Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    DEN230021

    Validate with FDA (Live)

    Device Name
    ensoETM
    Date Cleared
    2023-09-13

    (167 days)

    Product Code
    Regulation Number
    870.5720
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    DEN140018, K170009

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ECD01 model of ensoETM™ is indicated as follows:

    The ensoETM™ is a thermal regulating device, intended to:

      1. connect to a Gaymar Medi-Therm III Conductive Hyper/Hypothermia System or Stryker Altrix Precision Temperature Management System to control patient temperature.
      1. connect to a Gaymar Medi-Therm III Conductive Hyper/Hypothermia System or Stryker Altrix Precision Temperature Management System to reduce the likelihood of ablation-related esophageal injury resulting from radiofrequency cardiac ablation procedures, and
      1. provide gastric decompression and suctioning.
        The ECD02 model of ensoETM™ is indicated as follows:

    The ensoETM™ is a thermal regulating device, intended to:

      1. connect to a Gentherm/Cincinnati Sub-Zero Blanketrol II or Blanketrol III Hyper-Hypothermia System to control patient temperature.
      1. connect to a Gentherm/Cincinnati Sub-Zero Blanketrol II or Blanketrol III Hyper-Hypothermia Systemto reduce the likelihood of ablation-related esophageal injury resulting from radiofrequency cardiac ablation procedures, and
      1. provide gastric decompression and suctioning.
    Device Description

    The ensoETM is a multi-lumen silicone tube that is placed in the esophagus in a manner similar to a standard orogastric tube. The primary function of the ensoETM™ is to connect to an external heat exchanger to either (1) cool or warm a patient or (2) cool the esophagus to reduce the likelihood of thermal injurv after radiofrequency (RF) cardiac ablation procedures.

    The external heat exchanger circulates warmed or chilled water through the two outer lumens of the ensoETMTM. One outer lumen serves as the water inflow, while the other serves as the water outflow. The distal end of the ensoETM™ contains an opening between the two outer lumens, allowing the water to circulate back to the external heat exchanger without contacting the patient.

    The central lumen of the ensoETM™ is open to the stomach to allow for gastric decompression and suction.

    The ensoETM™ is primarily composed of standard medical-grade silicone. The ensoETMIM is a single-use, disposable, non-implantable device with an intended duration of use of 72 hours or less.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the studies performed for the ensoETM™ device, based on the provided text:

    Acceptance Criteria and Device Performance for ensoETM™

    Note: The provided text details performance criteria for various non-clinical tests and outcomes of clinical studies. It doesn't explicitly state "acceptance criteria" in a singular, consolidated table for the overall device's effectiveness. Instead, the "acceptance criteria" are embedded within the description of each test or study objective. For the purpose of this response, I will synthesize these into a comprehensive table, drawing from the study descriptions.

    The primary clinical acceptance criterion for the device's new indication (reducing ablation-related esophageal injury) is implied by the clinical trial results showing a reduction in esophageal lesions, although statistical significance was not always met in individual RCTs and the pooled analysis showed a non-significant reduction.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test/Study TypeAcceptance Criteria (from text)Reported Device Performance (from text)
    Biocompatibility/MaterialsNon-cytotoxic (ISO 10993-5)Test articles found non-cytotoxic.
    Non-sensitizing (ISO 10993-10)Test articles found non-sensitizing.
    Non-irritating (ISO 10993-10)Test articles found non-irritating.
    Shelf Life/SterilityNon-sterile device acceptable based on RWD evaluation and complaint data.Device provided non-sterile (acceptable).
    Device and packaging functionality over 3 years (accelerated aging).Performance testing on devices which had completed 3-year accelerated aging demonstrated that the device and packaging materials did not experience significant degradation.
    MR Compatibility (Model ECD02)Magnetically induced displacement angle and torque do not pose significant risk in 1.5 T or 3.0 T MR environments.Results indicate effects do not pose significant risk.
    Maximum temperature rise ≤ 2.61 ℃ in 1.5 T or 3.0 T MR system at 2.0 W/kg SAR for 15 minutes.Maximum temperature rise was 2.61 ℃.
    Bench Performance - Tensile TestBond strength of ≥15 N.All tests met the specified acceptance criteria.
    Bench Performance - Burst Strength Test>15 psi.All tests met the specified acceptance criteria.
    Bench Performance - Ultimate Material Strength Test≥ 5 Mpa.All tests met the specified acceptance criteria.
    Bench Performance - Leakage TestNo visible evidence of leaks.All tests met the specified acceptance criteria.
    Bench Performance - Vacuum Resistance TestDevice does not collapse under 200 mmHg of negative pressure.All tests met the specified acceptance criteria.
    Bench Performance - End Cap Axial Deflection TestRequired force to bend is less than that needed for NeoMed enteral feeding connector.All tests met the specified acceptance criteria.
    Bench Performance - Dimensional TestingAccurate dimensions for Extruded Tube, Extension Tube, Proximal End Cap, Distal End Cap.All tests met the specified acceptance criteria.
    Bench Performance - Compatibility Testing (Flow Rate)> 31 L/hr.All tests met the specified acceptance criteria.
    Bench Performance - Compatibility Testing (Temperature Accuracy)Within ± 1.0 ℃.All tests met the specified acceptance criteria.
    Bench Performance - Compatibility Testing (Alarm Trigger)Alarms at ≤ 29 ℃ and ≥ 45 ℃ and when flow occluded.All tests met the specified acceptance criteria.
    Bench Performance - Compatibility Testing (System Pressures under Occlusive situation)< 9.0 psi.All tests met the specified acceptance criteria.
    Animal Study - Primary Outcome (Esophageal Integrity)No impact on esophageal mucosa tissue or adjacent organs after experimental temperature modification.Results demonstrated no impact on esophageal mucosa tissue or adjacent organs; no direct effect on epithelial lining or deeper layers.
    Animal Study - Secondary Outcome (Temperature Control)Success in: - Reducing swine body temperature by 4°C from baseline. - Maintaining reduced body temperature for 24 hours. - Rewarming to baseline temperature at up to 0.5°C per hour. (Success defined as attaining and maintaining goal temps: 39℃ at baseline, 35℃ at reduced.)Met success criteria for reducing, maintaining, and rewarming body temperature in swine.
    Clinical Trials - Esophageal Injury Reduction(Implied: Reduction in incidence/severity of esophageal injury compared to control)Pooled data: Odds ratio = 0.55, 95% CI = (0.22, 1.35), p = 0.22 (non-significant difference). Individual RCTs showed mixed results (e.g., St. George's showed significantly less EDELs in device group, Penn showed higher EDELs in device group but less severe injury).
    Clinical Trials - Safety (Adverse Events)Acceptable adverse event profile with no unexpected safety findings compared to control.No unexpected safety findings; comparable serious and other adverse event rates between device and control arms.
    Cross-sectional Study - Primary Endpoint (AEF Rates)(Implied: Reduction in AEF rates after adoption of ensoETM™ cooling.)Data presented but FDA found significant limitations; not used as primary evidence for safety and effectiveness.
    Cross-sectional Study - Secondary Endpoint (AF Recurrence Rates)(Implied: Reduction in AF recurrence rates after adoption of ensoETM™ cooling.)At 12 months, 72.2% of cooled patients free of arrhythmia vs. 58.2% of pre-cooling patients (absolute difference of 14%, p=0.03).

