Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K260292

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2026-02-27

    (29 days)

    Product Code
    Regulation Number
    870.5175
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K081645, K183174, K170997, K151649, K822572, K092761

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HOTWIRE™ is indicated for creation of an atrial septal defect in the heart.

    Device Description

    HOTWIRE™ Kit (901XXX):

    The HOTWIRE™ is a sterile, single-use guidewire device that delivers radiofrequency (RF) power in a monopolar mode to a distal electrode segment for the creation of an atrial septal defect in the heart. The HOTWIRE™ is intended to be used in conjunction with compatible third-party intravascular sheaths and/or dilators, such as the Agilis NXT Steerable Introducer, St. Jude Medical/Abbott (K081645), and third-party RF electrosurgical generator(s), such as the Valleylab FT10 Electrosurgical Platform (K151649) which utilizes a commercially-available Patient Return Electrode (PRE) which is in compliance with IEC 60601-2-2, such as the Valleylab E7507DB (K822572).

    The HOTWIRE™ is comprised of a stainless steel core wire. The main body of the wire is jacketed with an insulating polymer that provides electrical insulation and facilitates smooth movement of the device through vascular dilators and/or sheaths. The floppy distal segment of the wire has an atraumatic tip with an uninsulated stainless-steel coil, which serves as an electrode, and also provides fluoroscopic and echogenic visualization. A tungsten marker coil at the tip provides additional radiopacity. The stiff body of the HOTWIRE™ provides support for advancing wire-guided devices into the left atrium after the distal segment has traversed the septum. The proximal insulated portion of the wire has visual markers that align the electrode tip with third-party transseptal sheaths and/or dilators. A portion of the proximal wire is uninsulated for placement of an included Adapter Pin that connects to hand pieces used with compatible third-party RF electrosurgical generators.

    HOTWIRE™ + Handpiece Kit (902XXX):

    The HOTWIRE™ + Handpiece Kit is a sterile, single-use guidewire device that delivers radiofrequency (RF) power in a monopolar mode to a distal electrode segment for the creation of an atrial septal defect in the heart. The HOTWIRE™ + Handpiece Kit is intended to be used in conjunction with compatible third-party intravascular sheaths and/or dilators, such as the Agilis NXT Steerable Introducer, St. Jude Medical/Abbott (K081645), and the HOTWIRE™ System RF Generator, which utilizes a commercially-available Patient Return Electrode (PRE) which is in compliance with IEC 60601-2-2, such as the Valleylab E7507DB (K822572).

    The HOTWIRE™ is comprised of a stainless steel core wire. The main body of the wire is jacketed with an insulating polymer that provides electrical insulation and facilitates smooth movement of the device through vascular dilators and/or sheaths. The floppy distal segment of the wire has an atraumatic tip with an uninsulated stainless-steel coil, which serves as an electrode, and also provides fluoroscopic and echogenic visualization. A tungsten marker coil at the tip provides additional radiopacity. The stiff body of the HOTWIRE™ provides support for advancing wire-guided devices into the left atrium after the distal segment has traversed the septum. The proximal insulated portion of the wire has visual markers that align the electrode tip with third-party transseptal sheaths and/or dilators. A portion of the proximal wire is uninsulated for placement of an included handpiece that connects with the HOTWIRE™ System RF Generator.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K221044

    Validate with FDA (Live)

    Date Cleared
    2022-05-05

    (27 days)

    Product Code
    Regulation Number
    870.1280
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K170997

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AcQGuide® VUE Steerable Sheath with electrodes is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart.

    The AcQGuide® VUE deflection facilitates catheter positioning.

    The electrodes help facilitate visualization of the sheath when used with a compatible localization system, such as the AcQMap System.

    Device Description

    The AcQGuide® VUE Steerable Sheath, Model 900201 is a single use, percutaneous catheter introducer designed to provide additional maneuverability to interventional catheters that are advanced through the sheath and into the right or left chambers of the heart. The distal portion of the sheath is comprised of a composite structured single lumen shaft. At the proximal end, an ergonomic handle provides torque and active deflection, a hemostasis valve allows safe introduction of an interventional catheter, and a side port provides access for aspiration, fluid flushes and fluid/medication infusions. There is a cable exiting from the handle of the sheath to provide connection for the distal electrodes. This cable terminates with four standard 2mm pins that are connected to the Auxiliary Interface Box or Auxiliary Catheter Cable of the AcQMap High Resolution Imaging and Mapping System, Model 900100 or 900000. The pin connections are assigned by the user for electrode identification (D1, 2, 3, 4) to the corresponding channel on the AcQMap System 900100. No designated order of pin connection is required. The AcQGuide® VUE Steerable Sheath includes a valve bypass tool as an optional accessory to facilitate the insertion of a loop or circular catheter.

    The dilator is designed to introduce the Steerable Sheath into the vasculature and into chambers of the heart. The dilator has a smooth tapered tip and provides a smooth transition to the round edge of the non-traumatic tip of the sheath. The dilator is able to track over a 0.035″ guidewire. The hub section is attached to the shaft. It is a standard female luer made of high-density polyethylene (HDPE) material.

