510(k) Data Aggregation

    K Number
    K212211

    Validate with FDA (Live)

    Date Cleared
    2022-02-09

    (209 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K191560, K172090, K161901

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Turnpike catheters: The Turnpike catheters are intended to access discrete regions of the coronary and/or peripheral vasculature. They may be used to facilitate placement and exchange of guidewires and to subselectively infuse/ deliver diagnostic and therapeutic agents, and to assist in crossing de novo coronary chronic total occlusions (CTO).

    GuideLiner V3 catheter: GuideLiner catheters are intended to be used in conjunction with guide catheters to access discrete regions of the coronary and/or peripheral vasculature. to facilitate placement of interventional devices, and to assist in crossing de novo coronary chronic total occlusions (CTO).

    TrapLiner catheter: The TrapLiner catheter is intended for use in conjunction with guide catheters to access discrete regions of the coronary and/or peripheral vasculature, to facilitate placement of interventional devices, to facilitate the exchange of an interventional device while maintaining the position of a guidewire within the vasculature, and to assist in crossing de novo coronary chronic total occlusions (CTO).

    Device Description

    Turnpike catheters: The Turnpike catheters are single lumen catheters designed for use in the coronary and peripheral vasculature. The catheter shaft is constructed of two polymer layers that encapsulate a braid and a duallayer coil. The distal 60 cm of the Turnpike catheters are hydrophilic coated, and all models are compatible with 0.014" guidewires and 5F guide catheters. The turnpike catheters are available in four configurations: Turnpike, Turnpike Spiral, Turnpike Gold, and Turnpike LP.

    GuideLiner V3 catheters: The GuideLiner V3 catheter is a rapid-exchange guide extension catheter designed for use in the coronary and peripheral vasculature. It is available in five sizes - 5F, 5.5F, 6F, 7F, and 8F. All sizes of the GuideLiner V3 catheter have a 150 cm working length, consisting of a 125 cm long stainless steel pushwire shaft followed distally by a 25 cm long full-round, silicone-wiped guide extension segment. The distal 17 cm of the 125 cm pushwire shaft is covered with a semicircular shaped polymer that meets the proximal end of the full-round guide extension segment.

    TrapLiner catheters: The TrapLiner catheter is a rapid-exchange guide extension catheter with a trapping balloon on the distal end of the pushrod. The stainless steel hypotube pushrod is covered on the distal end by a semi-circular polymer ('half-pipe') and transitions to a hydrophilic coated full-round polymer guide extension section. There are two radiopaque marker bands on the guide extension segment, one on the distal tip and one on the collar. The Nylon trapping balloon (3.1 mm diameter) is located proximal to the half-pipe and has a single radiopaque gold marker under the proximal end of the balloon is reflowed, and an adhesive is applied on each end. The TrapLiner catheter has an over-molded Nylon hub on the proximal end to facilitate balloon inflation.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria or a study designed to prove a device meets acceptance criteria in the context of an AI/ML device. The document is an FDA 510(k) premarket notification for various catheters (Turnpike, GuideLiner V3, TrapLiner) and details their indications for use, technological characteristics, and a clinical study conducted to evaluate their safety and effectiveness for assisting in crossing de novo coronary chronic total occlusions (CTO).

    The clinical study described is a prospective, multi-center, single-arm study of 150 subjects. It evaluates the safety and effectiveness of the catheters and specialized guidewires in patients undergoing CTO-PCI, not an AI device. The primary endpoint was "procedure success through discharge or 24 hours post-procedure, whichever came first," with specific components for angiographic visualization of guidewire placement and absence of in-hospital MACE.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance, or details about sample size for a test set, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, ground truth type, or training set information as requested, because this information is not present in the provided text for an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172090

    Validate with FDA (Live)

    Date Cleared
    2017-10-20

    (101 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K161901

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    GuideLiner catheters are intended to be used in conjunction with guide catheters to access discrete regions of the coronary and/or peripheral vasculature, and to facilitate placement of interventional devices.

    Device Description

    The GuideLiner V3 catheter is a rapid-exchange guide extension catheter designed for use in the coronary and peripheral vasculature. It is available in five sizes – 5F, 5.5F, 7F, and 8F. All sizes of the GuideLiner V3 catheter have a 150 cm working length, consisting of a 125 cm long stainless steel pushwire shaft followed distally by a 25 cm long full-round, silicone-wiped guide extension segment. The distal 17 cm of the 125 cm pushwire shaft is covered with a semi-circular polymer that meets the proximal end of the full-round guide extension segment. The GuideLiner V3 catheter has two platinumiridium marker bands; the distal marker band is located at the distal tip and the proximal marker band is located near the collar. The GuideLiner V3 catheter also has two positioning marks located 95 cm (single mark) and 105 cm (double mark) from the distal tip.

    AI/ML Overview

    This document, K172090, describes a 510(k) premarket notification for a medical device (GuideLiner V3 Catheter), not a study assessing the performance of an AI/ML-based medical device. Therefore, the information required to answer your questions regarding acceptance criteria and study proving a device meets these criteria is not present in the provided text.

    The document is a clearance letter from the FDA, stating that the GuideLiner V3 Catheter is substantially equivalent to legally marketed predicate devices. It lists the types of performance and biocompatibility tests conducted to demonstrate this substantial equivalence. However, it does not provide specific quantitative acceptance criteria or detailed study results (like sample sizes, ground truth establishment, expert qualifications, etc.) that would be typical for an AI/ML device performance study.

    Here's why the requested information cannot be extracted:

    • Acceptance Criteria & Performance Table: The document states that "The results of the verification tests met the specified acceptance criteria," but it does not define what those criteria were (e.g., specific thresholds for kink radius, bend diameter, etc.) nor does it report the actual measured performance values for these tests.
    • Sample Size and Data Provenance: This information is not provided for the various performance and biocompatibility tests mentioned. These are likely engineering and bench tests, not clinical studies with patient data.
    • Experts for Ground Truth & Adjudication Method: "Ground truth" in the context of this device would likely refer to engineering specifications or established physical properties. There's no mention of experts establishing visual "ground truth" or adjudication, as this is not an image-based diagnostic AI device.
    • MRMC Comparative Effectiveness Study: This is irrelevant here as the device is a physical catheter, not an AI assisting human readers.
    • Standalone Performance: The performance tests mentioned are inherently "standalone" for the device's physical and mechanical properties.
    • Type of Ground Truth: For a physical medical device like a catheter, "ground truth" refers to established engineering standards, material properties, and functional specifications, not expert consensus on medical images or pathology.
    • Training Set Sample Size & Ground Truth Establishment: This applies to AI/ML models. The GuideLiner V3 Catheter is a physical medical device, not an AI/ML product. Therefore, there's no "training set" or "ground truth for training" in the context of machine learning.

    The document confirms that various tests were performed (e.g., Kink Radius, Bend Diameter, Biocompatibility tests like Cytotoxicity, Sensitization) and that the device "passed" or "met" the specified acceptance criteria. However, it does not provide the specifics of these criteria or the numerical results.