510(k) Data Aggregation

    K Number
    K243789

    Validate with FDA (Live)

    Date Cleared
    2025-09-04

    (269 days)

    Product Code
    Regulation Number
    870.1290
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K173288

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LIBERTY® Endovascular Robotic System is intended for use in the remote delivery and manipulation of guidewires and catheters, and remote manipulation of guide catheters, to facilitate navigation to anatomical targets in the peripheral vasculature.

    The LIBERTY® Endovascular Robotic System is not intended for coronary or neurointerventional procedures.

    Device Description

    The LIBERTY® Endovascular Robotic System (LIBERTY System) is a single-use, sterile, disposable device designed to help physicians navigate guidewires and catheters in peripheral endovascular procedures. It allows physicians to control the devices remotely using a handheld controller while enabling a physician to remain seated and away from the X-ray radiation source, thus reducing radiation exposure.

    Key components:

    • Bedside Robotic Drive: This motorized unit grips and maneuvers the interventional devices.
    • Handheld Remote Control: This wireless controller allows the physician to remotely advance, retract, and rotate the guidewire and microcatheter by communicating with the Robotic Drive.
    • Mounting Arm: This arm securely attaches the Robotic Drive to the patient's bed near the insertion point.
    • Accessories: These include hemostatic valves and an extension tube for managing fluids during the procedure.

    The system is intended to be used in peripheral endovascular procedures. Following manual insertion of an introducer sheath and guide catheter, the system is utilized via the handheld remote control that operates the robotic drive from a distance and enables remote positioning of guidewire and catheter (microcatheter) devices at a desired point inside the peripheral vasculature. Once the interventional devices are placed at the target site using the LIBERTY Robotic System, the devices are used manually consistent with their cleared indications for use for the interventional procedure.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K202275

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-12-11

    (122 days)

    Product Code
    Regulation Number
    870.1290
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K173288

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CorPath GRX System is intended for use in the remote delivery and manipulation of guidewires and rapid exchange catheters, and remote manipulation of guide catheters during percutaneous coronary and vascular procedures.

    Device Description

    The CorPath GRX System is intended to allow physicians to deliver and manipulate commercially available guidewires, rapid exchange catheters and guide catheters during percutaneous coronary and vascular procedures. During the use of the CorPath GRX System, the physician controls the movement and maneuvering of the devices using intuitive controls under independent angiographic fluoroscopy visual guidance using computer controlled movements while in a seated position away from the radiation source.

    The CorPath GRX System is composed of the following two functional sub-units:

      1. Bedside Unit Which consists of the Extended Reach Arm, Robotic Drive and Single-use Cassette
      1. Remote Workspace Which consists of the Control Console, angiographic monitor(s), hemodynamic monitors, X-ray foot pedal, and optional Interventional Cockpit.

    Commercially available guidewires, rapid exchange catheters, and guide catheters are loaded into the Singleuse Cassette. By using the joysticks or the Control Console touch screen, the physician can control the Robotic Drive to advance, retract, and rotate the guidewire, advance and retract the rapid exchange catheter, and advance, retract, and rotate the guide catheter. The Robotic Drive and Control Console communicate via a single communication cable.

    In addition, the CorPath GRX Software contains the following functionality for automated movements (also referred to as the technIQ automated movements), of the interventional devices:

    • . Rotate on Retract - When selected, rotates the guidewire a set amount upon retraction of the quidewire joystick to facilitate redirection of the quidewire while it is being navigated to the target location (previously cleared under CorPath GRX System, K173806).
    • Wiggle - When selected, this movement enables a small clockwise and counterclockwise rotation of the guidewire while advancing to assist in navigation.
    • . Spin - When selected, this movement will enable a large clockwise and counterclockwise rotation of the quidewire while advancing to assist in lesion crossing.
    • . Constant Speed - When selected, the guidewire or device joysticks will advance and retract at a constant speed of either 2mm/second or 5mm/second depending on the speed selected by the operator.
    • . Dotter - When selected, this movement will enable a linear back and forth motion of the device when advancing to assist in lesion crossing and delivery of therapy.
    AI/ML Overview

    This document describes a 510(k) premarket notification for the CorPath GRX System, a robotic system for percutaneous coronary and vascular procedures. The submission focuses on software changes introducing four additional automated features (spin, wiggle, dotter, and constant speed) to the already cleared CorPath GRX System.

