Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K192036

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-08-29

    (30 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K163549, K173470, K182233

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ClotTriever Thrombectomy System consists of the ClotTriever Sheath. The ClotTriever Thrombectomy System is indicated for:

    • · The non-surgical removal of soft thrombi and emboli from blood vessels.
    • · Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.
      The ClotTriever Thrombectomy System is intended for use in the peripheral vasculature.
    Device Description

    The ClotTriever Thrombectomy System is a single-use, sterile medical device designed for use in the peripheral vasculature. The ClotTriever Thrombectomy System is comprised of the ClotTriever Sheath and the ClotTriever Catheter. The ClotTriever Sheath consists of a reinforced polymeric sheath equipped with a self-expanding distal mesh funnel, a flush/aspiration port, and a proximal hemostatic valve. A dilator is provided to aid insertion. Other provided accessories include the funnel loading tool and a Large Bore 60 cc syringe. The ClotTriever Catheter consists of reinforced polymeric coaxial shafts terminating in an expandable coring element and thrombus collection bag. Two ports terminating in stopcocks are provided for de-airing the catheter shafts. To aid in fluoroscopic visualization, the Sheath dilator and ClotTriever Catheter distal tips are radiopaque.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the ClotTriever Thrombectomy System, focusing on device modifications and non-clinical testing. It explicitly states that "Clinical testing was not required for the determination of substantial equivalence." This means that the information requested regarding acceptance criteria, study details, human reader performance, ground truth establishment for a training set, and sample sizes for training and test sets in a clinical context is not applicable to this submission, as no human clinical study was performed.

    The document primarily details non-clinical testing to demonstrate substantial equivalence to a predicate device.

    Here's a breakdown of the available information:

    1. A table of acceptance criteria and the reported device performance:

    The document broadly states: "Test results demonstrated that all acceptance criteria were met; therefore, the device conforms to established product specifications and intended use."

    While a specific table of acceptance criteria with numerical performance values is not provided, the following list represents the types of non-clinical tests conducted to meet internal specifications:

    Acceptance Criteria (Test Type)Reported Device Performance (Implied)
    Pouch Seal InspectionMet
    Dye Penetration InspectionMet
    Visual & Dimensional Inspections ClotTriever SheathMet
    Visual & Dimensional Inspections ClotTriever DilatorMet
    Visual & Dimensional Inspections Loading ToolMet
    Guidewire Compatibility Verification, ClotTriever DilatorMet
    Deployment Force ClotTriever Sheath from DilatorMet
    Retraction Force Dilator Thru ClotTriever SheathMet
    Insertion Verification ClotTriever Delivery Catheter thru SheathMet
    Retraction of the ClotTriever Catheter Thru ClotTriever SheathMet
    Kink Resistance/Radius Verification ClotTriever Sheath & DilatorMet
    Leakage Verification, ClotTriever SheathMet
    Leakage Verification, ClotTriever DilatorMet
    Fluid Test Thru ClotTriever Sheath/DilatorMet
    ClotTriever Dilator Retraction in Clot AnalogMet
    Simulated Use, Track and Rotation ClotTriever SheathMet
    Simulated Use, Track and Tensile ClotTriever DilatorMet
    Simulated Use, Track and Tensile ClotTriever SheathMet
    Insertion and Placement ResistanceMet
    Corrosion ResistanceMet

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in a clinical context as no human clinical study was performed. The non-clinical tests would have used physical samples of the device components. The document does not specify the number of samples for each non-clinical test or their provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as no human clinical study was conducted, and therefore no ground truth was established by medical experts for a test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no human clinical study was conducted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The device is a medical instrument (thrombectomy system), not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. The device is a medical instrument, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For the non-clinical testing, the "ground truth" would be the established engineering and performance specifications for each test (e.g., a specific force for deployment, a tolerance for dimensions, absence of leakage). These are defined internally by the manufacturer during device development.

