    Intended Use

    The Portable Oxygen Concentrator provides a high concentration of supplemental oxygen to adult patients requiring respiratory therapy on a prescriptive basis. It may be used at home, in institution, vehicle, train, airplane, boats and other transport modalities. This device is to be used as an oxygen supplement and is not intended to be life sustaining or life supporting.

    Users should follow their doctor's advice on setting the oxygen flow rate and should not adjust the flow rate without consulting a healthcare professional.

    Note: Patients should regularly consult with their physician to evaluate the need for adjustments in their oxygen therapy settings.

    Device Description

    The Portable Oxygen Concentrator is a Class II, low-risk medical device designed to provide a high-concentration oxygen supply (87%-95.5%) to adult patients requiring supplemental oxygen therapy as prescribed by a healthcare professional. It is intended for use at home, in institution, vehicle, train, airplane, boats and other transport modalities and complies with FAA regulations for in-flight use. The device is not intended for life-support or life-sustaining purposes.

    The Portable Oxygen Concentrator utilizes Pressure Swing Adsorption (PSA) technology, which extracts oxygen from ambient air by selectively adsorbing nitrogen through molecular sieve beds. Oxygen is delivered through a pulse dose mechanism, synchronizing oxygen release with the patient's inhalation cycle to optimize efficiency and minimize waste.

    The series consists of four models, each offering different pulse dose settings:

    • W-R1 (MAX): 1, 2, 3, 4, 5, 6, S
    • W-R1: 1, 2, 3, 4, 5, 6
    • W-R2: 1, 2, 3, 4, 5
    • W-R2 (Lite): 1, 2, 3, 4

    The device operates in pulse flow mode and supports multiple power sources, including 100–240V AC (50–60Hz) and a rechargeable lithium-ion battery (14.4V / 6500mAh). While the hardware supports 13.0–16.8V DC input, DC operation is not currently supported, as no DC accessories are provided or authorized. A single battery charge provides up to 4.5 hours of continuous use, ensuring flexibility across various environments.

    Designed for portability and efficiency, the W-R Series features a lightweight build (1.8 kg), low noise operation, and an intuitive LCD display. Its ergonomic and user-friendly design has been internationally recognized with six global design awards, including iF, Red Dot, and IDEA, for its usability, portability, and patient-centered innovation.

    The device is suitable for operation within a temperature range of -5°C to 40°C (23°F to 104°F), humidity levels of 5% to 90% (non-condensing), and atmospheric pressure from 54kPa to 106kPa. It can function at altitudes up to 5,000 meters (16,400 feet).

    The Portable Oxygen Concentrator consists of a casing, compressor, molecular sieve system, solenoid valve, battery, cooling fan, control board, and display screen.

    Note: The device does not include a nasal cannula; patients should purchase one separately. The oxygen outlet follows international standards, and recommended cannula specifications can be found in Section 2.12: Cannula Use of the user manual.

    AI/ML Overview

    The provided FDA 510(k) clearance letter is for a Portable Oxygen Concentrator. This device is not an AI/ML-enabled device. The information requested in the prompt (acceptance criteria, study details, sample size, ground truth, expert adjudication, MRMC studies, etc.) is typically associated with the rigorous evaluation of AI/ML software as a medical device (SaMD) or AI-enabled medical devices, especially those using diagnostic imaging.

    Medical devices like portable oxygen concentrators are evaluated primarily on their physical performance characteristics, safety standards, and conformity to established regulations for mechanical and electrical safety. The clearance letter outlines the non-clinical tests performed (safety and performance testing, biocompatibility testing, electrical safety and EMC testing, battery safety testing, and software verification/validation), but these are not the types of studies that involve expert readers, ground truth consensus, or MRMC comparative effectiveness specific to AI diagnostics.

    Therefore, I cannot extract the requested information (acceptance criteria in the context of an AI study, sample sizes for test/training sets, data provenance, expert adjudication, MRMC studies, standalone performance, or ground truth establishment relevant to AI/ML) directly from this document, as it pertains to a different type of medical device evaluation.

    To directly answer your request based on the provided document, which is for a Portable Oxygen Concentrator (not an AI/ML diagnostic device):

    The FDA 510(k) Clearance Letter for the Portable Oxygen Concentrator focuses on demonstrating substantial equivalence to a predicate device through:

    • Same Intended Use: Providing supplemental oxygen to adult patients on a prescriptive basis for respiratory therapy.
    • Similar Technological Characteristics: Utilizing breath detection, molecular sieve/pressure swing adsorption, and a pulse dose mechanism.
    • Performance Data: Presenting specifications like oxygen concentration, pulse volumes, sound levels, and mechanical/electrical safety.
    • Compliance with Recognized Standards: Adhering to various international IEC and ISO standards for medical electrical equipment, biocompatibility, and oxygen concentrators.

    The "acceptance criteria" for a device of this type are generally meeting the performance specifications and safety standards outlined in the non-clinical testing section, and demonstrating that any differences from the predicate device do not raise new questions of safety or effectiveness.

    Here's a breakdown of the closest equivalents to your requested categories, given the nature of the device and the document:


    1. A table of acceptance criteria and the reported device performance

    For a portable oxygen concentrator, acceptance criteria are generally related to its physical and performance specifications like oxygen purity, flow rates, noise levels, and battery life, rather than diagnostic accuracy metrics.

    FeatureAcceptance Criteria (Subject Device Specification)Reported Device Performance (from "Specifications" and "Performance" sections)
    Oxygen Concentration90% - 3% /+ 5.5% at all settings90% - 3% /+ 5.5% at all settings
    Inspiratory Trigger Sensitivity<0.12 cmH2O<0.12 cmH2O
    Maximum Outlet Pressure< 28.9 PSI (199 kPa)< 28.9 PSI (199 kPa)
    Nominal Sound Level (at setting 2)Comparable to predicate (39 dBA typical) and within safe limits for patient use. (Subject device: 42.3 dBA)42.3 dBA (at setting 2)
    Maximum Sound Pressure Level (at max flow setting S)Comparable to predicate (54 dBA) and within safe limits for patient use. (Subject device: 50.5 dBA)50.5 dBA (at maximum flow setting S)
    Battery DurationSufficient for portable use (Predicate: unspecified, but implies multiple hours)1.5-4.5 hrs
    Operating Temperature-5°C to 40°C (23°F to 104°F)23 to 104℉ (-5 to 40℃)
    Operating Humidity5% to 90%, non-condensing5% to 90%, non-condensing
    Operating Altitude0 to 5000 meters (0 to 16,400 ft)0 to 16,400 ft (0 to 5000 meters)
    Measurement UncertaintiesPulse volumes: ± 15% of rated volumePressure: ± 0.03 psig (General) / ± 0.05 cm H2O (Inspiratory Trigger Sensitivity)Oxygen concentration: ± 3% (not accounting for temperature, barometric pressure, and time from measurement device calibration)Stated as listed.
    Compliance with StandardsAdherence to relevant IEC and ISO standards (e.g., IEC 60601-1, ISO 80601-2-69, ISO 18562 series, IEC 62133-2)Demonstrated through non-clinical testing.

    Study Proving Acceptance Criteria:

    The study proving the device meets its acceptance criteria is primarily through the Non-Clinical Testing Summary described in Section VIII of the 510(k) Summary.

    • Safety and Performance Testing: This involved evaluations against ISO 80601-2-69:2020 and ISO 80601-2-67:2020 (standards for oxygen concentrators and oxygen-conserving equipment), and internal bench testing.
    • Biocompatibility Testing: Conducted per ISO 18562 series.
    • Electrical Safety and EMC Testing: Evaluated per IEC 60601 series standards.
    • Battery Safety Testing: Conducted per IEC 62133-2.
    • Software Verification and Validation: Conducted per FDA guidance.

    These tests collectively demonstrate that the device performs according to its specifications and is safe for its intended use, proving that it meets its "acceptance criteria" for a physical medical device.


