Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K210259

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-01-20

    (353 days)

    Product Code
    Regulation Number
    872.3275
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Parkell Self-Adhesive Cement is intended for the cementation of indirect restoratives including ceramic, zirconia, composite, and metal-based inlays, onlays, crowns, bridges, and posts.

    Device Description

    Parkell Self-Adhesive Cement is a two-component, dual-cured, self-adhesive resin cement. It is designed for luting of indirect restorations (inlays, onlays, crowns, bridges, and endodontic posts) made of metal alloys, composite resins, porcelain, lithium disilicate or zirconia. The material also exhibits high bond strength and is radiopaque to allow visibility of excess in sub-gingival areas. Parkell Self-Adhesive Cement will cure in 10 seconds with a dental curing light emitting blue light at 430 - 480 nm and a minimum intensity of 600 mW/cm². The product will self-cure in 3-4 minutes at mouth temperature. Parkell Self-Adhesive Cement will be provided in 5 mL auto-mix syringes. The device comprises catalyst and base components, which are combined just prior to use via a static (1:1) mixing tip which is affixed to the dual chamber syringe.

    AI/ML Overview

    The provided document is a 510(k) summary for a dental cement (Parkell Self-Adhesive Cement). It aims to demonstrate substantial equivalence to a predicate device. This type of submission relies heavily on bench testing and biocompatibility studies, rather than clinical performance (human trials) for its primary evidence.

    Therefore, the concept of "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context pertains to the non-clinical performance and physical properties of the dental cement, as well as its biocompatibility, when compared to the established standards and the predicate device.

    Here's a breakdown of the requested information based on the document:

    1. Table of Acceptance Criteria and Reported Device Performance & 6. Standalone Performance:

    The document implicitly defines acceptance criteria by comparing the subject device's performance to the predicate device's performance, or to generally accepted values for dental cements (e.g., "Pass/Fail Criteria"). The performance listed is for the standalone (algorithm only without human-in-the-loop performance) of the material itself.

    PropertyPredicate Device Performance / Acceptance Criteria (Pass/Fail Criteria)Parkell Self-Adhesive Cement Performance
    Physical Properties
    Film Thickness (Microns)≤ 50 microns≤ 50 microns
    Working time (seconds) @ 35 C≤ 60 seconds≤ 60 seconds
    Setting time (minutes) @ 35 C≤ 10 minutes≤ 10 minutes
    Flexural strength LC (MPa)≥ 50 MPa≥ 50 MPa
    Flexural Strength SC (MPa)≥ 50 MPa≥ 50 MPa
    Compressive Strength LC (MPa)≥ 200 MPa≥ 200 MPa
    Water Sorption (ug/mm³)≤ 40 ug/mm³≤ 40 ug/mm³
    Water Solubility (ug/mm³)≤ 7.5 u/mm³≤ 7.5 u/mm³
    Radiopacity (% Al)≥ 100 %Al≥ 100 %Al
    Depth of Cure (mm)≥ 1mm≥ 1mm
    Diametral Tensile Strength (MPa)≥ 30 MPa≥ 30 MPa
    Barcol Hardness after 20 minutes @35c (MPa)> 50 MPa> 50 MPa
    Barcol Hardness after 20 secs Light Cure each surface (MPa)> 60 MPa> 60 MPa
    Bond Strength (MPa) to Substrates (LC = Light Cure, SC = Self-Cure, Treatment indicated where applicable)
    Dentin LC (None)≥ 6 MPa≥ 6 MPa
    Dentin LC (PBA)≥ 10 MPa≥ 10 MPa
    Dentin SC (None)≥ 3 MPa≥ 3 MPa
    Enamel LC (None)≥ 10 MPa≥ 10 MPa
    Enamel LC (H3PO4)≥ 10 MPa≥ 10 MPa
    Enamel SC (None)≥ 3 MPa≥ 3 MPa
    Lithium Disilicate LC (Silane)≥ 5 MPa≥ 5 MPa
    Lithium Disilicate LC (HF)N/A (Predicate did not have this indication)≥ 10 MPa
    Zirconia LC (SB)≥ 10 MPa≥ 10 MPa
    Zirconia LC (PBA)N/A (Predicate did not have this indication)≥ 10 MPa
    Zirconia LC (HF)N/A (Predicate did not have this indication)≥ 10 MPa
    Zirconia SC (SB)≥ 5 MPa≥ 5 MPa
    Porcelain LC (SB)≥ 10 MPa≥ 10 MPa
    Porcelain LC (Silane)N/A (Predicate did not have this indication)≥ 10 MPa
    Porcelain LC (HF)N/A (Predicate did not have this indication)≥ 10 MPa
    Porcelain SC (HF)N/A (Predicate did not have this indication)≥ 5 MPa
    Palladium LC (SB)≥ 10 MPa≥ 10 MPa
    Titanium LC (SB)≥ 10 MPa≥ 10 MPa
    Titanium SC (SB)≥ 5 MPa≥ 5 MPa
    Cobalt LC (SB)≥ 10 MPa≥ 10 MPa
    Cobalt SC (SB)≥ 5 MPa≥ 5 MPa

