Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K183200

    Validate with FDA (Live)

    Date Cleared
    2019-01-09

    (51 days)

    Product Code
    Regulation Number
    886.5916
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K160859

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iSee (roflufocon D, roflufocon E) Daily Wear Orthokeratology Contact Lenses are indicated for daily wear in an orthokeratology fitting program for the temporary reduction of myopia of up to 5.00 diopters in non-diseased eyes. To maintain the orthokeratology effect of myopia reduction, lens wear must be continued on a prescribed wearing schedule.

    Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.

    Device Description

    The iSee (roflufocon D, roflufocon E) Daily Wear Orthokeratology Contact Lenses are are lathe cut from one of the following hydrophobic, FDA Group #3 fluoro-silicone acrylate materials: roflufocon D or roflufocon E-with the following properties: Refractive Index, Light Transmission (clear), Light Transmission (tinted), Wetting Angle (Dynamic contact receding angle), Specific Gravity, Oxygen Permeability (Dk) ISO/FATT Method, Visitint lenses contain one or more of the following color additives conforming to: 21 CFR Part 73 & 74, Subpart D. The lenses are available in the following lens parameters: Base Curve, Center Thickness, Diameter, Spherical Power, Secondary Curves, Peripheral Curves, Multifocal Power.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for the iSee Daily Wear Orthokeratology Contact Lenses. It does not contain any information about acceptance criteria or a study proving the device meets acceptance criteria in the context of an AI/ML medical device.

    Instead, the document focuses on demonstrating substantial equivalence to an already marketed predicate device (OPTIMUM GP OK Daily Wear Contact Lens). The basis for this equivalence is a comparison of technological characteristics, materials, production methods, and intended use.

    Here's a breakdown of why this document doesn't provide the requested AI/ML device study information:

    • Non-Clinical and Clinical Studies (Section VII): This section explicitly states:
      • "The non-clinical performance data to establish the safety and effectiveness of contact lenses manufactured from roflufocon D and roflufocon E has been addressed by reference to the predicate devices."
      • "Clinical performance data to demonstrate the safety and effectiveness of contact lenses manufactured from roflufocon D and roflufocon E have been addressed in previous applications."

    This means that C&E GP Specialists, Inc. did not conduct new independent studies for this 510(k) submission. They are relying on existing data for the materials and the predicate device to establish safety and effectiveness.

    Therefore, I cannot provide the requested information from this document for an AI/ML device since it is for a contact lens and references prior data rather than presenting a new study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171077

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-08-01

    (112 days)

    Product Code
    Regulation Number
    886.5916
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K160859, K161100

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HEXA100 (hexafocon A) SPHERICAL Rigid Gas Permeable (RGP) Contact Lens is indicated for daily wear for the correction of refractive error in aphakic and not aphakic persons with nondiseased eyes with myopia or hyperopia.

    The HEXA100 (hexafocon A) TORIC Rigid Gas Permeable (RGP) Contact Lens is indicated for daily wear for the correction of refractive error in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 10.00 diopters.

    The HEXA100 (hexafocon A) MULTIFOCAL/BIFOCAL Rigid Gas Permeable (RGP) Contact Lens is indicated for daily wear for the correction of refractive error in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters.

    The HEXA100 (hexafocon A) IRREGULAR CORNEA Daily Wear Contact Lens may be prescribed in otherwise non-diseased eyes that require a rigid gas permeable lens for the management of irregular corneal conditions such as; keratoconus, pellucid marginal degeneration or following penetrating keratoplasty or refractive (e.g. LASIK) surgery.

    The HEXA100 (hexafocon A) ORTHOKERATOLOGY contact lenses are indicated for daily wear in an orthokeratology fitting program for the temporary reduction of myopia of up to 5.00 diopters in non-diseased eyes. To maintain the orthokeratology effect of myopia reduction, lens wear must be continued on a prescribed wearing schedule.

    Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.

    Device Description

    The HEXA100 (hexafocon A) Daily Wear Contact Lens is manufactured from a machine latheable rigid gas permeable material composed of siloxanyl fluoromethacrylate copolymer that is tinted for visibility and available with or without an ultraviolet (UV) light absorber.