    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not applicable in the context of an AI/ML test set. For a portable oxygen concentrator, testing involves physical devices. The non-clinical tests would have been performed on a representative sample of manufactured devices (e.g., a certain number of units for durability, environmental, and performance tests). The document does not specify the exact number of physical units tested, as this is standard engineering validation rather than a clinical study with patient data.
    • Data Provenance: Not applicable in the context of AI/ML data. This device is tested through engineering bench testing, not clinical data sets. The manufacturer is Nanjing Yinuoji Medical Technology Co., Ltd. in China, so device testing would likely be performed there or at accredited labs globally. The data is prospective in the sense that physical devices were manufactured and then tested, but not in the "retrospective/prospective data collection" sense typically associated with AI/ML studies where patient data is gathered.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. "Ground truth" in the context of AI/ML is about establishing a definitive diagnosis or label for medical images or patient data. For a portable oxygen concentrator, "ground truth" is established by direct physical measurement (e.g., using calibrated sensors to measure oxygen concentration, flow rates, sound levels) and adherence to engineering and safety standards by qualified engineers and technicians. No clinical "experts" in the sense of radiologists or pathologists establishing ground truth for diagnostic accuracy are involved.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. This is a concept for reconciling disagreements among human readers or labelers in AI/ML dataset creation. Device performance testing relies on objective measurements against established engineering specifications and international standards, not subjective adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. MRMC studies are specifically designed to assess the impact of AI algorithms on human diagnostic performance (e.g., how AI assistance affects radiologists' accuracy). This device does not have an AI component that assists human readers in diagnosis.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • While the device has software, "standalone performance" in the AI/ML context refers to the algorithmic performance in diagnostic tasks without human input. Here, the "standalone performance" is the device's ability to mechanically and electronically produce and deliver oxygen according to its specifications, which was tested via Internal Bench Testing and compliance with ISO standards.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • As explained in #3, "ground truth" for this device comes from objective physical measurements against predefined engineering specifications and international medical device performance and safety standards. For example, oxygen concentration is measured by an oxygen analyzer, flow rates by flow meters, sound by calibrated sound level meters, etc.

    8. The sample size for the training set

    • Not applicable. This device is not an AI/ML algorithm that requires a "training set" of data. Its functionality is based on established mechatronic and chemical-physical principles (Pressure Swing Adsorption).

    9. How the ground truth for the training set was established

    • Not applicable, as there is no "training set" for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K230052

    Validate with FDA (Live)

    Device Name
    Inogen Rove 6
    Manufacturer
    Date Cleared
    2023-06-30

    (175 days)

    Product Code
    Regulation Number
    868.5440
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    GCE Zen-O Portable Oxygen Concentrator Model RS-00500.,K162433

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Inogen Rove 6 Portable Oxygen Concentrator provides a high concentration of supplemental oxygen to patients requiring respiratory therapy on a prescriptive basis. It may be used in the home, institution, and transport modalities. This device is to be used as an oxygen supplement and is not intended to be life sustaining or life supporting.

    Device Description

    The Inogen Rove 6 Portable Oxygen Concentrator (Inogen Rove 6) is a Class 2, low risk, portable oxygen generator that provides a high concentration of supplemental oxygen to patients requiring respiratory therapy on a prescriptive basis. It may be used in home, institution, vehicle, and other transport modalities. It is used with a nasal cannula to channel oxygen from the device to the patient. The concentrator and the nasal cannula are non-sterile. The Inogen Rove 6 Portable Oxygen Concentrator is capable of continuous use in a home, institution, vehicle, and various mobile environments. Power options include 100 - 240 V-AC (50-60Hz) power supply, rechargeable battery packs, or a 13.5 -15.0 V-DC power cable. The Inogen Rove 6 Portable Oxygen Concentrator uses molecular sieve/pressure swing adsorption technology. Ambient air is drawn through particle filters by a compressor and forced through molecular sieve beds, which adsorb nitrogen and allow oxygen to pass. The airflow is then changed, and nitrogen is desorbed from the molecular sieve, allowing it to adsorb again during the next cycle. Oxygen is collected in an accumulator reservoir. Waste nitrogen is exhausted back into the room. A series of sieve beds, manifolds and precision valves, sensors and embedded software are used to control the cycle to make the system function. Oxygen is delivered to the patient on a pulse dose basis in precise amounts during the inhalation part of the breathing cycle. This conserver technology eliminates waste of unused oxygen at other times in the breathing cycle when it is not needed. The Inogen Rove 6 Portable Oxygen Concentrator senses the beginning of the inhalation cvcle and releases a specified dose of oxygen enriched gas from the accumulator reservoir, through a final filter, into the connected nasal cannula and on to the patient. The Inogen Rove 6 Portable Oxygen Concentrator utilizes Bluetooth technology that pairs the portable oxygen concentrator to a mobile device or tablet using the Inogen Connect App.

