Found 2 results

510(k) Data Aggregation

    K Number
    K180526

    Validate with FDA (Live)

    Device Name
    EIDON FA
    Manufacturer
    Date Cleared
    2018-07-16

    (139 days)

    Product Code
    Regulation Number
    886.1570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K142047, K112330, K172649

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CenterVue EIDON FA is a confocal scanning ophthalmoscope indicated for color, infrared and autofluorescence imaging and fluorescein angiography of a human retina with or without the use of a mydriatic agent.

    Device Description

    The CenterVue EIDON FA has been derived from the CenterVue EIDON, a retinal imaging device cleared under K142047. EIDON FA is a scanning ophthalmoscope which uses LED light to capture confocal images of the retina. In particular, EIDON FA uses infrared light to obtain infraredreflectance images, white light to obtain color images and blue light to obtain auto-fluorescence and fluorescence images. EIDON FA can be used with or without pharmacological dilation.

    The CenterVue EIDON FA is part of a family of devices, which includes three models: EIDON FA, EIDON AF, and EIDON. EIDON is the base model, which features the following imaging modalities: infrared reflectance, color and red-free. EIDON AF adds autofluorescence imaging to the base model. The EIDON FA is the fully featured device, which adds fluorescein angiography to the capabilities of the EIDON AF and encompasses the features and functionality of the other models.

    EIDON FA operates as a standalone unit, running a dedicated software application, is intended for prescription use only, and includes:

    1. an optical head, including a removable lens cap;
    2. a patient head-rest, including a removable front-rest;
    3. a patient chin rest;
    4. a base, including a touch-screen device (tablet with magnetic holder and USB cable), USB joystick and an external power supply.

    EIDON FA operates based on the following principles:
    a) An illumination system consisting of an infrared (IR) LED (825-875 nm and 940 nm), a white LED (440-650 nm), a blue LED (440-475 nm) and a green LED, illuminates the patient eye with the following functions:

    • the IR LED with central wavelength at 850 nm allows the capture of IR photos. The patient's retina is uniformly illuminated by a line in a horizontal direction. Along the optical path there is an oscillating mirror which scans the line in order to illuminate the retina with a field of view of 60°.
    • Two IR LEDs with central wavelength at 940 nm are seen from the eye in a free viewing system. The two LEDs are equally shifted with respect to the machine's optical axis. The LEDS are switched on during all exams in order to enable pupil tracking.
    • The white and blue LEDs allow the capture of color photos. The retina is uniformly illuminated by a line in the horizontal direction. Along the optical path an oscillating mirror scans the line in order to illuminate the retina with a field of view of 60°.
    • The blue LED is also used to capture auto-fluorescence and fluorescence retinal images;
    • The green LED is used as fixation target.
      b) An imaging system including a barrier filter (high-pass with a cutoff at 500 nm) stops back reflected light from the retina and allows fluorescent light to be detected for imaging. A focusing lens is included in the imaging path to achieve optimal focusing on a CMOS camera.
      c) An anterior segment alignment system is included, using two cameras and the two IR LEDs. The LEDs illuminate the anterior segment by diffusion, whereas allow a stereoscopic reconstruction of the pupil's position to be obtained with respect to the instrument's front lens.

    EIDON FA interacts with the patient by directing infrared, white, blue (for imaging) and green (for fixation) illumination into the patient's eye. The chin-rest and head-rest are the only parts of the device that contact the patient. The chin-rest includes a patient proximity sensor and is motorized for height adjustment.

    AI/ML Overview

    The provided text describes the regulatory information and technical specifications of the CenterVue EIDON FA device, focusing on its substantial equivalence to predicate devices for FDA clearance. However, it does not contain specific acceptance criteria, comprehensive study details, or performance data in the structured format requested. The clinical performance data section is very brief and high-level.

    Therefore, I cannot fully complete the requested table and answer all points accurately based solely on the provided text. I will extract and present the available information, and explicitly state where information is missing.


    Acceptance Criteria and Device Performance Study for CenterVue EIDON FA

    The provided documentation describes the CenterVue EIDON FA as a confocal scanning ophthalmoscope indicated for color, infrared, autofluorescence imaging, and fluorescein angiography of the human retina. The substantial equivalence determination relies on comparisons to predicate devices (CenterVue EIDON, Kowa VX-20, and Heidelberg Spectralis HRA²).

    1. Table of Acceptance Criteria and Reported Device Performance

    Based on the provided text, the acceptance criteria are not explicitly stated as numerical thresholds or specific performance metrics. Instead, the focus is on demonstrating "substantial equivalence" to predicate devices, particularly regarding imaging capability and safety. The performance is reported qualitatively.

    Criterion TypeAcceptance Criteria (Not explicitly stated in quantitative terms)Reported Device Performance (Summary from text)
    Imaging QualityImplied: Images produced by EIDON FA in all modalities (color, infrared, autofluorescence, fluorescein angiography) must be "similar" to or demonstrate "effectiveness" comparable to the respective predicate devices for their specific imaging functions. The document states "no additional concerns for safety and effectiveness" in comparison to predicates."The comparison showed that EIDON FA provides in all modalities images that are similar to those of the mentioned devices." This is a qualitative assessment.
    SafetyImplied: The device must meet electrical safety, EMC, light hazard, and general safety standards (e.g., IEC 60601-1, IEC 60601-1-2, ISO 15004-1, ISO 15004-2). Any technological differences with predicates (e.g., higher power LEDs for AF/FA) must not introduce new safety concerns."The device complies with the IEC 60601-1:2005 and IEC 60601-1-2:2007 standards." "The device complies with the ISO 15004-1:2006 standard for ophthalmic instruments and with the ISO 15004-2:2007 standard for light hazard protection." "No additional concerns for safety and effectiveness."
    SoftwareImplied: Software must be verified and validated according to FDA guidance and industry standards, with a focus on safety given its "moderate" level of concern."Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff... The software for this device was considered as a 'moderate' level of concern... The software also complies with the IEC 62304 standard."
    Fundus Camera Std.Implied: The device must comply with relevant fundus camera standards."The device complies with the ISO 10940:2009 standard for fundus cameras."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document states, "Images of eyes with and without diagnosed pathology were included in the comparison." However, it does not specify the number of images or patients used in the clinical comparison.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The study implicitly appears to be retrospective, as it involves comparing EIDON FA images with existing images from predicate devices.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • Number of Experts: This information is not provided in the document.
    • Qualifications of Experts: This information is not provided in the document. The assessment is described as a "comparison" of images, but who performed this comparison (e.g., a panel of ophthalmologists/radiologists or internal staff) and their qualifications are not detailed.

    4. Adjudication Method for the Test Set

    • The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for the clinical image comparison. The assessment seems to be a general comparison.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a MRMC study comparing human readers with and without AI assistance was not mentioned or performed. This submission is for an imaging device, not an AI-powered diagnostic algorithm designed to assist human readers. The clinical performance data section describes a comparison of image quality between the EIDON FA device and predicate devices.

    6. Standalone (Algorithm Only) Performance

    • This question is not applicable as the EIDON FA is an imaging device, not an AI algorithm. Its "performance" refers to the quality of the images it produces, which are then interpreted by a human clinician.

    7. Type of Ground Truth Used

    • The "ground truth" for the clinical comparison was implicitly the diagnosed pathology of the eyes. The study included "images of eyes with and without diagnosed pathology." The comparison was about the similarity of images captured by EIDON FA to those from predicate devices, rather than a direct diagnostic accuracy assessment against a gold standard for a specific condition.

    8. Sample Size for the Training Set

    • This question is not applicable, as the device is an imaging system and not an AI algorithm that requires a training set in the typical machine learning sense. The "training" of the device is its manufacturing and calibration process, not data-driven machine learning.

    9. How the Ground Truth for the Training Set Was Established

    • This question is not applicable for the same reason as point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K180293

    Validate with FDA (Live)

    Device Name
    Retia 2
    Manufacturer
    Date Cleared
    2018-05-30

    (118 days)

    Product Code
    Regulation Number
    886.1570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K172649

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CenterVue RETIA 2 is a confocal scanning ophthalmoscope indicated for infrared and auto-fluorescence imaging of a human retina with or without the use of a mydriatic agent.

