Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K163607

    Validate with FDA (Live)

    Date Cleared
    2017-12-15

    (359 days)

    Product Code
    Regulation Number
    870.2340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K160840, K160876, K102830

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ECG Acquisition Systems is intended to acquire a resting, diagnostic 12-lead ECG for display and subsequent upload to vhECG Pro, which installed in iPad or iPhone. This enables clinicians, or trained care personnel who are acting on the orders of a licensed physician, to acquire, process, display, store, and print diagnostic 12-lead ECGs.

    The ECG Acquisition Systems is for use on adult and pediatic populations, diseased or non-diseased and is not intended for use on neonatal (birth to 28 days) or infants (29 days up to 2 years). The device is not for use as a vital signs physiological monitor.

    The subject device is intended to be used for clinical use rather than home use. It can be used in a professional healthcare facility, or a clinical environment such as clinics. It cannot be used in transport environment, such as ambulance.

    The ECG Acquisition Systems provides uninterpreted 12-lead ECG data and is not to be a sole means of diagnosis.

    Device Description

    The ECG Acquisition Systems is a hand-held, battery powered, 12-lead resting electrocardiograph system with vhECG Pro software. The ECG Acquisition Systems consists of ECG Acquisition Recorder, Patient Cable and vhECG Pro software. ECG data is transferred via Bluetooth® 4.0 Technology and displayed on mobile screen in real-time before, during and after acquisition. With vhECG Pro, resting ECGs can be recorded, viewed and printed in various formats. Besides, ECGs can also be uploaded to vhCloud ( cloud platform) by vhECG Pro, they can be reviewed by other medical staff authorized via its iPhone or iPad in which vhCloud is installed. The device is intended to provide reference for medical diagnosis, not intended for a replacement of diagnosis of clinicians.

    AI/ML Overview

    The provided text is a 510(k) Summary for the "ECG Acquisition Systems" device. It describes the device, its intended use, comparison to a predicate device, and results of nonclinical tests. However, it explicitly states that clinical data is not applicable for this submission and does not contain any information about a study proving the device meets specific acceptance criteria based on clinical performance metrics (like sensitivity, specificity, or reader improvement with AI).

    Therefore, I cannot provide the requested information regarding acceptance criteria and a study that proves the device meets those criteria, as no such clinical performance study is described in the provided document. The document focuses on demonstrating substantial equivalence through nonclinical testing (electrical safety, EMC, QoS, wireless coexistence, ECG performance, patient cable performance, and software verification/validation) and comparison of technological characteristics to a predicate device, not on diagnostic accuracy or efficacy studies involving human readers or AI.

    The response below reflects what could be extracted if such a study were present, but indicates the absence of this information in the provided text.


    This document outlines the substantial equivalence of the "ECG Acquisition Systems" to a predicate device for FDA 510(k) clearance. It specifies the device's technical characteristics and nonclinical test results. However, it explicitly states that "Clinical data is not applicable for this submission." Therefore, the provided text does not contain information about acceptance criteria for clinical performance (e.g., diagnostic accuracy, sensitivity, specificity, or human reader improvement with AI) or a study that proves the device meets such criteria through clinical evaluation.

    The document primarily relies on nonclinical testing and comparison to a predicate device to demonstrate substantial equivalence, rather than a clinical performance study.

    Here's what can be gathered, alongside the confirmation that the requested clinical performance study details are absent:

    1. Table of Acceptance Criteria and Reported Device Performance:

    As no clinical performance study is described, there are no acceptance criteria for metrics like sensitivity, specificity, or other diagnostic performance indicators. The "acceptance criteria" discussed in this document pertain to adherence to technical standards and nonclinical test results.

    CategoryAcceptance Criteria (Standard/Guidance)Reported Device Performance
    Electric SafetyIEC 60601-1: 2012Pass
    EMCIEC 60601-1-2: 2007; 47 CFR PART 15 Subpart C; ESTI EN (various)Pass
    QoS TestingGuidance for Industry and FDA Staff: Radio Frequency Wireless Technology in Medical DevicesPass
    Wireless Coexistence47 CFR PART 15 Subpart C; EN 300328 V1.8.1Pass
    ECG PerformanceIEC 60601-2-25Pass
    Patient Cable PerformanceAAMI ANSI EC53Pass
    Software V&V TestingGuidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"Pass

    2. Sample size used for the test set and the data provenance:
    No clinical test set data is described. The nonclinical tests refer to compliance with technical standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    Not applicable, as no clinical performance test set or ground truth establishment by experts is described.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
    Not applicable.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    No MRMC study was done, as the device provides "uninterpreted 12-lead ECG data and is not to be a sole means of diagnosis" and does not involve AI assistance for interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    Not applicable, as the device primarily acquires and displays raw ECG data, and does not provide automated interpretation (explicitly states: "The ECG Acquisition Systems provides uninterpreted 12-lead ECG data and is not to be a sole means of diagnosis.").

