Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K213857

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-10-14

    (308 days)

    Product Code
    Regulation Number
    870.1415
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K161772, K182035, K190925, K203329

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HeartFlow Analysis is an AI-based medical device software for the clinical quantitative and qualitative analysis of previously acquired Computed Tomography DICOM data for patients with suspected coronary artery disease. It provides anatomic data, plaque identification and characterization, as well as the calculations of FFRCT, a coronary physiological simulation, computed from simulated pressure, velocity and blood flow information obtained from a 3D computer model generated from static coronary CT images. The HeartFlow Analysis is intended to support the risk assessment and functional evaluation of coronary artery disease.

    The HeartFlow Analysis is provided to support qualified clinicians to aid in the evaluation and risk assessment of coronary artery disease. The HeartFlow Analysis is intended to be used by qualified clinicians in conjunction with the patient's clinical history, symptoms, and other diagnostic tests, as well as the clinician's professional judgment.

    Device Description

    The HeartFlow Analysis is an Al-based medical device software developed for the clinical quantitative and qualitative analysis of CT DICOM data. It is a tool for the analysis of CT DICOM-compliant cardiac images and data, to assess the anatomy and function of the coronary arteries in the risk stratification and evaluation of coronary artery disease.

    The software displays coronary and functional information using graphics and text, including computed and derived quantities of percent stenosis, plaque volumes, blood flow, pressure and velocity, to aid the clinician in the assessment and treatment planning of coronary artery disease.

    The HeartFlow Analysis is performed on previously physician-acquired image data and is unrelated to acquisition equipment and clinical workstations.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text does not explicitly state specific acceptance criteria with numerical targets or thresholds. It generally discusses "validation studies including stress testing, and repeatability testing to ensure the safety and effectiveness of the device" and that "results concluded the device was acceptable for use."

    However, based on the context of the device and its intended use, we can infer general performance areas. Since no specific acceptance criteria are given, the "Reported Device Performance" column will reflect the general conclusion from the document.

    Criteria Area (Inferred)Acceptance Criteria (Not Explicitly Stated)Reported Device Performance
    Safety and EffectivenessDevice is safe and effective for its intended use.Validation studies included stress testing and repeatability testing. Medical device design validation has been completed, encompassing testing and evaluation using previously acquired diagnostic images from HeartFlow-sponsored clinical trials. Results concluded the device was acceptable for use.
    Accuracy of FFRct CalculationFFRct calculations are accurate.The device computes FFRct, a coronary physiological simulation, from simulated pressure, velocity, and blood flow information obtained from a 3D computer model generated from static coronary CT images. No specific accuracy metrics are provided in this document, but implies accuracy through its intended use and general validation.
    Plaque Identification/CharacterizationPlaque identification and characterization is accurate.The device provides plaque identification and characterization, as well as anatomic data. Supported by comparison to the Autoplaque predicate device. Implies accuracy through intended use and general validation.
    Anatomic Data ExtractionAccurate extraction of anatomic data.The device extracts anatomic data to aid in risk assessment and functional evaluation. Implies accuracy through intended use and general validation.

    2. Sample Size for Test Set and Data Provenance

    • Sample Size for Test Set: The document states that testing and evaluation used "previously acquired diagnostic images received through HeartFlow sponsored clinical trials." However, a specific number for the test set sample size is not provided.
    • Data Provenance: The data was "previously acquired diagnostic images received through HeartFlow sponsored clinical trials." This suggests the data is retrospective (already acquired) and likely originates from various clinical trial sites, but specific countries are not mentioned.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    The document does not specify the number of experts used or their qualifications for establishing ground truth for the test set.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method (e.g., 2+1, 3+1).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    The provided text does not mention if an MRMC comparative effectiveness study was done, nor does it specify an effect size of human readers' improvement with AI vs. without AI assistance. The focus is on the device's standalone validation.

    6. Standalone (Algorithm Only) Performance

    Yes, a standalone (algorithm only) performance assessment was done. The document states:

    • "The software was designed, developed, tested and validated according to written procedures."
    • "Validation studies included stress testing, and repeatability testing to ensure the safety and effectiveness of the device."
    • "Medical device design validation has been completed. Medical device design included testing and evaluation using previously acquired diagnostic images received through HeartFlow sponsored clinical trials."
    • "Summaries of pre-clinical studies were reviewed as part of a prior predicate review (K161772, the original predicate of K182035/K190925/K203329). The results concluded the device was acceptable for use."

    This indicates that the device's performance was evaluated independently without human intervention during the "testing and evaluation" phase described, proving its standalone capabilities.

    7. Type of Ground Truth Used

    The document implies that the ground truth was established through clinical diagnosis and evaluation of the "previously acquired diagnostic images" from HeartFlow-sponsored clinical trials. While it doesn't explicitly state "expert consensus," this is the most likely method for establishing ground truth in clinical trials concerning coronary artery disease diagnoses from images. No mention of pathology or outcomes data as direct ground truth is made in this specific excerpt for the validation studies mentioned.

    8. Sample Size for the Training Set

    The document states: "The core technology remains unchanged from the primary predicate and continues to be trained using deep learning (AI and machine learning) since 2015, to incorporate learnings from the volumes of CT data and studies."

    However, a specific sample size for the training set is not provided. It only mentions "volumes of CT data and studies."

    9. How the Ground Truth for the Training Set Was Established

    The document implies that the ground truth for the training set was established through "learnings from the volumes of CT data and studies." Similar to the test set, this would likely involve expert interpretation and analysis of the CT data used for training the deep learning algorithms, reflecting accepted clinical diagnoses and findings within those studies. No further details are given.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203329

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-01-08

    (57 days)

    Product Code
    Regulation Number
    870.1415
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K161772, K182035

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HeartFlow Analysis is a coronary physiologic simulation software for the clinical quantitative and qualitative analysis of previously acquired Computed Tomography DICOM data for clinically stable symptomatic patients with coronary artery disease. It provides the calculations of FFR-cf a mathematically derived quantity, computed from simulated pressure, velocity and blood flow information obtained from a 3D computer model generated from static coronary CT images. The HeartFlow Analysis is intended to support the functional evaluation of coronary artery disease.

    The HeartFlow Analysis is provided to support qualified clinicians to aid in the evaluation and assessment of coronary arteries. The HeartFlow Analysis is intended to be used by qualified clinicians in conjunction with the patient's clinical history, symptoms, and other diagnostic tests, as well as the clinician's professional judgment.

    Device Description

    The HeartFlow Analysis is a coronary physiological simulation software developed for the clinical quantitative and qualitative analysis of CT DICOM data. It is a tool for the analysis of CT DICOM-compliant cardiac images and data, to assess the anatomy and function of the coronary arteries.

    The software displays the anatomy combined with function using graphics and text, including computed and derived quantities of blood flow, pressure and velocity, to aid the clinician in the assessment and treatment planning of coronary artery disease.

    The HeartFlow Analysis is performed on previously physician-acquired image data and is unrelated to acquisition equipment and clinical workstations.

    AI/ML Overview

    The provided document is a 510(k) summary for the HeartFlow Analysis (FFRct v3) device. It describes the device and claims substantial equivalence to a predicate device (HeartFlow FFRct v2.Planner, K190925). However, it does not contain the detailed acceptance criteria or the specific study results that directly prove the device meets said criteria for FFRct v3.

    The document states: "Summaries of pre-clinical studies were reviewed as part of a prior predicate review (K161772, the predicate of K182035 that is the predicate of K190925). The results concluded the device was acceptable for use." This refers to studies for previous versions of the device, not specifically for "FFRCT v3" and its new solver.

    The document claims: "Changes to flow and distribution calculations within the Gen3 solver do not raise new questions of safety and effectiveness." and "Results of all current and previously referenced testing conclude the device is acceptable for use." This implies that the new version is acceptable based on the previous version's validation studies and that the changes are not significant enough to warrant new clinical studies to prove effectiveness.