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: The document does not explicitly state the numerical sample size (e.g., number of specimens tested for each property). It only indicates that "Performance and Physical Properties" tests were performed.
    • Data Provenance: The data is generated from non-clinical (bench) testing. The country of origin of the data is not specified, but the submission is to the U.S. FDA by a company based in New York. The testing would be considered prospective in the sense that it was conducted specifically for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • Number of Experts: This is not applicable to this type of submission. The "ground truth" for the performance and physical properties of the dental cement is established through standardized laboratory test methods (e.g., ISO standards like ISO 4049, ISO 11405, etc.) and direct object measurement, not expert consensus or interpretation.
    • Qualifications of Experts: Not applicable, as expert interpretation is not the primary means of establishing "ground truth" for material properties.

    4. Adjudication Method for the Test Set:

    • Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or image-based diagnostic studies involving human readers/interpreters to resolve discrepancies. For bench testing of material properties, the results are quantitative measurements against defined criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is a submission for a medical device (dental cement), not an AI-assisted diagnostic or imaging device. Therefore, no MRMC study, human readers, or AI assistance is involved.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

    • Quantitative measurements against established international standards and defined criteria. For physical properties, the ground truth is derived from the reproducible results of standardized laboratory tests. For biocompatibility, the ground truth is assessed against the requirements of ISO 10993 series of standards.

    8. The Sample Size for the Training Set:

    • Not applicable. This is a medical device submission based on bench testing of material properties, not an AI/machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set was Established:

    • Not applicable, as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K210504

    Validate with FDA (Live)

    Date Cleared
    2021-06-11

    (109 days)

    Product Code
    Regulation Number
    872.3275
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K161042, K153476

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cementation of ceramic and composite inlays, onlays and laminate veneers with less than 2mm thickness

    Device Description

    PANAVIA Veneer LC is a light-cure adhesive resin cement system. It consists of the light-cure cement paste (PANAVIA Veneer LC paste), CLEARFIL CERAMIC PRIMER PLUS, K-ETCHANT Syringe, and PANAVIA V5 Try-in Paste. CLEARFIL Universal Bond Quick or PANAVIA V5 Tooth Primer can be chosen for tooth treatment.

    The cement paste is a light-cure, resin based material which provides color stability and has radiopacity equal to or greater than 1mm aluminum. It is indicated for inlays, onlays and laminate veneer restorations made of ceramic and composite resin. It is supplied in Kuraray's ergonomic syringe and dispensed via an angled applicator tip (16G) into an inlay cavity or onto an onlay or a laminate veneer. It is available in 4 shades; Universal (A2), Clear, Brown (A4) and White. It is classified as a Type 2 and Class 2 (Group 1) material by ISO 4049. Shade adaptation is available to check by PANAVIA V5 Try-in Paste before cementation. This is the new registration application for the subject device and there have not been any prior submissions regarding the subject device.

    We are seeking clearance for PANAVIA Veneer LC paste in this submission.

    Concerning Try-in Paste which is included as components of the PANAVIA V5, CLEARFIL Universal Bond Quick, CLEARFIL CERAMIC PRIMER PLUS and K-ETCHANT Syringe, the specification of those 4 components were omitted for the reasons set forth below.

    We already have had 510(k) clearance of Try-in Paste of PANAVIA V5(510(k) Number: 150704), CLEARFIL Universal Bond Quick (510(k) Number: 161042), CLEARFIL CERAMIC PRIMER PLUS (510(k) Number: 150703) and K-ETCHANT Syringe (510(k) Number: 133078).

    AI/ML Overview

    This document describes the marketing authorization for the PANAVIA Veneer LC dental cement. Here's a breakdown of the acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    SectionRequirementPANAVIA Veneer LC (Subject device)PANAVIA V5 (Primary predicate device)CLEARFIL MAJESTY ES Flow (Predicate device)
    5.2.2 Film thickness, luting materials< 50 μm.COMPLIESCOMPLIESCOMPLIES
    5.2.7 Sensitivity to ambient lightThe material shall remain physically homogeneous.COMPLIESCOMPLIESCOMPLIES
    5.2.8 Depth of cureShade type 1 and Predicate device: Not less than 1.5mm. Shade type 2: Not less than 1.0mm.COMPLIESCOMPLIESCOMPLIES
    5.2.9 Flexural strengthEqual to or greater than 50 MPa.COMPLIESCOMPLIESCOMPLIES
    5.2.10 Water sorption and solubility≤ 40 μg/mm³.COMPLIESCOMPLIESCOMPLIES
    ≤ 7.5 μg/mm³.COMPLIESCOMPLIESCOMPLIES
    5.4 Color stability after irradiation and water sorptionNo more than a slight change in color shall be observed for luting materials.COMPLIESCOMPLIESCOMPLIES
    5.5 Radio-opacityEqual to or greater than that of the same thickness of aluminium and no less than 0.5 mm of any value claimed by the manufacturer.COMPLIESCOMPLIESCOMPLIES

    Bond Strength Test

    Adherend surface (Material composition)CriteriaSubject devicePrimary predicate device (PANAVIA V5)
    Tooth dentinIn-house standardCOMPLIESCOMPLIES
    Tooth enamelCOMPLIESCOMPLIES
    CeramicCOMPLIESCOMPLIES
    Composite resinCOMPLIESCOMPLIES
    Hybrid ceramicCOMPLIESCOMPLIES

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes used for each individual test within the comparative study according to ISO 4049: 2009 or the bond strength test. The data provenance is implied to be from internal studies conducted by Kuraray Noritake Dental Inc., as indicated by "In-house standard" for the bond strength criteria. The country of origin of the device manufacturer is Japan. The studies are assumed to be prospective in nature, conducted specifically for the purpose of this 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This document does not describe the use of human experts or ground truth establishment in the context of diagnostic interpretation or clinical decision-making. The "ground truth" here refers to objective physical and mechanical properties measured and compared against established ISO standards and in-house standards. Therefore, information regarding experts and their qualifications for establishing ground truth is not applicable to this type of device and study.

    4. Adjudication Method for the Test Set

    Not applicable. The studies involve objective measurements of material properties against predefined standards, not subjective interpretations requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This device is a dental cement, and its performance is evaluated based on physical and mechanical properties, not by human readers interpreting medical images or data. Therefore, an MRMC study and effect size of AI assistance are not relevant.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. This device is a dental cement, not an algorithm or AI-powered system. The studies described evaluate the intrinsic properties of the material.

    7. Type of Ground Truth Used

    The ground truth used for the subject device's performance evaluation against acceptance criteria is based on:

    • International Standards: Specifically, ISO 4049: 2009 (Dentistry - Polymer-based restorative and materials).
    • In-house Standards: For the bond strength test.
    • Comparative Data: Performance of the predicate devices (PANAVIA V5 and CLEARFIL MAJESTY ES Flow) serve as a comparative ground truth to demonstrate substantial equivalence.

    8. Sample Size for the Training Set

    Not applicable. This device is a dental cement, not a machine learning model. Therefore, there is no "training set" in the context of artificial intelligence.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    DEN180005

    Validate with FDA (Live)

    Device Name
    OsteoDetect
    Date Cleared
    2018-05-24

    (108 days)

    Product Code
    Regulation Number
    892.2090
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K142823,K161042,K180800

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OsteoDetect analyzes wrist radiographs using machine learning techniques to identify and highlight distal radius fractures during the review of posterior-anterior (PA) and lateral (LAT) radiographs of adult wrists.

    Device Description

    OsteoDetect is a software device designed to assist clinicians in detecting distal radius fractures during the review of posterior-anterior (PA) and lateral (LAT) radiographs of adult wrists. The software uses deep learning techniques to analyze wrist radiographs (PA and LAT views) for distal radius fracture in adult patients.

    AI/ML Overview

    1. Table of Acceptance Criteria and Reported Device Performance

    Standalone Performance

    Performance MetricAcceptance Criteria (Implicit)Reported Device Performance (Estimate)95% Confidence Interval
    AUC of ROCHigh0.965(0.953, 0.976)
    SensitivityHigh0.921(0.886, 0.946)
    SpecificityHigh0.902(0.877, 0.922)
    PPVHigh0.813(0.769, 0.850)
    NPVHigh0.961(0.943, 0.973)
    Localization Accuracy (average pixel distance)Small33.52 pixelsNot provided for average distance itself, but standard deviation of 30.03 pixels.
    Generalizability (AUC for all subgroups)High≥ 0.926 (lowest subgroup - post-surgical radiographs)Not explicitly provided for all, but individual subgroup CIs available in text.

