510(k) Data Aggregation

    K Number
    K183288

    Validate with FDA (Live)

    Date Cleared
    2019-07-30

    (246 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K133929, K170903, K172079

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, back, arm, leg, foot, due to strain from exercise or normal household and work activities.

    Device Description

    Transcutaneous Electrical Nerve Stimulator is a product that adopts modern electronic science and technology to deliver electric pulses generated to the user's skin through the electrodes.

    There are three models of Transcutaneous Electrical Nerve Stimulator which are KTR-206, KTR-208 and KTR-209. Their technical parameters are slightly different, but they share the basically same characteristics: 1) They are small, exquisite and portable; 2) various modes to satisfy different demands, applicable to a wider range of people; 3) wonderful electric pulse combination. 0~16 levels can be adjusted and chosen according to personal preference; 4) LCD display make the operation simple and easy; 5) integrated design of the body is easy for function operation and simple in practical use; 6) battery power display; 7) dual channel output. user can cover more treatment areas. For KTR-208 and KTR-209, the two channels are independently controlled for intensity adjustment which is more convenient to use.

    The Transcutaneous Electrical Nerve Stimulator is mainly composed of the host and electrode patches. And it uses AAA batteries for power supply. To start therapy, first insert batteries, then paste the electrode patches onto painful areas and press on/off button to power on. The modes and intensity can be selected according to needs. And the current status is displayed on LCD.

    AI/ML Overview

    The provided text is a 510(k) Summary for a Transcutaneous Electrical Nerve Stimulator (TENS) device. This document is a premarket submission to the FDA demonstrating that the new device is substantially equivalent to a legally marketed predicate device.

    Here's an analysis of the provided text, addressing your specific questions, keeping in mind that a 510(k) typically focuses on substantial equivalence rather than a full clinical trial to establish new clinical efficacy. Therefore, some information, like detailed acceptance criteria from a clinical study or human reader performance, may not be explicitly present as they aren't usually required for a 510(k) for this type of device.

    1. Table of Acceptance Criteria and the Reported Device Performance

    For a TENS device, acceptance criteria in a 510(k) submission primarily revolve around safety, electrical performance, and substantial equivalence to a predicate device, rather than clinical efficacy metrics (like pain reduction improvement percentage). The document establishes "substantial equivalence" as the primary acceptance criterion.

    The table below summarizes the comparison to the predicate device, which serves as the de-facto performance standard for substantial equivalence. The "Acceptance Criteria" are implied by the predicate values and compliance with recognized standards.

    ItemAcceptance Criteria (Implied by Predicate/Standards)Reported Device Performance (Targeted Device)Outcome
    Regulatory Information
    Regulation number21 CFR 882.589021 CFR 882.5890Same
    Regulation descriptionTranscutaneous electrical nerve stimulator for pain reliefTranscutaneous electrical nerve stimulator for pain reliefSame
    Product codeNUHNUHSame
    ClassIIIISame
    Indications for UseTemporary relief of pain associated with sore or aching musclesTemporary relief of pain associated with sore and aching muscles in the shoulder, back, arm, leg, foot, due to strain from exercise or normal household and work activities.Similar
    Patient populationAdultAdultSame
    Location for useOTCOTCSame
    Basic Unit Specifications
    Leakage currentCompliant with IEC 60601-2-10 (Battery operated: N/A)N/A (Battery operated)Same
    Software/Firmware/Microprocessor Control?YesYesSame
    Automatic Overload tripYes (Predicate: Yes)No (Predicate: Yes)Different
    Automatic no-load tripYesYesSame
    Patient override control methodOn/Off buttonOn/Off buttonSimilar
    Indicator display -On/Off statusYesYesSame
    Indicator display -Low batteryYesYesSame
    Automatic Shut OffYesYesSame
    Housing material and constructionABSABSSame
    Compliance with voluntary standardsIEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 60601-1-11IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 60601-1-11Similar
    Compliance with 21CFR 898YesYesSame
    Output Specifications
    WaveformBiphasic, Pulsed symmetric, square wave (Predicate is various)Biphasic, Pulsed symmetric, square waveSimilar
    Net Charge (per pulse)00Same
    Maximum Average Current ($500\Omega$)<50mA (Compliant with IEC 60601-2-10)KTR-206: 8.04mA; KTR-208: 6.89mA; KTR-209: 12.39mADifferent (but compliant)
    Max current density ($500\Omega$)~0.188-4.8 mA/cm² (Predicate range)KTR-206: 0.26mA/cm²; KTR-208: 0.22mA/cm²; KTR-209: 0.4mA/cm²Similar
    Max power density ($500\Omega$)~0.00752-1.38 W/cm² (Predicate range)KTR-206: 0.001W/cm²; KTR-208: 0.0008W/cm²; KTR-209: 0.0025W/cm²Similar
    Pulse frequency1-150Hz (Predicate range)KTR-206: 1 Hz-108Hz; KTR-208: 1 Hz-109Hz; KTR-209: 1 Hz-110HzSimilar
    Pulse duration50-250μs (Predicate range)KTR-206: 84μs-134μs; KTR-208: 82μs-128μs; KTR-209: 80μs-224μsSimilar

    Notes on Differences and Compliance:

    • Number of output modes: Targeted device has 5, predicate has 8, reference devices have more. The submission argues this difference doesn't affect safety and effectiveness as the output parameters of each mode were tested and passed relevant IEC standards (Note 1).
    • Appearance, weight, dimensions: Differ from predicate but are deemed insignificant and do not affect safety and effectiveness (Note 2).
    • Maximum Average Current: Different from predicate but all targeted device values (8.04mA, 6.89mA, 12.39mA) are < 50mA, complying with IEC 60601-2-10 requirements (Note 3), therefore not affecting safety and effectiveness.

    Study Proving Device Meets Acceptance Criteria:

    The study proving the device meets the acceptance criteria is primarily a non-clinical study involving various tests to demonstrate compliance with recognized electrical safety and performance standards, and comparison of technical characteristics to a legally marketed predicate device. The goal is to establish "Substantial Equivalence (SE)" to the predicate, K140168.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not applicable in the context of this 510(k) summary regarding a TENS device. This document describes non-clinical testing of the device itself (electrical parameters, safety, software verification) rather than testing on human subjects for clinical efficacy. There is no "test set" of patient data mentioned for evaluation of diagnostic or therapeutic performance in humans.
    • Data Provenance: The data provenance is from non-clinical bench testing of the new Transcutaneous Electrical Nerve Stimulator device and comparison of its specifications to the predicate and reference devices. This is not patient data from a specific country or retrospective/prospective study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. There is no "ground truth" related to patient outcomes or diagnoses in this 510(k) summary. The "ground truth" for the non-clinical tests would be the specifications and requirements outlined in the referenced standards (e.g., IEC 60601-1, IEC 60601-1-2) which are established by expert consensus in regulatory and technical committees.

    4. Adjudication method for the test set

    • Not applicable. As there is no clinical test set with patient data requiring interpretation or consensus, no adjudication method like 2+1 or 3+1 was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is a TENS device, which directly delivers electrical stimulation for pain relief. It is not an AI-powered diagnostic or assistive technology that human "readers" would use. Therefore, an MRMC comparative effectiveness study involving AI assistance is entirely outside the scope of this device and submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable in the sense of an algorithm performance study. The device is a physical TENS unit. Its "standalone" performance is measured by its electrical output characteristics and compliance with safety standards, as detailed in the technical comparison and non-clinical tests.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" in this context is established by international consensus standards for medical electrical equipment, particularly those related to TENS devices (e.g., IEC 60601 series). These standards define safety and performance requirements that the device must meet. Additionally, the technical specifications of the predicate device serve as a benchmark for demonstrating substantial equivalence.

    8. The sample size for the training set

    • Not applicable. This device is not an algorithm that requires a "training set" of data.

    9. How the ground truth for the training set was established

    • Not applicable. As there is no training set for an algorithm, there is no ground truth established for one.
    Ask a Question

    Ask a specific question about this device

    K Number
    K181955

    Validate with FDA (Live)

    Device Name
    WiEMSpro
    Date Cleared
    2019-02-12

    (204 days)

    Product Code
    Regulation Number
    890.5850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K133225, K170903

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    WIEMSPRO is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The WIEMSPRO is not intended to be used in conjunction with therapy or treatment of medical conditions of any kind. None of WIEMSPRO training programs is designed for injured or ailing muscles and its use on such muscles is contraindicated.

    WIEMSPRO is a machine with electronic muscle stimulation based on EMS technology. Regarding is specifically designed as an addition to other sports and for training muscles. It must be used for only healthy muscles and clients, not for rehabilitation purposes.

    WIEMSPRO electrical impulses allow the triggering of action potentials on motoneurones of motor nerves (excitations). These excitations of motoneurones are transmitted to the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration, duration, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles.

    Device Description

    The device described in this submission is an electro-medical device intended for stimulating healthy muscles in order to improve or facilitate muscle performance. It is designed for personal training performances.

    WIEMSero is a device with electronic muscle stimulation based on EMS technology. Regarding its use, the device is specifically designed as an addition to other sports and for training muscles. It must be used for only healthy muscles and clients, not for rehabilitation purposes. The device must be used over user's cotton undergarments and is therefore not patient-contacting

    WIEMSPRO system cannot be used while the user is in motion or lifting weights.

    AI/ML Overview

    The provided document is an FDA 510(k) premarket notification for the WIEMSPRO device, a powered muscle stimulator. It focuses on demonstrating substantial equivalence to previously cleared predicate devices rather than proving a specific performance metric against acceptance criteria through a clinical study for an AI/ML powered device.

    Therefore, the requested information regarding acceptance criteria, study design, sample sizes, expert involvement, and ground truth establishment, which are typical for clinical validation of AI/ML devices, is not present in this document.

    The document discusses bench testing to determine conformance to performance specifications and requirements, particularly for electrical safety, electromagnetic compatibility, and FCC radio frequency testing. It states that "No clinical studies are submitted."

    Given the document's content, here's an attempt to answer the questions based on what is provided and to highlight what information is missing:

    Acceptance Criteria and Device Performance (Based on Technical Specifications and Bench Testing)

    Since this is a 510(k) for a powered muscle stimulator, the "acceptance criteria" discussed are primarily related to meeting established electrical and safety standards for this class of device, and demonstrating equivalence to predicate devices based on technical characteristics. There isn't an explicit table of "AI performance metrics" and their acceptance thresholds as would be common for AI/ML device submissions.

    The device performance is reported implicitly by stating compliance with various IEC and FCC standards, and noting the technical specifications in comparison to predicate devices.

    Table of "Acceptance Criteria" (Implied by Compliance and Equivalence) and Reported Performance:

    Implied "Acceptance Criteria" CategorySpecific Criteria (Based on Standards & Comparison)Reported Device Performance / Compliance
    Electrical SafetyCompliance with IEC 60601-1 (general requirements for basic safety and essential performance of medical electrical equipment), IEC 60601-2-10 (particular requirements for nerve and muscle stimulators), and related standards (e.g., ANSI/AAMI ES60601-1). This includes aspects like patient leakage current, power source safety, and automatic shut-off mechanisms.- Passed "bench testing to determine conformance to performance specifications and requirements taking account of its intended use."- "Electrical safety (including particular requirements for the basic safety and essential performance of nerve and muscle stimulator and for medical electrical equipment)" was performed and showed correct operation.- Complies with IEC 60601-1-2:2007, IEC 60601-2-10:2012, IEC 60601-1-6 Edition 3.1 2013-10, ANSI/AAMI ES60601-1:2005 / A2:2010.
    Electromagnetic Compatibility (EMC)Compliance with IEC 60601-1-2 (requirements for electromagnetic compatibility of medical electrical equipment) and FCC Part 15 (unintentional radiators).- "Electromagnetic compatibility" testing was performed and showed correct operation.- Complies with IEC 60601-1-2:2007 and FCC 47 CFR Part 15 §15.107 and §15.109.
    Software Life Cycle ProcessesCompliance with IEC 62304 (medical device software – software life cycle processes).- Complies with IEC 62304:2006.
    Risk ManagementCompliance with ISO 14971 (application of risk management to medical devices).- Complies with ISO 14971:2007.
    Technical Equivalence to PredicatesWIEMSPRO's technological characteristics (e.g., classification, product code, regulation number, panel, class, intended use, power source, output modes, current/voltage, waveform, safety features) should be substantially equivalent to the predicate devices (E-Fit EF-1280 and Compex WirelessUSA) or any differences should not affect safety and effectiveness.- Comparison table provided, indicating "Same" or "Similar" for most characteristics.- Notable differences (e.g., higher max output current, but lower current density compared to one predicate; absence of TENS function present in one predicate) are argued not to affect safety and effectiveness. "Maximum output current value is higher... however maximum output current density... is much lower than that of Compex Wireless USA... none of the differences spotted impact the equivalence..."

    Study Details (Information Not Applicable or Not Provided for this Device Type/Submission)

    The following points are primarily relevant for AI/ML device submissions involving clinical performance evaluation, which is not the nature of this 510(k) submission.

    1. Sample sizes used for the test set and the data provenance:
      • Not Applicable / Not Provided. This document describes "bench testing" for electrical safety and EMC, and a comparison to predicate devices, not a clinical study with a test set of patient data for AI performance.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      • Not Applicable. No clinical test set requiring expert-established ground truth is detailed.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
      • Not Applicable. No clinical test set requiring adjudication is detailed.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      • Not Applicable / No. This device is a muscle stimulator, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was performed or described.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      • Not Applicable. This is a physical medical device, not an AI algorithm. Its performance is evaluated through engineering and safety standards, not standalone algorithmic performance.
    6. The type of ground truth used (expert concensus, pathology, outcomes data, etc.):
      • Not Applicable. For the bench testing, the "ground truth" would be established engineering standards and test procedures for electrical and electromagnetic performance.
    7. The sample size for the training set:
      • Not Applicable. This is not an AI/ML device that undergoes training on a data set.
    8. How the ground truth for the training set was established:
      • Not Applicable. No training set exists for this device.
    Ask a Question

    Ask a specific question about this device