    Therefore, many of the requested details, such as specific acceptance criteria for FFRct v3, reported device performance data, sample sizes, ground truth establishment for a new study, and multi-reader multi-case study results for this specific version, are not provided in this document.

    Based only on the provided text, here's what can be extracted and what is missing or implied:

    1. Table of acceptance criteria and the reported device performance:

      Acceptance CriteriaReported Device Performance (for FFRct v3)
      Not explicitly stated for FFRct v3 in this document. The document refers to prior predicate review studies (K161772) that concluded the device was acceptable for use, implying prior acceptance metrics were met by older versions.The document states: "Changes to flow and distribution calculations within the Gen3 solver do not raise new questions of safety and effectiveness." and "Results of all current and previously referenced testing conclude the device is acceptable for use." This implies that the performance of FFRct v3, with its Gen3 solver, is considered equivalent to and meets the same acceptable standards as its predicates, without providing new quantitative performance metrics for FFRct v3 itself.
    2. Sample sizes used for the test set and the data provenance:

      • Test set sample size: Not specified for FFRct v3. The document refers to "previously acquired diagnostic images received through HeartFlow sponsored clinical trials" for earlier versions of the device.
      • Data provenance: Not specified for FFRct v3. For previous versions, it's implied the data came from "HeartFlow sponsored clinical trials," but country of origin or retrospective/prospective nature is not detailed for any specific version.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not specified for FFRct v3. This information might be detailed in the "pre-clinical studies" mentioned for the predicate devices (K161772), but those details are not included in this summary.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not specified for FFRct v3.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not specified for FFRct v3. The document mentions the device is intended "to support qualified clinicians to aid in the evaluation and assessment," but no MRMC study details or effect sizes are provided.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • The core claim of the HeartFlow Analysis (FFRct) is to "compute FFRCT, a mathematically derived quantity." This inherently describes a standalone algorithmic function. However, the document also states, "Clinician review and assessment of analysis prior to use as supplemental diagnostic aid." This suggests the standalone performance is part of a human-in-the-loop clinical workflow, but results for standalone performance are not explicitly detailed.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Not specified for FFRct v3. For the underlying FFRCT technology, the ground truth for FFR (Fractional Flow Reserve) is typically invasive coronary angiography with pressure wire measurements. This is only implied from the known context of FFRct validity, but not explicitly stated in this document for the validation of FFRct v3.
    8. The sample size for the training set:

      • Not specified for FFRct v3, or any previous versions addressed in this summary. The document mentions "Medical device design included testing and evaluation using previously acquired diagnostic images," which would likely include training data, but no specifics are given.
    9. How the ground truth for the training set was established:

      • Not specified for FFRct v3, or any previous versions discussed in this summary.

    In summary, the provided FDA 510(k) summary for HeartFlow Analysis (FFRct v3) primarily focuses on establishing substantial equivalence to a predicate device based on technological characteristics and intended use. It does not provide specific new study data, acceptance criteria, sample sizes, or ground truth details for the current FFRct v3 version, instead referring to earlier predicate device reviews where such studies were reportedly conducted. The implication is that the changes in FFRct v3 (specifically the Gen3 solver) are not significant enough to require new comprehensive clinical validation studies to prove safety and effectiveness beyond what was already established for its predecessors.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190925

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-08-15

    (128 days)

    Product Code
    Regulation Number
    870.1415
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K161772

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HeartFlow Analysis is a coronary physiologic simulation software for the clinical quantitative and ysis of previously acquired Computed Tomography DICOM data for clinically stable symptomatic patients with coronary artery disease. It provides FFRCT, a mathematically derived quantity, computed from simulated pressure, velocity and blood flow information obtained from a 3D computer model generated from static coronary CT images. FFRCT analysis is intended to support the functional evaluation of coronary artery disease.

    The results of this analysis are provided to support qualified clinicians to aid in the evaluation and assessment of coronary arteries. The results of HeartFlow FFRCT are intended to be used by qualified clinicians in conjunction with the patient's clinical history, symptoms, and other diagnostic tests, as well as the clinician's professional judgment.

    Device Description

    The HeartFlow Analysis is a coronary physiological simulation software developed for the clinical quantitative and qualitative analysis of CT DICOM data. It is a tool for the analysis of CT DICOMcompliant cardiac images and data, to assess the anatomy and function of the coronary arteries.

    The software displays the anatomy combined with function using graphics and text, including computed and derived quantities of blood flow, pressure and velocity, to aid the clinician in the assessment (diagnosis and treatment planning) of coronary artery disease.

    HeartFlow FFR analyses are performed on previously physician-acquired image data and are unrelated to acquisition equipment and clinical workstations.

    The new planner feature is also software, and uses as input the anatomic FFRct model, and an idealized model generated from the FFRct model. Just as a CFD solution is run on the anatomic FFRct model to get the color-coded FFRct Analysis, the planner feature runs a reduced order CFD solution for user selected combinations of the anatomic FFRct model and the idealized model.

    AI/ML Overview

    Here's a breakdown of the HeartFlow Analysis device's acceptance criteria and the study information, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided 510(k) summary does not explicitly list quantitative acceptance criteria for the device's performance. Instead, it focuses on demonstrating that the device's new "Planner" feature, which uses a reduced-order CFD solution for user-selected combinations of anatomic and idealized models, produces FFRct results equivalent to those achieved with the existing FFRct solver for a given modified anatomy.

    Therefore, accepting this interpretation, here's a table based on the information provided:

    Acceptance CriterionReported Device Performance
    For a given modified anatomy (combinations of anatomic and idealized model), the FFRct results achieved with the Delta-solver are equivalent to those achieved with the FFRct solver.The design verification test (DVT) ensured this equivalency. The results of all current and previously referenced testing conclude the device is acceptable for use.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample size used for the test set specifically for the Planner feature. It mentions "testing of various FFRct model modifications to represent a variety of vessel and lesion morphologies and their idealized state."

    Regarding data provenance, the document states: "Summaries of pre-clinical studies were reviewed as part of a prior predicate review (K161772, the predicate to K182035). The results concluded the device was acceptable for use. The applicability of the clinical data is not effected by the changes proposed under the predicate K182035 nor this 510(k). No additional pre-clinical data is being provided with this submission." This indicates that previous clinical data was used, but details on country of origin or retrospective/prospective nature are not provided for the current submission's evaluation.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    The document does not provide information on the number of experts or their qualifications used to establish ground truth for the test set of the Planner feature. The assessment appears to be a technical comparison between the two CFD solvers rather than a clinical ground truth adjudicated by experts.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for the test set, as the evaluation focuses on the technical equivalency of two computational solvers rather than subjective human assessment.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No MRMC comparative effectiveness study is mentioned for the current submission's evaluation of the Planner feature. The text highlights that "No additional pre-clinical data is being provided with this submission," suggesting the focus is on the technical changes rather than a new clinical trial.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone evaluation was performed. The "design verification test (DVT)" focused on comparing the FFRct results produced by the Delta-solver (used by the "Planner" feature) with those from the FFRct solver (the existing algorithm) for a given modified anatomy. This is an algorithm-only comparison.

    7. The Type of Ground Truth Used

    For the evaluation of the Planner feature, the "ground truth" was effectively the output of the established FFRct solver for a modified anatomy. The goal was to demonstrate that the new "Delta-solver" within the Planner feature could produce equivalent results to this existing, validated computational model.

    8. The Sample Size for the Training Set

    The document does not provide information about the sample size of the training set.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide information about how the ground truth for the training set was established. The submission focuses on the new feature of the HeartFlow Analysis software and references previous predicate reviews for the underlying FFRct technology.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1