    Device Description

    The CenterVue RETIA 2 is a modification of the CenterVue EIDON (K142047). RETIA 2 is a scanning ophthalmoscope which uses infrared light to obtain infrared-reflectance images of the retina and blue light to obtain auto-fluorescence images of the retina. RETIA 2 operates as a standalone unit, running a dedicated software application, intended for prescription use only, and includes: 1. Optical head, including a removable lens cap; 2. Patient head-rest, including removable front-rest; 3. Patient chin rest; 4. Base, including touch-screen device (tablet with magnetic holder and USB cable), USB joystick and an external power supply. RETIA 2 operates based on the following principles: a) An illumination system consisting of infrared LEDs (850nm, 940nm), a blue LED (450nm) and a green LED illuminates the patient eye with the following functionality: - The IR LED with a centroid wavelength of 850 nm allows the capture of IR photos. The patient's retina is uniformly illuminated by a line in a horizontal direction. An oscillating mirror scans the line in order to illuminate the retina with a field of view of 60°. - Two IR LEDs with a centroid wavelength of 940 nm are seen from the eye in a free viewing system. The two LEDs are equally shifted with respect to the machine's optical axis. The LEDS are switched on during all exams in order to enable pupil tracking. - . The blue LED (450 nm) is used to illuminate the retina and capture fluorescence retinal images. The retina is uniformly illuminated by a line in a horizontal direction. An oscillating mirror scans the line in order to illuminate the retina with a field of view of 60°. - The green LED is used as fixation target. b) An imaging system including a barrier filter (high-pass with a cutoff at 500 nm) stops back-reflected light from the retina and only allows fluorescent light to be detected for imaging. A focusing lens is included in the imaging path to achieve optimal retinal focusing on a CMOS camera having a resolution of 5 megapixels. c) An anterior segment alignment system is included, using two cameras and the two IR LEDs. The LEDs illuminate the anterior segment by diffusion, whereas the cameras allow a stereoscopic reconstruction of the pupil's position to be obtained with respect to the instrument's front lens. RETIA 2 interacts with the patient by directing infrared- and blue-wavelength illumination for imaging purposes and green illumination for fixation purposes into the patient's eye. The chin-rest are the only parts of the device that contact the patient. The chin-rest includes a patient proximity sensor and is motorized for height adjustment.

    AI/ML Overview

    The document describes the CenterVue RETIA 2, a confocal scanning ophthalmoscope, and its comparison to predicate devices, specifically the CenterVue EIDON (K142047) and Heidelberg SPECTRALIS (K172649). The study primarily focuses on demonstrating that the RETIA 2 provides infrared and autofluorescence images that are similar to those of the predicate devices.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state numerical acceptance criteria in terms of sensitivity, specificity, or other quantitative metrics for image quality or diagnostic performance. Instead, it relies on a qualitative comparison.

    Acceptance Criteria (Implied)Reported Device Performance
    Produce infrared images similar to the predicate device (CenterVue EIDON)."The comparison showed that RETIA 2 provides infrared... images that are similar to those of the mentioned devices."
    Produce autofluorescence images similar to the reference device (Heidelberg SPECTRALIS)."The comparison showed that RETIA 2 provides... autofluorescence images that are similar to those of the mentioned devices."
    Continued conformance with relevant safety and performance standards.ISO 15004-2 (Light Hazard Protection): RETIA 2 fulfills requirements for a group 1 determination (no potential light hazard). ISO 15004-1: Complies with the standard for ophthalmic instruments. IEC 60601-1:2005 and IEC 60601-1-2:2007: Complies for electrical safety and electromagnetic compatibility. IEC 62304: Software complies with the standard. ISO 10940: Complies with applicable requirements for Fundus Cameras.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the exact sample size for the test set (number of images or patients). It states: "Images of eyes with and without diagnosed pathology were included in the comparison."
    The data provenance (country of origin, retrospective/prospective) is not mentioned.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    The document does not provide this information. It only states that images "with and without diagnosed pathology were included," implying that some form of diagnosis or ground truth was available, but it doesn't describe how it was established or by whom.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not specify the adjudication method used for the comparison.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no mention of a multi-reader multi-case (MRMC) comparative effectiveness study, nor is there any discussion of AI assistance or its impact on human reader improvement. The study described is a device-to-device comparison based on image similarity.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device itself is a standalone imaging system. The "performance data - Clinical" section describes a comparison of images generated by the RETIA 2 against images from predicate devices. This represents the standalone performance of the device in generating images, as there's no mention of a human in the loop for interpreting the images for the purpose of this comparison study. However, it's not an "algorithm only" study as the device is a piece of hardware.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The document mentions "images of eyes with and without diagnosed pathology." This suggests that the ground truth for image classification (e.g., presence or absence of pathology) was based on pre-existing diagnoses, likely from clinicians or medical records. However, the exact nature of this "diagnosis" (e.g., expert consensus, pathology reports, clinical outcomes) is not further detailed.

    8. The sample size for the training set

    The document does not mention a training set, as the study described is a device comparison for image similarity, not an AI model requiring training data.

    9. How the ground truth for the training set was established

    Not applicable, as no training set for an AI model is described.