510(k) Data Aggregation

    K Number
    K213411

    Validate with FDA (Live)

    Date Cleared
    2021-12-22

    (64 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K161109, K163502, K162474

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    iSR'obot Mona Lisa 2.0 is a user-controlled, stereotaxic accessory intended to guide physicians in the planning and positioning of insertion tools, such as a third-party needle or a probe, during image-guided diagnostic and interventional procedures in conjunction with the guidance of transrectal ultrasound involving the prostate gland in a clinical setting. Examples of such procedures include, but are not limited to, image fusion for diagnostic clinical examinations and procedures, soft tissue biopsies, and soft tissue ablations.

    The iSR'obot Mona Lisa 2.0 provides 2D and 3D visualization of Ultrasound images and the ability to fuse and register these images with those from other imaging modalities such as Magnetic Resonance, etc. It also provides the ability to display a simulated image of an insertion tool on a computer monitor screen, the target organ, and the current and projected future path of the insertion tool taking into account patient movement. Other software features include multiplanar reconstruction, segmentation, image measurements, 2D/3D image registration, and reporting.

    Device Description

    The iSR'obot Mona Lisa 2.0 is user-controlled, stereotaxic accessory intended to guide physicians in the planning and positioning of insertion tools, such as a needle or a probe, during image-guided diagnostic and interventional procedures involving the prostate gland in a clinical setting. The device displays the 2D live image feeds from commercially available ultrasound systems and constructs 3D ultrasound image stacks. The system allows the importation of an MRI image to create a model of patient prostate by providing fusion between Ultrasound and Magnetic Resonance Imaging (MRI). The system is compatible with commercially available ultrasound systems, transrectal ultrasound bi-plane probes, and commercially available needle devices. Other software features include multi-planar reconstruction, segmentation, image measurements and 2D/3D image registration.

    AI/ML Overview

    The provided text for iSR'obot Mona Lisa 2.0 (K213411) focuses on demonstrating substantial equivalence to a predicate device and details non-clinical testing. It explicitly states, "There is no pre-market clinical investigation submitted for iSR'obot Mona Lisa 2.0." Therefore, there is no information available in this document regarding acceptance criteria for clinical performance, the results of a clinical study proving the device meets these criteria, or details related to human readers, sample sizes for test/training sets, or ground truth establishment in a clinical context.

    However, the document does describe non-clinical system performance testing using phantoms. I can extract information related to those non-clinical tests.

    Non-Clinical System Performance Testing (using phantom)

    1. Acceptance Criteria and Reported Device Performance (Non-Clinical)

    Acceptance Criteria (Implied from Testing Purpose)Reported Device Performance (Non-Clinical)
    Ability to accurately position the needle after system calibration.Testing conducted to demonstrate that after the system has been calibrated, the system is able to accurately position the needle. (Pass / Meets criteria implied)
    Ability to adjust generated prostate models and planned needle position to compensate for patient movement during the procedure.Testing conducted to demonstrate the adjustment of the generated prostate models and planned needle position to compensate for the patient movement during the procedure. (Pass / Meets criteria implied)
    Ability to facilitate multi-needle positioning.Testing conducted to demonstrate multi-needle positioning. (Pass / Meets criteria implied)

    Note: The document only states that these tests were "conducted to demonstrate" the listed functionalities. It does not provide specific numerical results or quantitative acceptance thresholds for these non-clinical tests in the provided summary. A "Pass / Meets criteria implied" is inferred based on the submission being cleared and the statement that "successful verification and non-clinical testing" was performed.

    2. Sample Sizes Used for the Test Set and Data Provenance:

    • Sample Size (Test Set): Not specified in the provided text. The tests were conducted using a "phantom."
    • Data Provenance: The tests were non-clinical, using a "phantom." Therefore, there is no patient data provenance (country, retrospective/prospective).

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    • Number of Experts: Not applicable for non-clinical phantom testing.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set:

    • Adjudication Method: Not applicable for non-clinical phantom testing.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • Was an MRMC study done? No. The document explicitly states, "There is no pre-market clinical investigation submitted for iSR'obot Mona Lisa 2.0."
    • Effect size of human readers improving with AI vs. without AI assistance: Not applicable, as no MRMC study was conducted.

    6. Standalone (Algorithm only without human-in-the-loop performance) Study:

    • Was a standalone study done? Not explicitly mentioned as a separate formal study, but the "Non-clinical System Performance Testing" on a phantom would represent an algorithm/system-only performance test in a controlled environment, demonstrating the device's capabilities without direct human clinical intervention during the specified tasks (e.g., automated positioning accuracy). However, the device is marketed as "user-controlled, stereotaxic accessory intended to guide physicians," implying a human-in-the-loop for its clinical use.

    7. Type of Ground Truth Used:

    • Ground Truth Type: For the non-clinical phantom testing, the ground truth would typically be the precisely known or measured locations/trajectories within the phantom, established by engineering specifications, physical measurements, or other metrology techniques. The document refers to the system's "ability to accurately position the needle" and "adjustment... to compensate for the patient movement," which suggests a known target or reference from which accuracy/adjustment can be measured against.

    8. Sample Size for the Training Set:

    • Sample Size (Training Set): Not applicable or not specified in the provided text, as this document focuses on regulatory submission for a marketed device based on non-clinical testing and substantial equivalence, not detailing the development or training of specific AI components if they exist within the system beyond the mechanical guidance.

    9. How the Ground Truth for the Training Set was Established:

    • Ground Truth Establishment (Training Set): Not applicable or not specified.
    Ask a Question

    Ask a specific question about this device

    K Number
    K203659

    Validate with FDA (Live)

    Date Cleared
    2021-07-08

    (205 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K161109, K163502

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    iSR'obot Mona Lisa 1.0 is intended for use by a trained urologist or physician to perform the computer-assisted transperineal prostate biopsy under transrectal ultrasound guidance. The device serves as a biopsy needle guide only. It shall be used in conjunction with a third- party ultrasound machine and endorectal probe that supports B-Mode, and a thirdparty prostate biopsy gun and needle. The insertion of biopsy needle will be done by urologist. The patient is administered general anesthesia and placed in a lithotomy position.

    Device Description

    The iSR'obot Mona Lisa 1.0 is able to display the 2D live image feeds from commercially available ultrasound systems and also construct 3D ultrasound image stacks. The system allows the importation of an MRI image to create a model of patient prostate by providing fusion between Ultrasound and Magnetic Resonance Imaging (MRI). The system is compatible with commercially available ultrasound systems, transrectal ultrasound bi-plane probes, and commercially available needle devices. Other software features include multi-planar reconstruction, segmentation, image measurements, and 2D/3D image registration. The device is a user-controlled, stereotaxic accessory intended to guide the urologist or physician in the planning and positioning of a needle during ultrasound-guided transperineal prostate procedures such as biopsies in the operating room.

    The iSR'obot Mona Lisa 1.0 comprises a workstation, robotic navigation module (comprising of robotic arm and bed rail stabilizer), and disposables.

    AI/ML Overview

    The Biobot Surgical Pte Ltd iSR'obot Mona Lisa 1.0 device did not