Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K193663

    Validate with FDA (Live)

    Date Cleared
    2021-03-10

    (435 days)

    Product Code
    Regulation Number
    870.4360
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K983272, K170029

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FloPump 57mL Centrifugal Pump is a device that uses a method other than revolving rollers to pump the blood through an extracorporeal circuit for periods lasting less than 6 hours for the purpose of providing either:

    i) Full or partial cardiopulmonary bypass (i.e., circuit includes an oxygenator) during open surgical procedures on the heart or great vessels; or

    ii) Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory pathway necessary for open surgical procedures on the aorta or vena cava.

    Device Description

    The FloPump 57mL is a single use disposable centrifugal pump has an inlet that draws blood from a patient and an outlet that pushes the blood out, where it then passes through an oxygenator and back to the patient. The pump mates with the Sorin Group Stöckert Centrifugal Pump Console and rotates the internal impeller using a magnetic driver. The FloPump 57mL is a non-occlusive pump. The pump has a spinning rotor with flow channels which imparts rotary motion to the incoming blood, directing it through a spiral housing to the outflow port. The FloPump 57mL is part of the extracorporeal circuit, and is therefore in contact with the patient's blood while circulating. The FloPump 57mL does not have any other patient contact.

    AI/ML Overview

    This document describes the FloPump 57mL Centrifugal Pump, a non-roller type blood pump intended for use in extracorporeal circuits for periods less than 6 hours.

    Here's an analysis of the acceptance criteria and supporting studies as presented in the document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly list specific quantitative "acceptance criteria" with corresponding "reported device performance" in a structured table as might be expected for an AI/CAD device. Instead, it states that non-clinical testing was performed and the results were "Substantially equivalent to predicates" for a range of performance characteristics.

    Performance CharacteristicAcceptance Criteria (Implicit)Reported Device Performance
    Flow curvesSubstantially equivalent to predicate devicesSubstantially equivalent to predicates
    Heat generationSubstantially equivalent to predicate devicesSubstantially equivalent to predicates
    Prime volumeSubstantially equivalent to predicate devicesSubstantially equivalent to predicates
    Air handlingSubstantially equivalent to predicate devicesSubstantially equivalent to predicates
    HemolysisSubstantially equivalent to predicate devicesSubstantially equivalent to predicates
    ReliabilitySubstantially equivalent to predicate devicesSubstantially equivalent to predicates
    BiocompatibilitySubstantially equivalent to predicate devicesSubstantially equivalent to predicates
    SterilizationSterilization process results in a SAL of 10^-6The sterilization process results in a SAL of 10^-6
    Packaging durabilityNo signs of damage and functioned as intended following testingNo signs of damage and functioned as intended following testing
    Shelf-lifeNo signs of damage and functioned as intended following testingNo signs of damage and functioned as intended following testing

    2. Sample Size Used for the Test Set and Data Provenance:

    The provided text does not contain any information regarding sample sizes for a standalone (test set) for clinical performance, as it explicitly states that clinical testing was not required. The testing mentioned is non-clinical, likely laboratory-based or bench testing. Therefore, there's no information on data provenance (country of origin, retrospective/prospective) related to a clinical test set.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Since no clinical testing was required or performed, there is no information on experts establishing ground truth for a clinical test set. The "ground truth" for the non-clinical tests would have been established by engineering specifications, validated test methods, and comparison against the performance of predicate devices.

    4. Adjudication Method for the Test Set:

    Given that no clinical testing requiring human interpretation or judgment was performed, there is no information on an adjudication method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and Effect Size:

    No MRMC comparative effectiveness study was done as clinical testing was not required for this device. This device is a mechanical pump, not an AI/CAD diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This is not applicable to the FloPump 57mL Centrifugal Pump. This device is a physical, mechanical blood pump, not an algorithm or AI system. The "standalone" performance here refers to the device's functional performance on its own, which was evaluated through the non-clinical tests listed.

    7. The Type of Ground Truth Used:

    For the non-clinical tests, the "ground truth" was based on:

    • Engineering specifications and design requirements: For parameters like priming volume, flow rates, and maximum rated pressure.
    • Established test methods and standards: For sterilization (SAL of 10^-6), biocompatibility, and perhaps reliability testing.
    • Performance of predicate devices: The primary method for proving substantial equivalence was demonstrating similar performance to legally marketed predicate devices for flow curves, heat generation, prime volume, air handling, hemolysis, and reliability. This implies that the accepted performance range of the predicate devices served as the "ground truth" for comparison.

    8. The Sample Size for the Training Set:

    Not applicable. This device is a mechanical blood pump, not an AI/ML algorithm that requires a "training set" of data.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable. As stated above, this is a mechanical device, not an AI/ML system requiring a training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1