    2. Sample Sizes and Data Provenance

    Clinical Trial Data:

    • Test set size:
      • Riverside: 6 patients (3 control, 3 ensoETM)
      • University of Pennsylvania (Penn): 44 patients completed (22 device, 22 control)
      • St. George's University Hospitals (St. George's): 120 patients underwent endoscopy (60 control, 60 ensoETM) from a total of 188 patients
      • Pooled Analysis: 166 patients (83 control, 83 ensoETM)
    • Data provenance: Prospective, randomized controlled trials (RCTs). The specific countries are not explicitly named, but university hospitals suggest the US and potentially UK (St. George's University Hospitals).

    Cross-sectional Study (Primary Analysis):

    • Test set size: 25,186 patients (10,962 before cooling, 14,224 after cooling).
    • Data provenance: Retrospective cohorts, from 25 hospital systems (30 separate hospitals). Countries are not specified.

    Cross-sectional Study (Secondary Analysis):

    • Test set size: 513 patients (253 before cooling, 260 after cooling).
    • Data provenance: Retrospective review, a subset of the primary analysis, from two healthcare systems (three hospitals). Countries are not specified.

    3. Number of Experts and Qualifications for Ground Truth

    Clinical Trials (Esophageal Injury Ground Truth):

    • Number of experts: Not explicitly stated, but endoscopy was performed by clinicians at each site. The "scoring methodology" (Table 1) suggests a classification system used by the endoscopists.
    • Qualifications of experts: Implied to be medical professionals qualified to perform and interpret endoscopy findings, likely gastroenterologists or cardiologists with experience in post-ablation endoscopy. Specific years of experience are not mentioned.

    Cross-sectional Study (AEF & AF Recurrence Ground Truth):

    • Number of experts: Data were "extracted by site investigators" for the primary AEF endpoint, and "any of four electrophysiologists" for the secondary AF recurrence endpoint.
    • Qualifications of experts: Site investigators and electrophysiologists, specialized in cardiac procedures and data record review.

    4. Adjudication Method for the Test Set

    The text does not explicitly describe a formal adjudication method (e.g., 2+1, 3+1) for the clinical trial or cross-sectional study ground truth establishment.

    • For the clinical trials, it describes "scoring methodology for each of the sites and the common score which was used to combine the results," implying a standardized grading applied by the clinicians performing the endoscopy. It does not mention independent central review or consensus adjudication for the lesions.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted to evaluate how much human readers improve with AI vs. without AI assistance. The ensoETM™ is a physical medical device (temperature regulation device), not an AI-powered diagnostic or assistive tool for human readers/clinicians. Therefore, this type of study is not applicable.

    6. Standalone (Algorithm Only) Performance

    No, a standalone (algorithm only without human-in-the-loop performance) study was not conducted. The ensoETM™ is a physical device used during a medical procedure, not an algorithm.

    7. Type of Ground Truth Used

    • Clinical Trials (Esophageal Injury): Ground truth was established by expert assessment via endoscopy, using a defined injury grading scheme (e.g., Zargar grading, as seen in the Riverside study, and the common scoring scheme provided in Table 1).
    • Cross-sectional Study (AEF & AF Recurrence): Ground truth was based on retrospective medical records review for reported AEFs and AF recurrence rates following PVI.

    8. Sample Size for the Training Set

    The document does not describe a "training set" in the context of an algorithm or AI development. The studies mentioned (bench, animal, clinical trials, cross-sectional) are for validation and evaluation of the physical device's safety and effectiveness.

    9. How Ground Truth for the Training Set Was Established

    As there was no "training set" in the AI/algorithm sense, this question is not applicable. The device's design and initial performance were likely informed by general medical knowledge, engineering principles, and previous versions of the device, but not through an explicit "training set" for ground truth establishment in this context.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1