510(k) Data Aggregation

    K Number
    K211290

    Validate with FDA (Live)

    Date Cleared
    2021-08-03

    (97 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K210754,K170021,K193214,K190324

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OsteoCentric 2.4mm and 3.0mm Cannulated Screw are intended for fixation of fractures and non-unions of small bones and small bone arthrodesis. Examples include, but are not limited to scaphoid and other carpal fractures, metacarpal and phalangeal fusions, osteotomies, and bunionectomies.

    The OsteoCentric 3.5mm Cannulated Screws are intended for fixation of small bones and small bone fragments, such as fractures of the metatarsals, arthrodesis of the carpals and phalanges, steochnondritis dissecans, and ligament fixation.

    The OsteoCentric 4.5mm Cannulated Screws are intended for fracture fixation of long bones and long bone fragments.

    The OsteoCentric 6.5mm to 8.0mm Cannulated Screws are intended for fracture fixation of long bones and long bone fragments, long bone osteotomies, femoral neck fractures, slipped capital femoral epiphyses as an adjunct to treatment with a dynamic hip screw (DHS) in basilar neck fractures, tibial plateau fractures, ankle arthrodesis, pediatric femoral neck fractures, intercondylar femur fractures, SI joint disruptions, fixation of pelvis and iliosacral joints, and subtalar arthrodesis.

    The OsteoCentric 7.0mm and 8.0mm Cannulated Nut is indicated for fracture fixation where fracture compression is desired, and where insertion of the nut-and-bolt system does not increase the risk of morbidity due to the need for surgical access to both sides of the compressed area. The device may also be used when compression between separate bones is desired (e.g. syndesmosis) as long as there is not increased morbidity associated with the insertion of a nutand-bolt system, as discussed above.

    Device Description

    The OsteoCentric Technologies Cannulated Fasteners and Nuts consists of cannulated screws in a variety of lengths and diameters to accommodate different anatomic sizes of patients. The screws and nuts are provided non-sterile. The screws and nuts are manufactured from Stainless Steel per ASTM F138 or from Titanium per ASTM F136 or F1295. The purpose of this submission is to modify the indications for use.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the device:

    Regarding Acceptance Criteria and Supporting Study:

    The document does not contain any explicit acceptance criteria or a study that proves the device meets those criteria.

    Instead, the document is a 510(k) Premarket Notification for OsteoCentric Technologies' Cannulated Fasteners and Nuts. The core of this submission is to demonstrate substantial equivalence to previously cleared predicate devices, rather than proving performance against specific acceptance criteria through new testing.

    The key statement that clarifies this is:

    "No performance testing was required, as no physical changes were made to the devices, the ranges of sizes are comparable to predicate devices with the same indications, and no additional testing is required for the indications modifications proposed."

    Therefore, most of the requested information regarding acceptance criteria, performance data, sample sizes, and expert reviews is not applicable or not present in this specific regulatory submission. The substantial equivalence argument relies on the existing clearance of predicate devices and the lack of significant changes to the subject device that would warrant new performance testing.

    However, based on the information provided, we can infer some details related to the regulatory pathway and the basis for clearance:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Stated or Implied)Reported Device Performance
    Material Composition:
    Stainless Steel per ASTM F138Device is manufactured from Stainless Steel per ASTM F138
    Titanium per ASTM F136 or F1295Device is manufactured from Titanium per ASTM F136 or F1295
    Technological Characteristics:
    Indications for Use identical to predicatesAsserted to be identical to predicates with minor modifications not raising new safety/effectiveness issues.
    Materials of manufacture identical to predicatesAsserted to be identical to predicates.
    Principles of operation identical to predicatesAsserted to be identical to predicates.
    Sizes comparable to predicatesAsserted to be comparable to predicate devices with the same indications.
    Safety and Effectiveness:
    Not raise new issues of safety and effectiveness compared to predicatesConcluded that minor differences do not raise new issues of safety and effectiveness.
    No physical changes made to the devicesNo physical changes were made to the devices.
    No additional testing required for proposed indications modificationsNo additional testing was required for the proposed indications modifications.

    Important Note: The "acceptance criteria" here are inferred from the substantial equivalence argument. The document explicitly states "No performance testing was required," meaning there are no new, specific quantitative performance metrics or studies reported in this submission to meet new acceptance criteria. The criteria for clearance are met by demonstrating equivalence to devices that already met prior acceptance criteria.


    2. Sample size used for the test set and the data provenance

    • Not Applicable. The document explicitly states: "No performance testing was required." Therefore, there was no 'test set' in the context of new performance data for this submission. The substantial equivalence is based on the characteristics of the device being analogous to already cleared predicate devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. As no new performance testing was conducted, there was no test set requiring ground truth established by experts within the scope of this submission. The ground truth for the predicate devices' original clearances would have been established, but that information is not part of this document.

    4. Adjudication method for the test set

    • Not Applicable. No new test set or performance study was conducted.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This device is a medical implant (cannulated fasteners and nuts), not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant to this device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This device is a medical implant, not an algorithm or software. Standalone algorithm performance is not relevant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not Applicable. No new ground truth was established for this submission, as no new performance testing was conducted. The clearance is based on substantial equivalence to predicate devices, which would have had their performance and safety established through various means (including historical usage, engineering principles, and potentially prior testing) during their initial clearance processes.

    8. The sample size for the training set

    • Not Applicable. This device is a physical medical implant, not an AI or machine learning model that requires a "training set."