    AI/ML Overview

    This is information extracted from a medical device 510(k) submission.


    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Inogen Rove 6 Portable Oxygen Concentrator are based on comparison to its predicate device (Inogen Rove 4) and a reference device (GCE Zen-O), as well as compliance with relevant international standards.

    Acceptance Criteria CategorySpecific Criteria (Predicate/Reference)Reported Device Performance (Inogen Rove 6)
    Indications for UseThe Inogen Rove 4 Portable Oxygen Concentrator provides a high concentration of supplemental oxygen to patients requiring respiratory therapy on a prescriptive basis. It may be used in home, institution, and transport modalities. (K222086)Provides a high concentration of supplemental oxygen to patients requiring respiratory therapy on a prescriptive basis. It may be used in home, institution, vehicle, and other transport modalities. Substantially equivalent to predicate.
    Fundamental Scientific TechnologyBreath detection technology, Molecular Sieve/pressure swing adsorption technology (Predicate and Reference)Breath detection technology, Molecular Sieve/pressure swing adsorption technology. Substantially equivalent.
    Patient UsePatients requiring supplemental oxygen (Predicate). Adult patients with chronic pulmonary diseases such as chronic bronchitis, emphysema, asthma, or lung cancer, those in the terminal stage of cancer, or any patient requiring supplemental oxygen (Reference).Patients requiring respiratory therapy on a prescriptive basis. Substantially equivalent to predicate.
    Oxygen Concentration90% -3%/+6% at all settings (Predicate and Reference)90% -3%/+6% at all settings. Substantially equivalent.
    Output FlowPredicate (Inogen Rove 4): - Setting 1 (10-40 BPM): 7.0-21.0 ml/breath - Setting 4 (10-40 BPM): 21.0-84.0 ml/breath - Max total volume per minute: 840 ml/min. Reference (GCE Zen-O): - Pulse mode settings 1-6 (15-40 BPM): ml/breath varies from 5.25 ml/breath at setting 1, 40 BPM to 66ml/breath at setting 6, 15-35 BPM. Max total volume per minute: e.g., setting 6 at 35 breath/min = 2310 ml/min.Inogen Rove 6: - Setting 1 (10-40 BPM): 5.25-21.0 ml/breath - Setting 6 (10-40 BPM): 31.5-126.0 ml/breath - Max total volume per minute: 1260 ml/min. Similar, with 2 additional flow settings delivering a higher maximum output. The maximum output is similar to Zen-O setting 6 at 20 breaths/min (1320 ml/min).
    Maximum Outlet Pressure< 22 PSI (129 kPa) ± 10% (Predicate, Inogen Rove 4) 20.5 PSI (Reference, GCE Zen-O)<28.9 PSI (199 kPa). Test results show actual outlet pressures in the range of 24-25 PSI. Similar, but higher than predicate. This higher pressure is still below the FAA maximum oxygen pressure restriction and similar to another cleared device (Invacare Platinum Mobile Oxygen Concentrator K160630, <28.5 PSI).
    Performance StandardsCompliance with various IEC 60601 series and ISO 80601 series standards (as listed for predicate and reference devices). Usability testing per IEC 62366-1 (only mentioned for subject device, not predicate/reference explicitly for performance).Full compliance with IEC 60601-1:2012, IEC 60601-1-2:2012, IEC 60601-1-6:2020, IEC 60601-1-8:2012, IEC 60601-1-11:2015, ISO 80601-2-69:2020, ISO 80601-2-67:2020, and IEC 62366-1. Substantially equivalent to predicate/reference in standards.
    BiocompatibilityExternally Communicating, Tissue, Permanent Duration (>30 days) per ISO 18562-2: 2017 (Particulate matter) and ISO 18562-3:2017 (Volatile organic compounds) (Predicate). Externally Communicating, Tissue, Limited duration (< 24h) per ISO 10993-1 (Reference).Externally Communicating, Tissue, Permanent Duration (>30 days) per ISO 18562-2: 2017 (Particulate matter) and ISO 18562-3:2017 (Volatile organic compounds). Similar to predicate.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for the test sets in non-clinical testing. It only mentions "Inogen performed testing to demonstrate and support safety and effectiveness when compared to the predicate and the applicable standards."

