Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K222848

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-01-20

    (121 days)

    Product Code
    Regulation Number
    882.5600
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K162539

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. The pRESET® Thrombectomy Device is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received thrombolytic therapy. Endovascular therapy with the device should be started within 6 hours of symptom onset.
    2. The pRESET® Thrombectomy Device is indicated to restore blood flow by removing thrombus from a large intracranial vessel in patients experience stroke within 8 hours of symptom onset. Patients who are ineligible for thrombolytic therapy or who fail thrombolytic therapy are candidates for treatment.
    Device Description

    The pRESET® Thrombectomy Device is designed to restore blood flow in the neurovasculature by mechanical removal of thrombus in patients experiencing acute ischemic stroke due to large vessel occlusion with thrombus. The device is designed for use in large vessels of the neurovasculature such as the internal carotid artery (ICA) and the middle cerebral artery (MCA). The device is supplied sterile and intended for single use only.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets those criteria, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Endpoint)Reported Device Performance (pRESET) (Intent-to-Treat Population)Predicate Device (Solitaire) (Intent-to-Treat Population)Difference (pRESET minus Solitaire)1-Sided 95% Confidence Interval (Lower, Upper Bound)Met/Not Met (Based on Thresholds)
    Primary Effectiveness: 90 Day mRS ≤ 2 (Proportion of subjects with global disability mRS ≤ 2 at 90 days after index procedure)95 (54.91%)96 (57.49%)-2.57%-11.42%, 6.28%Met (Lower bound -11.42% > -12.5%)
    Primary Safety: 24 Hour sICH (Proportion of subjects with device- or procedure-related symptomatic intracerebral hemorrhage within 24 hours)0 (0.00%)2 (1.20%)-1.20%-2.58%, 0.19%Met (Upper bound 0.19% < 5%)
    Secondary Effectiveness: eTICI ≥ 2b50 (Proportion of subjects with eTICI 2b50 or greater flow in target vessel post-procedure with ≤ 3 passes)146 (84.39%)149 (89.22%)-4.83%-10.84%, 1.19%Met (Lower bound -10.84% > -12.1%)

    Note: The "Met/Not Met" column is inferred based on the stated "a priori threshold" for each criterion. The document explicitly states that the pRESET device was "demonstrated to be non-inferior" for these outcomes.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set:
      • Intent-to-Treat (ITT) Population: 340 subjects (173 pRESET, 167 Solitaire)
      • Per Protocol (PP) Population: 266 subjects (138 pRESET, 128 Solitaire)
      • As Treated (AT) Population: 322 subjects (166 pRESET, 156 Solitaire)
    • Data Provenance: Prospective, multicenter, randomized controlled clinical trial conducted across 24 sites in the US (n=19) and Germany (n=5).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    The document does not explicitly state the number or qualifications of experts used to establish the ground truth for parameters like eTICI scores or mRS assessments. However, it's a clinical trial, implying that such assessments would be made by qualified medical professionals (e.g., neurologists, interventional neuroradiologists) at each of the 24 participating sites. The ground truth for neurological and functional evaluations, and imaging interpretations (e.g., ASPECTS, core infarct volume, eTICI), would have been established by these clinical trial personnel.

    4. Adjudication Method for the Test Set

    The document does not explicitly describe an adjudication method (like 2+1, 3+1) for the test set. In a multicenter clinical trial investigating medical devices, independent core labs or central adjudication committees are often used for key endpoints, especially for imaging and neurological outcomes, to ensure consistency and reduce bias. However, this level of detail is not provided in the summary.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This study is a clinical trial comparing two thrombectomy devices (pRESET vs. Solitaire), not comparing human readers with and without AI assistance. The "readers" in this context would be the clinicians performing the procedures and assessing outcomes, not interpreting images for AI.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, this is a study of a physical medical device (thrombectomy device), not an algorithm or AI system. Therefore, no standalone algorithm performance study was relevant or performed.

    7. The Type of Ground Truth Used

    The ground truth used in the clinical trial includes:

    • Clinical Outcomes Data: Modified Rankin Scale (mRS) at 90 days (patient disability), symptomatic intracerebral hemorrhage (sICH) within 24 hours.
    • Angiographic Data: eTICI (Expanded Thrombolysis in Cerebral Infarction) scores indicating blood flow restoration.
    • Imaging Data: ASPECTS score (visual assessment of ischemic changes on CT), core infarct volume (from MRI-DWI or CTP).

    These are established by clinical assessments and imaging interpretations by medical professionals.

    8. The Sample Size for the Training Set

    The document describes a clinical trial (PROST) evaluating the pRESET Thrombectomy Device against the Solitaire Revascularization Device. This is a comparative effectiveness study, not a machine learning study. Therefore, there is no "training set" in the context of an algorithm or AI. The term "training set" is not applicable here.

    9. How the Ground Truth for the Training Set was Established

    As there is no training set for an AI/algorithm, this question is not applicable. The data collected was for a clinical trial to demonstrate safety and effectiveness, and the outcomes were assessed using standard clinical and imaging measures.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1