510(k) Data Aggregation

    K Number
    K180871

    Validate with FDA (Live)

    Date Cleared
    2018-09-14

    (164 days)

    Product Code
    Regulation Number
    868.5630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K170916

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Portable Nebulizer (MBPN002) is a vibrating mesh nebulizer system designed to aerosolize liquid medications for inhalation by the patient. The device may be used with pediatric (5 years and older), defined by the prescribed medication, and adult patients in hospital/institutional settings, home care use, schools, and long term care facilities.

    Device Description

    The Portable Nebulizer (MBPN002) is a small, handheld general purpose nebulizer which utilizes vibrating mesh technology to generate aerosol.

    AI/ML Overview

    The provided text describes a 510(k) submission for a modified Portable Nebulizer (MBPN002). This submission does not include a study involving AI, human readers, or a test set as would typically be described for an AI/ML device.

    The "acceptance criteria" discussed are primarily related to performance specifications and compliance with standards for a medical device, rather than the performance of an AI algorithm based on a test set. The "study" mentioned is a set of non-clinical performance tests and verifications to demonstrate that the modified device is substantially equivalent to its predicate.

    Here's an analysis based on the provided text, addressing the requested points where applicable, and noting when information is not present:


    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The acceptance criteria here are drawn from the performance specifications and the "Comparison to Predicate" table, which highlight what the device should do, and the "reported device performance" is essentially the confirmation that it meets these as identical to the predicate or through specific tests.

    Acceptance Criteria (from Predicate/Modified Device Specifications)Reported Device Performance (as demonstrated by testing/comparison)
    Indications for use remain identical: Aerosolize liquid medications for inhalation by pediatric (5+ years) and adult patients in hospital/institutional, home care, school, and long-term care facilities.Identical to predicate K170916.
    Patient Population remains identical: Pediatric (5+ years) and adult patients.Identical to predicate K170916.
    Environment of Use remains identical: Hospital/institutional, home care, schools, and long-term care facilities.Identical to predicate K170916.
    Contraindications remain identical: None.Identical to predicate K170916.
    Principle of Operation remains identical: Vibrating mesh technology.Identical to predicate K170916.
    Aerosolization remains identical: Continuous during inhalation and exhalation.Identical to predicate K170916.
    Compressed gas source: None needed.Identical to predicate K170916.
    Reservoir volume: 6 ml.Identical to predicate K170916.
    Nebulization rate: > 0.25 ml/min.Identical to predicate K170916. Particle characterization and nebulization rate are unchanged as the main unit operates at a consistent 15V.
    Duration of Use: Single patient, multi-use.Identical to predicate K170916.
    Nebulizer components cleanable: Yes.Identical to predicate K170916. Cleaning and disinfection validated.
    Software driven: Yes (firmware).Identical to predicate K170916.
    Power source: 2 "AA" batteries OR external Class II IEC 60601-1 compliant supply.The modified device adds external power, but "no change to device performance" is noted due to controlled 15V operation. Electrical safety and EMC testing confirmed compliance with relevant standards (ANSI/AAMI/ES 60601-1, IEC 60601-1-2, IEC 60601-1-11).
    Power consumption: < 1.5 W.Identical to predicate K170916.
    Weight: 74 gm w/o batteries.Identical to predicate K170916.
    Dimensions (mm): 77 x 41 x 73.Identical to predicate K170916.
    Operating Conditions: 10 to 40°C / 30-85% RH.Identical to predicate K170916.
    Storage Conditions: -20 to + 70°C / 20-75% RH.Identical to predicate K170916.
    User interface: On/Off switch, LED indicators.Identical to predicate K170916.
    Standards met: ANSI/AAMI/ES60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11.Electrical safety and EMC testing confirmed compliance (ANSVAAMI/ES 60601-1:2005+A1:2012; IEC 60601-1-2:2014; IEC 60601-1-11:2015).
    Biocompatibility (ISO 10993-1): External Communicating (Indirect gas pathway), Tissue/Bone/Dentin communicating (permanent), Surface Contact (Mucosal membrane, permanent).Cytotoxicity, Sensitization, Irritation, Sub-chronic Systemic toxicity, Acute Systemic Toxicity, Leachable and Extractables, Gas Emission (VOC, Inorganic compounds, PM2.5) concluded biocompatible.
    Cleaning and Disinfection: Effective manual cleaning and thermal disinfection.Validation confirmed effective cleaning and disinfection by listed methods.
    Particle Characterization (MMAD): Equivalent to predicate.Verified after two years of real-time aging using cascade impactor and laser light scattering.

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not applicable or not provided in the context of an AI/ML device test set. The submission describes non-clinical testing of the physical medical device.

    • No specific "test set" in terms of patient data or images is mentioned.
    • The tests performed are for electrical safety, EMC, shelf-life, useful-life, durability, cleaning, particle characterization, and biocompatibility, using physical samples of the device and materials.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not applicable or not provided. "Ground truth" in the context of expert consensus on data is not relevant to this type of device submission. Device performance was evaluated against engineering specifications and regulatory standards by qualified testing facilities or personnel.


    4. Adjudication Method for the Test Set

    This information is not applicable or not provided. Adjudication methods are typically used for establishing ground truth in clinical studies involving interpretation, which is not what this submission describes.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If so, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This information is not applicable or not provided. This is not an AI-powered device, and therefore no MRMC study, human reader improvement, or AI assistance is mentioned.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable or not provided. This is not an algorithmic device.


    7. The Type of Ground Truth Used

    For this medical device, the "ground truth" refers to established engineering specifications, regulatory standards (e.g., IEC 60601 series, ISO 10993-1), and the performance characteristics of the legally marketed predicate device (K170916).
    The tests performed (electrical safety, EMC, biocompatibility, particle characterization, cleaning/disinfection validation) were conducted against predefined, measurable criteria and standards.


    8. The Sample Size for the Training Set

    This information is not applicable or not provided. There is no mention of a training set as this is not an AI/ML device.


    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable or not provided. There is no mention of a training set or ground truth establishment for such.