    Intended Use

    The ENSEAL X1 Tissue Sealers are bipolar electrosurgical instruments for use with an electrosurgical generator. They are intended for use during open or laparoscopic surgical procedures to cut and seal vessels, and to cut, grasp and dissect tissue during surgery.

    Indications for use include open and laparoscopic general, gynecological, urologic, thoracic, and ENT surgical procedures or any procedure where vessel ligation (cutting and sealing), tissue grasping, dissection, and division of vessels, lymphatics, and tisue bundles is performed (e.g. bowel resections, hysterectomies, gall bladder procedures, Nissen Fundoplication, adhesiolysis, and oophorectomies). The devices can be used on vessels up to and bundles as large as will fit in the jaws of the instruments.

    The ENSEAL X1 Tissue Sealers have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use this system for these procedures.

    Device Description

    The ENSEAL X1 Tissue Sealer instrument is a sterile. single patient use surgical instrument to coagulate and transect vessels up to and including 7 mm in diameter, tissue and/or vascular bundles. This device is for soft tissue only. The instrument consists of a grip housing assembly, a rotating shaft, a moveable jaw, and a knife. The instrument shaft can be rotated 360° to facilitate visualization and enable easy access to targeted tissue. The jaws are in a normally-opened position and can be partially or fully closed by squeezing the closing handle. The jaws are designed for grasping and holding targeted tissue when clamped. The ENSEAL X1 Tissue Sealer has separate seal and cut capabilities. The lower jaw of the ENSEAL X1 Tissue Sealer can be used in the open or closed position to deliver energy based on the electrode configuration and jaw design. Bipolar energy is delivered when the SEAL button or the MIN foot pedal is pressed. The power cord is permanently attached to the device and connects the instrument to the generator. The ENSEAL X1 Tissue Sealer instrument is designed for use exclusively with the Ethicon Generator G11 (GEN11) software version 2016-1 or later, packaged separately.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study conducted to prove that the ENSEAL X1 Tissue Sealers meet these criteria. The device is an electrosurgical instrument for cutting and sealing vessels and tissues.

    Acceptance Criteria and Reported Device Performance

    The core performance acceptance criterion for the ENSEAL X1 Tissue Sealers is its vessel seal performance, measured by burst pressure. Other criteria relate to electrical safety, sterility, mechanical integrity, and biocompatibility.

    Acceptance CriteriaReported Device Performance
    Vessel Seal Performance
    Burst Pressure240 mmHg (on vessels up to 7 mm)
    Maximum Indicated Vessel Size7 mm
    Electrical Safety & EMC
    IEC 60601-1-2 (EMC)Complies
    IEC 60601-1 & IEC 60601-2-2 (Electrical Safety)Complies
    Sterility/Shelf-Life
    Sterility Assurance Level (SAL)$10^{-6}$ (using Ethylene Oxide per ISO 11135)
    Shelf-Life5-years
    BiocompatibilityEvaluated based on ISO 10993-1 and FDA guidance
    Mechanical Integrity
    Axial Jaw Retention StrengthDemonstrated ability to perform as expected; risks minimized
    Compression System Stress TestDemonstrated ability to perform as expected; risks minimized
    Hinge Pin Weld StrengthDemonstrated ability to perform as expected; risks minimized

    Study Proving Device Meets Acceptance Criteria

    The study conducted to prove the device meets acceptance criteria involved a combination of bench testing and animal studies.

    1. Sample Size and Data Provenance:

      • Bench Testing: Used porcine arteries for sealed vessel burst pressure testing. The exact number of samples is not explicitly stated, but it is indicated that "Data generated from the bench testing met the predetermined acceptance criteria."
      • Acute Animal Testing: Performed in four animals for each device (subject and predicate) targeting arteries, veins, and vessel pedicles less than or equal to 7mm in diameter. Data provenance is from an acute porcine study.
      • Chronic/Survival Animal Testing: Performed in ten animals for each device (subject and predicate) targeting arteries, veins, and vessel pedicles less than or equal to 7mm in diameter. Data provenance is from a 30 (±2) day chronic survival study in animals (implied porcine).
      • The studies were pre-clinical (animal and bench), not human clinical trials. The country of origin for the data is not specified but is implicitly associated with the manufacturer's testing facilities.
    2. Number of Experts and Qualifications for Ground Truth (Test Set):

      • This section is not applicable as the studies described are pre-clinical bench and animal studies (not human-in-the-loop diagnostic studies). Ground truth was established through physical measurements (e.g., burst pressure) and direct observation of tissue effects in animals.
    3. Adjudication Method for the Test Set:

      • This section is not applicable for the type of pre-clinical studies conducted. The evaluation involved objective measurements and observations, not expert consensus on interpretations.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No, an MRMC comparative effectiveness study was not done. The studies were focused on the device's technical performance and comparative equivalence to a predicate device in an animal model, not on human reader performance with or without AI assistance.
    5. Standalone Performance:

      • Yes, the studies essentially evaluated the "standalone" performance of the algorithm/device. The bench testing directly measured the device's ability to seal vessels and achieve a certain burst pressure. The animal studies assessed its performance in vivo (tissue effects, sealing/dividing vessels) independent of human interpretive input. The description directly evaluates the device's inherent functional capabilities.
    6. Type of Ground Truth Used:

      • The ground truth for the performance evaluations was based on:
        • Physical Measurements: E.g., burst pressure (mmHg) for vessel sealing, precise diameter measurements of vessels used (up to 7mm).
        • Direct Observation/Pathological Assessment: Evaluation of tissue effects, ability to seal and divide vessels, and assessment of outcomes in animal models (both acute and chronic survival, looking at tissue health and healing post-procedure).
    7. Sample Size for the Training Set:

      • This concept is not applicable to this type of device and study. The ENSEAL X1 Tissue Sealers are a physical electrosurgical instrument, not an AI/machine learning algorithm requiring a "training set" of data. The "training" for such a device involves engineering design, manufacturing, and preclinical validation.
    8. How the Ground Truth for the Training Set Was Established:

      • This concept is not applicable for the reasons stated above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K201696

    Validate with FDA (Live)

    Date Cleared
    2021-01-22

    (214 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K172580

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ENSEAL X1 Tissue Sealers are bipolar electrosurgical instruments for use with an electrosurgical generator. They are intended for use during open or laparoscopic surgical procedures to cut and seal vessels, and to cut, grasp and dissect tissue during surgery.

    Indications for use include open and laparoscopic general, gynecological, urologic, thoracic, and ENT surgical procedures or any procedure where vessel ligation (cutting and sealing), tissue grasping, dissection, and division of vessels, lymphatics, and tisue bundles is performed (e.g. bowel resections, gall bladder procedures, Nissen Fundoplication, adhesiolysis, and oophorectomies). The devices can be used on vessels up to and including 7 mm and bundles as large as will fit in the jaws of the instruments.

    The ENSEAL X1 Tissue Sealers have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use this system for these procedures.

    Device Description

    The ENSEAL X1 Tissue Sealer instrument is a sterile, single patient use surgical instrument to coagulate and transect vessels up to and including 7mm in diameter, tissue and/or vascular bundles. This device is for soft tissue only. The instrument consists of a grip housing assembly, a rotating shaft, a moveable jaw, and a knife. The instrument shaft can be rotated 360° to facilitate visualization and enable easy access to targeted tissue. The jaws are in a normally-opened position and can be partially or fully closed by squeezing the closing handle. The jaws are designed for grasping and holding targeted tissue when clamped. The ENSEAL X1 Tissue Sealer has separate seal and cut capabilities. The lower jaw of the ENSEAL X1 Tissue Sealer can be used in the open or closed position to deliver energy based on the electrode configuration and jaw design. Bipolar energy is delivered when the SEAL button or the MIN foot pedal is pressed. Pressing the CUT button advances the knife the length of the jaws to cut the targeted tissue. The power cord is permanently attached to the device and connects the instrument to the generator. The ENSEAL X1 Tissue Sealer instrument is designed for use exclusively with the Ethicon Generator G11 (GEN11) software version 2016-1 or later, packaged separately.

    AI/ML Overview

    The provided document describes the substantial equivalence of the "ENSEAL X1 Tissue Sealer" to a predicate device, the "ENSEAL G2 Tissue Sealer." The acceptance criteria and the study that proves the device meets those criteria are primarily focused on performance data from bench and animal testing rather than a clinical trial with human subjects.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly list a table of "acceptance criteria" alongside specific numerical results for all tests in a single cohesive table. However, it states that "Data generated from the bench testing met the predetermined acceptance criteria." The primary measurable acceptance criterion mentioned is for vessel seal performance.

    Acceptance CriterionReported Device Performance (ENSEAL X1 Tissue Sealer)
    Sealed Vessel Burst Pressure240 mmHg Burst Pressure
    Ability to seal and divide vessels up to and including7 mm (demonstrated in bench and animal testing)
    BiocompatibilityMet ISO 10993-1 and FDA guidance
    Electromagnetic Compatibility (EMC) and Electrical SafetyComplies with IEC 60601-1-2, IEC 60601-1, and IEC 60601-2-2
    Sterility Assurance Level (SAL)10^-6 (validated using Ethylene Oxide per ISO 11135)
    Shelf-Life5 years
    Mechanical Performance (Axial Jaw Retention Strength, Compression System Stress Test, Hinge Pin Weld Strength)Demonstrated ability to perform as expected and minimized risk of mechanical failures
    Performance in acute porcine studyPerformed as well as the legally identified predicate device
    Performance in chronic/survival porcine studyPerformed as well as the historical performance of the legally identified predicate device

    2. Sample Size Used for the Test Set and Data Provenance

    • Bench Testing (Vessel Burst Pressure):

      • Sample Size: Not explicitly stated as a number of vessels, but mentioned as "Porcine arteries were used."
      • Data Provenance: Porcine (animal tissue). Retrospective (if referring to previously obtained tissue) or Prospective (if tissue was freshly obtained for the study) is not specified.
    • Acute Animal Testing:

      • Sample Size: Four animals for each device (ENSEAL X1 Tissue Sealer and predicate device).
      • Data Provenance: Porcine (animal model). Prospective (as these were conducted tests).
    • Chronic/Survival Animal Testing:

      • Sample Size: Ten animals for each device (ENSEAL X1 Tissue Sealer and predicate device).
      • Data Provenance: Porcine (animal model). Prospective (as these were conducted tests).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not mention the involvement of human experts to establish "ground truth" in the context of radiologists or similar interpretations for the animal or bench studies. The testing methodologies (e.g., burst pressure measurements, observation of tissue effects) inherently provide the "ground truth" for the performance criteria.

    4. Adjudication Method for the Test Set

    Not applicable, as the data collection from bench and animal studies does not typically involve human adjudication in the way clinical image interpretation studies do. The measurements and observations are direct.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. The document explicitly states: "This premarket notification does not rely on human clinical trial data to demonstrate substantial equivalence." Thus, no MRMC study was conducted.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is a medical device (electrosurgical instrument), not an AI algorithm. Its performance is evaluated directly through physical and biological tests, not through an algorithm's output.

    7. The Type of Ground Truth Used

    • Bench Testing: Direct physical measurements (e.g., force, pressure, visual observation of sealing/cutting) on porcine vessels.
    • Animal Testing: Direct observation and evaluation of tissue effects (e.g., vessel sealing, tissue dissection, complications) in live porcine models, both acutely and chronically. The "ground truth" here is the biological and mechanical outcome in the animal models.

    8. The Sample Size for the Training Set

    Not applicable. The device is a physical instrument, not an AI/ML algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for a physical electrosurgical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180403

    Validate with FDA (Live)

    Date Cleared
    2018-05-23

    (98 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K172580

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ENSEAL X1 Tissue Sealers are bipolar electrosurgical instruments for use with an electrosurgical generator. They are intended for use during open or laparoscopic surgical procedures to cut and seal vessels, and to cut, grasp and dissect tissue during surgery.

    Indications for use include open and laparoscopic general, gynecological, urologic, thoracic, plastic and reconstructive, and ENT surgical procedures or any procedure where vessel ligation (cutting and sealing), tissue grasping, dissection, and division of vessels, lymphatics, and tissue bundles is performed (e.g. bowel resections, hysterectomies, gall bladder procedures, Nissen Fundoplication, adhesiolysis, and oophorectomies). The devices can be used on vessels up to and including 7 mm and bundles as large as will fit in the jaws of the instruments.

    The ENSEAL X 1 Tissue Sealers have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use this system for these procedures

