20 NmNo difference
    Upper limit specification for Guidance Arm Translation Speed: 1.25 m/s1.25 m/sNo difference
    Biocompatibility: Yes (ISO 10993-1, -5, -10, -11, -12)Yes (ISO 10993-1, -5, -10, -11, -12)No difference
    Sterilization: SteamSteamNo difference

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document provides limited specific numbers for test set sample sizes and provenance:

    • Total System Accuracy (K173402): Evaluated with "a typodont as simulation of a patient with three osteotomies per typodont in four locations (Upper Right / Upper Left / Lower Right / Lower Left)." This implies multiple osteotomies on each typodont, but the exact number of typodonts is not specified.
    • Guidance Arm Accuracy / Repeatability (K161399): "collecting 27 data points in spaces within two work volumes (54 total points) against a calibrated CMM."
    • Other tests: Descriptions like "Run through of Typical Use Case," "Testing of all potential boundary parameters," and "Simulating all error messages and pop-ups" do not provide specific numerical sample sizes.
    • Data Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective, beyond indicating "simulated use" on typodonts/in a simulated clinical environment.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Simulated Use (End User Validation) (K173402 and K161399): Performed by "Surgeons" (K173402) and "End User" (K161399). The number of surgeons/end users and their specific qualifications (e.g., years of experience, specialization) are not specified in the provided text.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not describe any adjudication method used for establishing ground truth in the reported tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The document refers to the Neocis Guidance System as a "computerized navigational system" and a "stereotaxic medical device that guides surgeons." It is a surgical guidance device, not an AI-powered diagnostic or interpretive tool that assists "human readers." Therefore, an MRMC comparative effectiveness study involving human readers (as typically seen in diagnostic imaging AI) is not applicable to this type of device, and no such study is mentioned. The system assists surgeons during the surgical procedure rather than enhancing their interpretation of images.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device is designed for human-in-the-loop operation, as it "guides surgeons" and "is not intended to replace the surgeon's judgment." The reported accuracy tests (System Lateral Accuracy, System Depth Accuracy, System Angular Accuracy, Guidance Arm Accuracy / Repeatability) likely represent the standalone technical performance of the guidance system's robotic/mechanical components, separate from the surgeon's manual actions, but within the context of a simulated surgical environment where the system provides guidance. The document does not explicitly present "standalone algorithm performance" in a way that separates algorithmic output from its interaction with the mechanical guidance system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the reported accuracy tests appears to be physical measurements against established benchmarks or calibrated equipment.

    • Total System Accuracy (K173402): Accuracy was evaluated, implying a comparison against a known, intended plan or precise measurements on the typodont.
    • Guidance Arm Accuracy / Repeatability (K161399): Evaluated "against a calibrated CMM" (Coordinate Measuring Machine), which provides highly accurate physical measurements.
    • End User Calibration Verification (K161399): "Dimensional analysis and verification of Calibration Materials."

    8. The sample size for the training set

    The document does not provide any information regarding a training set or its sample size. This type of surgical guidance system typically relies on computational geometry, kinematics, and control theory rather than machine learning models that require labeled training data in the same way as an AI diagnostic algorithm.

    9. How the ground truth for the training set was established

    Since no training set is mentioned for an AI/machine learning model, this question is not applicable. The device's functionality is based on established engineering principles and calibration processes.

    Ask a Question