Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K190535

    Validate with FDA (Live)

    Date Cleared
    2019-08-06

    (155 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BiMobile Dual Mobility System is indicated for patients with mobility-limiting diseases, fractures or defects of the hip joint or proximal femur, which cannot be treated by conservative or osteosynthetic procedures. The BiMobile Dual Mobility System is indicated for the following conditions:

    • · Primary and secondary osteoarthritis
    • · Rheumatoid arthritis
    • · Correction of functional deformities
    • Avascular necrosis
    • Femoral neck fractures
    • · Revision after implant loosening dependent on bone mass and quality
    • · Dislocation risks
      The device is intended for cemented and cementless use.
    Device Description

    The BiMobile cup system is a dual mobility acetabular implant system. The system is consisting of wrought cobalt chromium molybdenum alloy (CoCrMo acc. to ISO 5832-12) metal shells with a highly polished inner surface, which are available in three different outer surface modifications. There is the cemented version with a glass blasted surface of concentric grooves, and two cementless versions where a pressfit generating macrostructure is covered by either a PlasmaLink (titanium plasma spray acc. to ISO 5832-2/ASTM F1580) or TiCaP (titanium plasma spray and calcium phosphate acc. to ISO 5832-2/ASTM F1580 and F1609) coating. For each size, from 42 to 70, a dedicated liner made of Ultra-high-molecular-weight polyethylene (conventional UHMWPE acc. to ISO 5834-2/ASTM F648) is available. These can be combined with ceramic or CoCr alloy femoral heads of sizes 22 mm (for Size 42-46) and 28 mm (for size 48-70). The purpose of this submission is to add acetabular liners in Vitamin E highly crosslinked UHMWPE (E-Dur) acc. to ASTM F2695, ASTM F2565 to the BiMobile Dual Mobility System. This premarket notification does not include the femoral components. The BiMobile Dual Mobility System is compatible with previously cleared femoral heads (K171273, K161840, K953653, K931571, and K920756), femoral stems and neck segments (K183141, K161840, K151008, K142187, and K953653,) of Waldemar Link GmbH & Co. KG.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called the "BiMobile Dual Mobility System - E-Dur Inserts." It outlines the device, its intended use, and its comparison to predicate devices, focusing on the justification for substantial equivalence.

    However, the document does not contain the specific information required to answer your request about acceptance criteria and the study that proves the device meets them, especially in the context of AI/ML or diagnostic performance. The provided text details non-clinical performance testing for a hip implant system, which involves material characterization, biocompatibility testing, wear analysis, and mechanical testing. These are typical assessments for orthopedic implants to demonstrate their safety and effectiveness in a mechanical and biological sense, not for evaluating a diagnostic or AI-driven system's performance.

    Therefore, many of your specific questions cannot be answered from the provided text.

    Here's what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance:

    The document mentions several types of non-clinical performance testing. While it states that the results "demonstrate that the device is as safe, as effective, and substantially equivalent to the predicate devices," it does not provide specific numerical acceptance criteria or reported performance values for each test (e.g., a specific wear rate limit or a minimum axial disassembly force). It only lists the types of tests performed.

    Acceptance CriteriaReported Device Performance
    Not specified in the document for individual tests.The results of non-clinical performance testing demonstrate that the device is as safe, as effective, and substantially equivalent to the predicate devices. (Specific performance values not provided)

    Regarding questions 2-9 about test sets, ground truth, expert adjudication, MRMC studies, and training sets, the document does not provide any relevant information. This is because the submission is for a physical medical implant, not a diagnostic device or an AI/ML system.

    Here's why these questions are not applicable based on the provided text:

    • 2. Sample sized used for the test set and the data provenance: Not applicable. The tests performed are non-clinical (material, wear, mechanical) and do not involve patient data or diagnostic test sets.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth, in the context of diagnostic or AI performance, is not relevant for this type of device.
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for non-clinical mechanical/material testing.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a hip implant, not a diagnostic or AI-assisted interpretation device.
    • 6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable.
    • 7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable.
    • 8. The sample size for the training set: Not applicable.
    • 9. How the ground truth for the training set was established: Not applicable.

    In summary: The provided FDA 510(k) clearance letter and summary are for a mechanical medical device (a hip implant system). The "performance testing" described is non-clinical, mechanical, and material-based, aimed at demonstrating the physical properties and durability of the implant. It does not involve AI/ML performance, diagnostic accuracy, human readers, or patient-specific ground truth data in the way your questions are framed.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183141

    Validate with FDA (Live)

    Date Cleared
    2019-02-08

    (87 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K142187, K161840

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LINK ® MP® Monoblock is indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur which cannot be treated by conservative or osteosynthetic procedures

    The LINK® MP® Monoblock is indicated for the following conditions:

    • Revision of loosened femoral prosthesis components involving extensive bone resorption of the proximal femur and widening of the medullary cavity or marked thinning of proximal femoral cortical bone the fixation of a new standard prosthesis is not possible

    • Revision of loosened femoral prosthesis components by peri-/subprosthetic fracture

    The device is intended for cementless use.

    Device Description

    The LINK® MP® Monoblock Hip Prosthesis is a straight hip prosthesis stem. The hip prosthesis is made of Titanium Aluminum Vanadium alloy (Ti-6Al-4V). The prosthesis stems are available in a range of sizes, lengths, CCD angles and offsets. Additional features include a tapered stem with a microporous glass-blasted surface.

    The LINK® MP® Monoblock Hip Prosthesis is designed to be used in conjunction with femoral heads that are manufactured from Cobalt Chromium Molybdenum casting alloy (CoCrMo) or ceramic (BIOLOX® forte and BIOLOX® delta) via 12/14 morse taper (K161840). In total joint use, the femoral head articulates aqainst LINK® BiMobile Cup System (K171273). The LINK® MP® Monoblock Hip Prosthesis provides cementless fixation to the bone.