    The data provenance is not specified regarding country of origin or whether it was retrospective or prospective. It refers to "Bench Testing" and "testing at a nationally registered test laboratory," implying laboratory-based testing rather than clinical data from human subjects.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the given document. The testing described is non-clinical bench testing and compliance with performance standards, which does not typically involve expert ground truth establishment in the same way clinical studies or image-based AI studies do.


    4. Adjudication Method for the Test Set

    This information is not applicable/provided. The document describes non-clinical bench testing against technical specifications and performance standards, not a study requiring adjudication of human-read results.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    This information is not applicable/provided. The Inogen Rove 6 Portable Oxygen Concentrator is a hardware medical device (portable oxygen generator), not an AI-assisted diagnostic or interpretative tool. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance would not have been performed for this device.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable/provided. As explained above, this device is not an AI algorithm. Its performance is evaluated through bench testing of its mechanical and electrical functions and oxygen delivery specifications. The device does contain "embedded software" (page 4), and "Software verification and validation" (page 19) was performed, indicating standalone testing of the software components. However, this is not an "algorithm only" study in the context of AI performance, but rather software testing for the device's operational control.


    7. The Type of Ground Truth Used

    For the non-clinical testing, the "ground truth" is defined by:

    • Established Technical Specifications: Such as oxygen concentration levels (90% -3%/+6%), pulse volumes (± 15% of rated volume), pressure values, sound levels, etc.
    • International Standards: Specifically, ISO 80601-2-69, ISO 80601-2-67, various parts of IEC 60601, and IEC 62366-1.
    • Comparison to Predicate and Reference Devices: Demonstrating "substantial equivalence" to the performance and characteristics of previously cleared devices, particularly for aspects like oxygen delivery mode, purity, and flow.

    This is primarily performance data and compliance with objective technical and regulatory standards, not expert consensus, pathology, or outcomes data in a clinical sense.


    8. The Sample Size for the Training Set

    This information is not applicable/provided. No AI model requiring a "training set" in the context of machine learning is described for this device. The software validation mentioned refers to verification of the device's embedded control software.


    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable/provided, as no AI model with a training set is described.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222086

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-12-09

    (147 days)

    Product Code
    Regulation Number
    868.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K111885, K162433

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Inogen Rove™ 4 Portable Oxygen Concentrator provides a high concentration of supplemental oxygen to patients requiring respiratory therapy on a prescriptive basis. It may be used in home, institution and transport modalities.

    This device is to be used as an oxygen supplement and is not intended to be life sustaining or life supporting.

    Device Description

    The Inogen Rove™ 4 Portable Oxygen Concentrator is a portable oxygen generator that provides a high concentration of supplemental oxygen to patients requiring respiratory therapy on a prescriptive basis. It may be used in home, institution and transport modalities. It is used with a nasal cannula to deliver oxygen from the device to the patient. This device is to be used as an oxygen supplement and is not intended to be life sustaining or life supporting.

    The Inogen Rove 4 Portable Oxygen Concentrator is capable of continuous use in a home, institution, and various mobile environments. Power options include 100 - 240 V-AC (50-60Hz) power supply, rechargeable battery packs, or a 13.5 -15.0 V-DC power cord.

    The Inogen Rove 4 Portable Oxygen Concentrator uses molecular sieve/pressure swing adsorption technology. Ambient air is drawn through particle filters by a compressor and forced through molecular sieve beds, which adsorb nitrogen and allow oxygen to pass. The airflow is then changed and nitrogen is desorbed from the molecular sieve, allowing it to adsorb again during the next cycle. Oxygen is collected in an accumulator reservoir. Waste nitrogen is exhausted back into the room. A series of sieve beds, manifolds, valves, sensors, and software are used to control the cycle to make the system function.

    Oxygen is delivered to the patient on a pulse dose basis in pre-set amounts during the inhalation part of the breathing cycle. This conserver technology eliminates waste of unused oxygen at other times in the breathing cycle when it is not needed. Inogen Rove 4 Portable Oxygen Concentrator senses the beginning of the inhalation cycle and releases a specified dose of oxygen enriched gas.

