510(k) Data Aggregation

    Why did this record match?
    Reference Devices :

    K171798, K171721

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neurological Theray Devices - Accessories including Garment electrodes which are intended to be used with legally marketed electrical stimulating devices such as transcutaneous electrical nerve stimulators or powered muscle stimulators. The cutaneous electrodes will deliver the stimulation signals generated by the stimulator to the body surface with which they are in contact. The Garment electrodes including glove, socks, sleeve, knee belt and back belt which are made up of silver or conductive silicone rubber.

    Device Description

    The Neurological Therapy Devices - Accessories -Conductive fiber(Model:Electrotherapy device Electrode for Knee, Electrotherapy device Electrode for Elbow, Electrotherapy device Electrotherapy device Electrode socks.Neurological physiotherapy devices-accessories), are conductive garments that are made from material which is knitted from primarily nylon yarn with minor amounts of lycra, spandex to achieve a stretch fabric for a snug garment fit polyester for tactile qualities and polyester sheathed carbon yarn and pure silver coated nylon fibers to provide conductivity. The knitting follows standard knitting procedures with the conductive knitted material being fabricated into Neurological Therapy Devices - Accessories garments which are provided in multiple sizes of various garment configurations including; gloves, bands, socks, and sleeves.

    The Neurological Therapy Devices - Accessories-conductive silicone rubber(Model:Electrotherapy device Electrode belt for back of body, Electrotherapy device Electrode belt for Abdominal Muscle TrainerToning, Electrotherapy device Electrode belt for Body joints,Electrotherapy device Electrode belt for arm of body,Electrotherapy device Electrode belt for back of body,Electrotherapy device Electrode belt for shoulder of body,Neurological physiotherapy devices-accessories), is made of conductive silicone rubber coated in conductive garments material, has excellent conductive performance and good flexibility and plasticity, can adapt to the needs of different shapes and sizes of electrodes.

    The stretch characteristics of the material provide sufficient elasticity to ensure firm surface contact with the skin. The design of the devices is such that they can be used skin by reversing the surface contacting the skin.

    The Neurological Therapy Devices - Accessories utilize conductive silicone rubber as a conduit for current transmission. When current passes through the garment electrodes, the conductive silicone rubber transmit the current to the surface of the body, thus enabling the transmission of current. The entire surface of the Neurological Therapy Devices - Accessories is very conductive having a resistance of less than 7 ohms per inch. This provides low current density with uniform current distribution to enable efficient use of these Neurological Therapy Devices - Accessories for use in TENS (transcutaneous electrical nerve stimulators) and NMES (powered muscle stimulator devices).

    The Neurological Therapy Devices - Accessories are non- sterile external devices which are designed for single patient, for multiple uses and are intended for Prescription Use and/or OTC use with FDA Cleared TENS and NMES class II devices.

    AI/ML Overview

    The provided FDA 510(k) summary (K232995) describes the acceptance criteria and the study conducted to demonstrate that the Neurological Therapy Devices - Accessories meet these criteria.

    Here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    Material (Biocompatibility)Compliant with ISO 10993-5:2009 (Cytotoxicity)Compliant
    Compliant with ISO 10993-23:2021 (Skin Sensitization)Compliant
    Compliant with ISO 10993-10:2021 (Skin Irritation)Compliant
    Electrical PerformanceResistivity less than 7 ohms per inchLess than 7 ohms per inch (through multiple washings)
    Uniform delivery of low doses of current to the skinAchieved during bench testing
    Durability/ReusableMaintain inherent conductivity with multiple washingsBench testing confirmed no significant adverse effect on conductivity after multiple washings
    Intended UseTo be used with legally marketed electrical stimulating devices (TENS/NMES) to deliver stimulation signalsThe device is designed and demonstrated to work with such devices.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: The document explicitly states "No clinical testing was performed" (Section 8). Therefore, there isn't a "test set" in the context of human subjects or clinical data. The performance was evaluated through non-clinical bench testing. The sample sizes for the materials tested in biocompatibility and electrical performance bench tests are not specified in this summary.
    • Data Provenance: The company is Shenzhen Roundwhale Technology Co., Ltd. in China. The testing would presumably have been conducted in China or by laboratories contracted by the company, adhering to international standards. The data is non-clinical (bench testing).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not applicable. As no clinical testing was performed, there was no "test set" requiring expert ground truth establishment in a clinical context. The performance was assessed against technical specifications and international standards in a laboratory setting.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. There was no clinical test set requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an accessory (an electrode) for electrotherapy devices, not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study related to human reader performance with or without AI assistance is irrelevant and was not conducted.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm-only device. It is a physical accessory (electrode). The "standalone" performance refers to the device's electrical conductivity and biocompatibility, which were assessed via bench testing as described.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is based on:

    • International Standards: Specifically, ISO 10993 series for biocompatibility testing (ISO 10993-5, ISO 10993-10, ISO 10993-23).
    • Technical Specifications: Defined electrical performance criteria (e.g., resistivity less than 7 ohms per inch).
    • Comparison to Predicate Devices: Demonstrating similarity in design, materials, function, and performance to legally marketed devices.

    8. The sample size for the training set

    Not applicable. As this device is a physical cutaneous electrode and not an AI/machine learning model, there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203158

    Validate with FDA (Live)

    Device Name
    BioWave BioWraps
    Date Cleared
    2021-07-08

    (259 days)

    Product Code
    Regulation Number
    882.1320
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K171721, K944487

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BioWave BioWraps are cutaneous electrodes to be used with legally marketed BioWave branded neurostimulators. The knitted garment electrodes are non-sterile reusable prescription-use and OTC conductive garments that are intended to deliver the stimulation signals generated by the stimulator to the body surface with which they are in contact. These body parts include hand/wrist, elbow, foot/ankle, knee, and lower back.

    Device Description

    The BioWave BioWraps are washable, wrappable bands made from a stretchable neoprene-like outer and silver fiber conductive inner fabric. The electrodes are secured to the skin through the wrapping of the bands. The BioWraps are intended for use on the hand/wrist, elbow, foot/ankle, knee, and lower back. When used alongside a conductive cream, the electrodes provide a low current density with uniform current distribution. The BioWraps are to be used alongside cleared BioWave neurostimulator devices (the BioWavePRO® RX (K052289), BioWaveHOME® RX (K152437), and BioWaveGO® OTC (K180943)).

    AI/ML Overview

    The BioWave BioWraps are cutaneous electrodes designed for use with BioWave neurostimulators. The provided document is a 510(k) premarket notification and primarily focuses on demonstrating substantial equivalence to predicate devices, rather than a detailed clinical study proving the device meets specific acceptance criteria in a comparative effectiveness study with human readers.

    Here's an analysis of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" in the format of a clinical trial endpoint. Instead, it describes performance characteristics for substantial equivalence to predicate devices.

    Performance CharacteristicAcceptance Criteria (Implied by Substantial Equivalence)Reported Device Performance (BioWave BioWraps)
    ConductivitySimilar to predicate devicesHighly conductive, less than 1.27 ohms resistance per inch (similar or less than predicates)
    ResistivitySimilar to predicate devicesMeets design controls
    ImpedanceSimilar to predicate devicesMeets design controls; 1.27 ohms resistance per inch (compared to predicate's 7 ohms/inch and 2 ohms/inch)
    Uniform Current DeliveryConsistent with predicate devicesUniform delivery of low doses of current
    ReusabilityDemonstrated by predicate devicesDuly demonstrated, no significant adverse effect on conductivity after multiple washings
    BiocompatibilityCompliant with ISO 10993-5 and 10993-10Compliant with ISO 10993-5 and 10993-10 (cytotoxicity, sensitization, cutaneous reactivity met established acceptance criteria)

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: Not applicable. The "study" described is bench testing, not a clinical trial with a test set of patients or medical images.
    • Data Provenance: Not applicable for clinical data. The testing described is laboratory-based bench testing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • Not applicable. Ground truth as typically understood in medical device AI (e.g., expert consensus on diagnosis) is not relevant for this type of bench testing. The "ground truth" here is the adherence to engineering specifications and international standards.

    4. Adjudication Method for the Test Set:

    • Not applicable. There was no clinical test set requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • No, an MRMC comparative effectiveness study was not done. This document describes a 510(k) submission for cutaneous electrodes, which does not typically involve such studies. The device is not an AI-powered diagnostic tool, but rather an accessory to a neurostimulator. It does not involve human readers for interpretation in the way AI in radiology would.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    • Not applicable. This device is not an algorithm. The performance described is the physical and electrical characteristics of the garment electrodes themselves.

    7. The Type of Ground Truth Used:

    • The "ground truth" for the bench testing was based on engineering specifications, material science principles, and compliance with international standards (e.g., ISO 10993 for biocompatibility) and comparisons to the performance characteristics of legally marketed predicate devices.

    8. The Sample Size for the Training Set:

    • Not applicable. This is not an AI/ML device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable. No training set was used.