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
    Not applicable, as no clinical performance study requiring ground truth is described.

    8. The sample size for the training set:
    Not applicable, as the document does not describe a machine learning model or a training set.

    9. How the ground truth for the training set was established:
    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171943

    Validate with FDA (Live)

    Date Cleared
    2017-09-25

    (89 days)

    Product Code
    Regulation Number
    870.2340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K131503, K160876

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the 3-Channel Electrocardiograph is to acquire ECG signals from adult and pediatric patients (beginning at birth through 21 years of age) through body surface ECG electrodes. The electrocardiograph is intended to be used only in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the 3-Channel Electrocardiograph can help users to analyze and diagnose heart disease. However, the ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only.

    Device Description

    SE-3 Series (including SE-3, SE-300A, SE-300B, and SE-301) 3-channel electrocardiograph gathers ECG signals of 12 leads simultaneously. While all 12 leads are recorded and analyzed, printouts will only have 3 leads on a page, with 4 pages in series to output all 12 leads. It displays the operation menu, ECG parameters as well as electrocardiograms.

    AI/ML Overview

    This document describes a 510(k) premarket notification for Electrocardiographs Models SE-3, SE-300A, SE-300B, and SE-301 by Edan Instruments, Inc. It seeks to demonstrate substantial equivalence to previously cleared predicate devices.

    Here's an analysis of the acceptance criteria and supporting studies based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as quantitative targets in terms of diagnostic accuracy metrics. Instead, the document relies on demonstrating that the proposed devices (SE-3, SE-300A, SE-300B, and SE-301) perform comparably to the predicate devices (SE-3, SE-300A, SE-300B - K091513, and reference devices SE-601C - K131503, SE-12 Series - K160876). The comparison is primarily focused on technical specifications and safety standards.

    The closest to "performance" in a diagnostic sense relates to the ECG interpretation features.

    Acceptance Criteria CategorySpecific Criteria (Implicitly "comparable to predicate")Reported Device Performance
    Intended UseSame as predicate (acquire ECG from adult and pediatric patients, analyze and diagnose heart disease on an advisory basis).Same
    Safety SpecificationsCompliance with relevant IEC standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-2-25).Complies with IEC 60601-1:2005/A1:2012, IEC 60601-1-2:2007 (for all models). SE-301 specifically mentions compliance to IEC 60601-1:2005/A1:2012 and EN 60601-1:2006/A1:2013, IEC 60601-1-2:2007, EN 60601-1-2:2007/AC:2010, IEC 60601-2-25:2011.
    BiocompatibilityCompliance with ISO 10993-1 for surface-contacting devices (<24 hours).Passes Cytotoxicity, Skin Sensitization, Skin Irritation tests.
    Bench TestingPass tests per IEC 60601-2-25:2011.All results show pass.
    ECG Interpretation FeaturesPerformance comparable to predicate devices' SEMIP algorithm (V1.6 and V1.7 versions).V1.7 algorithms for SE-3 series were tested for pediatrics and results were comparable to V1.8. V1.8 in SE-301 is the same as SE-12 Series. Improvements in V1.7 for RVH, MI, T wave abnormal, and ventricular preexcitation measurement accuracy.
    Software Verification & ValidationCompliance with FDA Guidance for "moderate" level of concern software.Software verification and validation testing conducted, documentation provided, device considered "moderate" level of concern, demonstrating proper function.
    Electrical Performance (e.g., HR Accuracy, Noise, CMRR)Maintain or improve upon predicate specifications.Many specifications are "Same" as predicate (e.g., HR range, accuracy, noise). Some are "Different" with improved values (e.g., A/D resolution, input circuit current, DC offset voltage, frequency response, CMRR, sampling frequency).

    2. Sample Size Used for the Test Set and Data Provenance

    The document states, "ECG interpretation features were also validated by database testing." However, no specific sample size for this test set is provided. The data provenance (e.g., country of origin, retrospective/prospective) is also not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This detailed information is not provided in the document. It only mentions "ECG interpretation features were also validated by database testing" without elaborating on how the ground truth of that database was established or by whom.

