510(k) Data Aggregation

    K Number
    K182537

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-03-08

    (175 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K171138, K172951

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RADMAX Digital Imaging Software, used together with a digital X-ray image processing system designed for acquiring images and processing acquired images. The main features of the software are controlling and interfacing the detector, acquiring images after X-ray, storing acquired images, managing data, and image processing. The RADMAX Digital Imaging Software, is not intended for the acquisition of mammographic image data.

    The X-ray generator is not part of the RADMAX Digital Imaging Software, device.

    If the X-ray generator does not allow interfacing with external software the RADMAX Digital Imaging Software device cannot be interfaced with X-ray Generator.

    However, when using third-party generator, use the AED function to acquire the image by sensing the X-ray photon. When using the DRGEM Corporation, generator (models GXR, GXR-U), the RADMAX Digital Imaging Software can only select or change values of X-ray exposure parameters (kVP, mAs or kVP, mA or density).

    Device Description

    The RADMAX Digital Imaging Software from DRGEM Corporation, is a digital Xray image processing system designed for acquiring images and processing acquired images. The software can be used together with a digital X-ray detector and or an X-Ray generator. The main features of the software are controlling and interfacing the detector, acquiring images after X-ray, storing acquired images, managing data, and image processing.

    The X-ray generator or digital detector is not part of the RADMAX Digital lmaging Software, device. The RADMAX Digital Imaging Software does not control exposure or electrical charge and X-ray calibration. If the X-ray generator does not allow interfacing with external software like the RADMAX Digital Imaging Software device, then the software cannot be interfaced with X-ray Generator. The RADMAX Digital Imaging Software can only select or change values of X-ray exposure parameters (KVp. mA second or kVp. mAs) according to the defined value determined by each X-ray company.

    The RADMAX Digital Imaging Software, is not intended for the acquisition of mammographic image data.

    The RADMAX Digital Imaging Software device is not intended for the acquisition of mammographic image data and is meant to be used by qualified medical personnel only. The Users must be qualified to create and diagnose radiological image data.

    The main functions of the RADMAX Imaging Software are as follows:

    • . Acquisition and storage of digital X-ray images from a digital X-ray Detector.
    • Input Study information (patient information, exam information). .
    • . Management of stored (archived) images.
    • Image processing for enhancement of archived images. ●
    • Review of stored images. ●
    • Editing of images. ●
    • DICOM conformance (e.g. DICOM Storage, DICOM Work list, DICOM Print, etc.) ●
    • For a DR system (X-ray machine and generator and Digital X-rav detector, etc.) or a need to interface with installed X-ray system, the:
      • Ability to configure X-ray exposure condition (kVp, mA, Sec etc) for various body parts and positions.
      • -Communication between the Generator Console and the RADMAX device.

    The X-ray generator control function depends on the X-ray Generator company. The X-ray generator is not part of the RADMAX Digital Imaging Software device since the RADMAX device can only interface and control the Generator by the algorithm provided by the X-ray Company. The RADMAX device can only select or change values of X-ray exposure parameters (kVp, mA second or kVp; mAs) according to the defined value determined by each X-ray company.

    The RADMAX Digital Imaging Software, device does not control exposure or electrical charge and or calibration of the X-ray equipment. If the X-ray generator does not allow interfacing with an external software like the RADMAX Imaging Software, then RADMAX, cannot be interfaced with the X-rav Generator.

    AI/ML Overview

    The provided text describes the RADMAX Digital Imaging Software, focusing on its substantial equivalence to a predicate device (K123650: ARIX RAD ACQUISITION CONSOLE). Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't present a formal table of quantitative acceptance criteria with reported numerical performance values in the way a diagnostic AI would. Instead, it uses a comparative table to demonstrate "substantial equivalence" to a predicate device (K123650: ARIX RAD ACQUISITION CONSOLE) across various functionalities. The "acceptance criteria" here are essentially that the new device's functionality must be "Same" or, if "different," must have "No impact on safety or efficacy" and "no new potential or increased safety risks."