    8. The sample size for the training set

    Not applicable, as no algorithm or human reader training set was required for this device's regulatory submission based on non-clinical testing.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182531

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-10-10

    (26 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K173470, K163549

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ClotTriever Thrombectomy System consists of the ClotTriever Catheter and ClotTriever Sheath. The ClotTriever Thrombectomy System is indicated for:
    The non-surgical removal of soft thrombi and emboli from blood vessels.Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.
    The ClotTriever Thrombectomy System is intended for use in the peripheral vasculature.

    Device Description

    The ClotTriever Thrombectomy System is a single-use, sterile medical device designed for use in the peripheral vasculature. The ClotTriever Thrombectomy System is comprised of the ClotTriever Sheath and the ClotTriever Catheter. The ClotTriever Sheath consists of a polymeric sheath equipped with a self-expanding distal mesh funnel, a flush/aspiration port, and a proximal hemostatic valve. A dilator is provided to aid insertion. Other provided accessories include the funnel loading tool and a 60 cc syringe. The ClotTriever Catheter consists of polymeric coaxial shafts terminating in an expandable coring element and thrombus collection bag. At the proximal end of the catheter is a handle used to enable tensioning of the coring element. Two ports terminating in stopcocks are provided for de-airing the catheter shafts. To aid in fluoroscopic visualization, the dilator and ClotTriever Catheter distal tips are radiopaque.

    AI/ML Overview

    The provided document is a 510(k) summary for the ClotTriever Thrombectomy System, which is a medical device for removing thrombi and emboli. It describes non-clinical testing performed to demonstrate substantial equivalence to a predicate device, but it explicitly states that clinical testing was not required. Therefore, the document does not contain information about a study proving the device meets acceptance criteria related to clinical performance or human reader effectiveness.

    However, based on the non-clinical testing section, I can provide information about the acceptance criteria and performance as described for those tests.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Visual & Dimensional InspectionsNot explicitly stated, implied to be conformity to design specifications.All acceptance criteria were met.
    Retraction Force of HandleNot explicitly stated, implied to be conformity to design specifications.All acceptance criteria were met.
    Syringe Leak TestingNot explicitly stated, implied to be conformity to design specifications.All acceptance criteria were met.
    Syringe Vacuum TestingNot explicitly stated, implied to be conformity to design specifications.All acceptance criteria were met.
    Syringe Tensile TestingNot explicitly stated, implied to be conformity to design specifications.All acceptance criteria were met.
    Clot burden Removal Performance TestNot explicitly stated, implied to be conformity to design specifications.All acceptance criteria were met.
    Simulated Use, Tensile TestNot explicitly stated, implied to be conformity to design specifications.All acceptance criteria were met.

    2. Sample size used for the test set and the data provenance:
    For the non-clinical tests listed, the sample sizes are not explicitly provided in this 510(k) summary. The data provenance is generally "non-clinical testing" conducted by the manufacturer, Inari Medical, Inc. The document does not specify country of origin for the data (as it's laboratory testing, this is typically internal to the manufacturer). It is, by definition, a prospective testing of the device prototypes/units.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    Not applicable. These are non-clinical engineering and performance tests, not clinical evaluations requiring expert interpretation of medical images or patient outcomes.

    4. Adjudication method for the test set:
    Not applicable for these non-clinical tests.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    No. This device is a thrombectomy system (a mechanical device), not an AI-powered diagnostic tool. Clinical testing was also not performed or required for this particular submission.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
    No. This is a physical medical device, not a standalone algorithm.

    7. The type of ground truth used:
    For the non-clinical tests, the "ground truth" would be established engineering specifications, performance standards, and physical measurements/observations during the tests. For example, for "Syringe Leak Testing," the ground truth would be the absence of leaks under specified conditions.

    8. The sample size for the training set:
    Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established:
    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173672

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-02-05

    (67 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K163549

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Flow Triever Retrieval/Aspiration System consists of the FlowTriever Catheter, Aspiration Guide Catheter, and Retraction Aspirator. The FlowTriever Retrieval/Aspiration System is indicated for:
    -The non-surgical removal of emboli and thrombi from blood vessels.
    -Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.
    The FlowTriever Retrieval/Aspiration System is intended for use in the peripheral vasculature.

