    Device Name
    VISUREF 600
    Date Cleared
    2026-03-18

    (107 days)

    Product Code
    Regulation Number
    886.1850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K162684

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VISUREF 600 is intended to be used to measure the refractive power, corneal radii (Keratometry) and to assist in pre- and post-examination of the human eye.
    VISUREF 600 is a combined autorefractor and keratometer.

    The device supports the following measurements in eye examination of the anterior eye segment:

    • Assessment of the opacity of the lens via retro-illumination.
    • Measurement of tear film break-up time by recording the time between the last complete blink and the first appearance of dark spots in the fluorescein-infilled tear film.
    • Imaging of the meibomian glands located in the upper and lower eyelids using infrared illumination.
    Device Description

    VISUREF 600 is an autorefraction and keratometry device. It can measure the objective refraction of the human eye, including sphere, cylinder, axis and higher order aberrations and the keratometry, including the central and peripheral corneal curvature.

    Additional features are available to complement main use cases of the eye examination device. Those features include:

    • Visual acuity test including variations of contrast and glare
    • Color image observation with optional use of color filters
    • Opacity observation
    • Meibomian gland assessment
    • Measurement of tear film breakup time
    • Measurement of pupil and white-to-white diameter

    The main components are:

    • Chinrest to support patient fixation
    • Measurement unit including light emitting diodes and sensors for measurement as well as target images for the patient to look at
    • User interface display and joystick to control the device operation
    • Printer to print out measurement results
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K180839

    Validate with FDA (Live)

    Device Name
    AT 030
    Date Cleared
    2018-06-07

    (69 days)

    Product Code
    Regulation Number
    886.1930
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K133476, K162684, K013402, K100035

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AT 030 is an appliance that serves to measure intraocular pressure, according to the Goldman method. Assessment of IOP may contribute to the diagnosis, and aid in the management of glaucoma.

    Device Description

    The AT 030 applanation tonometer serves for measuring intraocular pressure, following the principle introduced by Professor Goldmann.

    The AT 030 consists of the applanation tonometer with a measuring prism. The measuring prism, which includes a built-in optical doubling system, can be attached to the tonometer via the extensible measuring prism holder. The prism contacts the patient's cornea and the applanation surface is observed through the stereomicroscope of the slit lamps. The rotary knob of the graduated drum allows the user to vary the applied force. A scale on the tonometer indicates a value that can be converted into the intraocular pressure by using the conversion table supplied in the Instruction for Use.

    The AT 030 applanation tonometer is compatible with the following ZEISS slit lamps: SL 120 (K133476), SL 130 (K133476), SL 220 (K162684), LSL VISULAS 532s (K013402), and LSL VISULAS 532s vite (K100035). The AT 030 works solely mechanically.

    AI/ML Overview

    This device is a Tonometer, not an AI/ML device. Therefore, the requested information about acceptance criteria for AI performance metrics, sample sizes for test and training sets, ground truth establishment methods (expert consensus, pathology, outcomes data), multi-reader multi-case studies, and standalone algorithm performance is not applicable.

    However, I can provide the available information regarding the device's performance based on the provided text.

    Here's a summary of the device performance and relevant testing:

    1. Table of acceptance criteria and the reported device performance

    FeatureAcceptance Criteria (Predicate AT 900)Reported Device Performance (AT 030)
    Operation principleGoldmann TonometerGoldmann Tonometer
    TypeManual contact tonometerManual contact tonometer
    IndicationIntraocular Pressure (IOP) measurementIntraocular Pressure (IOP) measurement
    Type of Pressure TransducerMeasuring PrismMeasuring Prism
    Measuring PrismArticle#: HSC 1000855 Cleared in K981432 (Identical/unmodified)Article#: HSC 1000855 (Identical/unmodified)
    Prism MaterialAcrylic (PMMA)Acrylic (PMMA)
    Measurement techniqueRequires maintaining a uniform applanation of the surface of the cornea.Requires maintaining a uniform applanation of the surface of the cornea.
    Applanation surface diameter3.06 mm3.06 mm
    Measurement Range0 – 80 mm Hg2 – 80 mm Hg
    Measurement readingReading of the value, multiplying x 10 yields the IOP in mm Hg.Reading the value, multiplying x 10 yields the IOP in mm Hg.
    Accuracy and repeatabilityThe measurement deviation in the measuring prism is in the measuring range from 0 - 58.84 mN and amounts to a maximum of ±1.5% and to a minimum of ±0.49 mN of the nominal value.The measurement deviation in the measuring prism is in the measuring range from 0 – 58.84 mN and amounts to a maximum of ±1.5% and to a minimum of ±0.49 mN of the nominal value.

    The device (AT 030) claims substantial equivalence to the predicate device (Haag-Streit Goldmann manual tonometer AT 900) based on sharing the same intended use and technological characteristics, including the same operational principle, type, indication, pressure transducer, measuring prism, prism material, measurement technique, applanation surface diameter, and accuracy/repeatability. The only noted difference in the table is the lower bound of the measurement range (2-80 mmHg for AT 030 vs. 0-80 mmHg for AT 900).

    Specific study and testing:

    • Bench Testing: The AT 030 has been tested and shown to comply with the applicable requirements of ANSI Z 80.10-2014 Ophthalmic Instruments - Tonometers. It meets the requirements of a reference tonometer as specified in Annex A of ANSI Z80.10.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable. The device is a mechanical tonometer, and the performance demonstrated is through bench testing against a standard, not a clinical test set with patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. Ground truth for a mechanical device is established through calibrated measurement standards, not expert consensus.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is a mechanical device, not an AI device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the bench testing, the ground truth is based on the requirements and specifications defined in the ANSI Z 80.10-2014 Ophthalmic Instruments - Tonometers standard, likely involving calibrated measurement references.

    8. The sample size for the training set

    • Not applicable. This is a mechanical device, not an AI device.

    9. How the ground truth for the training set was established

    • Not applicable. This is a mechanical device, not an AI device.