    In the HEXA100 (hexafocon A) Daily Wear Contact Lens with UV Blocker, a Benzophenone UV blocking monomer is used to block UV radiation. The UV blocking for HEXA100 averages > 98% in the UVB range of 280nm - 315nm and >84% in the UVA range of 316 - 380nm.

    The HEXA100 (hexafocon A) Daily Wear Contact Lens incorporates a visibility tint to make the lens more visible for handling. The tinted lenses contain one or more of the following color additives: D&C Green No. 6, C.I. Solvent Yellow No. 18, D&C Violet No. 2 and D&C Red No. 17.

    The HEXA100 (hexafocon A) Daily Wear Contact Lens may be prescribed in a daily wear orthokeratology fitting program for the temporary reduction of myopia of up to 5.00 diopters in nondiseased eyes.

    The HEXA100 (hexafocon A) Daily Wear Contact Lens may be optionally treated to incorporate Tangible "10 Hydra-PEG-which is a thin polyethylene glycol (PEG)-based polymer that is covalently bonded to the surface of the contact lens and is designed to enhance the surface properties of the contact lens while retaining the mechanical properties of the underlying material. When treated with Tangible 10 Hydra-PEG, the underlying material (hexafocon A) is encapsulated in a thin layer of polymer that results in measurable improvement of wettability (sessile drop contact angle) compared to untreated lenses. The resulting layer is hydrophilic and approximately 30mm in thickness.

    The HEXA100 (hexafocon A) Daily Wear Contact Lenses may be packaged and shipped "dry" or "wet" in a polypropylene contact lens case. The primary container for shipping the HEXA100 (hexafocon A) Daily Wear Contact Lenses is the PolyVial Contact Lens Case. When shipped "wet", the HEXA100 (hexafocon A) Daily Wear Contact Lenses may be packaged and shipped in the Unique pH contact lens care system by Menicon Co., Ltd. The active ingredients in Unique pH solution are Edetate Disodium 0.01% and Polyquaternium 10.0011%.

    The HEXA100 (hexafocon A) Daily Wear Contact Lens is available in the Spherical. Toric. Multifocal/Bifocal, Irregular Cornea (Scleral) design configurations, within the following lens parameters:
    • Chord Diameter: 7.0mm to 22.0mm

    • Center Thickness: Varies
    • Base Curve: 5.0mm to 8.0mm
    • Spherical Powers: -20.00 Diopters to +20.00 Diopters
    • Toric Powers: up to -10.00 Diopters
    • Add Powers: up to +4.00 Diopters

    The HEXA100 (hexafocon A) Daily Wear Contact Lens for daily wear orthokeratology is available in following lens parameters:
    ● Chord Diameter: 6.5 to 11.5 mm
    ● Center Thickness: 0.10 to 0.70 mm

    • Base Curve (BC): 6.5 to 11.0 mm
      ● Secondary Curves: 0.10 to 2.0 mm (flatter or steeper than BC)
      ● Peripheral Curves: 0.10 to 2.0 mm (flatter or steeper than BC)
    • Spherical Powers: -10.00 to +3.00 D
    AI/ML Overview

    Acceptance Criteria and Device Performance for HEXA100 (hexafocon A) Daily Wear Contact Lens

    This submission establishes substantial equivalence for the HEXA100 (hexafocon A) Daily Wear Contact Lens by comparing its performance to predicate devices. The acceptance criteria are essentially demonstrated equivalence to these established devices through physical, chemical, and biological testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    The device demonstrates performance equivalent or superior to predicate devices across various physical and chemical properties and biocompatibility. The acceptance criteria for the HEXA100 (hexafocon A) Daily Wear Contact Lens are implicitly set by the performance characteristics of its predicate devices and internal validation for safety and stability.