    4. Adjudication Method for the Test Set

    The adjudication method for establishing ground truth from the test set is not specified.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study is not mentioned. The submission focuses on demonstrating substantial equivalence based on technical specifications, safety, and a "database testing" for interpretation features, rather than human reader performance with and without AI assistance. Therefore, no effect size of human readers improving with AI assistance is reported.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The statement "ECG interpretation features were also validated by database testing" strongly suggests a standalone algorithm performance evaluation was conducted, where the algorithm's interpretations were compared against a database. However, details of this standalone study (metrics, specific results beyond "pass," sample size) are not provided.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The document does not explicitly state the type of ground truth used for the "database testing" of the ECG interpretation features. It can be inferred that it would be against a known or expert-adjudicated database of ECG recordings, but the specific nature of that ground truth (e.g., expert consensus, comparison to gold standard diagnoses, or other clinical data) is not detailed.

    8. The Sample Size for the Training Set

    The document does not report the sample size for the training set used to develop or train the SEMIP algorithm versions.

    9. How the Ground Truth for the Training Set Was Established

    The document does not describe how the ground truth for the training set was established. It mentions that "the criteria of some diseases are optimized to improve automatic diagnostic accuracy including RVH, MI, T wave abnormal and optimization of some terms and optimize ventricular preexcitation measure method to improve accuracy" for SEMIP V1.7, implying some form of optimization based on clinical data, but the method of ground truth establishment is not detailed.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170995

    Validate with FDA (Live)

    Date Cleared
    2017-06-02

    (60 days)

    Product Code
    Regulation Number
    870.2340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K160876

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SE-18 18-lead electrocardiograph is intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only.

    Device Description

    The SE-18 acquires and displays an 18 leads waveforms, which can also be printed by an integrated thermal printer with effective recording width 216mm. Digital filtering techniques similar to those incorporated on the SE-1515 and SE-12 have been used in SE-18: including Anti-baseline drift filter, AC filter (50/60Hz), EMG Filter and Low pass Filter, which can help the user to record a higher quality ECG. The EDAN Instruments "Smart ECG Measurement and interpretation program" (SEMIP) is included in this machine. The SEMIP program is completely integrated and provides the clinician with a detailed analysis of the ECG signal to aid in the interpretation of the ECGs. The recorded ECG can be saved in flash memory or sent to a PC by Ethernet or WIFI. During the examination, there are no substances delivered to and/or extracted from the patient. SE-18's function block diagram is shown as Fig 1. It's made up of an external ECG Sampling Box, SE-18 Main Control Board with RAM, FLASH, Ethernet module, WIFI module and USB module, a thermal printer module, a power supply module, a keyboard, and 15' 1024*768 display module.

    AI/ML Overview

    The provided text describes the 510(k) submission for the Edan Instruments, Inc. SE-18 Electrocardiograph. It establishes substantial equivalence to predicate devices (EDAN Instrument, Inc. Electrocardiograph, models SE-1515 and SE-12 Series) based on similar design features, performance specifications, and non-clinical testing.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a "table of acceptance criteria" in the format of pass/fail thresholds against which device performance is directly measured. Instead, it compares the specifications of the proposed device (SE-18) against the predicate devices (SE-1515 and SE-12 Series) in a series of tables. The acceptance criteria are implicitly met if the proposed device's performance characteristics are "Same" as or demonstrate "minor differences" that "do not raise different questions of safety or effectiveness" compared to the predicate devices.

    Below is a summary drawn from the provided comparison tables, focusing on a few key performance parameters. It consolidates information from Tables 1 and 2-2.

