510(k) Data Aggregation

    K Number
    K202631

    Validate with FDA (Live)

    Date Cleared
    2020-12-08

    (88 days)

    Product Code
    Regulation Number
    890.5500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K162782

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iHelmet Laser Comb is indicated to promote hair growth in males with androgenic alopecia who have Norwood-Hamilton classifications of Ha-V or females with androgenic alopecia who have Ludwig-Savin Classifications of I-4, II-1, II-2, or frontal and both with Fitzpatrick Skin Phototypes I-IV.

    Device Description

    The iHelmet Laser Comb is a comb-shaped low level laser therapy (LLLT) device that emits laser light with the intention to promote hair growth. The device provides distributed laser to the scalp at 650+/-10nm while the comb teeth simultaneously part the user's hair to ensure the laser light reaches the user's scalp. The Laser Hair Therapy is designed as handheld product, and it consists of main unit, the charging dock and a power cable, as well as it is powered by the built-in rechargeable lithium battery. The device is one-button operated, and has an audible timer that automatically turns the lasers off after the treatment is completed.

    AI/ML Overview

    This document is a 510(k) Pre-Market Notification, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a novel clinical study with acceptance criteria for a new AI device. Therefore, much of the requested information regarding AI device acceptance criteria and study design is not available in this document.

    However, I can extract the relevant information from the provided text about the device and its claimed equivalence.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't provide specific acceptance criteria in terms of performance metrics (like sensitivity, specificity, accuracy) for a new AI algorithm. Instead, it demonstrates an acceptance of "substantial equivalence" based on various technical and safety benchmarks compared to a predicate device.

    Acceptance Criteria (Demonstration of Substantial Equivalence)Reported Device Performance (iHelmet Laser Comb)
    Intended Use (promote hair growth in specific populations with androgenic alopecia)Matches predicate: "The iHelmet Laser Comb is indicated to promote hair growth in males with androgenic alopecia who have Norwood-Hamilton classifications of IIa-V or females with androgenic alopecia who have Ludwig-Savin Classifications of I-4, II-1, II-2, or frontal and both with Fitzpatrick Skin Phototypes I-IV."
    Product Code (OAP)Matches predicate: OAP
    Classification (Class II)Matches predicate: Class II
    Location for use (OTC)Matches predicate: OTC
    Type of Laser (Visible red light-emitting diodes)Matches predicate: Visible red light-emitting diodes
    Power (<5mW)Matches predicate: <5mW
    Applicable people (Norwood-Hamilton IIa-V (males); Ludwig-Savin I-4, II-1, II-2, or frontal (females))Matches predicate
    Applicable skin (Fitzpatrick Skin Phototypes I-IV)Matches predicate
    Appearance design (Comb)Matches predicate: Comb
    Safety Compliance (IEC 60601-1-2; IEC 60601-1; IEC 60601-1-11; IEC 60825-1)Complied with: IEC 60601-1-2; IEC 60601-1; IEC 60601-1-11; IEC 60825-1
    Biocompatibility Compliance (ISO 10993-5; ISO 10993-10)Complied with: ISO 10993-5; ISO 10993-10
    Light Class (Difference explained not to raise safety/effectiveness issue)Subject Device: Class 1; Predicate Device: Class 3R. Note 1 states "they all meet the IEC 60825-1 requirements. So the differences will not raise any safety or effectiveness issue."
    Amount of Light-emitting diodes (Slight differences explained not to raise safety/effectiveness issue)Subject Device: 12-16 pcs (model dependent); Predicate Device: 12 pcs. Note 2 states "they are very similar to the predicate device and comply with the regulation requirements, so the differences will not raise any safety or effectiveness any issue."
    Wavelength (Slight differences explained not to raise safety/effectiveness issue)Subject Device: 650nm±10nm, 624 nm (model dependent); Predicate Device: 655±5nm. Note 2 states "they are very similar to the predicate device and comply with the regulation requirements, so the differences will not raise any safety or effectiveness any issue."
    Treatment time (Slight differences explained not to raise safety/effectiveness issue)Subject Device: 6-11 minutes per treatment (model dependent); Predicate Device: 8 minutes per treatment. Note 2 states "they are very similar to the predicate device and comply with the regulation requirements, so the differences will not raise any safety or effectiveness any issue."