    PropertyAcceptance Criteria (Based on Predicate Devices and Safety Standards)HEXA100 (hexafocon A) Reported Performance (Uncoated)HEXA100 (hexafocon A) Reported Performance (Hydra-PEG Coated)
    Biocompatibility & ToxicologyNon-toxic, non-irritating (GLP-compliant)Not toxic, not irritatingNot toxic, not irritating
    Physicochemical, Mechanical, Optical PropertiesSubstantially equivalent to predicate devicesSubstantially equivalent to predicatesSubstantially equivalent to predicates
    Bioburden Levels<100 cfu/lens initially and after 30 days storageBelow acceptance criteriaBelow acceptance criteria
    Physical Parameter StabilityStable after 30 days storage in Unique pHStable after 30 days storageStable after 30 days storage
    Compatibility with Care SolutionsPhysically compatible with currently marketed solutionsPhysically compatiblePhysically compatible
    Material (USAN name)hexafocon A (Boston XO)hexafocon Ahexafocon A
    Water Content (%)<1%<1%<1%
    Specific Gravity (g/cm³)1.265 (Boston XO)1.2661.266
    Oxygen Permeability (Dk)108 x 10⁻¹¹ (Boston XO)113 x 10⁻¹¹118 x 10⁻¹¹
    Wettability (Sessile Drop Contact Angle)96.4° (Boston XO) / Roflufocon C, D, E specific values (Optimum GP with HPT, OPTIMUM GP OK)96.4°49.1° (significant improvement)
    UV Absorber/Blocker AvailabilityYes, optional (similar to predicates)Yes, OptionalYes, Optional
    Includes Hydra-PEG Surface CoatingYes (similar to predicate Optimum GP with HPT)No (for uncoated)Yes, Optional
    Refractive Index (dry)Not explicitly stated for predicates, but consistency expected1.41361.4134
    Light Transmission (@380-780nm)Not explicitly stated for predicates, but >90% generally expected>91%>94%
    UVA Transmission (@316-380nm)<16% (demonstrates UV blocking)<16%<16%
    UVB Transmission (@280-315nm)<2% (demonstrates UV blocking)<2%<2%

    2. Sample Size for Test Set and Data Provenance

    The provided document does not explicitly state a sample size for a "test set" in the context of clinical trials as it relies heavily on non-clinical testing and established clinical safety of the material and coating.

    • Non-clinical Testing: Conducted "in vitro and in vivo" for toxicology and biocompatibility. All toxicology tests were conducted in accordance with GLP regulations.
    • Wettability Data (Sessile Drop Contact Angle): The table indicates "n=30" for the average sessile drop contact angle measurements.
    • Data Provenance: The document does not specify the country of origin for the non-clinical testing data. It also does not specify whether the data is retrospective or prospective, although GLP (Good Laboratory Practice) regulations generally imply prospective and controlled studies for toxicology.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    The concept of "experts" establishing ground truth as typically understood in AI/imaging studies (e.g., radiologists) is not applicable here. This submission focuses on the material's physical, chemical, and biological properties, and its substantial equivalence to predicate contact lenses. The "ground truth" for these tests is established by scientific methodologies and accepted standards (e.g., ISO for Dk, GLP for toxicology).

    4. Adjudication Method for the Test Set

    Adjudication methods (e.g., 2+1, 3+1) are not applicable here. These methods are typically used in clinical studies where human interpretation of medical images or patient outcomes requires consensus among multiple experts. The testing performed for this device involved laboratory measurements and analyses of material properties and biological responses, not subjective interpretations requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC comparative effectiveness study was not performed, nor is it applicable for this device. Such studies are typically conducted for AI-assisted diagnostic devices where the interaction between human readers (e.g., radiologists) and AI algorithms is being evaluated for improved diagnostic accuracy or efficiency. The HEXA100 contact lens is a passive medical device, not an AI diagnostic tool.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    A standalone algorithm-only performance study was not performed, nor is it applicable. This device is a contact lens, not an algorithm.

    7. Type of Ground Truth Used

    The "ground truth" for this submission is based on:

    • Scientific and Engineering Standards: Measurements of physical properties (e.g., refractive index, specific gravity, oxygen permeability) and chemical properties.
    • Biological Standards: Results from in vitro and in vivo biocompatibility and toxicology tests, often governed by ISO standards (e.g., ISO 10993 for biological evaluation of medical devices).
    • Predicate Device Performance: The established safety and effectiveness profiles of legally marketed predicate devices serve as the benchmark for demonstrating substantial equivalence.

    8. Sample Size for the Training Set

    A "training set" is not applicable for this device. This is not a machine learning or AI device that requires a training set. The manufacturing process of contact lenses (lathe-cut, custom manufactured) and the material properties are characterized through traditional scientific and engineering methods.

    9. How the Ground Truth for the Training Set Was Established

    As there is no training set, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1