Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K231183

    Validate with FDA (Live)

    Date Cleared
    2023-08-25

    (121 days)

    Product Code
    Regulation Number
    878.5000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K172149

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Non-absorbable Surgical Polyester Suture is indicated for use in general soft tissue approximation and/or ligation, but not for use in ophthalmic, cardiovascular, and neurological procedures. The device is limited to use where short term wound support (7-10 days) is required and can be left in place for a maximum of 10 days.

    Device Description

    The subject device is a coated, braided, non-absorbable synthetic surgical suture composed of polyethylene terephthalate which is supplied sterile. The suture is coated with bees wax and dyed green. The color additive is D&C Green 6. The subject device will be offered in diameters ranging from USP size 6-0 through 2 and available in length varying from 45cm to 150cm with or without needles attached.

    AI/ML Overview

    This document (K231183) is a 510(k) Premarket Notification for a Non-absorbable Surgical Polyester Suture. It seeks to demonstrate substantial equivalence to a predicate device (K172149). The information provided focuses on the physical and biological compatibility of the suture, rather than the performance of a software or AI-driven medical device in a diagnostic context. Therefore, many of the requested categories (e.g., sample size for test set, data provenance, number of experts for ground truth, adjudication method, MRMC study, standalone performance, training set size) are not applicable to this type of device submission.

    Here's an analysis based on the provided text, addressing the applicable points:


    1. Table of Acceptance Criteria and Reported Device Performance

    The device performance is demonstrated by compliance with various ISO, USP, and ASTM standards. The "acceptance criteria" are implied by these standards, meaning the device must "Pass" or "Comply" with the requirements of each standard.

    Test CategorySpecific Test / StandardAcceptance Criteria (Implied)Reported Device Performance
    Physical PerformanceUSP <861> Sutures - DiameterComply with USP <861>Comply with USP <861>
    USP <871> Sutures - Needle AttachmentComply with USP <871>Comply with USP <871>
    USP <881> Tensile StrengthComply with USP <881>Comply with USP <881>
    Length (of suture)Not less than 95.0% of the length stated on the labelMet the criteria (stated in Analysis 3)
    Packaging & SterilityASTM F88/F88M-15 Seal Strength of Flexible Barrier MaterialsComply with ASTM F88/F88M-15Comply with ASTM F88/F88M-15
    ASTM F1929-15 Detecting Seal Leaks by Dye PenetrationComply with ASTM F1929-15Comply with ASTM F1929-15
    USP <85> Bacterial Endotoxins TestComply with USP <85>Comply with USP <85>
    ISO 10993-7:2008 Ethylene oxide sterilization residualsAcceptable Residual Levels(Not explicitly stated Pass/Fail, but implied by compliance with ISO 10993-1)
    BiocompatibilityISO 10993-5 CytotoxicityPassPass
    ISO 10993-10 SensitizationPassPass
    ISO 10993-10 Intracutaneous ReactivityPassPass
    ISO 10993-11 Acute systemic toxicityPassPass
    USP <151> Pyrogen TestPassPass (also stated as Comply to USP<151>)
    ISO 10993-11 Subacute Systemic ToxicityPassPass
    ISO 10993-3 Bacterial Reverse MutationPassPass
    ISO 10993-3 Chromosome AberrationPassPass
    ISO 10993-3 Gene MutationPassPass
    ISO 10993-6 ImplantationPassPass
    ASTM F756-17 HemolysisPassPass

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. For non-clinical tests like those listed (USP, ASTM, ISO standards), specific sample sizes and testing methodologies are defined within the respective standards. The document only states that the tests were "conducted" and "complies" or "passed." The data provenance (country of origin, retrospective/prospective) is also not specified, as these are lab-based tests rather than patient studies.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not Applicable. This document describes the testing of a physical medical device (suture) against established performance and biocompatibility standards. "Ground truth" in the context of expert consensus, typical for diagnostic software or AI, is not relevant here. The "ground truth" for physical device performance is defined by the objective measurement criteria within the specified international and national standards (e.g., USP monographs for suture diameter, tensile strength).


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not Applicable. As described above, this is not a study involving human interpretation of diagnostic data that would require an adjudication method.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not Applicable. An MRMC study is relevant for diagnostic software performance, particularly AI-driven tools. This submission is for a physical surgical suture.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not Applicable. This is not an algorithm or software device.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" (or more accurately, the reference standard) for this device is based on established objective measurements and criteria defined by validated international and national standards (e.g., USP Monographs for physical properties, ISO 10993 series for biocompatibility). These standards specify methodologies and acceptance limits that the device must meet.


    8. The sample size for the training set

    Not Applicable. This is not an AI or machine learning device that requires a training set. The device is a physical product.


    9. How the ground truth for the training set was established

    Not Applicable. As this device does not use a training set, no ground truth needs to be established for it.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1