    The Inogen Rove 4 Portable Oxygen Concentrator utilizes Bluetooth technology that pairs the portable oxygen concentrator to a mobile device or tablet using the Inogen Connect App.

    AI/ML Overview

    The Inogen Rove™ 4 Portable Oxygen Concentrator's acceptance criteria and the study that proves its performance are described below.

    1. Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / SpecificationPredicate Device: Belluscura X-PLO₂R PerformanceSubject Device: Inogen Rove 4 Performance (Reported as 'Similar' or specific values)Comparison
    K NumberK203086K222086N/A
    Product CodeCAWCAWSubstantially equivalent
    CFR21 CFR 868.544021 CFR 868.5440Substantially equivalent
    Classification22Substantially equivalent
    Indications for UseSupplemental oxygen for patients requiring high concentrations, portable, continuous use in home, institutional, and travel/mobile environments. Not life-sustaining.High concentration of supplemental oxygen to patients requiring respiratory therapy on a prescriptive basis. Used in home, institution, and transport modalities. Not life-sustaining or life-supporting.Substantially equivalent.
    PrescriptiveYesYesSubstantially equivalent
    Fundamental scientific technologyBreath detection, Molecular Sieve/pressure swing adsorption.Breath detection, Molecular Sieve/pressure swing adsorption.Substantially equivalent
    Patient useAdult patient onlyAdult patient onlySubstantially equivalent
    User/Patient Interface - TypeUser interface panelUser interface panelSubstantially equivalent
    User/Patient Interface - DisplayLCD Display (numbers, text, symbols)LCD Display (numbers, symbols)Substantially equivalent. Setting, battery, and auditory alarm status are displayed.
    User/Patient Interface - Alarm IndicatorYellow "Alarm/Warning" triangle symbol on LCD display (ISO 60601-1-8 compliant)Yellow LED on UIP above "Alarm/Warning" triangle symbol that illuminates (ISO 60601-1-8 compliant)Substantially equivalent. Both use "Alarm/Warning" triangle symbol.
    User/Patient Interface - Breath Detect NotificationBreath detect icon pops up on LCD display.Green LED on UIP illuminates.Substantially equivalent. Both indicate when a breath is detected and oxygen pulse delivered.
    User/Patient Interface - Auditory BuzzerAudible beeps for alarm/status change (ISO 60601-1-8 compliant)Audible beeps for alarm/status change (ISO 60601-1-8 compliant)Substantially equivalent.
    User/Patient Interface - Sieve bedsUser replaceable using cartridge pull handle.User replaceable using M6 hex Allen key.Substantially Equivalent. Both allow user replacement, but with different tools.
    User/Patient Interface - Air Intake FilterClean once per week.Clean particle filters once per week.Substantially equivalent.
    Optional accessoriesCarry Bag, Strap, Backpack, External Battery ChargerCarry Bag, Backpack, External Battery ChargerSubstantially equivalent. (Strap is missing from subject device list, but generally comparable).
    Mobile ApplicationNoInogen Connect Mobile Application (iOS/Android) for viewing settings, battery, alerts.Substantially equivalent. Inogen Rove 4 allows information on mobile device and concentrator display. Predicate only on display.
    Operating SystemSoftware monitoredSoftware monitoredSubstantially Equivalent
    Bluetooth TechnologyNoYes (BLE Connection to Android or iPhone with Inogen Connect App)Substantially Equivalent. (See Mobile Application comparison).
    AC/DC Power Adapter100-240V, 50/60Hz AC.100-240V, 50/60Hz AC.Substantially equivalent.
    DC Power Cable12-volt DC outlet connection.12-volt DC outlet connection.Substantially equivalent.
    CannulaOff the shelf 7' nasal cannula.User supplied off-the-shelf.Substantially equivalent.
    Battery4 or 8-cell lithium battery, slide-in attachment.4 or 8-cell lithium battery, slide-in attachment.Substantially equivalent.
    Size (4-cell battery)7.0"H, 7.3"W, 2.9"D7.5"H, 6.0"W, 2.7"DSimilar: Inogen Rove 4 is smaller.
    Principle of OperationPressure Swing Adsorption (PSA) technology, pulse dose upon inhalation detection.Molecular sieve / pressure swing adsorption technology, pulse dose upon inhalation detection.Substantially equivalent.
    Operating ConditionsPortable. Home, institutional, travel/mobile environments.Portable. Home, institution and transport modalities.Substantially equivalent.
    Oxygen Delivery ModePulse DosePulse DoseSubstantially equivalent.
    Output Flow (Total Volume/Minute)200 ml/min (Setting 1) to 800 ml/min (Setting 4) (at 15-40 BPM)Higher volumes (e.g., 21.0 ml/min at 10 BPM for Setting 1)Similar volumes. Inogen Rove 4 has higher volumes per minute for each setting.
    Oxygen Purity82% to 92% at all settings87% to 96% at all settingsSimilar. Inogen Rove 4 has higher oxygen concentrations per minute for each setting.
    Maximum Outlet Pressure10psig (41.37 kPa)< 22 PSI, 18.7 PSI (129 kPa) ± 10%Inogen Rove 4 has a higher maximum outlet pressure. Differences are discussed in Section VIII.
    Performance StandardsIEC 60601-1, -1-2, -1-6, -1-8, -1-11, ISO 80601-2-69, ISO 80601-2-67, IEC 62366-1IEC 60601-1:2012, -1-2: 2020, -1-6:2020, -1-8:2012, -1-11:2015, ISO 80601-2-69:2020, ISO 80601-2-67:2020, IEC 62366-1Similar (updated versions of standards for subject device).
    BiocompatibilityExternally Communicating, Tissue, Permanent Duration (>30 days). ISO 18562-2:2017 (Particulate matter), ISO 18562-3:2017 (VOCs).Externally Communicating, Tissue, Permanent Duration (>30 days). ISO 18562-2:2017 (Particulate matter), ISO 18562-3:2017 (VOCs).Substantially equivalent.