    Some of the key design attributes of the Steerable Sheath include:

    • Deflectability
    • Hemostasis
    • Kink-resistance
    • Visibility under fluoroscopy
    • Easily flushed during the procedure
    • Single-handed operation
    • Biocompatible materials
    • Sterile, single use
    • Combined three-way stopcock
    AI/ML Overview

    The AcQGuide® VUE Steerable Sheath is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart. Its deflection facilitates catheter positioning, and its electrodes aid in visualization with a compatible localization system like the AcQMap System.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    Based on the provided information, the acceptance criteria are largely implied by successful completion of various nonclinical tests, demonstrating the device's conformance to specifications and comparable safety and effectiveness to the predicate device.

    Acceptance Criteria CategorySpecific Criteria (Implied)Reported Device Performance
    Packaging & Shelf-LifeMaintains sterility and product integrity over shelf-life.Passed. Shelf life is 24 months, differing slightly but not raising new safety/effectiveness concerns compared to predicate's 18 months.
    SterilizationAchieves a sterility assurance level (SAL) of 10^-6 via Ethylene Oxide (EtO).Passed. Subjected to the identical EtO sterilization process as the predicate, meeting SAL of 10^-6. Confirmed suitable for existing EtO processing cycle. Non-pyrogenic.
    BiocompatibilityBiocompatible for external communicating device, with circulating blood contact for ≤24 hours.Passed. Conducted in accordance with ISO 10993-1:2018 and FDA Guidance. Categorized as external communicating device, circulating blood contact ≤24 hours. Materials similar to predicate.
    Dimensional SpecificationsAll device components meet required dimensions.Passed. Ensured through dimensional testing.
    Visual InspectionAll components free of defects.Passed. Ensured through visual inspection.
    Leak TestingPackaging meets specifications.Passed. Ensured through leak testing.
    Functional & CompatibilityMeets functional and compatibility requirements.Passed. Ensured through functional and compatibility testing.
    Mechanical TestingComponents withstand expected forces.Passed. Ensured through mechanical testing.
    Handle Torque TestingWithstands expected torques.Passed. Ensured through handle torque testing.
    Physician Simulated UseConsistent performance during intended use.Physician simulated use in an animal model was conducted.
    Electrical SafetyComplies with IEC 60601-1 for basic safety and essential performance.Passed. Assessed for Defibrillation Proof Applied Parts, Dielectric Voltage Withstand, and Sheath Leakage Current.
    Electromagnetic Compatibility (EMC)Complies with IEC 60601-1-2 for electromagnetic disturbances.Passed. Qualification testing of the AcQMap System with applied parts containing cables and Platinum-Iridium Electrodes, and with general-purpose electrophysiology catheters.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document primarily describes nonclinical (bench) testing. For these tests, specific sample sizes are not explicitly provided in the summary. The studies are prospective in nature, as they involve testing newly manufactured devices to verify their performance against established specifications. The data provenance is Acutus Medical, Inc., Carlsbad, CA, U.S.A., implying the testing was conducted or overseen by the company.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    The document does not mention the use of experts to establish a "ground truth" for the nonclinical test sets in the way it would for a clinical study with image interpretation or disease diagnosis. The "truth" for these bench tests is derived from engineering specifications, industry standards (e.g., ISO, IEC), and established medical device testing protocols. The "physician simulated use in an animal model" likely involved qualified medical professionals, but their specific number, qualifications, and their role in establishing a "ground truth" (beyond observing performance) are not detailed.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    No adjudication method is mentioned. For the nonclinical bench tests described, the outcomes are objective and based on meeting predefined specifications, rather than subjective interpretations requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was conducted or mentioned. The device described is a steerable sheath with electrodes for visualization, not an AI-powered diagnostic tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This refers to a medical device's performance without direct human intervention, especially applicable to AI algorithms. The AcQGuide® VUE Steerable Sheath is a physical medical device that inherently requires human-in-the-loop operation by a physician. Therefore, a "standalone algorithm only" performance study is not applicable to this device. Its performance is evaluated in relation to its mechanical, electrical, and material properties, and its interaction with a compatible localization system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the nonclinical testing described, the "ground truth" is defined by:

    • Engineering Specifications: For dimensional, mechanical, and functional tests.
    • International Standards: Such as ISO 10993-1 for biocompatibility, ISO 11135 and AAMI TIR28 for sterilization, and IEC 60601-1/1-2 for electrical safety and EMC.
    • Predefined Performance Metrics: For tests like leak testing, torque, and compatibility.
    • Physiological Response: In the animal model for physician simulated use.

    8. The sample size for the training set

    This question is not applicable. The AcQGuide® VUE Steerable Sheath is a physical medical device, not a software algorithm that requires a training set in the context of machine learning or artificial intelligence.

    9. How the ground truth for the training set was established

    This question is not applicable, as there is no "training set" for this physical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1