    Device Description

    The ENSEAL X1 Tissue Sealer, Curved Jaw Articulating instrument is a sterile, single patient use surgical instrument to coagulate and transect vessels up to and including 7 mm in diameter, tissue and/or vascular bundles. This device is for soft tissue only. The instrument consists of a grip housing assembly, a rotating shaft, a moveable iaw, and a knife. The instrument shaft can be rotated 360° to facilitate visualization and enable easy access to targeted tissue. The instrument shaft, normally in the straight position, can be articulated using the articulation wheel to gain additional access to tissue and facilitate additional angles across tissue and/or vascular bundles. The jaws are in a normally-opened position and can be partially or fully closed by squeezing the closing handle. The jaws are designed for grasping and holding targeted tissue when clamped. The ENSEAL X1 Tissue Sealer, Curved Jaw Articulating has separate seal and cut capabilities. The lower jaw of the ENSEAL X1 Tissue Sealer, Curved Jaw Articulating can be used in the open or closed position to deliver energy based on the electrode configuration and jaw design. Bipolar energy is delivered when the SEAL button or the MIN foot pedal is pressed. Pressing the CUT button advances the knife the length of the jaws to cut the targeted tissue. The power cord is permanently attached to the device and connects the instrument to the generator. The ENSEAL X1 Tissue Sealer, Curved Jaw Articulating instrument is designed for use exclusively with the Ethicon Generator G11 (GEN11) software version 2016-1 or later, packaged separately.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the ENSEAL X1 Tissue Sealer, Curved Jaw Articulating, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred from study description)Reported Device Performance
    BiocompatibilityCompliance with ISO 10993-1 and FDA guidance for biological evaluation.Device shares materials with a previously cleared device (K172580) which met these criteria; no new testing was required.
    Electrical Safety & EMCCompliance with IEC 60601-1-2 (EMC) and IEC 60601-1 & IEC 60601-2-2 (Electrical Safety).The system complies with specified IEC standards.
    Sterilization/Shelf-LifeSterility Assurance Level (SAL) of 10^-6 via Ethylene Oxide per ISO 11135.Validated to achieve SAL of 10^-6.
    Shelf-LifeDesignated shelf-life of 1 year.Achieves designated 1-year shelf-life.
    Mechanical TestingAbility to perform under worst-case scenarios for: Axial Jaw Retention, Jaw Strength, Hinge Pin Weld Strength, Tip Compression, Jaw Gap, and Impedance (for both 37 cm and 45 cm shaft lengths).Demonstrated ability to perform as well as the predicate device for all specified mechanical tests. Both 37 cm and 45 cm shafts were found equivalent.
    Bench Testing (Vessel Burst Pressure)Ability to seal and divide vessels up to 7 mm effectively. Data generated must meet "predetermined acceptance criteria" (specific values not provided).Data generated met the predetermined acceptance criteria.
    Acute Animal Study (Tissue Effects)No difference in tissue effects compared to the predicate device when sealing and dividing vessels up to and including 7 mm.Demonstrated no difference in tissue effects and ability to seal and divide vessels up to 7mm as well as the predicate device.
    Survival Animal Study (Tissue Effects)No difference in tissue effects compared to the predicate device after 30 days when sealing and dividing vessels up to and including 7 mm (arteries, veins, vessel pedicles).Demonstrated no difference in tissue effects and ability to seal and divide vessels up to 7mm as well as the predicate device after 30 days.

    2. Sample Size Used for the Test Set and Data Provenance

    • Bench Testing: The document states "Porcine arteries were used in this testing" but does not specify the sample size (number of arteries/vessels tested). The data is likely prospective, as it's part of a device validation study. The country of origin is not specified, but typically such studies are conducted in the country of the manufacturer or a contract research organization.
    • Acute Animal Study: "an acute porcine study" was conducted. Specific sample size (number of animals or applications) is not provided. The data is prospective.
    • Survival Animal Study: "a chronic survival study" was conducted. Specific sample size (number of animals or applications) is not provided. The data is prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not mention the use of human experts to establish ground truth for the test sets in the traditional sense of diagnostic AI. The ground truth for this medical device (electrosurgical sealer) is established through technical measurements (e.g., burst pressure, mechanical force) and biological outcomes (e.g., tissue effects, vessel patency/occlusion, healing) observed in the animal studies. These evaluations are typically performed by veterinarians, histopathologists, and engineers, whose qualifications are implied by their roles in such studies, but not explicitly stated.

    4. Adjudication Method for the Test Set

    Not applicable in the context of this device study. Adjudication methods like "2+1" or "3+1" are relevant for expert consensus in image interpretation or clinical diagnosis, not for physical device performance testing.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is specifically relevant for assessing the diagnostic performance of AI algorithms in medical imaging, usually comparing human readers with and without AI assistance. This document describes the validation of a surgical electrosurgical device, not a diagnostic AI system.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Again, this question is not applicable. The device is a physical electrosurgical instrument, not an algorithm. Its performance is inherent to its design and how it interacts with tissue.

    7. The Type of Ground Truth Used

    • Biocompatibility: Compliance with established biological safety standards (ISO 10993-1).
    • Electrical Safety & EMC: Compliance with international electrical safety and electromagnetic compatibility standards (IEC 60601 series).
    • Sterilization/Shelf-Life: Achieved sterility assurance level (SAL).
    • Mechanical Testing: Measured physical properties (e.g., force, dimension) against engineering specifications and comparison to the predicate device.
    • Bench Testing: Measured burst pressure of sealed vessels (a quantitative biological outcome) and qualitative assessment of vessel division, compared against predetermined acceptance criteria.
    • Acute Animal Study: Histopathological and gross evaluation of tissue effects (e.g., seal integrity, thermal damage) in live porcine models, compared to the predicate device.
    • Survival Animal Study: Long-term histopathological and gross evaluation of tissue effects and healing after 30 days in live porcine models, compared to the predicate device. The "ground truth" here is the biological response and healing process observed by experts.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI algorithm that undergoes "training." Its design and performance are based on engineering principles, materials science, and preclinical testing.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for this type of medical device.