    2. Sample size used for the test set and the data provenance:

    • This document describes safety and performance testing against electrical and biocompatibility standards, not a clinical trial with a "test set" in the context of an AI algorithm.
    • The tests mentioned in section 7 ("Test Summary") are bench tests against specific IEC and ISO standards (e.g., IEC 60601-1 for electrical safety, IEC 60825-1 for laser safety, ISO 10993-5/10 for biocompatibility). These standards define test methods and acceptance criteria for the device's physical and electrical characteristics, not for clinical performance or AI algorithm validation.
    • Therefore, there is no information about a "sample size" of patients or data, nor data provenance (country of origin, retrospective/prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This document does not describe the establishment of ground truth by clinical experts for a test set of an AI algorithm. The predicate device and the subject device are low-level laser therapy devices, not AI-powered diagnostic or treatment algorithms.

    4. Adjudication method for the test set:

    • Not applicable. This document does not describe a test set or an adjudication method for an AI algorithm.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This document does not describe an AI device, an MRMC study, or human reader performance.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This document is for a physical medical device (laser comb) and not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable. For the purpose of establishing substantial equivalence, the "ground truth" refers to the established safety and performance of the predicate device and compliance with recognized safety standards for the subject device. There is no AI algorithm requiring a specific type of clinical ground truth.

    8. The sample size for the training set:

    • Not applicable. This document does not describe an AI device or a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. This document does not describe an AI device, a training set, or the establishment of ground truth for it.

    In summary: The provided document is a 510(k) summary for a low-level laser therapy device (iHelmet Laser Comb), demonstrating its substantial equivalence to a predicate device. It addresses technical safety and performance aspects, not the validation of an AI algorithm with clinical performance metrics. Therefore, most of the questions related to AI-specific study design (sample size, data provenance, expert ground truth, MRMC studies, standalone performance, training sets) are not applicable to the content of this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180617

    Validate with FDA (Live)

    Device Name
    Hair Up
    Manufacturer
    Date Cleared
    2018-06-01

    (85 days)

    Product Code
    Regulation Number
    890.5500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K141567, K162782

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hair Up is intended to promote hair growth in males with androgenetic alopecia who have Norwood Hamilton Classifications of IIa to V and Fitzpatrick Classification of Skin Phototypes I to IV, and in females with androgenetic alopecia who have Ludwig-Savin Classifications of I-II and Fitzpatrick Classification of Skin Phototypes I to IV.

    Device Description

    The Hair Up is basically an identical device with the 510(k)-cleared device Hair Up (K172968). A new 510(k) clearance is applied this time with a new indications for use (IFU). Subject device, Hair up and its primary predicate device consist of 21 red visible laser diodes (LDs) and 30 red light emitting diodes (LEDs) configured within an outer helmet and protective inner liner. The use of LDs and LEDs provides full coverage of the upper 1/3 of the head which is commonly covered with stylized hair. The system will automatically pause therapy until the correct head position is re-established. At the end of the therapy cycle, the subject device Hair Up signals that therapy is complete and the device is ready to be powered down.

    AI/ML Overview

    The provided text is a 510(k) summary for the "Hair Up" device, an infrared lamp intended to promote hair growth. It does not contain information about acceptance criteria or a study proving the device meets specific performance metrics in terms of effectiveness in promoting hair growth.

    The document primarily focuses on demonstrating substantial equivalence to legally marketed predicate devices through a comparison of technical characteristics and indications for use. The "Performance Data" section describes non-clinical tests related to safety, electrical requirements, and biological evaluation, not a clinical study on hair growth.

    Therefore, I cannot provide the requested table and detailed study information based on the given input. The categories below are answered based on the information that is available in the document, and where information is not present, it is explicitly stated.


    Acceptance Criteria and Device Performance Study Information for Hair Up

    This 510(k) summary for the "Hair Up" device does not contain a clinical study with acceptance criteria and reported device performance related to its primary intended use of promoting hair growth. The document focuses on demonstrating substantial equivalence to predicate devices through technical comparisons and non-clinical safety testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance CriteriaReported Device Performance
    Hair GrowthNot provided in this documentNot provided in this document

    2. Sample size used for the test set and the data provenance

    • Sample Size (Test Set): Not applicable. No clinical test set is described for evaluating hair growth performance.
    • Data Provenance: Not applicable.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. No clinical test set is described with expert-established ground truth for hair growth.

    4. Adjudication method for the test set

    • Not applicable. No clinical test set is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done as the device is an infrared lamp for hair growth, not an AI-based diagnostic or imaging device used by human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. The "Hair Up" is a physical device (infrared lamp), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not applicable for hair growth performance. The non-clinical tests mentioned (wavelength, power, safety, EMC, biological evaluation) would have their own defined test conditions and standards, but these are not 'ground truth' in the context of clinical efficacy for hair growth.

    8. The sample size for the training set

    • Not applicable. The device is a physical infrared lamp, not an AI or algorithm that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable.