510(k) Data Aggregation

    K Number
    K251322

    Validate with FDA (Live)

    Date Cleared
    2025-07-25

    (87 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K231966 LOGIQ E10, K223832 Vivid S70N, K161588 Vscan Extend, K220940 EchoPAC Software Only/EchoPAC Plug-in

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Venue, Venue Go, Venue Fit and Venue Sprint are general purpose diagnostic ultrasound systems for use by qualified and trained healthcare professionals or practitioners that are legally authorized or licensed by law in the country, state or other local municipality in which he or she practices, for ultrasound imaging, measurement, display and analysis of the human body and fluid. The users may or may not be working under supervision or authority of a physician. Users may also include medical students working under the supervision or authority of a physician during their education / training.

    Venue, Venue Go and Venue Fit are intended to be used in a hospital or medical clinic. Venue, Venue Go and Venue Fit clinical applications include: abdominal (GYN and Urology), thoracic/pleural, ophthalmic, Fetal/OB, Small Organ (including breast, testes, thyroid), Vascular/Peripheral vascular, neonatal and adult cephalic, pediatric, musculoskeletal (conventional and superficial), cardiac (adults and pediatric), Transrectal, Transvaginal, Transesophageal, Intraoperative (vascular) and interventional guidance (includes tissue biopsy, fluid drainage, vascular and non-vascular access). Modes of operation include: B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse and Combined modes: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD, B/CWD, B/Color/CWD.

    The Venue Sprint is intended to be used in a hospital, medical clinic, home environment and road/air ambulance. Venue Sprint clinical applications include: abdominal (GYN and Urology), thoracic/pleural, ophthalmic, Fetal/OB, Small Organ (including breast, testes, thyroid), Vascular/Peripheral vascular, neonatal and adult cephalic, pediatric, musculoskeletal (conventional and superficial), cardiac (adults and pediatric, 40 kg and above) and interventional guidance (includes free hand tissue biopsy, fluid drainage, vascular and non-vascular access). Modes of operation include: B, M, PW Doppler, Color Doppler and Harmonic Imaging.

    Device Description

    Venue, Venue Go, Venue Fit and Venue Sprint are general-purpose diagnostic ultrasound systems intended for use by qualified and trained healthcare professionals to evaluate the body by ultrasound imaging and fluid flow analysis.

    The systems utilize a variety of linear, convex, and phased array transducers which provide high imaging capability, supporting all standard acquisition modes.

    The systems have a small footprint that easily fits into tight spaces and positioned to accommodate the sometimes-awkward work settings of the point of care user.

    The Venue is a mobile system, the Venue Go and Venue Fit are compact, portable systems that can be hand carried using an integrated handle, placed on a horizontal surface, attached to a mobile cart or mounted on the wall. Venue, Venue Go and Venue Fit have a high-resolution color LCD monitor, with a simple, multi-touch user interface that makes the systems intuitive.

    The Venue Sprint is used together with the Vscan Air probes and provides the user interface for control of the probes and the needed software functionality for analysis of the ultrasound images and saving/storage of the related images and videos.

    The Venue, Venue Go, Venue Fit and Venue Sprint systems can be powered through an electrical wall outlet for long term use or from an internal battery for a short time with full functionality and scanning. A barcode reader and RFID scanner are available as additional input devices. The systems meet DICOM requirements to support users image storage and archiving needs and allows for output to printing devices.

    The Venue, Venue Go and Venue Fit systems are capable of displaying the patient's ECG trace synchronized to the scanned image. This allows the user to view an image from a specific time of the ECG signal which is used as an input for gating during scanning. The ECG signal can be input directly from the patient or as an output from an ECG monitoring device. ECG information is not intended for monitoring or diagnosis. Compatible biopsy kits can be used for needle-guidance procedures.

    AI/ML Overview

    The provided document, a 510(k) Clearance Letter and Submission Summary, primarily focuses on the substantial equivalence of the GE Healthcare Venue series of diagnostic ultrasound systems to previously cleared predicate devices. It specifically details the "Auto Bladder Volume (ABV)" feature as an AI-powered component and provides a summary of its testing.

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based only on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance (for Auto Bladder Volume - ABV)

    Acceptance CriteriaReported Device Performance
    At least 90% success rate in automatic caliper placement for bladder volume measurements when bladder wall is entirely visualized.Automatic caliper placement success rate: 95.09% (with a 95% confidence level)
    Performance demonstrated consistent across key subgroups including subjects with known BMI (healthy weight, obese, overweight).Healthy weight (18.5-24.9): 95.64%Obese (25-29.9): 95.59%Overweight (Over 30): 92.6%

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set (Verification Dataset) Sample Size: 1874 images from 101 individuals.
    • Data Provenance:
      • Country of Origin: USA and Israel.
      • Retrospective or Prospective: Not explicitly stated as either retrospective or prospective. However, the description of "data collected from several different Console variants" for training and verification suggests pre-existing data, which often leans towards a retrospective collection.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of Experts: Not explicitly stated. The document refers to "annotators" who performed manual annotation.
    • Qualifications of Experts: Not explicitly stated. The annotators are described as performing "manual annotation," implying they are skilled in this task, but specific qualifications (e.g., radiologists, sonographers, years of experience) are not provided.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not explicitly stated. The document mentions "annotators performed manual annotation," but does not detail if multiple annotators were used for each case or any specific adjudication process (e.g., 2+1, 3+1 consensus).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? No. The document states: "The subjects of this premarket submission, Venue, Venue Go, Venue Fit and Venue Sprint, did not require clinical studies to support substantial equivalence." The testing described for ABV is a standalone algorithm performance validation against established ground truth, not a comparative human-AI study.
    • Effect Size of Human Readers Improvement: Not applicable, as no MRMC study was performed.

    6. Standalone (Algorithm Only) Performance Study

    • Was a standalone study done? Yes. The "AI Summary of Testing" section describes a study for the Auto Bladder Volume (ABV) feature, which assesses the algorithm's "automatic caliper placement success rate" against manually established ground truth. This is a standalone performance evaluation of the algorithm.

    7. Type of Ground Truth Used (for ABV Test Set)

    • Ground Truth Type: Expert consensus/manual annotation. The document states: "Ground truth annotations of the verification dataset were obtained as follows: In all Training/Validation and Verification datasets, annotators performed manual annotation on images converted from DICOM files." They identified "landmarks, which represent the bladder edges," corresponding to standard measurement locations.

    8. Sample Size for the Training Set (for ABV)

    • Training Set Sample Size: Total dataset included 8,392 images from 496 individuals. Of these, 1,874 were used for the verification dataset, and "the rest" were used for training/validation. This implies the training/validation set would be 8392 - 1874 = 6518 images from the remaining individuals not included in the verification set.

    9. How the Ground Truth for the Training Set Was Established (for ABV)

    • Ground Truth Establishment: Similar to the verification dataset, "annotators performed manual annotation on images converted from DICOM files" for both Training/Validation and Verification datasets. They chose "4-6 images that represent different bladder volume status" for each individual and annotated "4 different landmarks" per view (transverse and longitudinal) representing bladder edges.
    Ask a Question

    Ask a specific question about this device

    K Number
    K250087

    Validate with FDA (Live)

    Device Name
    Vscan Air
    Date Cleared
    2025-05-01

    (107 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    Vscan Extend (K161588)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Vscan Air is a battery-operated software-based general-purpose ultrasound imaging system for use by qualified and trained healthcare professionals or practitioners that are legally authorized or licensed by law in the country, state or other local municipality in which he or she practices. The users may or may not be working under supervision or authority of a physician. Users may also include medical students working under the supervision or authority of a physician during their education / training. The device is enabling visualization and measurement of anatomical structures and fluid including blood flow.

    Vscan Air's pocket-sized portability and simplified user interface enables integration into training sessions and examinations in professional healthcare facilities (ex. Hospital, clinic, medical office), home environment, road/air ambulance and other environments as described in the user manual. The information can be used for basic/focused assessments and adjunctively with other medical data for clinical diagnosis purposes during routine, periodic follow-up, and triage.

    Vscan Air supports Black/ white (B-mode), Color flow (Color doppler), Pulsed wave Doppler mode, M-mode, combined (B + Color Doppler) and Harmonic Imaging modes with curved, linear and sector array transducers.

    With the curved array transducer of the dual headed probe solution, the specific clinical applications and exam types include: abdominal, fetal/obstetrics, gynecological, urology, thoracic/lung, cardiac (adult and pediatric, 40 kg and above), vascular/peripheral vascular, musculoskeletal (conventional), pediatrics, interventional guidance (includes free hand needle/catheter placement, fluid drainage, nerve block and biopsy).

    With the linear array transducer of the dual headed probe solution, the specific clinical applications and exam types include: vascular/peripheral vascular, musculoskeletal (conventional and superficial), small organs, thoracic/lung, ophthalmic, pediatrics, neonatal cephalic, interventional guidance (includes free hand needle/catheter placement, fluid drainage, nerve block, vascular access and biopsy).

    With the sector array transducer of the dual headed probe solution, the specific clinical applications and exam types include: cardiac (adult and pediatric, 40 kg and above), abdominal, fetal/obstetrics, gynecological, urology, thoracic/ lung, pediatrics, adult cephalic, interventional guidance (includes free hand needle/catheter placement, fluid drainage, nerve block and biopsy).

    Device Description

    Vscan Air™ is a battery-operated general-purpose diagnostic ultrasound imaging system for use by qualified and trained healthcare professionals. It enables ultrasound imaging guidance, visualization and measurement of anatomical structures and fluid.

    Vscan Air consists of an app which can be installed on Android™ or iOS devices, and 2 probes which use wireless technology for communication.

    Its pocket-sized portability and simplified user interface enable integration into training sessions and examinations in professional healthcare facilities (ex. Hospital, clinic, medical office), home environment, road/air ambulance and in other environments. The information can be used for basic/focused assessments and adjunctively with other medical data for clinical diagnosis purposes during routine, periodic follow-up, and triage assessments for adult, pediatric and neonatal patients. Vscan Air can also be useful for interventional guidance.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Vscan Air's Auto Bladder Volume feature, extracted from the provided FDA 510(k) clearance letter:

    Acceptance Criteria and Device Performance for Auto Bladder Volume

    Acceptance CriteriaReported Device Performance
    At least 90% success rate in automatic caliper placement for bladder volume measurements when bladder wall is entirely visualized.Automatic caliper placement success rate: 92.24% with a 90% confidence level. Consistent performance across key subgroups, e.g., BMI Overweight (>25): 92%, BMI Normal (<25): 95%.

    Study Details for Auto Bladder Volume (Test Set)

    1. Sample size used for the test set and data provenance:

      • Sample Size: 1,817 images from 142 individuals. Each individual was scanned in two views (Transverse and Longitudinal).
      • Data Provenance: The dataset included images from individuals of various ethnicities/countries, including USA, Germany, UK, Japan, and India. The study used retrospective data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document does not specify the exact number of experts, but states that the verification dataset was assessed by "experts for accuracy." It does not provide their specific qualifications (e.g., years of experience as radiologists).
    3. Adjudication method for the test set:

      • The document describes "annotators" performing manual annotation on images to establish ground truth. It does not explicitly mention a formal adjudication method like "2+1" or "3+1" for resolving disagreements among multiple annotators.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

      • No, an MRMC comparative effectiveness study comparing human readers with and without AI assistance was not done. This study solely focused on the standalone performance of the Auto Bladder Volume feature.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, a standalone performance study was done for the Auto Bladder Volume feature. The performance reported is that of the algorithm's automatic caliper placement.
    6. The type of ground truth used:

      • Expert Consensus/Manual Annotation. Ground truth annotations were obtained through manual placement of landmarks (bladder edges) by annotators on images, representing where measurement calipers would be placed.
    7. The sample size for the training set:

      • The total dataset included 4,014 images from 301 individuals. Since 1,817 images from 142 individuals were used for verification, the remaining 2,197 images from 159 individuals were used for training/validation (tuning).
    8. How the ground truth for the training set was established:

      • Similar to the verification dataset, "annotators performed manual annotation on images converted from vscanet files." These annotations involved identifying 4-6 images representing different bladder volume statuses, and on each view (transverse and longitudinal), marking 4 distinct landmarks representing the bladder edges.
    Ask a Question

    Ask a specific question about this device