    MRMC (Reader Study) Performance - Aided vs. Unaided Reads

    Performance MetricAcceptance Criteria (Implicit: Superiority of Aided)Reported Device Performance (OD-Aided)Reported Device Performance (OD-Unaided)95% Confidence Interval (OD-Aided)95% Confidence Interval (OD-Unaided)p-value for difference
    AUC of ROCAUC_aided - AUC_unaided > 00.8890.840Not explicitly given for AUCs themselves, but difference CI: (0.019, 0.080)Not explicitly given for AUCs themselves, but difference CI: (0.019, 0.080)0.0056
    SensitivitySuperior Aided0.8030.747(0.785, 0.819)(0.728, 0.765)Not explicitly given for individual metrics, but non-overlapping CIs imply significance.
    SpecificitySuperior Aided0.9140.889(0.903, 0.924)(0.876, 0.900)Not explicitly given for individual metrics, but non-overlapping CIs imply significance.
    PPVSuperior Aided0.8830.844(0.868, 0.896)(0.826, 0.859)Not explicitly given for individual metrics, but non-overlapping CIs imply significance.
    NPVSuperior Aided0.8530.814(0.839, 0.865)(0.800, 0.828)Not explicitly given for individual metrics, but non-overlapping CIs imply significance.

    2. Sample Size and Data Provenance for Test Set

    Standalone Performance Test Set:

    • Sample Size: 1000 images (500 PA, 500 LAT)
    • Data Provenance: Retrospective. Randomly sampled from an existing validation database of consecutively collected images from patients receiving wrist radiographs at the (b) (4) from November 1, 2016 to April 30, 2017. The study population included images from the US.

    MRMC (Reader Study) Test Set:

    • Sample Size: 200 cases.
    • Data Provenance: Retrospective. Randomly sampled from the same validation database used for the standalone performance study. The data includes cases from the US.

    3. Number of Experts and Qualifications for Ground Truth

    Standalone Performance Test Set and MRMC (Reader Study) Test Set:

    • Number of Experts: Three.
    • Qualifications: U.S. board-certified orthopedic hand surgeons.

    4. Adjudication Method for Test Set

    Standalone Performance Test Set:

    • Adjudication Method (Binary Fracture Presence/Absence): Majority opinion of at least 2 of the 3 clinicians.
    • Adjudication Method (Localization - Bounding Box): The union of the bounding box of each clinician identifying the fracture.

    MRMC (Reader Study) Test Set:

    • Adjudication Method: Majority opinion of three U.S. board-certified orthopedic hand surgeons. (Note: this was defined on a per-case basis, considering PA, LAT, and oblique images if available).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? Yes.
    • Effect Size (Improvement of Human Readers with AI vs. without AI assistance):
      • The least squares mean difference between the AUC for OsteoDetect-aided and OsteoDetect-unaided reads is 0.049 (95% CI, (0.019, 0.080)). This indicates a statistically significant improvement in diagnostic accuracy (AUC) of 4.9 percentage points when readers were aided by OsteoDetect.
      • Sensitivity: Improved from 0.747 (unaided) to 0.803 (aided), an improvement of 0.056.
      • Specificity: Improved from 0.889 (unaided) to 0.914 (aided), an improvement of 0.025.

    6. Standalone (Algorithm Only) Performance Study

    • Was a standalone study done? Yes.

    7. Type of Ground Truth Used

    Standalone Performance Test Set:

    • Type of Ground Truth: Expert consensus (majority opinion of three U.S. board-certified orthopedic hand surgeons).

    MRMC (Reader Study) Test Set:

    • Type of Ground Truth: Expert consensus (majority opinion of three U.S. board-certified orthopedic hand surgeons).

    8. Sample Size for Training Set

    The document does not explicitly state the sample size for the training set. It mentions "randomly withheld subset of the model's training data" for setting the operating point, implying a training set existed, but its size is not provided.

    9. How Ground Truth for Training Set Was Established

    The document does not explicitly state how the ground truth for the training set was established. It only refers to a "randomly withheld subset of the model's training data" during the operating point setting.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1