    Here's an analysis based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria with specific numerical targets. Instead, it states that "All testing has demonstrated that the device is substantially equivalent to the predicate device." The acceptance criterion is, therefore, demonstrating substantial equivalence to the predicate device (CorPath GRX System, K173806, and K173288).

    Acceptance CriteriaReported Device Performance
    Device functions as intended with new automated features (Spin, Wiggle, Dotter, Constant Speed)."All testing has demonstrated that the device is substantially equivalent to the predicate device." "The technIQ automated moves for the CorPath GRX System do not change the safety profile of the device as there is no change to delivery forces and the operator maintains complete control of all movements of the devices."
    No change in safety profile due to new automated features."The technIQ automated moves for the CorPath GRX System do not change the safety profile of the device as there is no change to delivery forces and the operator maintains complete control of all movements of the devices."
    Continued compliance with relevant regulations and performance standards.Demonstrated through various non-clinical tests (Functional Testing, Simulated Use Testing, Particulate Testing, Software Verification and Validation testing, Cybersecurity Penetration Testing).

    2. Sample size used for the test set and the data provenance

    The document mentions "Simulated Use Testing" and states that "Clinical data from the post-market PRECISION GRX Study was used to help demonstrate substantial equivalence." However, it does not specify the sample size for either the simulated use testing or the PRECISION GRX Study, nor does it provide details on the data provenance (e.g., country of origin, retrospective or prospective) for the clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The study appears to be focused on technical verification and validation, and a comparison against a predicate, rather than an expert-adjudicated performance study where "ground truth" would typically be established by multiple experts.

    4. Adjudication method for the test set

    This information is not provided in the document. Given the nature of the testing described (functional, simulated use, software verification), a formal adjudication method by multiple experts is unlikely to have been used in the same way it would be for a diagnostic AI study.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no indication that an MRMC comparative effectiveness study was done. The device is a "Steerable Catheter Control System," not an AI diagnostic tool, so the concept of "human readers improving with AI assistance" does not directly apply in the context of this submission. The "technIQ automated movements" are intended to assist the physician in manipulating devices, but the document does not include a comparative study of physician performance with vs. without these specific automated features.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device is a robotic system designed to be operated by a physician ("human-in-the-loop"). The "automated movements" are features within this system, not a standalone AI algorithm. Therefore, a standalone (algorithm only) performance study as typically understood for AI diagnostic devices was not done and would not be relevant for this type of device. The verification and validation testing would have involved the system with its automated features, likely within simulated use or benchtop environments.

    7. The type of ground truth used

    For the non-clinical laboratory tests (Functional, Simulated Use, Particulate, Software V&V, Cybersecurity), the "ground truth" would be established by engineering specifications, design requirements, and established testing protocols. For the "Clinical data from the post-market PRECISION GRX Study," the document does not specify the type of ground truth, but for a robotic intervention system, it would typically relate to procedural success, safety outcomes, and possibly ergonomic benefits, rather than a diagnostic "ground truth" like pathology.

    8. The sample size for the training set

    This information is not applicable/provided. The submission concerns software updates to a robotic control system with "automated movements," which are likely rule-based or control algorithms, rather than a machine learning model that requires a distinct "training set." Therefore, no training set size is mentioned.

    9. How the ground truth for the training set was established

    This information is not applicable/provided for the same reasons as #8. If the automated movements involve control algorithms rather than learned AI models, the "ground truth" for their development would be based on engineering principles, clinical requirements for device manipulation, and pre-defined operational parameters, not a labeled dataset.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180517

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-03-29

    (30 days)

    Product Code
    Regulation Number
    870.1290
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K173288

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CorPath GRX System is intended for use in the remote delivery and manipulation of guidewires and rapid exchange catheters, and remote manipulation of guide catheters during percutaneous coronary and vascular procedures.