    2. Sample size used for the test set and the data provenance

    The submission does not explicitly detail a "test set" in the context of clinical data for performance evaluation that would typically measure diagnostic accuracy. Instead, the testing described is primarily non-clinical bench testing to demonstrate safety and effectiveness against standards and comparison to a predicate device's specifications.

    Therefore, for aspects like pulse volume, oxygen purity, trigger sensitivity, alarms, software verification/validation, electrical/EMC/RFID, and biocompatibility, there isn't a "test set" sample size in the traditional sense of patient data. The "test set" would consist of the performance parameters measured on the device itself under various conditions.

    The provenance of this data is from bench testing conducted by the manufacturer and a nationally registered test laboratory. No specific country of origin for this testing is mentioned, nor is it categorized as retrospective or prospective clinical data, as it is technical performance testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable as the evaluation relies on objective performance measurements against established engineering standards and technical specifications (e.g., oxygen concentration, pressure), rather than expert interpretation of medical images or patient conditions to establish a "ground truth."

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable as the performance evaluation is based on objective measurements against engineering standards and specifications, not expert consensus on qualitative data.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable as the device is a portable oxygen concentrator, not an AI-assisted diagnostic or interpretative medical device. The submission describes a comparison between the subject device and a predicate device based on technical specifications and non-clinical testing, not a study involving human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable for the core function of the oxygen concentrator. The device does have "Software verification and validation" and a mobile app, but this does not imply a "standalone algorithm-only" performance evaluation in the context of diagnostic accuracy or a similar clinical endpoint. The mobile app's function is for displaying device settings and alerts to the user, not for standalone diagnostic interpretation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the device's performance evaluation is based on:

    • Established engineering standards: e.g., IEC 60601 series, ISO 80601 series, IEC 62366-1.
    • Objective physical measurements: such as oxygen purity, flow rates, output pressure, trigger sensitivity, battery life, sound levels, dimensions, and weight.
    • Predicate device specifications: used as a benchmark for comparison during the "Substantial Equivalence Discussion."

    There is no mention of expert consensus, pathology, or outcomes data as "ground truth" for this device.

    8. The sample size for the training set

    This information is not applicable. The submission describes the premarket notification for a medical device (portable oxygen concentrator) based on physical and technical performance testing, not a machine learning model that would require a "training set."

    9. How the ground truth for the training set was established

    This information is not applicable as there is no "training set" for a machine learning model in this submission.