    9. How the ground truth for the training set was established

    • Not Applicable. As explained above, there is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K210247

    Validate with FDA (Live)

    Date Cleared
    2021-04-30

    (91 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Predicate For
    Why did this record match?
    Reference Devices :

    K170021

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OsteoCentric Integrated Hip Fastener System is intended to treat stable and unstable intertrochanteric, subtrochanteric, and basilar neck fractures in which a stable medial buttress can be reconstructed.

    Device Description

    The OsteoCentric Integrated Hip Fastener System is intended for use in internal fixation of femoral neck fractures. The OsteoCentric Integrated Hip Fastem consist of plates and screws in a variety of lengths and diameters to accommodate different anatomic sizes of patients. The devices are provided non-sterile. All implantable devices are manufactured from Stainless Steel per ASTM F138.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called the "OsteoCentric Integrated Hip Fastener System." This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through clinical trials with strict acceptance criteria and performance metrics.

    Therefore, the document does not contain the kind of information typically requested for AI/ML device studies, such as:

    • A table of acceptance criteria and reported device performance (in terms of sensitivity, specificity, AUC, etc.)
    • Sample size used for the test set and data provenance (e.g., country of origin, retrospective/prospective)
    • Number of experts and their qualifications used to establish ground truth
    • Adjudication method
    • Multi-reader multi-case (MRMC) comparative effectiveness study
    • Standalone (algorithm-only) performance
    • Type of ground truth used (pathology, outcomes data, etc.)
    • Sample size for the training set
    • How ground truth for the training set was established

    Instead, this document describes an engineering analysis for a mechanical device.

    Here's what the document does provide regarding the "study" that proves the device meets the "acceptance criteria" (using these terms loosely in the context of a mechanical device submission):

    1. Acceptance Criteria and Reported Device Performance (Mechanical Context):

    The acceptance criteria here are implicitly demonstrating substantial equivalence to predicate devices in terms of mechanical properties for the intended use.

    Acceptance Criteria (Implicit)Reported Device Performance (Engineering Analysis Conclusion)
    Plate strength and stiffness sufficient for intended use and comparable to predicates.Plate strength and stiffness was compared by evaluating the material properties and moment of inertia at worst-case locations. The results of the engineering analysis confirmed that the strength, stiffness... of the OsteoCentric Integrated Hip Fastener System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.
    Screw strength sufficient for intended use and comparable to predicates.Screw strength was evaluated and compared to the predicates using the material properties, screw types and sizes. The results of the engineering analysis confirmed that the... strength... of the OsteoCentric Integrated Hip Fastener System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.
    Screw fixation sufficient for intended use and comparable to predicates (e.g., thread profile provides adequate fixation).Screw fixation was evaluated and compared to predicates using information about the thread profile. The results of the engineering analysis confirmed that the... fixation of the OsteoCentric Integrated Hip Fastener System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.
    Overall mechanical performance equivalent to legally marketed predicate devices."The overall technology characteristics and mechanical performance data lead to the conclusion that the OsteoCentric Integrated Hip Fastener System is substantially equivalent to the predicate device." "The results of the engineering analysis confirmed that the strength, stiffness and fixation of the OsteoCentric Integrated Hip Fastener System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices."
    Materials of manufacture (Stainless Steel per ASTM F138) are appropriate and similar to predicates.Implied by "All implantable devices are manufactured from Stainless Steel per ASTM F138" and "Materials of Manufacture" being identical between subject and predicates.
    Similar technological characteristics to predicates (e.g., modular stainless steel bone fixation system with similar interlocking features, plate angulation, lengths, widths, thicknesses, screw sizes, screw lengths)."The subject and predicate devices have similar technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are identical between the subject and predicates: Indications for Use, Materials of Manufacture." Also, "the subject plating/screw construct is also a modular stainless steel bone fixation system with similar interlocking features, plate angulation, plate lengths, plate widths, plate thicknesses, screw sizes and screw lengths."

    2. Sample size used for the test set and the data provenance:

    • Test Set: No discrete "test set" in the context of an AI/ML algorithm is mentioned. The assessment was based on an "engineering analysis" comparing physical properties and design features of the subject device to predicate devices. This involves material properties, moment of inertia, screw types/sizes, and thread profiles.
    • Data Provenance: Not applicable in the context of an engineering analysis of mechanical components. The data would come from specifications, material testing standards (e.g., ASTM F138), and design comparisons.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. Ground truth for mechanical properties is established through engineering principles, material science, and testing standards, not expert consensus in the medical imaging sense. The submission was prepared by Nathan Wright MS, an Engineer & Regulatory Specialist at Empirical Testing Corp.

    4. Adjudication method for the test set:

    • Not applicable. There was no clinical or interpretative "test set" requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is a mechanical bone fixation device, not an AI/ML diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a mechanical device.

    7. The type of ground truth used:

    • Engineering specifications and standards: The "ground truth" for this device's performance relies on established engineering principles, material science (e.g., Stainless Steel per ASTM F138 specifications), and comparative analysis against the known mechanical properties of legally marketed predicate devices. The aim is to show that the new device's mechanical integrity is equivalent or superior to existing devices for the same indications.

    8. The sample size for the training set:

    • Not applicable. There is no "training set" in the context of this type of mechanical device submission.

    9. How the ground truth for the training set was established:

    • Not applicable.