    Device Description

    The FlowTriever Retrieval/Aspiration System is a single-use over-the-wire catheter-based system for the minimally invasive treatment of thromboemboli in the peripheral vasculature. The system is comprised of three main components packaged separately:
    ● Aspiration Guide Catheter
    ● FlowTriever Catheter (available in 3 sizes: 6-10 mm, 11-14 mm, and 15-18 mm)
    ● Retraction Aspirator
    The FlowTriever Catheter is inserted through the Aspiration Guide Catheter and advanced to the thrombus. Self-expanding wireform disks are deployed to engage thrombus by retracting the outer delivery catheter. The hand-lever operated Retraction Aspirator simultaneously aspirates fluids and retracts the FlowTriever Catheter with thrombus into the Aspiration Guide Catheter to capture clot and restore blood flow.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Inari Medical FlowTriever Retrieval/Aspiration System. It focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing. Here's a breakdown of the requested information based on the text:

    1. Table of Acceptance Criteria and the Reported Device Performance

    The document states that "Test results demonstrated that all acceptance criteria were met," and "This testing FlowTriever demonstrated compliance with relevant product specifications." However, it does not explicitly list the specific quantitative acceptance criteria for each test or the exact performance values achieved. It only lists the types of tests performed.

    Test CategorySpecific TestAcceptance Criteria (Not explicitly stated in text)Reported Device Performance (Not explicitly stated in text)
    Material/Component IntegrityVisual and Dimensional InspectionsImplied: Within specified tolerancesImplied: Met specified tolerances
    Functional Performance (Catheter)Guidewire Compatibility VerificationImplied: Compatible with designated guidewiresImplied: Demonstrated compatibility
    Leakage Verification, AGC Hemostasis Valve And Accessory DevicesImplied: No leakage beyond acceptable limitsImplied: No leakage detected
    Air Leakage From AGC Hemostasis Valve During Syringe PullbackImplied: No air leakage beyond acceptable limitsImplied: No air leakage detected
    Air Leakage During Syringe PullbackImplied: No air leakage beyond acceptable limitsImplied: No air leakage detected
    Vacuum TestingImplied: Achieved and maintained specified vacuum levelsImplied: Met specified vacuum requirements
    Retraction Force TestingImplied: Within specified force rangeImplied: Within specified force range
    Push Button ForceImplied: Within specified force rangeImplied: Within specified force range
    Retraction Aspirator Device RetractionImplied: Consistent and effective retractionImplied: Demonstrated consistent and effective retraction
    Simulated Use, Track and TensileImplied: Met performance requirements for simulated use, tracking, and tensile strengthImplied: Met performance requirements
    Stopcock Torque TestingImplied: Within specified torque rangeImplied: Within specified torque range
    Burst Testing of Hemostasis Valve and StopcockImplied: Withstood specified burst pressuresImplied: Withstood specified burst pressures

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify the exact sample sizes used for each non-clinical test. The data provenance is non-clinical testing, meaning it involves engineering and bench testing, not patient data. Therefore, concepts like country of origin for data or retrospective/prospective studies do not apply in this context.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not applicable as the study described is non-clinical testing of a medical device, not a diagnostic or AI-driven system requiring expert ground truth for interpretation of outcomes. The "ground truth" here is adherence to engineering specifications and performance standards.

    4. Adjudication Method for the Test Set

    This information is not applicable for the reasons stated in point 3. The testing involves objective measurements against predefined specifications, not subjective human assessment requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. The device is a physical medical device (embolectomy catheter), not an AI-driven diagnostic or assistive system that would involve human readers. Clinical testing was explicitly stated as "not required for the determination of substantial equivalence."

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical testing was based on established product specifications, engineering standards, and the intended use of the device. The tests were designed to verify that the modified device met these predefined criteria.

    8. The Sample Size for the Training Set

    This information is not applicable. There is no "training set" in the context of this non-clinical engineering and performance testing for a physical medical device. This is not an AI/machine learning study.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1