    ItemFunctionalitySubject Device: RADMAX Digital Imaging SoftwarePredicate K123650 ARIX RAD Acquisition ConsoleIf different, Impact on Safety and or EfficacyReported Device Performance (as per document)
    1Computer & Operating SystemMicrosoft Windows 10 (minimum)Microsoft Windows XP/Windows 7 (minimum)Yes, there is a difference. Differences are in the processor speeds, Windows operating system, amount of RAM memory, monitors, and hard drive space requirements. These differences are due to the state of computer technology that was available at the time the predicate device was developed. There is "No impact on safety or efficacy" and there are no new potential or increased safety risks.Meets modern OS requirements; deemed equivalent in safety/efficacy
    2Intended useAcquiring, viewing, editing and storing radiographs and related patient imagesSameNo differenceSame as predicate
    3Intended userradiologistradiologistNo differenceSame as predicate
    4Network10/100/100 EthernetSameNo differenceSame as predicate
    5MonitorTwo color display adaptors 1920x108019-inch monitor using 1280x1024Yes, there is a difference. See item 1 above.Meets modern display requirements; deemed equivalent in safety/efficacy
    6User interaction/inputMouse, keyboard, touch monitorSameNo differenceSame as predicate
    7Multi-userAvailable, but at a time, only one user can use itSameNo differenceSame as predicate
    8Import / export imagesYesYesNo differenceSame as predicate
    9Acquisition devicesDigital X-Ray detectorsDigital X-Ray detectorsNo differenceSame as predicate
    10Imaging interfacesDetector dependentDetector dependentYes, there is a difference. Subject device uses Flat panel detectors cleared by FDA: K171138, & K172951Compatible with modern FDA-cleared flat panel detectors; deemed equivalent in safety/efficacy
    11Image organizationYes. Patient ID, Name, study instance UIDSameNo differenceSame as predicate
    12Image search availableYesYesNo differenceSame as predicate
    13Image storageYesYesNo differenceSame as predicate
    14Database softwareMySQLMySQLNo differenceSame as predicate
    15Image viewingYesYesNo differenceSame as predicate
    16Image measurementYesYesNo differenceSame as predicate
    17Image annotationYesYesNo differenceSame as predicate
    18Image operationsYesYesNo differenceSame as predicate
    19SecurityYes (Priority by user)SameNo differenceSame as predicate
    20DICOM 3.0 compatibilityYesYesNo differenceSame as predicate
    21Generator ControlYesYesNo differenceSame as predicate
    22Generator Control ProtocolGenerator dependentGenerator dependentNo differenceSame as predicate
    23RAW Image data processingYesYesNo differenceSame as predicate
    24Post image data processingYesYesNo differenceSame as predicate
    25RIS code managerYesYesNo differenceSame as predicate
    26WorklistYesYesNo differenceSame as predicate
    27Patient size/LateralityYesYesNo differenceSame as predicate
    28Display radiographic technique, kV, mA, ms, mAsYesYesNo differenceSame as predicate
    29Thumbnail viewingYesYesNo differenceSame as predicate
    30LoginYesYesNo differenceSame as predicate
    31New patient manual registerYesYesNo differenceSame as predicate
    32X-Ray generator windowYesYesNo differencesSame as predicate
    33Bucky selectionYesYesNo differenceSame as predicate
    34Body partYesYesNo differenceSame as predicate
    35Generator status displayYesYesNo differenceSame as predicate
    36Image resetYesYesNo differenceSame as predicate
    37panningYesYesNo differenceSame as predicate
    38Magnify glassYesYesNo differenceSame as predicate
    39Fit imageYesYesNo differenceSame as predicate
    40Image StitchingYesYesNo differenceSame as predicate
    41Series/Image listYesYesNo differenceSame as predicate

    2. Sample Size Used for the Test Set and Data Provenance:

    The document mentions "nonclinical testing results are provided in the 510(k)" and that "The Validation Test Plan was designed to evaluate all input functions, output functions, and actions performed by the RADMAX Digital Imaging Software and followed the process documented in the System Validation Test Plan." However, it does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective) for any testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    The document does not provide information on the number of experts used or their qualifications for establishing ground truth. The device functionality is primarily compared against a predicate device based on technical specifications rather than diagnostic performance against clinical ground truth.

    4. Adjudication Method for the Test Set:

    The document does not describe any adjudication method like 2+1 or 3+1. The testing appears to be focused on system validation and functionality comparison, not on inter-rater agreement for diagnostic interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

    No MRMC comparative effectiveness study was mentioned in the provided text. The document focuses on demonstrating substantial equivalence in features and functionality, not on how the software improves human reader performance or diagnostic accuracy.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

    The document explicitly states: "Diagnosis is not performed by the software but by Radiologists, Clinicians and or referring Physicians." This indicates the device is not a standalone diagnostic algorithm; it's an imaging software for acquisition, processing, and management. Therefore, a standalone performance study as typically understood for an AI diagnostic algorithm was not performed or applicable in this context.

    7. The Type of Ground Truth Used:

    Given that the device is primarily a software for image acquisition, processing, and management, the "ground truth" for its validation appears to be functional correctness and adherence to technical specifications rather than clinical ground truth (like pathology results or physician consensus on a diagnosis). The "predetermined acceptance criteria" mentioned in the nonclinical testing section would likely relate to whether the software performs its intended functions as designed (e.g., correct image acquisition, storage, processing, display of controls).

    8. The Sample Size for the Training Set:

    The document does not mention any training set size. This is consistent with the device being a general imaging software rather than a machine learning or AI-driven diagnostic tool that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Since there's no mention of a training set, there's no information on how its ground truth would have been established.