    AI/ML Overview

    The provided text is related to an FDA 510(k) premarket notification for a medical device, the "LINK® MP® Monoblock Hip Prosthesis." This document concerns the regulatory clearance of a physical medical device (an orthopedic implant), not an AI/ML software.

    Therefore, the requested information about acceptance criteria, study details, sample sizes, ground truth establishment, expert qualifications, adjudication methods, and MRMC studies related to AI/ML performance is not applicable to this document.

    The document states:

    • "Non-clinical performance testing and analysis were provided, including bench testing." This refers to physical tests on the implant, not a study involving data or AI.
    • "There was no clinical performance testing required for this device." This explicitly states that human clinical trials were not part of this submission for demonstrating substantial equivalence.

    In summary, the provided content does not contain the information requested about AI/ML device performance or the study details typically associated with it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171273

    Validate with FDA (Live)

    Date Cleared
    2018-01-19

    (263 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K161840, K953653, K142187, K151008

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LINK® BiMobile™ Dual Mobility System is indicated for patients with mobility-limiting diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures.

    The LINK® BiMobile™ Dual Mobility System is indicated for the following conditions:

    • · Primary and secondary coxarthrosis
    • · Osteoarthritis
    • Necrosis of the femoral head
    • Femoral neck fractures
    • · Revision after implant loosening
    • · Acetabular dysplasia

    The device is intended for cemented and cementless use.

    Device Description

    The BiMobile™ cup system is a dual mobility acetabular implant system. The system is consisting of wrought cobalt chromium molybdenum alloy (CoCrMo acc. to ISO 5832-12) metal shells with a highly polished inner surface which are available in three different outer surface modifications. There is the cemented version with a glass blasted surface of concentric grooves, and two cementless versions where a pressfit generating macrostructure is covered by either a PlasmaLink® (titanium plasma spray acc. to ISO 5832-2/ASTM F1580) or TiCaP® (titanium plasma spray and calcium phosphate acc. to ISO 5832-2/ASTM F1580 and F1609) coating. For each size, from 42 to 70, a dedicated liner made of Ultra-high-molecular-weight polyethylene (conventional UHMWPE acc. to ISO 5834-2/ASTM F648) is available. These can be combined with ceramic or CoCr alloy femoral heads of sizes 22 mm (for Size 42-46) and 28 mm (for size 48-70).

    This premarket notification does not include the femoral components. The LINK® BiMobile™ Dual Mobility System is compatible with the FDA cleared LINK® femoral stems of the SP-CL® / LCU® (K161840); and the Lubinus SPII® (K953653) hip systems, with the MP® Reconstruction Prosthesis (K142187) and MEGASYSTEM-C® (K151008) neck segments and with the femoral heads made of cobaltchromium and ceramic.

    AI/ML Overview

    The provided text is related to a 510(k) premarket notification for a medical device called the LINK® BiMobile™ Dual Mobility System, which is an acetabular cup for hip replacement.

    Based on the provided text, there is no information about acceptance criteria or a study proving that the device meets specific performance criteria in the context of typical AI/ML device evaluations (e.g., sensitivity, specificity, accuracy). This document is a regulatory submission for a traditional medical device, not a software as a medical device (SaMD) or an AI/ML-powered device.

    The "Performance Data" section explicitly states:

    • "Non-clinical performance testing and analysis were provided including bench testing, coating characterization and endotoxin testing."
    • "Bench testing included: Wear test per ISO 14242-1 and -2; Particles Analyses per ISO 17853 and ASTM F1877; Liner Press-in testing; Lever-out testing, Axial Disassembly testing and Range of Motion analyses of the LINK® BiMobile™ Dual Mobility System."
    • "The coating was evaluated per FDA Guidance Documents entitled "Guidance for Industry on the Testing of Metallic Plasma Sprayed Coatings on Orthopedic Implants to Support Reconsideration of Postmarked Surveillance Requirements" and "510(k) Information Needed for Hydroxyapatite Coated Orthopedic Implants"."
    • "Coating characterization included: Microstructural analysis; Mechanical testing and Metallurgical analysis."
    • "The results of non-clinical performance testing demonstrate that the device is as safe, as effective, and substantially equivalent to the predicate devices."
    • "There was no clinical performance testing required for this device."

    Therefore, I cannot populate the table or answer the specific questions related to acceptance criteria, AI/ML performance metrics, sample sizes for test/training sets, expert consensus, or MRMC studies, as these aspects are not present in the provided document. The device's "acceptance criteria" were met through non-clinical testing demonstrating substantial equivalence to predicate devices, rather than a clinical trial or AI/ML performance study.

    Summary of available information regarding the "study" for this device:

    1. A table of acceptance criteria and the reported device performance: Not provided in the context of AI/ML metrics. The performance data focuses on mechanical and material integrity (wear, particle analysis, press-in, lever-out, axial disassembly, range of motion, coating characteristics).
    2. Sample sized used for the test set and the data provenance: Not applicable in the context of an AI/ML test set. The "test set" refers to physical samples of the device and its components used for bench testing. The specific number of physical units tested is not detailed, nor is general "data provenance" as would be relevant for clinical data.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for mechanical testing is based on objective measurements against engineering specifications and international standards (ISO, ASTM).
    4. Adjudication method: Not applicable.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No, this was not done as it's a physical implant, not an AI assistance tool.
    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: No, this is not an algorithm-only device.
    7. The type of ground truth used: For physical performance tests, the "ground truth" refers to established engineering standards (ISO, ASTM) and material specifications.
    8. The sample size for the training set: Not applicable; there is no training set for this type of device.
    9. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1