    Device Description

    The CorPath GRX System is intended to allow physicians to deliver and manipulate commercially available guidewires, rapid exchange catheters and guide catheters during percutaneous coronary and vascular procedures. During the use of the CorPath GRX System, the physician maneuvers interventional devices using intuitive controls under independent angiographic fluoroscopy visual quidance using computer controlled movements while in a seated position away from the radiation source.

    The CorPath GRX System is composed of the following two functional sub-units:

      1. Bedside Unit Which consists of the Extended Reach Arm, Robotic Drive and Single-use Cassette
      1. Remote Workspace Which consists of the Control Console, angiographic monitor(s), hemodynamic monitors, X-ray foot pedal, and optional Interventional Cockpit.

    Commercially available guidewires, rapid exchange catheters, and guide catheters are loaded into the Singleuse Cassette. By using the joysticks or the Control Console touch screen, the physician can control the Robotic Drive to advance, retract, and rotate the guidewire, advance and retract the rapid exchange catheter, and advance, retrace, and rotate the guide catheter. The Robotic Drive and Control Console communicate via a single communication cable.

    AI/ML Overview

    The provided text is a 510(k) summary for the CorPath GRX System, which is a steerable catheter control system used for remote delivery and manipulation of guidewires, rapid exchange catheters, and guide catheters during percutaneous coronary and vascular procedures.

    This document describes a submission for a modificiation (K180517) to an already cleared device, the CorPath GRX System (K173288). The modification is specifically "limited to a new bedrail connection design for the Extended Reach Arm component."

    Based on the information provided, here's a breakdown of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (What was measured)Reported Device Performance (Result)
    Demonstration of substantial equivalence to predicate device (K173288) after modification (new bedrail connection for Extended Reach Arm component)."All testing has demonstrated that the device is substantially equivalent to the predicate devices." (Page 5) "The testing demonstrated that the device can be considered substantially equivalent to the predicate devices." (Page 5) "Based on the bench testing conducted, it is concluded that the CorPath GRX System is substantially equivalent to the predicate device, the CorPath GRX System (K173288, cleared February 15, 2018)." (Page 5)
    Performance Testing of Extended Reach Arm (specifically related to the new bedrail connection design)."Performance Testing Extended Reach Arm" was conducted, and the results contributed to the overall finding of substantial equivalence. The document doesn't provide specific quantitative metrics, but states the testing was successful. (Page 5)
    Compliance with IEC 60601-1 Safety Testing."IEC 60601-1 Safety Testing" was conducted, and the results contributed to the overall finding of substantial equivalence. The document doesn't provide specific quantitative metrics, but states the testing was successful. (Page 5)

    2. Sample Size Used for the Test Set and the Data Provenance:

    The document describes non-clinical laboratory tests rather than studies involving human subjects or real-world patient data. Therefore, the concepts of "sample size for the test set" and "data provenance (e.g., country of origin of the data, retrospective or prospective)" as typically applied to clinical or AI performance studies are not directly applicable here. The testing was performed on the device itself and its components.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    This information is not applicable. The study involved engineering and safety testing of a medical device, not a diagnostic or prediction task that would require expert-established ground truth on a test set.

    4. Adjudication Method for the Test Set:

    This information is not applicable for the same reasons as point 3.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No. This was a 510(k) submission for a robotic catheter control system, not an AI-assisted diagnostic or interpretation device that would involve human readers or MRMC studies. The device itself performs actions (delivery and manipulation of catheters) under human control.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    The device is inherently a "human-in-the-loop" system, as physicians maneuver interventional devices using controls. The testing described focuses on the device's mechanical, electrical, and safety performance following a modification, not on an "algorithm only" performance.

    7. The Type of Ground Truth Used:

    The "ground truth" for this type of submission is the functional and safety standards (e.g., IEC 60601-1) and the performance characteristics of the predicate device. The testing's purpose was to demonstrate that the modified device's performance meets these established benchmarks and remains substantially equivalent to the original cleared device despite the design change.

    8. The Sample Size for the Training Set:

    This information is not applicable. This is not an AI/machine learning device that involves a training set. The "training" for such a device would be its design, engineering, and manufacturing processes, culminating in verification and validation testing.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the same reason as point 8.