    Performance CharacteristicAcceptance Criteria (Predicate)Reported Device Performance (SE-18)Outcome (compared to Predicate)
    HR Recognition
    HR Range30 bpm ~300 bpm30 bpm ~300 bpmSame
    Accuracy± 1 bpm± 1 bpmSame
    ECG Unit (Electrical)
    Leads18 standard leads (SE-1515); Standard 12 leads (SE-12 Series)18 standard leadsSame (vs. SE-1515) / Different (vs. SE-12 Series, with explanation below)
    Sampling Frequency16kHz (SE-1515); 10k Hz (SE-12 Series)16kHzSame (vs. SE-1515) / Different (vs. SE-12 Series)
    A/D24 bits24bitsSame
    Resolution0.1575uV/LSB (SE-1515); 2.52uV/LSB (SE-12 Series)0.1575 μV/LSBSame (vs. SE-1515) / Different (vs. SE-12 Series)
    Frequency Response0.01~300Hz(-3dB) (SE-1515); 0.05 Hz ~ 150 Hz (-3 dB) (SE-12 Series)0.01~300Hz(-3dB)Same (vs. SE-1515) / Different (vs. SE-12 Series)
    Input Impedance≥100MΩ (10Hz) (SE-1515); ≥50 MΩ(10 Hz) (SE-12 Series)≥100M Ω (10Hz)Same (vs. SE-1515) / Different (vs. SE-12 Series)
    Noise≤12.5µVp-p≤12.5μVp-pSame
    Multichannel Crosstalk≤0.5mm≤0.5mmSame
    Filters
    DFT Filter settings0.01Hz/0.05Hz/0.32Hz/0.67Hz (SE-1515); 0.05Hz / 0.15Hz / 0.25Hz / 0.32Hz / 0.5Hz / 0.67Hz (SE-12 Series)0.01Hz/0.05Hz/0.32Hz/0.67HzSame (vs. SE-1515) / Different (vs. SE-12 Series)
    Lowpass Filter settings300Hz/270Hz/150Hz/100Hz/75Hz (SE-1515); 150Hz / 100Hz / 75Hz (SE-12 Series)300Hz/270Hz/150Hz/100Hz/75HzSame (vs. SE-1515) / Different (vs. SE-12 Series)
    CMRR≥123dB (AC OFF) (SE-1515); ≥115dB (AC Off) (SE-12 Series)≥140dB (AC On), ≥123dB (AC Off)Greater than (vs. SE-12 Series) / Same (vs. SE-1515 for AC Off)
    Pacemaker Detection
    Amplitude±750μV to ±700mV (SE-1515); ±2 to ±700 mV (SE-12 Series)±750μV to ±700mVSame (vs. SE-1515) / Different (vs. SE-12 Series)
    Width50us to 2.0ms (SE-1515); 0.1 to 2.0 ms (SE-12 Series)50µs to 2.0msSame (vs. SE-1515) / Different (vs. SE-12 Series)

    The document states that SE-18 has "minor differences" compared to the predicate devices and that these "do not raise different questions of safety or effectiveness." This implies that the observed differences are within acceptable limits or represent improvements that do not negatively impact safety or effectiveness.

    2. Sample size used for the test set and the data provenance

    The document explicitly states "Clinical data: Not applicable."
    For non-clinical data, it mentions Bench Testing was conducted per IEC 60601-2-25: 2011, and "all the results show pass." However, specific sample sizes for this general bench testing are not provided. The data provenance is not mentioned beyond being part of the manufacturer's internal testing. It is retrospective, as it was submitted for a 510(k) premarket notification.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable, as no clinical study or expert-based ground truth establishment is described for the device's performance other than that the device's interpretation statements are "offered to clinicians on an advisory basis only." For the safety and performance testing, standards like IEC 60601-1 and IEC 60601-2-25 are used, which do not typically involve human expert consensus for ground truth on device specifications like sampling frequency or input impedance.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as there is no described clinical study involving assessment by human readers requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The document states "Clinical data: Not applicable" and describes an electrocardiograph, not an AI-assisted diagnostic tool requiring an MRMC study. The "EMIP" program for interpretation is mentioned, but its effectiveness or impact on human readers is not studied or presented here.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    The device includes the "EDAN Instruments 'Smart ECG Measurement and interpretation program' (SEMIP)". While this is an algorithm, the document does not describe a standalone performance study of SEMIP's diagnostic accuracy. It states that "the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only," suggesting it's not intended for standalone diagnostic use. The focus is on the device's hardware and basic signal acquisition/processing capabilities meeting predicate equivalence.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical performance and safety specified, the ground truth is based on engineering specifications and adherence to international standards (e.g., IEC 60601-1, IEC 60601-2-25). For the SEMIP program, no explicit ground truth methodology for its interpretive statements is provided in this submission summary. Given its "advisory basis only" role, a rigorous ground truth validation for diagnostic accuracy might not have been required for this 510(k).

    8. The sample size for the training set

    Not applicable. The document does not describe a machine learning model that would require a training set. The SEMIP program is mentioned, but details on its development data or training set are not provided.

    9. How the ground truth for the training set was established

    Not applicable, as no training set for a machine learning model is described in the document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1